Arms Metro Health Center

CLIA Laboratory Citation Details

2
Total Citations
18
Total Deficiencyies
18
Unique D-Tags
CMS Certification Number 01D0858174
Address 712 25th Street North, Birmingham, AL, 35203
City Birmingham
State AL
Zip Code35203
Phone(205) 407-6900

Citation History (2 surveys)

Survey - July 12, 2023

Survey Type: Standard

Survey Event ID: BV2511

Deficiency Tags: D5445 D6047 D6056 D6057

Summary:

Summary Statement of Deficiencies D5445 CONTROL PROCEDURES CFR(s): 493.1256(d)(1)(2)(g) Unless CMS Approves a procedure, specified in Appendix C of the State Operations Manual (CMS Pub. 7), that provides equivalent quality testing, the laboratory must-- (d)(1) Perform control procedures as defined in this section unless otherwise specified in the additional specialty and subspecialty requirements at 493.1261 through 493.1278. (d)(2) For each test system, perform control procedures using the number and frequency specified by the manufacturer or established by the laboratory when they meet or exceed the requirements in paragraph (d)(3) of this section. (g) The laboratory must document all control procedures performed. This STANDARD is not met as evidenced by: Based on a review of Abbott iStat Blood Gas Quality Control (QC) records, a review of the Individualized Quality Control Plan (IQCP) documents, a review of Patient Log records, and an interview with Testing Personnel #1, the laboratory failed to follow the QC requirements specified by the laboratory's IQCP. This was noted one out of 22 months reviewed. The findings include: 1. A review of the iStat Blood Gas QC records revealed two levels of quality control were run on the following days: 11/1 /2022, 12/1/2022, and 2/17/2023. In January 2023 only the Level 2 control was run on 1/3/2023 with a notation, "No TriControl - Level 1 control on order". 2. A review of the iStat Blood Gas IQCP revealed at least two levels of the Tri-control should be performed each month. 3. A review of the January 2023 Patient Log revealed 14 patient tests were run. 4. During an interview on 7/12/2023 at 1:30 PM, Testing Personnel #1 confirmed the above findings. D6047 TECHNICAL CONSULTANT RESPONSIBILITIES CFR(s): 493.1413(b(8)(i) The procedures for evaluation of the competency of the staff must include, but are not Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 3 -- limited to direct observations of routine patient test performance, including patient preparation, if applicable, specimen handling, processing and testing. This STANDARD is not met as evidenced by: Based on a review of Personnel records, an interview with Testing Personnel #1, and an interview with the Laboratory Director (also the Technical Consultant), the Technical Consultant failed to assess competency of Testing Personnel by direct observation of routine patient testing. This was noted for two out of two competency assessments reviewed since the date of the last survey, 8/17/2021, to the date of the current survey, 7/12/2023. The findings include: 1. A review of Personnel records revealed a 6 month competency dated 8/2022 and an annual competency dated 10/11 /2022 for Testing Personnel #1. The forms included the six required elements of competency assessment, however, the forms were blank except for a signature from the Technical Consultant. 2. During an interview on 7/12/2023 at 2:00 PM, the Surveyor asked how competency assessments were performed by the Technical Consultant; Testing Personnel #1 explained she discussed how she performed patient testing with the Technical Consultant. Testing Personnel #1 confirmed her competency assessment had not included direct observations of routine patient test performance by the Technical Consultant. 3. During an interview on 7/27/2023 at 1:45 PM, the Laboratory Director/ Technical Consultant confirmed Testing Personnel #1 was not directly observed during assessment of competency. D6056 CLINICAL CONSULTANT CFR(s): 493.1415 The laboratory must have a clinical consultant who meets the qualification requirements of 493.1417 of this part and provides clinical consultation in accordance with 493.1419 of this part. This CONDITION is not met as evidenced by: Based on a review of Personnel records and interviews with the Laboratory Director and Testing Personnel #1, the laboratory failed to fill the position of Clinical Consultant (CC) with a qualified invidual, as required by CLIA for Moderate Complexity laboratories. This was noted from 2/20/2023 to the date of the current survey, 7/12/2023. The findings include: 1. Refer to D6057. D6057 CLINICAL CONSULTANT QUALIFICATIONS CFR(s): 493.1417 The clinical consultant must be qualified to consult with and render opinions to the laboratory's clients concerning the diagnosis, treatment and management of patient care. The clinical consultant must-- (a) Be qualified as a laboratory director under 493. 1405(b)(1), (2), or (3)(i); or (b) Be a doctor of medicine, doctor of osteopathy or doctor of podiatric medicine and possess a license to practice medicine, osteopathy or podiatry in the State in which the laboratory is located. This STANDARD is not met as evidenced by: Based on a review of Personnel records and interviews with the Laboratory Director and Testing Personnel #1, the laboratory failed to fill the position of Clinical Consultant (CC) with a qualified individual. This was noted from 2/20/2023 to the -- 2 of 3 -- date of the current survey, 7/12/2023. The findings include: 1. A review of Personnel records revealed educational documents for the Laboratory Director (who also acted as Technical Consultant) and Testing Personnel #1. There were no records available for any additional laboratory personnel on the day of the survey. 2. During an interview with Testing Personnel #1 on 7/12/2023 at 2:00 PM, the Surveyor asked who had assumed the position of Clinical Consultant, after the previous CC left the company. Testing Personnel #1 was unsure of who the current CC was, however, indicated it would be either the Laboratory Director or a Certified Registered Nurse Practioner (CRNP) on staff. The Laboratory Director qualified under 493.1405(b)(5), which does not meet the qualifications required to fill the position as the CC. 3. During an interview on 7/27/2023 at 1:45 PM, The Laboratory Director stated the following: "..I was unaware that they had not hired another doctor. Please let me know if the CRNP qualifies for Clinical Consultant". 4. On 7/28/2023 the Laboratory Personnel Report (Form CMS-209) was forwarded to the surveyor with the name of an individual who is a CRNP. The individual listed did not meet educational and licensing requirements for the Clinical Consultant position. -- 3 of 3 --

