CLIA Laboratory Citation Details
33D1043251
Survey Type: Special
Survey Event ID: 5DEZ11
Deficiency Tags: D0000 D2016 D2084 D2107 D2118 D6000 D6016 D0000 D2016 D2084 D2107 D2118 D6000 D6016
Summary Statement of Deficiencies D0000 Based on a proficiency testing (PT) desk review survey performed on January 16, 2025, the laboratory was found to be out of compliance based on the following CONDITION LEVEL DEFICIENCIES: D2016 - 42 C.F.R. 493.803 Condition: Successful participation. D6000 - 42 C.F.R. 493.1403 Condition: Laboratory Director, moderate complexity. D2016 SUCCESSFUL PARTICIPATION CFR(s): 493.803(a)(b)(c) (a) Each laboratory performing nonwaived testing must successfully participate in a proficiency testing program approved by CMS, if applicable, as described in subpart I of this part for each specialty, subspecialty, and analyte or test in which the laboratory is certified under CLIA. (b) Except as specified in paragraph (c) of this section, if a laboratory fails to participate successfully in proficiency testing for a given specialty, subspecialty, analyte or test, as defined in this section, or fails to take remedial action when an individual fails gynecologic cytology, CMS imposes sanctions, as specified in subpart R of this part. (c) If a laboratory fails to perform successfully in a CMS- approved proficiency testing program, for the initial unsuccessful performance, CMS may direct the laboratory to undertake training of its personnel or to obtain technical assistance, or both, rather than imposing alternative or principle sanctions except when one or more of the following conditions exists: (1) There is immediate jeopardy to patient health and safety. (2) The laboratory fails to provide CMS or a CMS agent with satisfactory evidence that it has taken steps to correct the problem identified by the unsuccessful proficiency testing performance. (3) The laboratory has a poor compliance history. This CONDITION is not met as evidenced by: Based on review of Centers for Medicare & Medicaid Services (CMS) Proficiency Testing (PT) Certification and Survey Provider Enhanced Reporting system (CASPER 0155D) and College of American Pathologists (CAP) PT summary reports, Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 3 -- the laboratory failed to successfully participate in the CMS approved PT program for consecutive testing events in the General Immunology specialty for the Alpha- fetoprotein (AFP) (Tumor Marker) test analyte in 2022, 2023, and 2024 and Immunoglobulin E (IgE) test analyte in 2024; Endocrinology subspecialty for the Cortisol, Human Chorionic Gonadotropin (hCG), and T3 Uptake test analytes in 2024; and the Toxicology subspecialty for the Lithium test analyte in 2023 and 2024 resulting in unsuccessful performance. Refer to D2084, D2107, and D2118. D2084 GENERAL IMMUNOLOGY CFR(s): 493.837(f) (f) Failure to achieve satisfactory performance for the same analyte or test in two consecutive testing events or two out of three consecutive testing events is unsuccessful performance. This STANDARD is not met as evidenced by: Based on CMS PT CASPER 0155D and CAP PT summary reports from 2022, 2023, and 2024, the laboratory failed to achieve satisfactory performance (80% or greater) for consecutive testing events in the General Immunology specialty for the AFP (Tumor Marker) and IgE test analytes. FINDINGS: 1. A review of the CASPER 155 report revealed the following unsatisfactory scores: General Immunology Specialty: 2022 Third Event = 60% 2023 First Event = 50% 2023 Third Event = 50% 2024 First Event = 0% 2024 Second Event = 0% AFP (Tumor Marker) Test Analyte: 2022 Third Event = 0% 2023 First Event = 0% 2023 Third Event = 0% 2024 First Event = 0% 2024 Second Event = 0% IgE Test Analyte: 2024 First Event = 0% 2024 Second Event = 0% 2. A review of the proficiency testing scores from CAP (2022, 2023 and 2024) confirmed the above findings. 3. It was noted that the laboratory claimed to discontinue AFP (Tumor Marker) and IgE analyte testing in response to the PT desk review performed October 21, 2024. D2107 ENDOCRINOLOGY CFR(s): 493.843(f) (f) Failure to achieve satisfactory performance for the same analyte or test in two consecutive testing events or two out of three consecutive testing events is unsuccessful performance. This STANDARD is not met as evidenced by: Based on CMS PT CASPER 0155D and CAP PT summary reports from 2024, the laboratory failed to achieve satisfactory performance (80% or greater) for consecutive testing events in the Endocrinology subspecialty for the Cortisol, hCG, and T3 Uptake test analytes. FINDINGS: 1. A review of the CASPER 155 report revealed the following unsatisfactory scores: Cortisol Test Analyte: 2024 First Event = 0% 2024 Second Event = 0% hCG Test Analyte: 2024 First Event = 0% 2024 Second Event = 0% T3 Uptake Test Analyte: 2024 First Event = 0% 2024 Second Event = 0% 2024 Third Event = 0% 2. A review of the proficiency testing scores from CAP (2024) confirmed the above findings. 