Acme Pediatric Providers, Pllc

CLIA Laboratory Citation Details

4
Total Citations
35
Total Deficiencyies
13
Unique D-Tags
CMS Certification Number 45D1006441
Address 500 North Sam Houston Blvd, Suite 2, San Benito, TX, 78586
City San Benito
State TX
Zip Code78586
Phone956 276-0400
Lab DirectorOFELIA RAMOS

Citation History (4 surveys)

Survey - October 8, 2024

Survey Type: Standard

Survey Event ID: ZDTB11

Deficiency Tags: D1001 D2007 D2010 D5417 D5813 D1001 D2007 D2010 D5417 D5813

Summary:

Summary Statement of Deficiencies D1001 CERTIFICATE OF WAIVER TESTS CFR(s): 493.15(e) Laboratories eligible for a certificate of waiver must-- (1) Follow manufacturers' instructions for performing the test; and (2) Meet the requirements in subpart B, Certificate of Waiver, of this part. This STANDARD is not met as evidenced by: Based on review of manufacturer's instructions for the Piccolo Comprehensive Metabolic Panel cassette, review of the laboratory's quality assurance reports from November 2023 and December 2023, review of patient test records from November 2023 to January 2024, and staff interview, the laboratory failed to following manufacturer's instructions by ensuring quality control testing was acceptable prior to testing patient samples on 3 of 4 test dates. The findings included: 1. The manufacturer's instructions for the Piccolo Comprehensive Metabolic Panel (March 2015 PN:400-7139-1 Rev. P) stated: "Do not report results if controls are outside the labeled limits." 2. A review of the laboratory's quality assurance report from November 2023 determined the laboratory identified quality control results for the analyte Total Bilirubin failed on 11/17/2023 and 12/01/2023. The quality assurance report stated: "Please consider reported patient tests to be invalid for procedures that had controls out of limits. And: a) Retrieve patient record and indicate that the test result was invalid. b) Make the correction, but don't white out or erase the previous result. You may make one line through it, but don't erase. c) Notify the requesting physician of the error and request permission to repeat the test. d) Keep a record in lab of what happened." 3. A review of the laboratory's quality assurance report from December 2023 determined the laboratory identified quality control results for the analyte Total Bilirubin failed on 12/08/2023. The quality assurance report stated: "Please invalidate patients performed 12/08 and 01/04/2024." 4. Quality control materials were with each new lot, shipment, or every 30 days on the Piccolo analyzer. Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 4 -- 5. A review of patient test records from November 17, 2023 to November 29, 2023 identified 64 patients were tested and reported after the quality control failure (see patient alias list 1). 6. A review of patient test records from December 8, 2023 to January 4, 2024 identified 80 patients were tested and reported after the quality control failure (see patient alias list 2). 7. The technical consultant confirmed the findings in an interview conducted on 10/08/2024 at 1140 hours in the break room. D2007 TESTING OF PROFICIENCY TESTING SAMPLES CFR(s): 493.801(b)(1) The samples must be examined or tested with the laboratory's regular patient workload by personnel who routinely perform the testing in the laboratory, using the laboratory's routine methods This STANDARD is not met as evidenced by: NOTE: THIS IS A REPEAT DEFICIENCY FROM THE SURVEY CONDUCTED 11/30/2022. Based on review of the laboratory's submitted CMS 209 Form, review of the laboratory's American Proficiency Institute's proficiency testing records from 2023 and 2024, and staff interview, the laboratory failed to ensure 3 of 4 testing personnel participated in proficiency testing. The findings included: 1. A review of the laboratory's submitted Form CMS 209 revealed the laboratory identified 5 testing testing personnel. Four of the five testing personnel were employed by the facility in 2023 and 2024 at the time proficiency testing was performed. 2. A review of the laboratory's American Proficiency Institute's proficiency testing records from 2023 (events 1, 2, and 3) and 2024 (events 1 and 2) revealed the laboratory failed to have documentation of 3 of 4 testing personnel participating in proficiency testing. They were (as listed on Form CMS 209): Testing personnel number 1 Testing personnel number 2 Testing personnel number 5 3. The technical consultant confirmed the findings in an interview conducted on 10/08/2024 at 0920 hours in the break room. D2010 TESTING OF PROFICIENCY TESTING SAMPLES CFR(s): 493.801(b)(2) The laboratory must test samples the same number of times that it routinely tests patient samples. This STANDARD is not met as evidenced by: Based on review of the laboratory's proficiency testing