Bergen Clinical Labs Inc

CLIA Laboratory Citation Details

3
Total Citations
55
Total Deficiencyies
24
Unique D-Tags
CMS Certification Number 31D0691343
Address 40 Engle St, Englewood, NJ, 07631
City Englewood
State NJ
Zip Code07631
Phone(201) 429-9733

Citation History (3 surveys)

Survey - September 9, 2025

Survey Type: Standard

Survey Event ID: KUCG11

Deficiency Tags: D5403 D5415 D5417 D5439 D5469 D5417 D5439 D5469 D6091 D6093 D6076 D6086 D6091 D5219 D5401 D5403 D5415 D5805 D5807 D6076 D6086 D5805 D5807 D6093

Summary:

Summary Statement of Deficiencies D5219 EVALUATION OF PROFICIENCY TESTING PERFORMANCE CFR(s): 493.1236(c)(2) (c)(2) Any test or procedure listed in subpart I of this part for which compatible proficiency testing samples are not offered by a CMS-approved proficiency testing program. This STANDARD is not met as evidenced by: Based on surveyor review of the Proficiency Testing (PT) records and interview with the Technical Supervisor (TS), the laboratory failed to verifiy the accuracy at least twice annually for Hematology testing performed on the Sysmex XP-300 analyzer in the calendar years 2023 and 2024. The findings include: 1. The laboratory did not verify accuracy of the following analytes. a) Mean Corpuscular Volume (MCV) b) Mean Corpuscular Hemoglobin (MCH) c) Mean Corpuscular Hemoglobin Concentration (MCHC) d) Red Cell Distribution Width - Standard Deviation (RDW- SD) e) Red Cell Distribution Width -Coefficient of Variations (RDW-CV) f) Mean Platelet Volume (MPV) 2. The TS confirmed on 9/9/25 at 11:00 am, the laboratory did not verify accuracy for the above mentioned analytes at least twice annually. D5401 PROCEDURE MANUAL CFR(s): 493.1251(a) (a) A written procedures manual for all tests, assays, and examinations performed by the laboratory must be available to, and followed by, laboratory personnel. Textbooks may supplement but not replace the laboratory's written procedures for testing or examining specimens. This STANDARD is not met as evidenced by: A) Based on surveyor review of the Procedure Manual, and interview with the Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 7 -- Technical Supervisor (TS), the laboratory failed to follow the "Periodic Procedures for Cartridges"from July 2020 to 9/9/25. The finding includes: 1. The PM stated "For acceptance of newly received cartridge lots, check the Temperature. If all windows are white or if only the A or B windows are blue or the 1 or 2 windows are red, then transit temperatures were satisfactory and the cartridges can be used" but there was no evidence the temperature was checked. 2. The TS confirmed on 9/9/25 at 11:00 am, the laboratory did not follow the procedure stated above. Note: This was previously cited 3/23/21. B) Based on surveyor review of the PM and interview with the TS, the laboratory failed to follow their procedures "Laboratory Quality Management Program" and "Procedures and responsibilities" from 3/23/21 to 9/9/25. The findings include: 1. The PM stated "Monthly: The laboratory performs a quality system audit on a monthly basis", "The monthly audit begins with a discussion of any problems raised by the laboratory personnel. It continues with a review of all the following documentation". a) Laboratory environmental logs. b) Refrigerator and freezer temperature logs. c) Instrument maintenance logs. d) Remedial and alert logs. e) Levey-Jennings quality control plots. 2. There was no evidence that remedial and alert logs where maintained for all specialties. 3. There was no evidence that Levey- Jennings and quality control plots were maintained for all specialites. 4. There was no documented evidence that the above mentioned procedure was performed. 5. The TS confirmed on 9/9/25 at 11:00 am, the laboratory did not follow the procedure stated above. C) Based on surveyor review of the PM and interview with the TS, the laboratory failed to follow their procedure "internal Quality Control" from 3/23/21 to 9 /9/25. The findings includes: 1. The PM stated "QC data are reviewed regularly by the supervisors and printed monthly for review by the laboratory director (LD)" 2. There was no documented evidence the LD performed a monthly review of QC for any testing that the laboratory performed. 3. The TS confirmed on 9/9/25 at 11:00 am, the laboratory did not follow the procedure stated above. 48354 D) Based on surveyor review of the PM, Quality Control (QC) and interview with the TS, the laboratory failed to follow their procedure "Starting a New Lot of Controls" for Hematology tests performed on the Sysmex XP 300 analyzer from 4/6/23 to 9/9/25. The findings include: 1. The PM stated "Parallel test new lots of control by analyzing the three levels of control a minimum of twice a day for 5 days prior to expiration of the previous lot." 2. Surveyor review of QC reports revealed the laboratory did not parallel tests new lots of QC twice a day for five days. 3. The TS confirmed on 9/9/25 at 1:15pm, the laboratory did not follow the PM. D5403 PROCEDURE MANUAL CFR(s): 493.1251(b) (b) The procedure manual must include the following when applicable to the test procedure: (b)(1) Requirements for patient preparation; specimen collection, labeling, storage, preservation, transportation, processing, and referral; and criteria for specimen acceptability and rejection as described in 493.1242. (b)(2) Microscopic examination, including the detection of inadequately prepared slides. (b)(3) Step-by- step performance of the procedure, including test calculations and interpretation of results. (b)(4) Preparation of slides, solutions, calibrators, controls, reagents, stains, and other materials used in testing. (b)(5) Calibration and calibration verification procedures. (b)(6) The reportable range for test results for the test system as established or verified in 493.1253. (b)(7) Control procedures. (b)(8)

