CLIA Laboratory Citation Details
33D0926623
Survey Type: Special
Survey Event ID: X3GX11
Deficiency Tags: D0000 D2016 D2084 D2085 D6000 D6016 D0000 D2016 D2084 D2085 D6000 D6016
Summary Statement of Deficiencies D0000 Based on a proficiency testing (PT) desk review survey performed on July 10, 2024, the laboratory was found to be out of compliance based on the following CONDITION LEVEL DEFICIENCIES: D2016 - 42 C.F.R. 493.803 Condition: Successful participation. D6000 - 42 C.F.R. 493.1403 Condition: Laboratory Director, moderate complexity. D2016 SUCCESSFUL PARTICIPATION CFR(s): 493.803(a)(b)(c) (a) Each laboratory performing nonwaived testing must successfully participate in a proficiency testing program approved by CMS, if applicable, as described in subpart I of this part for each specialty, subspecialty, and analyte or test in which the laboratory is certified under CLIA. (b) Except as specified in paragraph (c) of this section, if a laboratory fails to participate successfully in proficiency testing for a given specialty, subspecialty, analyte or test, as defined in this section, or fails to take remedial action when an individual fails gynecologic cytology, CMS imposes sanctions, as specified in subpart R of this part. (c) If a laboratory fails to perform successfully in a CMS- approved proficiency testing program, for the initial unsuccessful performance, CMS may direct the laboratory to undertake training of its personnel or to obtain technical assistance, or both, rather than imposing alternative or principle sanctions except when one or more of the following conditions exists: (1) There is immediate jeopardy to patient health and safety. (2) The laboratory fails to provide CMS or a CMS agent with satisfactory evidence that it has taken steps to correct the problem identified by the unsuccessful proficiency testing performance. (3) The laboratory has a poor compliance history. This CONDITION is not met as evidenced by: Based on Centers for Medicare & Medicaid Services (CMS) Proficiency Testing (PT) Certification and Survey Provider Enhanced Reporting system (CASPER 0155D) and American Proficiency Institute (API) PT summary reports, the laboratory failed to Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 3 -- successfully participate in the CMS approved PT program for the General Immunology subspecialty and Alpha Fetoprotein (AFP) test analyte. Refer to D2084 and D2085. D2084 GENERAL IMMUNOLOGY CFR(s): 493.837(f) Failure to achieve satisfactory performance for the same analyte or test in two consecutive testing events or two out of three consecutive testing events is unsuccessful performance. This STANDARD is not met as evidenced by: Based on review of the CASPER 0155D and API PT summary reports from 2023 and 2024, the laboratory failed to achieve satisfactory performance (80% or greater) in four of five events for the General Immunology subspecialty in the AFP test analyte. FINDINGS: 1. A review of the CASPER 155 report revealed the following unsatisfactory scores: AFP Test Analyte: 2023 First Event = 60% 2023 Second Event = 0% 2023 Third Event = 0% 2024 Second Event = 20% 2. A review of the proficiency testing scores from API (2023 and 2024) confirmed the above findings. D2085 GENERAL IMMUNOLOGY CFR(s): 493.837(g) Failure to achieve an overall testing event score of satisfactory performance for two consecutive testing events or two out of three consecutive testing events is unsuccessful performance. This STANDARD is not met as evidenced by: Based on review of the CMS PT CASPER 0155D and API PT summary reports from 2023 and 2024, the laboratory failed to achieve satisfactory performance (80% or greater) in four of five events for the General Immunology subspecialty. FINDINGS: 1. A review of the CASPER 155 report revealed the following unsatisfactory scores: General Immunology Subspecialty: 2023 First Event = 60% 2023 Second Event = 0% 2023 Third Event = 0% 2024 Second Event = 20% 2. A review of the proficiency testing scores from API (2023 and 2024) confirmed the above findings. D6000 MODERATE COMPLEXITY LABORATORY DIRECTOR CFR(s): 493.1403 The laboratory must have a director who meets the qualification requirements of 493. 