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access

Survey - August 17, 2021

Survey Type: Standard

Survey Event ID: SLW611

Deficiency Tags: D2000 D5217 D5400 D5401 D5407 D5413 D5431 D5447 D6000 D6013 D6015 D6020 D6033 D6034

Summary:

Summary Statement of Deficiencies D2000 ENROLLMENT AND TESTING OF SAMPLES CFR(s): 493.801 Each laboratory must enroll in a proficiency testing (PT) program that meets the criteria in subpart I of this part and is approved by HHS. The laboratory must enroll in an approved program or programs for each of the specialties and subspecialties for which it seeks certification. The laboratory must test the samples in the same manner as patients' specimens. For laboratories subject to 42 CFR part 493 published on March 14, 1990 (55 FR 9538) prior to September 1, 1992, the rules of this subpart are effective on September 1, 1992. For all other laboratories, the rules of this subpart are effective January 1, 1994. This CONDITION is not met as evidenced by: Based on a review of CMS-116 Clinical Laboratory Improvement Amendments (CLIA) Application for Certification and an interview with the Testing Personnel #3, the laboratory failed to enroll in a proficiency testing (PT) program for Routine Chemistry - Blood Gases (this is a moderate complexity, regulated Chemistry test). The laboratory was previously performing PPMP and waived testing before adding routine chemistry (iSTAT blood gases) in March 2021 (patient testing started in March 2021). The findings include: 1. A review of CMS-116 Clinical Laboratory Improvement Amendments (CLIA) Application for Certification revealed that the laboratory is performing Blood Gases on the iSTAT. 2. During an interview on 08/17 /2021 at 11:45 AM, Testing Personnel #3 confirmed the laboratory is not performing Proficiency Testing and they have not ordered PT for the iSTAT Blood Gases. D5217 EVALUATION OF PROFICIENCY TESTING PERFORMANCE CFR(s): 493.1236(c)(1) At least twice annually, the laboratory must verify the accuracy of any test or procedure it performs that is not included in subpart I of this part. Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 6 -- This STANDARD is not met as evidenced by: Based on a lack of Wet Prep records and an interview with the Chief Operating Officer, the laboratory failed to verify the accuracy of Vaginal Wet Mount at least twice annually. This was noted from 12/01/2020 to date of survey (08/17/2021). The findings include: 1. A review of the Vaginal Wet Mount records revealed a lack of accuracy verification being performed at least twice annually. 2. During an interview on 08/17/2021 at 12:05 PM, the Chief Operating Officer confirmed accuracy was not verified for Vaginal Wet Mount performed on the Microscope. D5400 ANALYTIC SYSTEMS CFR(s): 493.1250 Each laboratory that performs nonwaived testing must meet the applicable analytic systems requirements in 493.1251 through 493.1283, unless HHS approves a procedure, specified in Appendix C of the State Operations Manual (CMS Pub.7), that provides equivalent quality testing. The laboratory must monitor and evaluate the overall quality of the analytic systems and correct identified problems as specified in 493.1289 for each specialty and subspecialty of testing performed. This CONDITION is not met as evidenced by: Based on a review of i-STAT records and an interview with Testing Personnel #3, the surveyor determined the laboratory failed to monitor and evaluate the overall quality of i-STAT Blood Gases. This was noted from March 2021 to date of survey (08/17 /2021). The findings include: 1. Refer to D5407. 2. Refer to D5413. 3. Refer to D5431. D5401 PROCEDURE MANUAL CFR(s): 493.1251(a) A written procedures manual for all tests, assays, and examinations performed by the laboratory must be available to, and followed by, laboratory personnel. Textbooks may supplement but not replace the laboratory's written procedures for testing or examining specimens. This STANDARD is not met as evidenced by: Based on a review of the Procedure Manual and an interview with the Health Informatics and Quality Specialist, the laboratory failed to have a written procedure for Vaginal Wet Mount performed as Provider-Performed Microscopy (PPM). The findings include: 1. A review of the Procedure Manual revealed a lack of written procedures for Vaginal Wet Mount. 