Allied Laboratories, Inc

CLIA Laboratory Citation Details

2
Total Citations
7
Total Deficiencyies
7
Unique D-Tags
CMS Certification Number 05D0605512
Address 453 Ravendale Dr Ste B, Mountain View, CA, 94043
City Mountain View
State CA
Zip Code94043
Phone(650) 335-8336

Citation History (2 surveys)

Survey - August 12, 2024

Survey Type: Special

Survey Event ID: MGHN11

Deficiency Tags: D0000 D2107 D6000 D2016 D2108 D6016

Summary:

Summary Statement of Deficiencies D0000 A proficiency testing desk review survey was performed on 8/9/2024, the laboratory was found not in compliance with the following CONDITION LEVEL DEFICIENCIES D2016 - 42 C.F.R. 493.803 Condition: Successful participation [proficiency testing]; D6000 - 42 C.F.R. 493.1403 Condition: Laboratories performing moderate complexity testing; laboratory director. D2016 SUCCESSFUL PARTICIPATION CFR(s): 493.803(a)(b)(c) (a) Each laboratory performing nonwaived testing must successfully participate in a proficiency testing program approved by CMS, if applicable, as described in subpart I of this part for each specialty, subspecialty, and analyte or test in which the laboratory is certified under CLIA. (b) Except as specified in paragraph (c) of this section, if a laboratory fails to participate successfully in proficiency testing for a given specialty, subspecialty, analyte or test, as defined in this section, or fails to take remedial action when an individual fails gynecologic cytology, CMS imposes sanctions, as specified in subpart R of this part. (c) If a laboratory fails to perform successfully in a CMS- approved proficiency testing program, for the initial unsuccessful performance, CMS may direct the laboratory to undertake training of its personnel or to obtain technical assistance, or both, rather than imposing alternative or principle sanctions except when one or more of the following conditions exists: (1) There is immediate jeopardy to patient health and safety. (2) The laboratory fails to provide CMS or a CMS agent with satisfactory evidence that it has taken steps to correct the problem identified by the unsuccessful proficiency testing performance. (3) The laboratory has a poor compliance history. This CONDITION is not met as evidenced by: Based on review of the Certification and Survey Provider Enhanced Reporting (CASPER) - 0155D and College of American Pathologist/EXCEL (CAP) records (2023-2, 2023-3, 2024-1 and 2024-2), the laboratory failed to successfully participate Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 3 -- in a proficiency testing program approved by HHS for each specialty, subspecialty and analyte or test in which the laboratory is certified under CLIA, the laboratory failed to successfully participate in the subspecialty of Endocrinology for the analyte Human Chorionic Gonadotropin (HCG) resulting in subsequent unsuccessful performance. Refer to D2107 & D2108. D2107 ENDOCRINOLOGY CFR(s): 493.843(f) Failure to achieve satisfactory performance for the same analyte or test in two consecutive testing events or two out of three consecutive testing events is unsuccessful performance. This STANDARD is not met as evidenced by: Based on desk review of Certification and Survey Provider Enhanced Reporting (CASPER) Report 0155D Individual Laboratory Profile and CAP evaluation reports, the laboratory failed to achieve satisfactory performance for four of four proficiency events in 2023 and 2024 for the subspecialty of Endocrinology. The finding include 1. The laboratory received the following scores: 0% on the 2023 overall subspecialty of Endocrinology - first event; 0% on the 2023 overall subspecialty of Endocrinology - second event; 0% on the 2023 overall subspecialty of Endocrinology - third event; and 0% on the 2024 overall subspecialty of Endocrinology - first event. 2. A review of the 2023 and 2024 proficiency testing scores from CAP confirmed the above findings. D2108 ENDOCRINOLOGY CFR(s): 493.843(g) Failure to achieve an overall testing event score of satisfactory performance for two consecutive testing events or two out of three consecutive testing events is unsuccessful performance. This STANDARD is not met as evidenced by: Based on desk review of Certification and Survey Provider Enhanced Reporting (CASPER) Report 0155D Individual Laboratory Profile and CAP evaluation reports, the laboratory failed to achieve satisfactory performance for four consecutive proficiency events in 2023 and 2024 for the analyte HCG. The finding include 1. The laboratory received the following scores: 0% for the analyte - HCG on the first event of 2023; 0% for the analyte - HCG on the second event of 2023; 0% for the analyte - HCG on the third event of 2023; and 0% for the analyte - HCG on the first event of 2024. 2. A review of the 2023 and 2024 proficiency testing scores from CAP confirmed the above findings. D6000 MODERATE COMPLEXITY LABORATORY DIRECTOR CFR(s): 493.1403 The laboratory must have a director who meets the qualification requirements of 493. 1405 of this subpart and provides overall management and direction in accordance with 493.1407 of this subpart. This CONDITION is not met as evidenced by: -- 2 of 3 -- Based on a proficiency testing desk review of the CASPER 155D report CAP records for 2023-1, 2023-2, 2023-3 and 2024-1 events, the laboratory director failed to provide overall management and direction of the laboratory services. Refer to D6016. D6016 LABORATORY DIRECTOR RESPONSIBILITIES CFR(s): 493.1407(e)(4)(i) The laboratory director is responsible for the overall operation and administration of the laboratory, including the employment of personnel who are competent to perform test procedures, and record and report test results promptly, accurate, and proficiently and for assuring compliance with the applicable regulations. (e) The laboratory director must-- (e)(4)(i) Ensure that the proficiency testing samples are tested as required under Subpart H of this part; This STANDARD is not met as evidenced by: Based on a proficiency testing desk review of CASPER 155 report CAP records for 2023-1, 2023-2, 2023-3 and 2024-1 events, the laboratory director failed to ensure successful participation in an HHS approved proficiency testing program. Refer to D2107 & D2108. -- 3 of 3 --

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access

Survey - January 30, 2019

Survey Type: Standard

Survey Event ID: 9WLC11

Deficiency Tags: D6128

Summary:

Summary Statement of Deficiencies D6128 TECHNICAL SUPERVISOR RESPONSIBILITIES CFR(s): 493.1451(b)(9) The technical supervisor is responsible for evaluating and documenting the performance of individuals responsible for high complexity testing at least annually after the first year, unless test methodology or instrumentation changes, in which case, prior to reporting patient test results, the individual's performance must be reevaluated to include the use of the new test methodology or instrumentation. This STANDARD is not met as evidenced by: Based on review of documentation for annual competency evaluation(s) for the testing personnel and interview with the laboratory manager and technical supervisor, it was determined that the technical supervisor failed to evaluate and document the performance of individuals responsible for high complexity testing at least annually after the first year unless test methodology or instrumentation changes, in which case, prior to reporting patient test results, the individual's performance must be reevaluated to include the use of the new test methodology or instrumentation. The findings included: a. No competency evaluation documentation could be retrieved for testing personnel performing high complexity testing in 2017 and 2018, (493-1495; Testing personnel responsibilities). b. The laboratory manger and technical supervisor confirmed on 01/30/2019 14:30 (survey date) that the laboratory had no documentation to show that the testing personnel annual competency and evaluation was performance. c. Based on the laboratory's annual testing declaration submitted for 2019, the laboratory analyzed and reported 1,345,136 patient test results. Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 1 --

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access