Adult & Pediatric Dermatology Specialists, Pc

CLIA Laboratory Citation Details

4
Total Citations
8
Total Deficiencyies
7
Unique D-Tags
CMS Certification Number 07D0883635
Address 160 Hawley Lane, Trumbull, CT, 06611
City Trumbull
State CT
Zip Code06611
Phone(475) 265-5350

Citation History (4 surveys)

Survey - August 13, 2026

Survey Type: Standard

Survey Event ID: SU0A11

Deficiency Tags: D5413 D5473

Summary:

Summary Statement of Deficiencies D5413 TEST SYSTEMS, EQUIPMENT, INSTRUMENTS, REAGENT CFR(s): 493.1252(b) (b) The laboratory must define criteria for those conditions that are essential for proper storage of reagents and specimens, accurate and reliable test system operation, and test result reporting. The criteria must be consistent with the manufacturer's instructions, if provided. These conditions must be monitored and documented and, if applicable, include the following: (b)(1) Water quality. (b)(2) Temperature. (b)(3) Humidity. (b)(4) Protection of equipment and instruments from fluctuations and interruptions in electrical current that adversely affect patient test results and test reports. This STANDARD is not met as evidenced by: Based on surveyor observation, record review and staff interview, the laboratory failed to define and provide evidence of monitoring and documenting ambient room temperature and humidity requirements in the subspecialty of histopathology. Findings include: 1. Surveyor observation on 08/13/2026 at 10:45 AM of the laboratory's histopathology slide reading room revealed one of one 'Olympus BX45 Microscope' in use. 2. Record review on 08/13/2026 of the laboratory's equipment maintenance records for 2024 and 2025 revealed the lack of documentation of ambient room temperature and humidity levels for the histopathology slide reading room. 3. Record review on 08/13/2026 of the 'Olympus BX45 Microscope' operator's manual revealed the following operating environmental conditions: a. 'Ambient temperature: 5 to 40 degrees Celsius' b. 'Maximum relative humidity: 80%' 4. Staff interview on 08/13/2026 at 11:00 AM with the laboratory's practice manager confirmed the above findings. 5. The laboratory performs 4,635 tests annually in the subspecialty of histopathology. D5473 CONTROL PROCEDURES CFR(s): 493.1256(e)(2)(g) Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 2 -- (e)(2) Each day of use (unless otherwise specified in this subpart), test staining materials for intended reactivity to ensure predictable staining characteristics. Control materials for both positive and negative reactivity must be included, as appropriate. This STANDARD is not met as evidenced by: Based on record review and staff interview, the laboratory failed to document daily quality assessment of hematoxylin and eosin (H&E) stains, special stains, and immunohistochemistry (IHC) stains in the subspecialty of histopathology. Findings include: 1. Record review on 08/13/2026 of the laboratory's specimen transport logs for 2026 and 2025 revealed lack of documentation of daily quality assessment of hematoxylin and eosin (H&E) stains, special stains, and immunohistochemistry (IHC) stains. 2. Record review on 08/13/2026 of the laboratory's 'Quality Control Assurance' standard operating procedure revealed the following: a. 'Quality Control Procedure' i. 'Upon receipt and prior to interpretation, the dermatopathologist shall assess each slide for adequate staining quality and contrast'. b. Lack of documentation that the assessment of stain quality and contrast is required to be documented. 3. Staff interview on 08/13/2026 at 10:30 AM with the laboratory's practice manager confirmed the above findings. 4. The laboratory performs 4,635 tests annually in the subspecialty of histopathology. -- 2 of 2 --

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access

Survey - September 10, 2024

Survey Type: Standard

Survey Event ID: J8RZ11

Deficiency Tags: D5209 D5217 D5779

Summary:

Summary Statement of Deficiencies D5209 PERSONNEL COMPETENCY ASSESSMENT POLICIES CFR(s): 493.1235 As specified in the personnel requirements in subpart M, the laboratory must establish and follow written policies and procedures to assess employee and, if applicable, consultant competency. This STANDARD is not met as evidenced by: Based on record review and staff interview, the laboratory failed to establish competency assessment policy and procedures to assess competency for the regulatory responsibilities of the general supervisor (GS), technical supervisor (TS), technical consultant (TC) and clinical consultant (CC). Findings include: 1. Record review on 09 /10/2024 of the laboratory's 'Personnel Training/ Employee Proficiency Evaluations/ Monthly Meetings' binder revealed lack of documentation of competency assessment for the regulatory positions of GS, TS, TC and CC. 2. Record review on 09/10/2024 of the laboratory's policies and procedure revealed lack of documentation of an established competency assessment policy and procedures to assess competency for the regulatory responsibilities of the GS, TS, TC and CC. 3. Staff interview on 09/10 /2024 at 11:00 AM with testing personnel #1 confirmed the above findings. 4. The laboratory performs 413 tests annually in the subspecialty of mycology and 5,963 tests annually in the subspeciality histopathology. D5217 EVALUATION OF PROFICIENCY TESTING PERFORMANCE CFR(s): 493.1236(c)(1) At least twice annually, the laboratory must verify the accuracy of any test or procedure it performs that is not included in subpart I of this part. This STANDARD is not met as evidenced by: Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 2 -- Based on record review and staff interview the laboratory failed to follow their established policies and procedures to ensure that proficiency testing (PT) samples were performed twice annually in the subspecialty of histopathology. Findings Include: 1. Record review on 09/10/2024 of the laboratory's 'Monthly Quality Assurance/ Monthly Patient Quality Assurance/ Quality Control-Proficiency' binder revealed lack of documentation of PT samples being sent out for microscopic review /evaluations for 2023. 2. Record review on 09/10/2024 of the laboratory's 'Proficiency Testing: Mohs Micrographic Surgery Skin Specimen' procedure revealed 'Semi annually, the tech will send two cases containing the original slides and send it out for a microscopic examination by a Board Certified Dermatopathologist or Mohs surgeon'. 3. Staff interview on 09/10/2024 at 11:30AM with the testing personnel #1 confirmed the above findings. 4. The laboratory performs 5,963 samples annually in the subspecialty of histopathology. D5779

