Summary Statement of Deficiencies D2007 TESTING OF PROFICIENCY TESTING SAMPLES CFR(s): 493.801(b)(1) The samples must be examined or tested with the laboratory's regular patient workload by personnel who routinely perform the testing in the laboratory, using the laboratory's routine methods This STANDARD is not met as evidenced by: Based on surveyor review of the Proficiency Testing (PT), interview with Testing Personnel (TP), the laboratory failed to ensure that all TP who routinely perform Hematology testing participated in the American Proficiency Institute (API) PT in the calendar years 2019, 2020, 2021 and first and second events of 2022. The findings include: 1. A review of PT attestation records showed that two out of seven TP performed PT in the aforementioned time frame. 2. The TP confirmed on 9/13/22 at 2: 30 pm that PT events were not rotated between TP. D5403 PROCEDURE MANUAL CFR(s): 493.1251(b) The procedure manual must include the following when applicable to the test procedure: (1) Requirements for patient preparation; specimen collection, labeling, storage, preservation, transportation, processing, and referral; and criteria for specimen acceptability and rejection as described in 493.1242. (2) Microscopic examination, including the detection of inadequately prepared slides. (3) Step-by-step performance of the procedure, including test calculations and interpretation of results. (4) Preparation of slides, solutions, calibrators, controls, reagents, stains, and other materials used in testing. (5) Calibration and calibration verification procedures. (6) The reportable range for test results for the test system as established or verified in 493.1253. (7) Control procedures. (8)