policy, review of the laboratory's American Proficiency Institute's proficiency testing records from 2023 and 2024, and staff interview, the laboratory failed to ensure proficiency testing samples were tested the same number of times as patient samples for 1 of 5 events. The findings included: 1. A review of the laboratory's proficiency testing policy (approved 11/8/20218) stated: "PT specimens are to be treated the same as patient samples." Patient samples were routinely only tested once. 2. A review of the laboratory's American Proficiency Institute's proficiency testing records from 2023 (events 1, 2, and 3) and 2024 (events 1 and 2) revealed the laboratory tested proficiency testing samples multiple times on 1 of 5 events. The proficiency testing samples from 2024 Event 2 were tested twice. Sample: HSY-6 tested: 7/19/2024 10: 13 7/19/2024 10:23 Sample: HSY-7 tested: 7/19/2024 10:16 7/19/2024 10:25 Sample: HSY-8 tested: 7/19/2024 10:18 7/19/2024 10:26 Sample: HSY-9 tested: 7/19/2024 10: 20 7/19/2024 10:28 Sample: HSY-10 tested: 7/19/2024 10:21 7/19/2024 10:29 3. The -- 2 of 4 -- technical consultant confirmed the findings in an interview conducted on 10/08/2024 at 0920 hours in the break room. D5417 TEST SYSTEMS, EQUIPMENT, INSTRUMENTS, REAGENT CFR(s): 493.1252(d) Reagents, solutions, culture media, control materials, calibration materials, and other supplies must not be used when they have exceeded their expiration date, have deteriorated, or are of substandard quality. This STANDARD is not met as evidenced by: Based on review of the laboratory's quality control records from September 2024, review of the laboratory's patient test records from September 2024, and staff interview, the laboratory failed to ensure quality control material was not expired prior to use on 1 of 23 test days. The findings included: 1. A review of the laboratory's quality control records from September 2024 determined the laboratory tested expired quality control material prior to testing patients on 1 of 23 days. On September 24, 2024 the following expired control material was tested: Low control: Lot 2240501 expiration date: 9/23/2024 Normal control: Lot 24440502 expiration date: 9/23/2024 2. A review of patient test records from September 24, 2024 determined the following patients were tested: Sequence: 3888 Sequence: 3889 Sequence: 3890 Sequence: 3891 After the patients were tested, the laboratory calibrated the instrument and tested new lots of control material. The patient samples were not retested after the acceptable controls. 3. The technical consultant confirmed the findings in an interview conducted on 10/08/2024 at 0950 hours in the conference room. D5813 TEST REPORT CFR(s): 493.1291(g) The laboratory must immediately alert the individual or entity requesting the test and, if applicable, the individual responsible for using the test results when any test result indicates an imminently life-threatening condition, or panic or alert values. This STANDARD is not met as evidenced by: NOTE: THIS IS A REPEAT DEFICIENCY FROM THE SURVEY CONDUCTED 11/30/2022. Based on review of the laboratory's policies, review of patient test records from September 2024, review of the laboratory's critical value log from May 2024 to October 7, 2024, and staff interview, the laboratory failed to have documentation of the notification of 4 of 5 critical values. The finding include: 1. A review of the laboratory's policy titled "Critical Values" (approved by the laboratory director on 11/8/2018) revealed the laboratory had the following defined critical values: White Blood Cells under 2 or greater than 20.0 Hemoglobin under 7.5 or greater than 18 Hematocrit under 25 or greater than 55 Platelets under 50 or greater than 800 2. A review of the laboratory's policy titled "Reporting Critical Values" (approved by the laboratory director on 11/8/2028) revealed: "It is the policy of this laboratory to document the reporting of Critical Values. Document: - who was notified - when was the person notified - by who was the person notified." 3. A review of patient test records from September 2024 identified the following 5 results which met the laboratory's criteria as a critical result: Date Seq Value 09/03 3722 WBC: 20.4 09/03 3726 WBC: 44.4 09/23 3879 HBG: 18.8 09/24 3915 HCT: 567 09/24 3921 PLT: 19 4. A review of the laboratory's critical value log from September 2024 -- 3 of 4 -- determined the following 4 critical values were not documented: Date Seq Value 09 /03 3722 WBC: 20.4 09/23 3879 HBG: 18.8 09/24 3915 HCT: 567 09/24 3921 PLT: 19 5. An interview with the technical consultant on 10/08/2024 at 1100 hours in the break room - after his review of the records- confirmed the findings. Key Seq - Sequence WBC - white blood cell HGB - hemoglobin HCT - hematocrit PLT - platelet -- 4 of 4 --