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access

Survey - March 29, 2022

Survey Type: Standard

Survey Event ID: O3JM11

Deficiency Tags: D5407 D5413

Summary:

Summary Statement of Deficiencies D5407 PROCEDURE MANUAL CFR(s): 493.1251(d) Procedures and changes in procedures must be approved, signed, and dated by the current laboratory director before use. This STANDARD is not met as evidenced by: Based on record review and staff interview, the laboratory failed to ensure changes to the fungal culture procedure were updated, approved and signed by the laboratory director in the subspecialty of mycology prior to patient testing. Findings include: 1. Record review on 3/29/2022 of the laboratory's 'Obtaining/entering/reading /documenting results overview' procedure revealed: a. The procedure for 'taking /obtaining a fungal culture' was to inoculate the fungal culture media Sabouraud Dextrose Agar. b. After 2 weeks the provider does a visual reading of the Sabouraud Dextrose agar vial and results a positive or negative culture. c. Approval date of the procedure was 7/31/2018. 2. Surveyor observation on 3/29/2022 at 12:00 PM of the laboratory's inoculated fungal culture media revealed Remel Dermatube Dermatophyte Test Medium (DTM), Lot #: 367893 exp 9/22/2022 in use. 3. Record review on 3/29/2022 of the Remel DTM manufacturer product insert revealed to examine the media at regular intervals for a red color development. 4. Record review on 3/29/2022 of the laboratory's 'Form 21: Laboratory Test Requisition and Report Log' revealed: a. 2 shipments of Sabouraud Dextrose Agar with date received 6/20/18 and 1/24/2019. b. 4 shipments of DTM with date received 1/17/2020, 9/4/2020, 10/21 /2021 and 2/16/2022 (Lot #367893). 5. Staff interview with the laboratory director (LD) on 3/29/2022 at 12:10 PM confirmed the above findings. The LD disclosed he /she was aware the laboratory switched to DTM at some point and a positive result on DTM is the color change to red but did not update the procedure. 6 The laboratory performs 277 fungal cultures annually in the subspecialty of mycology. D5413 TEST SYSTEMS, EQUIPMENT, INSTRUMENTS, REAGENT Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 2 -- CFR(s): 493.1252(b) The laboratory must define criteria for those conditions that are essential for proper storage of reagents and specimens, accurate and reliable test system operation, and test result reporting. The criteria must be consistent with the manufacturer's instructions, if provided. These conditions must be monitored and documented and, if applicable, include the following: (1) Water quality. (2) Temperature. (3) Humidity. (4) Protection of equipment and instruments from fluctuations and interruptions in electrical current that adversely affect patient test results and test reports. This STANDARD is not met as evidenced by: Based on record review and staff interview, the laboratory failed to follow manufacturer instructions to monitor and document temperatures for proper testing performance of fungal cultures in the subspecialty of mycology. Findings include: 1. Surveyor observation on 3/29/2022 at 12:00 PM of the laboratory's inoculated fungal culture media revealed Remel Dermatube Dermatophyte Test Medium (DTM), Lot #: 367893 exp 9/22/2022 were in use and stored at room temperature. 2. Record review on 3/29/2022 of the Remel DTM product insert revealed to incubate DTM in ambient air at 25-30 degrees Celsius for up to 14 days. 3. Record review on 3/29/2022 of laboratory maintenance records revealed the lack of documentation of room temperature readings where inoculated DTM are stored during the 14 day testing period. 4. Staff interview with the laboratory director on 3/29/2022 at 12:10 PM confirmed the above findings. 5. The laboratory performs 277 fungal cultures annually in the subspecialty of mycology. -- 2 of 2 --

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access

Survey - July 24, 2018

Survey Type: Standard

Survey Event ID: NQZW11

Deficiency Tags: D6053

Summary:

Summary Statement of Deficiencies D6053 TECHNICAL CONSULTANT RESPONSIBILITIES CFR(s): 493.1413(b)(9) The technical consultant is responsible for evaluating and documenting the performance of individuals responsible for moderate complexity testing at least semiannually during the first year the individual tests patient specimens. This STANDARD is not met as evidenced by: Based on record review and staff interview, the laboratory failed to assess and document new testing personnel competency to perform fungal cultures in the subspecialty of mycology twice in the first year of performing patient testing. Findings include: 1. Record review of the CMS Laboratory Personnel Report 209 form on 7/24/18 revealed 1 new testing personnel was hired since the last survey. 2. Record review of the testing personnel competency records on 7/24/18 revealed the semiannual assessment for 1 of 1 new testing personnel was not available. 3. Record review of the laboratory's procedure manual on 7/24/18 revealed the lab did not have a policy in place to train and perform semiannual competency in the first year and annual thereafter for new testing personnel. 4. Staff interview with the laboratory director on 7/24/18 at 11:00 AM confirmed the above findings. Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 1 --

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access