3. It was noted that the laboratory claimed to discontinue Cortisol and hCG analyte testing in response to the PT desk review performed October 21, 2024. D2118 TOXICOLOGY -- 2 of 3 -- CFR(s): 493.845(f) (f) Failure to achieve satisfactory performance for the same analyte or test in two consecutive testing events or two out of three consecutive testing events is unsuccessful performance. This STANDARD is not met as evidenced by: Based on CMS PT CASPER 0155D and CAP PT summary reports from 2023 and 2024, the laboratory failed to achieve satisfactory performance (80% or greater) for consecutive testing events in the Toxicology subspecialty for the Lithium test analyte. FINDINGS: 1. A review of the CASPER 155 report revealed the following unsatisfactory scores: Toxicology Subspecialty: 2023 Third Event = 0% 2024 First Event = 0% 2024 Second Event = 0% Lithium Test Analyte: 2023 Third Event = 0% 2024 First Event = 0% 2024 Second Event = 0% 2. A review of the proficiency testing scores from CAP (2023 and 2024) confirmed the above findings. 3. It was noted that the laboratory claimed to discontinue Lithium analyte testing in response to the PT desk review performed October 21, 2024. D6000 MODERATE COMPLEXITY LABORATORY DIRECTOR CFR(s): 493.1403 The laboratory must have a director who meets the qualification requirements of 493. 1405 of this subpart and provides overall management and direction in accordance with 493.1407 of this subpart. This CONDITION is not met as evidenced by: Based on review of CMS PT CASPER 0155D and CAP PT summary reports from 2022, 2023, and 2024, the laboratory director (LD) failed to provide overall management and direction of the laboratory services. Refer to D6016. D6016 LABORATORY DIRECTOR RESPONSIBILITIES CFR(s): 493.1407(e)(4)(i) (e)(4)(i) The proficiency testing samples are tested as required under Subpart H of this part; This STANDARD is not met as evidenced by: Based on review of CMS PT CASPER 0155D and CAP PT 2022-3, 2023-1, 2023-3, 2024-1, and 2024-2 summary reports, the LD failed to ensure successful participation in an HHS approved proficiency testing program. Refer to D2084, D2107, and 2118. -- 3 of 3 --
Get full access to the detailed deficiency summary for this facility
Survey Type: Special
Survey Event ID: L8UJ11
Deficiency Tags: D2016 D2107 D6000 D6016 D6016 D0000 D2016 D2107 D6000
Summary Statement of Deficiencies D0000 Based on a proficiency testing (PT) desk review survey performed on October 21, 2024, the laboratory was found to be out of compliance based on the following CONDITION LEVEL DEFICIENCIES: D2016 - 42 C.F.R. 493.803 Condition: Successful participation. D6000 - 42 C.F.R. 493.1403 Condition: Laboratory Director, moderate complexity. D2016 SUCCESSFUL PARTICIPATION CFR(s): 493.803(a)(b)(c) (a) Each laboratory performing nonwaived testing must successfully participate in a proficiency testing program approved by CMS, if applicable, as described in subpart I of this part for each specialty, subspecialty, and analyte or test in which the laboratory is certified under CLIA. (b) Except as specified in paragraph (c) of this section, if a laboratory fails to participate successfully in proficiency testing for a given specialty, subspecialty, analyte or test, as defined in this section, or fails to take remedial action when an individual fails gynecologic cytology, CMS imposes sanctions, as specified in subpart R of this part. (c) If a laboratory fails to perform successfully in a CMS- approved proficiency testing program, for the initial unsuccessful performance, CMS may direct the laboratory to undertake training of its personnel or to obtain technical assistance, or both, rather than imposing alternative or principle sanctions except when one or more of the following conditions exists: (1) There is immediate jeopardy to patient health and safety. (2) The laboratory fails to provide CMS or a CMS agent