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access

Survey - November 30, 2022

Survey Type: Standard

Survey Event ID: RW4B11

Deficiency Tags: D2007 D5785 D5813 D5813 D0000 D2007 D5785

Summary:

Summary Statement of Deficiencies D0000 Noted deficiencies and plans of correction were discussed with the laboratory representative(s) at the exit conference. The facility was found to be in compliance with applicable Conditions of Participation in the CLIA program, and recertification is recommended. . D2007 TESTING OF PROFICIENCY TESTING SAMPLES CFR(s): 493.801(b)(1) The samples must be examined or tested with the laboratory's regular patient workload by personnel who routinely perform the testing in the laboratory, using the laboratory's routine methods This STANDARD is not met as evidenced by: Based on review of the laboratory's submitted Form CMS 209, review of the laboratory's American Academy of Family Physicians proficiency testing records from 2022, and staff interview, it was revealed the laboratory failed to ensure 2 of 5 testing personnel participated in proficiency testing. The findings include: 1. A review of the laboratory's submitted Form CMS 209 revealed the laboratory identified 5 testing testing personnel. 2. A review of the laboratory's American Academy of Family Physicians proficiency testing records from 2022 (events 1, 2, and 3) revealed the laboratory failed to have documentation of 3 of 5 testing personnel participating in proficiency testing. They were (as listed on Form CMS 209): Testing personnel number 2 Testing personnel number 3 Testing personnel number 4 3. The laboratory was asked to provide documentation of the identified personnel participating in proficiency testing. No documentation was provided. 4. An interview with the technical consultant on 11/30/2022 at 830 hours in the break room confirmed the findings. D5785

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access

Survey - September 21, 2021

Survey Type: Standard

Survey Event ID: 17LS11

Deficiency Tags: D0000 D2009 D0000 D2009

Summary:

Summary Statement of Deficiencies D0000 The laboratory was surveyed and found to be in compliance with the conditions of participation found in the CLIA regulations at 42 CFR 493 and recertification is recommended. D2009 TESTING OF PROFICIENCY TESTING SAMPLES CFR(s): 493.801(b)(1) The individual testing or examining the samples and the laboratory director must attest to the routine integration of the samples into the patient workload using the laboratory's routine methods. This STANDARD is not met as evidenced by: Based on review of the laboratory's American Academy of Family Physicians (AAFP) proficiency testing (PT) records, and confirmed in interview of facility personnel, the laboratory failed to provide documentation of the laboratory director signing two of eight attestation statements. The findings included: 1. Review of proficiency testing instructions from AAFP under Attestation Statement states, "We, the undersigned, recognizing that some special handling may be required due to the nature of proficiency testing materials, have a closely as practical, performed the analyses on these specimens in the same manner as regularly performed by patient samples. By signature below, the lab is authorizing results to be submitted to the regulatory /accrediting agencies." 2. Review of the laboratory's AAFP proficiency testing records from 2019 (events A, B, and C), 2020 (events A, B, and C), and 2021 (events A and B) found the laboratory director failed to sign the following attestation sheets. 2020 (event A) 2021 (event A) 2021 (event B) 3. An interview with the Technical Consultant on September 21, 2021 at 09:30 hours in the break room confirmed the findings. Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 1 --