1405 of this subpart and provides overall management and direction in accordance with 493.1407 of this subpart. This CONDITION is not met as evidenced by: Based on review of CMS PT CASPER 0155D and API PT summary reports for 2023 and 2024, the laboratory director (LD) failed to provide overall management and direction of laboratory services. Refer to D6016. D6016 LABORATORY DIRECTOR RESPONSIBILITIES CFR(s): 493.1407(e)(4)(i) -- 2 of 3 -- The laboratory director is responsible for the overall operation and administration of the laboratory, including the employment of personnel who are competent to perform test procedures, and record and report test results promptly, accurate, and proficiently and for assuring compliance with the applicable regulations. (e) The laboratory director must-- (e)(4)(i) Ensure that the proficiency testing samples are tested as required under Subpart H of this part; This STANDARD is not met as evidenced by: Based on review of CMS PT CASPER 0155D and API PT summary reports for 2023 and 2024, the LD failed to ensure successful participation in an HHS approved proficiency testing program. Refer to D2084 and D2085. -- 3 of 3 --
Get full access to the detailed deficiency summary for this facility
Survey Type: Special
Survey Event ID: DL4G11
Deficiency Tags: D2016 D2084 D2085 D6000 D2084 D2085 D6000 D6016 D6016
Summary Statement of Deficiencies D2016 SUCCESSFUL PARTICIPATION CFR(s): 493.803(a)(b)(c) (a) Each laboratory performing nonwaived testing must successfully participate in a proficiency testing program approved by CMS, if applicable, as described in subpart I of this part for each specialty, subspecialty, and analyte or test in which the laboratory is certified under CLIA. (b) Except as specified in paragraph (c) of this section, if a laboratory fails to participate successfully in proficiency testing for a given specialty, subspecialty, analyte or test, as defined in this section, or fails to take remedial action when an individual fails gynecologic cytology, CMS imposes sanctions, as specified in subpart R of this part. (c) If a laboratory fails to perform successfully in a CMS- approved proficiency testing program, for the initial unsuccessful performance, CMS may direct the laboratory to undertake training of its personnel or to obtain technical assistance, or both, rather than imposing alternative or principle sanctions except when one or more of the following conditions exists: (1) There is immediate jeopardy to patient health and safety. (2) The laboratory fails to provide CMS or a CMS agent with satisfactory evidence that it has taken steps to correct the problem identified by the unsuccessful proficiency testing performance. (3) The laboratory has a poor compliance history. This CONDITION is not met as evidenced by: Based on Proficiency Test (PT) desk review of the Center for Medicaid & Medicare Services (CMS) PT data reports and PT summary reports from American Proficiency Institute (API) PT program, the laboratory failed to participate successfully in CMS approved proficiency testing program, for the speciality General Immunology and the test analyte Alpha Fetoprotein (AFP). The following scores were assigned: Speciality General Immunology and AFP 2023 first event = 60% 2023 second event = 0% This is considered unsuccessful PT performance. Refer to D2084 and D2085 D2084 GENERAL IMMUNOLOGY Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 3 -- CFR(s): 493.837(f) Failure to achieve satisfactory performance for the same analyte or test in two consecutive testing events or two out of three consecutive testing events is unsuccessful performance. This STANDARD is not met as evidenced by: Based on PT desk review of the CMS PT data reports and PT summary reports from API PT program, the laboratory failed to participate successfully in CMS approved proficiency testing program, for the test analyte AFP. The following scores were assigned: AFP 2023 first event = 60% 2023 second event = 0% This is considered unsuccessful PT performance. D2085 GENERAL IMMUNOLOGY CFR(s): 493.837(g) Failure to achieve an overall testing event score of satisfactory performance for two consecutive testing events or two out of three consecutive testing events is unsuccessful performance. This STANDARD is not met as evidenced by: Based on PT desk review of the CMS PT data reports and PT summary reports from API PT program, the laboratory failed to participate successfully in CMS approved proficiency testing program, for the speciality General Immunology. The following scores were assigned: Speciality General Immunology 2023 first event = 60% 2023 second event = 0% This is considered unsuccessful PT performance. D6000 MODERATE COMPLEXITY LABORATORY DIRECTOR CFR(s): 493.1403 The laboratory must have a director who meets the qualification requirements of 493. 