2. During an interview on 08/17/2021 at 12:05 PM, the Health Informatics and Quality Specialist confirmed the laboratory did not have a procedure for the test above. D5407 PROCEDURE MANUAL CFR(s): 493.1251(d) Procedures and changes in procedures must be approved, signed, and dated by the current laboratory director before use. -- 2 of 6 -- This STANDARD is not met as evidenced by: Based on a review of the Procedure Manual and an interview with Testing Personnel #3, the Laboratory Director failed to approve the Procedure being used in the Laboratory by signature. The findings include: 1. A review of the Procedure Manual for the iSTAT System revealed the Laboratory Director did not approve the procedure by signature. 2. During an interview on 08/17/2021 at 12:15 PM, Testing Personnel #3 confirmed the Procedure reviewed during the survey was not signed by the Laboratory Director. D5413 TEST SYSTEMS, EQUIPMENT, INSTRUMENTS, REAGENT CFR(s): 493.1252(b) The laboratory must define criteria for those conditions that are essential for proper storage of reagents and specimens, accurate and reliable test system operation, and test result reporting. The criteria must be consistent with the manufacturer's instructions, if provided. These conditions must be monitored and documented and, if applicable, include the following: (1) Water quality. (2) Temperature. (3) Humidity. (4) Protection of equipment and instruments from fluctuations and interruptions in electrical current that adversely affect patient test results and test reports. This STANDARD is not met as evidenced by: Based on a review of the Procedure Manual for the i-STAT System, an observation by the surveyor during the walk through of laboratory areas, and an interview with Testing Personnel #3, the laboratory failed to monitor and document room temperature for the room where the i-STAT and cartridges are located. Also, the laboratory kept the cartridges out at room temperature longer than the manufacturer's recommendation. The findings include: 1. A review of the Procedure Manual for the i- STAT System revealed on page 1 "...Cartridges may be stored at room temperature (18 to 30 degrees Celsius or 64 to 86 degrees Fahrenheit) for 14 days. Cartridges should not be returned to the refrigerator once they have been at room temperature, and should not be exposed to temperatures above 30 degrees Celsius (86 degrees Fahrenheit)..." 2. During walk through of the room where the i-STAT is located it was observed by the surveyor there was one box of cartridges on the counter by the i- STAT. 3. During an interview conducted on 08/17/2021 at 11:35 AM, Testing Personnel #3 confirmed that the laboratory had not monitored room temperature in the room where the i-STAT and cartridges are kept. Also, Testing Personnel #3 stated the cartridges are kept at room temperature for two months. D5431 MAINTENANCE AND FUNCTION CHECKS CFR(s): 493.1254(a)(2) For unmodified manufacturer's equipment, instruments, or test systems, the laboratory must perform and document function checks as defined by the manufacturer and with at least the frequency specified by the manufacturer. Function checks must be within the manufacturer's established limits before patient testing is conducted. This STANDARD is not met as evidenced by: Based on a review of the iSTAT Electronic Simulator records, a review of the Procedure Manual for the iSTAT System, and an interview with Testing Personnel #3, the laboratory failed to perform and document function checks (Electronic Simulator) -- 3 of 6 -- every 8 hours. This was noted from March 2021 (when patient testing started) to August 2021 (date of Survey). The findings include: 1. A review of the Electronic Simulator Log revealed it was performed the following days and passed: 03/09/2021, 03/10/2021, 05/18/2021, 06/17/202, 06/22/2021, 07/07/2021, 07/13/2021, 07/20/2021, 07/27/2021, 08/03/202, and 08/10/2021. 