with satisfactory evidence that it has taken steps to correct the problem identified by the unsuccessful proficiency testing performance. (3) The laboratory has a poor compliance history. This CONDITION is not met as evidenced by: Based on review of Centers for Medicare & Medicaid Services (CMS) Proficiency Testing (PT) Certification and Survey Provider Enhanced Reporting system (CASPER 0155D) and College of American Pathologists (CAP) PT summary reports, Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 2 -- the laboratory failed to successfully participate in the CMS approved PT program for two of three consecutive testing events in the Endocrinology subspecialty for the TY (Thyroxine) test analyte in 2024, resulting in unsuccessful performance. Refer to D2107. D2107 ENDOCRINOLOGY CFR(s): 493.843(f) Failure to achieve satisfactory performance for the same analyte or test in two consecutive testing events or two out of three consecutive testing events is unsuccessful performance. This STANDARD is not met as evidenced by: Based on CMS PT CASPER 0155D and CAP PT summary reports from 2024, the laboratory failed to achieve satisfactory performance (80% or greater) for two consecutive testing events in the Endocrinology subspecialty for the TY test analyte. FINDINGS: 1. A review of the CASPER 155 report revealed the following unsatisfactory scores: Endocrinology Subspecialty: 2024 First Event = 42% 2024 Second Event = 40% TY Test Analyte: 2024 First Event = 0% 2024 Second Event = 0% 2. A review of the proficiency testing scores from CAP (2024) confirmed the above findings. D6000 MODERATE COMPLEXITY LABORATORY DIRECTOR CFR(s): 493.1403 The laboratory must have a director who meets the qualification requirements of 493. 1405 of this subpart and provides overall management and direction in accordance with 493.1407 of this subpart. This CONDITION is not met as evidenced by: Based on review of CMS PT CASPER 0155D and CAP PT summary reports from 2024, the laboratory director (LD) failed to provide overall management and direction of the laboratory services. Refer to D2016. D6016 LABORATORY DIRECTOR RESPONSIBILITIES CFR(s): 493.1407(e)(4)(i) The laboratory director is responsible for the overall operation and administration of the laboratory, including the employment of personnel who are competent to perform test procedures, and record and report test results promptly, accurate, and proficiently and for assuring compliance with the applicable regulations. (e) The laboratory director must-- (e)(4)(i) Ensure that the proficiency testing samples are tested as required under Subpart H of this part; This STANDARD is not met as evidenced by: Based on review of CMS PT CASPER 0155D and CAP PT 2024-1 and 2024-2 summary reports, the LD failed to ensure successful participation in an HHS approved proficiency testing program. Refer to D2107. -- 2 of 2 --
Get full access to the detailed deficiency summary for this facility
Survey Type: Special
Survey Event ID: I3PZ11
Deficiency Tags: D0000 D2016 D2096 D6000 D6016 D0000 D2016 D2096 D6000 D6016
Summary Statement of Deficiencies D0000 Based on a proficiency testing (PT) desk review survey performed on August 21, 2024, the laboratory was found to be out of compliance based on the following CONDITION LEVEL DEFICIENCIES: D2016 - 42 C.F.R. 493.803 Condition: Successful participation. D6000 - 42 C.F.R. 493.1403 Condition: Laboratory Director, moderate complexity. D2016 SUCCESSFUL PARTICIPATION CFR(s): 493.803(a)(b)(c) (a) Each laboratory performing nonwaived testing must successfully participate in a proficiency testing program approved by CMS, if applicable, as described in subpart I of this part for each specialty, subspecialty, and analyte or test in which the laboratory is certified under CLIA. (b) Except as specified in paragraph (c) of this section, if a laboratory fails to participate successfully in proficiency testing for a given specialty, subspecialty, analyte or test, as defined in this section, or fails to take remedial action when an individual fails gynecologic cytology, CMS imposes sanctions, as specified in subpart R of this part. (c) If a laboratory fails to perform successfully in a CMS- approved proficiency testing program, for the initial unsuccessful performance, CMS may direct the laboratory to undertake training of its personnel or to obtain technical assistance, or both, rather than imposing alternative or principle sanctions except when