1405 of this subpart and provides overall management and direction in accordance with 493.1407 of this subpart. This CONDITION is not met as evidenced by: Based on PT desk review of the CMS PT and API PT program records, the laboratory director failed to fulfill the laboratory director's responsibilities and ensure that the laboratory achieved a satisfactory performance and successfully participate in a PT program, approved by CMS, for the speciality for the speciality General Immunology and the test analyte AFP. Refer to D6016 D6016 LABORATORY DIRECTOR RESPONSIBILITIES CFR(s): 493.1407(e)(4)(i) The laboratory director is responsible for the overall operation and administration of the laboratory, including the employment of personnel who are competent to perform test procedures, and record and report test results promptly, accurate, and proficiently and for assuring compliance with the applicable regulations. (e) The laboratory director must-- (e)(4)(i) Ensure that the proficiency testing samples are tested as required under Subpart H of this part; -- 2 of 3 -- This STANDARD is not met as evidenced by: Based on PT desk review of the CMS PT and API PT program records, the laboratory director failed to fulfill the laboratory director's responsibilities and ensure that the laboratory achieved a satisfactory performance and successfully participate in a PT program, approved by CMS, for the speciality for the speciality General Immunology and the test analyte AFP. The following scores were assigned: Speciality General Immunology and AFP 2023 first event = 60% 2023 second event = 0% This is considered unsuccessful PT performance. -- 3 of 3 --
Get full access to the detailed deficiency summary for this facility
Survey Type: Special
Survey Event ID: 1WIY11
Deficiency Tags: D2016 D2084 D2085 D6000 D6016 D2016 D2084 D2085 D6000 D6016
Summary Statement of Deficiencies D2016 SUCCESSFUL PARTICIPATION CFR(s): 493.803(a)(b)(c) (a) Each laboratory performing nonwaived testing must successfully participate in a proficiency testing program approved by CMS, if applicable, as described in subpart I of this part for each specialty, subspecialty, and analyte or test in which the laboratory is certified under CLIA. (b) Except as specified in paragraph (c) of this section, if a laboratory fails to participate successfully in proficiency testing for a given specialty, subspecialty, analyte or test, as defined in this section, or fails to take remedial action when an individual fails gynecologic cytology, CMS imposes sanctions, as specified in subpart R of this part. (c) If a laboratory fails to perform successfully in a CMS- approved proficiency testing program, for the initial unsuccessful performance, CMS may direct the laboratory to undertake training of its personnel or to obtain technical assistance, or both, rather than imposing alternative or principle sanctions except when one or more of the following conditions exists: (1) There is immediate jeopardy to patient health and safety. (2) The laboratory fails to provide CMS or a CMS agent with satisfactory evidence that it has taken steps to correct the problem identified by the unsuccessful proficiency testing performance. (3) The laboratory has a poor compliance history. This CONDITION is not met as evidenced by: Based on a proficiency testing ( PT) desk review of the Center for Medicare and Medicaid Services (CMS) PT data reports and PT records from the American Proficiency Institute (API) PT program, the laboratory failed to participate and perform successfully in a PT program, for the specialty General Immunology and the test analyte Alpha-fetoprotein (AFP). The following scores were assigned: 2018 first event= 60% 2018 second event = 20% 2018 third event = 100% 2019 first event = 60% This is considered a repeatedly unsuccessful PT performance. Refer to D2084 and D2085. Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 3 -- D2084 GENERAL IMMUNOLOGY CFR(s): 493.837(f) Failure to achieve satisfactory performance for the same analyte or test in two consecutive testing events or two out of three consecutive testing events is unsuccessful performance. This STANDARD is not met as evidenced by: Based on a PT desk review of the CMS PT reports and PT records from the API PT program, the laboratory failed to participate and perform successfully in a PT program, approved by CMS, for the test analyte Alpha-fetoprotein (AFP). The following scores were assigned: 2018 first event= 60% 2018 second event = 20% 2018 third event = 100% 2019 first event = 60% This is considered a repeatedly unsuccessful PT performance. D2085 GENERAL IMMUNOLOGY CFR(s): 493.837(g) Failure to achieve an overall testing event score of satisfactory performance for two consecutive testing events or two out of three consecutive testing events is unsuccessful performance. This STANDARD is not met as evidenced by: Based on a PT desk review of the CMS PT reports and PT records from the API PT program, the laboratory failed to participate and perform successfully in a PT program, approved by CMS, for the specialty General Immunology. The following scores were assigned: for 2018 first event = 60% 2018 second event = 20% 2018 third event = 100% 2019 first event = 60% This is considered repeatedly unsuccessful PT performance. D6000 MODERATE COMPLEXITY LABORATORY DIRECTOR CFR(s): 493.1403 The laboratory must have a director who meets the qualification requirements of 493. 1405 of this subpart and provides overall management and direction in accordance with 493.1407 of this subpart. This CONDITION is not met as evidenced by: Based on PT desk review of the PT CMS data reports and API PT program records, the laboratory director failed to fulfill the laboratory director's responsibilities and ensure that the laboratory achieved a satisfactory performance and successfully participate in a PT program, for the speciality General Immunology and the test analyte Alpha-fetoprotein (AFP). Refer to D6016. D6016 LABORATORY DIRECTOR RESPONSIBILITIES CFR(s): 493.1407(e)(4)(i) The laboratory director is responsible for the overall operation and administration of the laboratory, including the employment of personnel who are competent to perform test procedures, and record and report test results promptly, accurate, and proficiently -- 2 of 3 -- and for assuring compliance with the applicable regulations. (e) The laboratory director must-- (e)(4)(i) Ensure that the proficiency testing samples are tested as required under Subpart H of this part; This STANDARD is not met as evidenced by: Based on PT desk review of the PT CMS data reports and API PT program records, the laboratory director failed to fulfill the laboratory director's responsibilities and ensure that the laboratory achieved a satisfactory performance and successfully participate in a PT program, for the speciality General Immunology and the test analyte Alpha-fetoprotein (AFP). The following scores were assigned: 2018 first event= 60% 2018 second event = 20% 2018 third event = 100% 2019 first event = 60% This is considered a repeatedly unsuccessful PT performance. -- 3 of 3 --
Get full access to the detailed deficiency summary for this facility
Survey Type: Special
Survey Event ID: XWU011
Deficiency Tags: D2016 D2084 D2085 D2098 D6000 D6016 D2016 D2084 D2085 D2098 D6000 D6016
Summary Statement of Deficiencies D2016 SUCCESSFUL PARTICIPATION CFR(s): 493.803(a)(b)(c) (a) Each laboratory performing nonwaived testing must successfully participate in a proficiency testing program approved by CMS, if applicable, as described in subpart I of this part for each specialty, subspecialty, and analyte or test in which the laboratory is certified under CLIA. (b) Except as specified in paragraph (c) of this section, if a laboratory fails to participate successfully in proficiency testing for a given specialty, subspecialty, analyte or test, as defined in this section, or fails to take remedial action when an individual fails gynecologic cytology, CMS imposes sanctions, as specified in subpart R of this part. (c) If a laboratory fails to perform successfully in a CMS- approved proficiency testing program, for the initial unsuccessful performance, CMS may direct the laboratory to undertake training of its personnel or to obtain technical assistance, or both, rather than imposing alternative or principle sanctions except when one or more of the following conditions exists: (1) There is immediate jeopardy to patient health and safety. (2) The laboratory fails to provide CMS or a CMS agent with satisfactory evidence that it has taken steps to correct the problem identified by the unsuccessful proficiency testing performance. (3) The laboratory has a poor compliance history. This CONDITION is not met as evidenced by: Based on a proficiency testing (PT) desk review of the Center for Medicare and Medicaid Services (CMS) PT reports and PT records from the American Proficiency Institute (API) PT program, the laboratory failed to participate and perform successfully in a PT program, approved by CMS, for the specialty General Immunology and test analyte Alpha-Feto Protein (AFP). The following scores were assigned: for General Immunology and AFP 2018 first event = 60% 2018 second event = 20% This is considered unsuccessful PT performance. Refer to D2084 and D2085. Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 3 -- D2084 GENERAL IMMUNOLOGY CFR(s): 493.837(f) Failure to achieve satisfactory performance for the same analyte or test in two consecutive testing events or two out of three consecutive testing events is unsuccessful performance. This STANDARD is not met as evidenced by: Based on a PT desk review of the CMS PT reports and PT records from the API PT program, the laboratory failed to participate and perform successfully in a PT program, approved by CMS, for the test analyte Alpha-Feto Protein (AFP). The following scores were assigned: for 2018 first event = 60% 2018 second event = 20% This is considered unsuccessful PT performance. D2085 GENERAL IMMUNOLOGY CFR(s): 493.837(g) Failure to achieve an overall testing event score of satisfactory performance for two consecutive testing events or two out of three consecutive testing events is unsuccessful performance. This STANDARD is not met as evidenced by: Based on a PT desk review of the CMS PT reports and PT records from the API PT program, the laboratory failed to participate and perform successfully in a PT program, approved by CMS, for the specialty General Immunology. The following scores were assigned: for 2018 first event = 60% 2018 second event = 20% This is considered unsuccessful PT performance. D2098 ENDOCRINOLOGY CFR(s): 493.843(a) Failure to attain a score of at least 80 percent of acceptable responses for each analyte in each testing event is unsatisfactory analyte performance for the testing event. This STANDARD is not met as evidenced by: Based on a PT desk review of the CMS PT reports and PT records from the API PT program, the laboratory failed to participate and perform successfully in a PT program, approved by CMS, for the test analyte Cortisol. The following scores were assigned: for 2018 second event = 40% This is considered unsatisfactory PT performance. D6000 MODERATE COMPLEXITY LABORATORY DIRECTOR CFR(s): 493.1403 The laboratory must have a director who meets the qualification requirements of 493. 1405 of this subpart and provides overall management and direction in accordance with 493.1407 of this subpart. This CONDITION is not met as evidenced by: -- 2 of 3 -- Based on PT desk review of the CMS PT and API PT program records, the laboratory director failed to fulfill the laboratory director's responsibilities and ensure that the laboratory achieved a satisfactory performance and successfully participate in a PT program, approved by CMS, for the speciality General Immunology and test analytes AFP and Cortisol. Refer to D6016. D6016 LABORATORY DIRECTOR RESPONSIBILITIES CFR(s): 493.1407(e)(4)(i) The laboratory director is responsible for the overall operation and administration of the laboratory, including the employment of personnel who are competent to perform test procedures, and record and report test results promptly, accurate, and proficiently and for assuring compliance with the applicable regulations. (e) The laboratory director must-- (e)(4)(i) Ensure that the proficiency testing samples are tested as required under Subpart H of this part; This STANDARD is not met as evidenced by: Based on PT desk review of the CMS PT and API PT program records, the laboratory director failed to fulfill the laboratory director's responsibilities and ensure that the laboratory achieved a satisfactory performance and successfully participate in a PT program, approved by CMS, for the speciality General Immunology and test analytes AFP and Cortisol. General Immunology and AFP 2018 first event = 60% 2018 second event = 20% This is considered unsuccessful PT performance. Cortisol The following scores were assigned: for 2018 second event = 40% This is considered unsatisfactory PT performance. -- 3 of 3 --