2. A review of the Procedure Manual for the iSTAT System revealed on page 19 "...Verify the performance of each handheld analyzer or Blood Analysis Module in the i-STAT System using the internal or external Electronic Simulator every 24 hours of use, or as needed for regulatory compliance. In the USA, verification is required every 8 hours for blood gases, ..." 3. During an interview on 08/17/2021 at 11:15 AM, Testing Personnel #3 confirmed the Electronic Simulator was being run each week with Quality Control and was not aware it needed to be run every 8 hours with patient testing. D5447 CONTROL PROCEDURES CFR(s): 493.1256(d)(3)(i)(g) Unless CMS Approves a procedure, specified in Appendix C of the State Operations Manual (CMS Pub. 7), that provides equivalent quality testing, the laboratory must-- At least once a day patient specimens are assayed or examined perform the following for-- Each quantitative procedure, include two control materials of different concentrations; (g) The laboratory must document all control procedures performed. This STANDARD is not met as evidenced by: Based on a review of the iSTAT Blood Gases quality control and an interview with the Testing Personnel #3, the laboratory failed run controls at least once a day patient specimens are performed or implement an Individualized Quality Control Plan (IQCP) for iSTAT Blood Gases. This was noted from March 2021 when patient testing started to August 2021 (day of survey). The findings include: 1. A review of the quality control for iSTAT Blood Gases revealed quality control was being performed once a week: 03/29/2021, 04/07/2021, 04/09/2021, 04/12/2021, 04/20/2021, 04/27/2021, 05 /04/2021, 05/10/2021, 05/19/2021, 05/25/2021, 06/01/2021, 06/08/2021, 06/17/2021, 06/22/2021, 06/29/2021, 07/07/2021, 07/13/2021, 07/20/2021, 07/27/2021, 08/03 /2021, and 08/10/2021. 2. During an interview on 08/17/2021 at 11:15 AM, Testing Personnel #3 confirmed the Laboratory was performing Quality Control weekly and an IQCP had not been written for iSTAT Blood Gases. Also, Testing Personnel #3 confirmed patient testing is usually performed 5 days a week. D6000 MODERATE COMPLEXITY LABORATORY DIRECTOR CFR(s): 493.1403 The laboratory must have a director who meets the qualification requirements of 493. 1405 of this subpart and provides overall management and direction in accordance with 493.1407 of this subpart. This CONDITION is not met as evidenced by: Based on review of validation records for the i-STAT Blood Gases, lack of blood gases Proficiency Testing records, quality control records for the i-STAT Blood Gases, and an interview with Testing Personnel #3, the Laboratory Director failed to fulfill the Laboratory Director's responsibilities. This was noted from March 2021 to the survey date (08/17/2021). The findings include: 1. Refer to D6013. 2. Refer to D6015. 3. Refer to D6020. -- 4 of 6 -- D6013 LABORATORY DIRECTOR RESPONSIBILITIES CFR(s): 493.1407(e)(3)(ii) The laboratory director is responsible for the overall operation and administration of the laboratory, including the employment of personnel who are competent to perform test procedures, and record and report test results promptly, accurate, and proficiently and for assuring compliance with the applicable regulations. (e) The laboratory director must-- (e)(3) Ensure that-- (e)(3)(ii) Verification procedures used are adequate to determine the accuracy, precision, and other pertinent performance characteristics of the method; This STANDARD is not met as evidenced by: Based on a review of the validation records for the iSTAT Blood Gases and interview with Testing Personnel (TP) #3, the surveyor determined the Laboratory Director failed to ensure the validation was reviewed and approved before patient testing started. This was noted for one of one new instruments installed (patient testing started in March of 2021). The findings include: 1. A review of the validation records for the iSTAT Blood Gases revealed the validation was performed March 9-10, 2021 and no indication of a signature of the Laboratory Director. 