one or more of the following conditions exists: (1) There is immediate jeopardy to patient health and safety. (2) The laboratory fails to provide CMS or a CMS agent with satisfactory evidence that it has taken steps to correct the problem identified by the unsuccessful proficiency testing performance. (3) The laboratory has a poor compliance history. This CONDITION is not met as evidenced by: Based on review of Centers for Medicare & Medicaid Services (CMS) Proficiency Testing (PT) Certification and Survey Provider Enhanced Reporting system (CASPER 0155D) and College of American Pathologists (CAP) PT summary reports, Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 3 -- the laboratory failed to successfully participate in the CMS approved PT program for three consecutive testing events in the Routine Chemistry subspecialty for the Amylase test analyte as well as two of three consecutive testing events for the CK (Creatine Kinase), Total test analyte in 2023 and 2024, resulting in unsuccessful performance. Refer to D2096. D2096 ROUTINE CHEMISTRY CFR(s): 493.841(f) Failure to achieve satisfactory performance for the same analyte or test in two consecutive testing events or two out of three consecutive testing events is unsuccessful performance. This STANDARD is not met as evidenced by: Based on CMS PT CASPER 0155D and CAP PT summary reports from 2023 and 2024, the laboratory failed to achieve satisfactory performance (80% or greater) for three consecutive testing events in the Routine Chemistry subspecialty for the Amylase test analyte as well as two consecutive testing events in the Routine Chemistry subspecialty for the CK, Total test analyte. FINDINGS: 1. A review of the CASPER 155 report revealed the following unsatisfactory scores: Amylase Test Analyte: 2023 Third Event = 0% 2024 First Event = 0% 2024 Second Event = 0% CK, Total Test Analyte: 2024 First Event = 0% 2024 Second Event = 0% 2. A review of the proficiency testing scores from CAP (2023 and 2024) confirmed the above findings. D6000 MODERATE COMPLEXITY LABORATORY DIRECTOR CFR(s): 493.1403 The laboratory must have a director who meets the qualification requirements of 493. 1405 of this subpart and provides overall management and direction in accordance with 493.1407 of this subpart. This CONDITION is not met as evidenced by: Based on review of CMS PT CASPER 0155D and CAP PT summary reports from 2023 and 2024, the laboratory director (LD) failed to failed to provide overall management and direction of the laboratory services. Refer to D2016. D6016 LABORATORY DIRECTOR RESPONSIBILITIES CFR(s): 493.1407(e)(4)(i) The laboratory director is responsible for the overall operation and administration of the laboratory, including the employment of personnel who are competent to perform test procedures, and record and report test results promptly, accurate, and proficiently and for assuring compliance with the applicable regulations. (e) The laboratory director must-- (e)(4)(i) Ensure that the proficiency testing samples are tested as required under Subpart H of this part; This STANDARD is not met as evidenced by: -- 2 of 3 -- Based on review of CMS PT CASPER 0155D and CAP PT 2023-3, 2024-1, and 2024- 2 summary reports, the LD failed to ensure successful participation in an HHS approved proficiency testing program. Refer to D2096. -- 3 of 3 --
Get full access to the detailed deficiency summary for this facility
Survey Type: Standard
Survey Event ID: U6N811
Deficiency Tags: D5413
Summary Statement of Deficiencies D5413 TEST SYSTEMS, EQUIPMENT, INSTRUMENTS, REAGENT CFR(s): 493.1252(b) The laboratory must define criteria for those conditions that are essential for proper storage of reagents and specimens, accurate and reliable test system operation, and test result reporting. The criteria must be consistent with the manufacturer's instructions, if provided. These conditions must be monitored and documented and, if applicable, include the following: (1) Water quality. (2) Temperature. (3) Humidity. (4) Protection of equipment and instruments from fluctuations and interruptions in electrical current that adversely affect patient test results and test reports. This STANDARD is not met as evidenced by: Based on surveyor's review of laboratory's temperature policy, review of I and II refrigerator temperature logs and direct observation of I & II refrigerators and an interview with the technical consultant, the laboratory failed to take