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access

Survey - October 17, 2018

Survey Type: Standard

Survey Event ID: QUW811

Deficiency Tags: D0000 D1001 D5400 D5403 D5411 D6056 D6057 D0000 D1001 D5400 D5403 D5411 D6056 D6057

Summary:

Summary Statement of Deficiencies D0000 The laboratory was surveyed on October 17, 2018. The facility was not in compliance with the following CLIA Conditions at 42 CFR part 493: D5400 - 42 C.F.R. 493.1250 Condition: Analytic Systems D6056 - 42 C.F.R. 493.1415 Condition: Laboratories performing moderate complexity testing - Clinical Consultant Laboratories issued a certificate of compliance must meet the following conditions: (a) Notify HHS or its designee within 30 days of any changes in- (1) Ownership; (2) Name; (3) Location; (4) Director; or (5) Technical supervisor (laboratories performing high complexity only). Review of the "Stock Purchase Agreement" revealed the agreement was "Signed December 6, 2016, but is effective as of October 17, 2016 (the "Effective Date"). The laboratory did not notify CLIA of a change in ownership within 30 days of its changes. The laboratory did not notify CLIA of a laboratory director change within 30 days of its changes. D1001 CERTIFICATE OF WAIVER TESTS CFR(s): 493.15(e) Laboratories eligible for a certificate of waiver must-- (1) Follow manufacturers' instructions for performing the test; and (2) Meet the requirements in subpart B, Certificate of Waiver, of this part. This STANDARD is not met as evidenced by: Based on surveyor observations, review of manufacturer's instructions, and confirmed in interview of facility personnel, the laboratory failed to follow the manufacturer's instructions to monitor revised expiration dates on i-STAT Chemistry 8 test cartridges. The findings were: 1. Direct observation made in the laboratory on October 17, 2018 at 09:20 hours revealed 3 Abbott i-STAT Chemistry 8 test cartridges on the laboratory counter. The cartridges did not have a revised expiration date on them. According to testing personnel two (as listed on Form CMS 209), the cartridges had been at room temperature for, "About three days." 2. Review of the manufacturer's Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 4 -- instructions on the package labeling for the Abbott i-STAT Chemistry 8 test cartridges revealed room temperature storage is 14 days. 3. An interview with testing person 2 (as listed on Form CMS 209) on October 17, 2018 at 09:05 hours in the laboratory confirmed the findings. When asked if the laboratory documents the date the cartridges are put at room temperature, she stated, "No." Key: CMS - Centers for Medicare and Medicaid Services D5400 ANALYTIC SYSTEMS CFR(s): 493.1250 Each laboratory that performs nonwaived testing must meet the applicable analytic systems requirements in 493.1251 through 493.1283, unless HHS approves a procedure, specified in Appendix C of the State Operations Manual (CMS Pub.7), that provides equivalent quality testing. The laboratory must monitor and evaluate the overall quality of the analytic systems and correct identified problems as specified in 493.1289 for each specialty and subspecialty of testing performed. This CONDITION is not met as evidenced by: Based on review of manufacturer's instructions, review of quality control records, review of patient records, and confirmed in interview of facility personnel, the laboratory failed to monitor overall quality of its analytic systems as evidenced by: 1. The laboratory failed to resolve flags on CBCs prior to their release to the healthcare provider. (refer to D5403). This is a repeat deficiency. 2. The laboratory failed to follow the manufacturer's instructions to re-analzye samples with system messages prior to their release to the healthcare provider. (refer to D5411) D5403 PROCEDURE MANUAL CFR(s): 493.1251(b) The procedure manual must include the following when applicable to the test procedure: (1) Requirements for patient preparation; specimen collection, labeling, storage, preservation, transportation, processing, and referral; and criteria for specimen acceptability and rejection as described in 493.1242. (2) Microscopic examination, including the detection of inadequately prepared slides. (3) Step-by-step performance of the procedure, including test calculations and interpretation of results. (4) Preparation of slides, solutions, calibrators, controls, reagents, stains, and other materials used in testing. (5) Calibration and calibration verification procedures. (6) The reportable range for test results for the test system as established or verified in 493.1253. (7) Control procedures. (8)

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access