2. During an interview on 08/17/2021 at 12:15 PM, TP #3 confirmed patient testing started in March of 2021 and the Laboratory Director did not sign the validation as indication of review and approval. D6015 LABORATORY DIRECTOR RESPONSIBILITIES CFR(s): 493.1407(e)(4) The laboratory director is responsible for the overall operation and administration of the laboratory, including the employment of personnel who are competent to perform test procedures, and record and report test results promptly, accurate, and proficiently and for assuring compliance with the applicable regulations. (e) The laboratory director must-- (e)(4) Ensure that the laboratory is enrolled in an HHS approved proficiency testing program for the testing performed. This STANDARD is not met as evidenced by: Based on a lack of Proficiency Testing records and an interview with Testing Personnel #3, the Laboratory Director failed to ensure the laboratory was enrolled in an approved proficiency testing program for Blood Gases. This noted from March 2021 to the survey date (08/17/2021). The findings include: 1. Refer to D2000. D6020 LABORATORY DIRECTOR RESPONSIBILITIES CFR(s): 493.1407(e)(5) The laboratory director is responsible for the overall operation and administration of the laboratory, including the employment of personnel who are competent to perform test procedures, and record and report test results promptly, accurate, and proficiently and for assuring compliance with the applicable regulations. (e) The laboratory director must-- (e)(5) Ensure that the quality control program is established and maintained to assure the quality of laboratory services provided. This STANDARD is not met as evidenced by: -- 5 of 6 -- Based on the i-STAT Blood Gases Quality Control and an interview with Testing Personnel #3, the Laboratory Director failed to ensure at least 2 levels of Quality Control were performed each day of patient testing or implement an IQCP (Individualized Quality Control Plan). This noted from March 2021 to the survey date (08/17/2021). The findings include: 1. Refer to D5447. D6033 TECHNICAL CONSULTANT-MODERATE COMPEXITY CFR(s): 493.1409 The laboratory must have a technical consultant who meets the qualification requirements of 493.1411 of this subpart and provides technical oversight in accordance with 493.1413 of this subpart. This CONDITION is not met as evidenced by: Based on a review of CMS-209 Laboratory Personnel Report and an interview with the Informatics and Quality Specialist, the laboratory failed to have a technical consultant who meets the qualification requirements. The laboratory was previously performing PPMP and waived testing before adding routine chemistry (iSTAT blood gases) in March 2021. The findings include: 1. Refer to D6034. D6034 TECHNICAL CONSULTANT QUALIFICATIONS CFR(s): 493.1411 The laboratory must employ one or more individuals who are qualified by education and either training or experience to provide technical consultation for each of the specialties and subspecialties of service in which the laboratory performs moderate complexity tests or procedures. The director of a laboratory performing moderate complexity testing may function as the technical consultant provided he or she meets the qualifications specified in this section. This STANDARD is not met as evidenced by: Based on a review of CMS-209 Laboratory Personnel Report and an interview with the Informatics and Quality Specialist, the laboratory failed to have a technical consultant who meets the qualification requirements. The laboratory was previously performing PPMP and waived testing before adding routine chemistry (iSTAT blood gases) in March 2021 (patient testing started in March 2021). The findings include: 1. A review of CMS-209 Laboratory Personnel Report revealed that the laboratory failed to fill the technical consultant position. 2. During an interview on 08/17/2021 at 10:30 AM, the Informatics and Quality Specialist confirmed the current Laboratory Director does not have laboratory experience to qualify as the Technical Consultant and they have not filled the position with a qualified individual. -- 6 of 6 --

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access