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access

Survey - March 23, 2021

Survey Type: Standard

Survey Event ID: JMWZ11

Deficiency Tags: D3037 D3037 D5221 D5401 D5415 D5421 D5439 D5469 D5775 D6074 D6013 D6074 D5221 D5401 D5415 D5421 D5439 D5469 D5775 D5791 D6013 D5791

Summary:

Summary Statement of Deficiencies D3037 RETENTION REQUIREMENTS CFR(s): 493.1105(a)(4) Proficiency testing records. Retain all proficiency testing records for at least 2 years. This STANDARD is not met as evidenced by: Based on surveyor review of Proficiency Testing (PT) records and interview with the Technical Consultant (TC), the laboratory failed to retain all records from Nonchemistry and Chemistry PT events performed with the American Association of Bioanalysts in the calendar year 2020. The findings include: 1. Graded results were missing for nonchemistry event Q3 2020. 2. The work records were not retained in the 2020 PT events Q1, Q2 and Q3. 3. The TC confirmed on 3/23/2020 at 9:30 am that PT records were not retained. D5221 EVALUATION OF PROFICIENCY TESTING PERFORMANCE CFR(s): 493.1236(d) All proficiency testing evaluation and verification activities must be documented. This STANDARD is not met as evidenced by: Based on surveyor review of the Proficiency Testing (PT) records and interview with the Technical Consultant (TC), the laboratory failed to review and evaluate results when they received an unacceptable score in Chemistry tests performed with the American Association of Bioanalysts in event Q2 for the calendar year 2020. The findings include: 1. The laboratory received an "out of grading range" for specimen 8 Glucose in - 2020 Q2. 2. There was no documented evidence that the laboratory investigated the failures. 3. The TC confirmed on 3/23/2021 at 10:00 am that the laboratory did not review and document an evaluation of unacceptable PT result. Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 5 -- D5401 PROCEDURE MANUAL CFR(s): 493.1251(a) A written procedures manual for all tests, assays, and examinations performed by the laboratory must be available to, and followed by, laboratory personnel. Textbooks may supplement but not replace the laboratory's written procedures for testing or examining specimens. This STANDARD is not met as evidenced by: Based on surveyor review of the Procedure Manual, and interview with the Technical Consultant (TC), the laboratory failed to follow the "Periodic Procedures for Cartridges" procedure. From July 2020 to the date of survey. The finding includes: 1. The PM stated "For acceptance of newly received cartridge lots, check the Temperature. If all windows are white or if only the A or B windows are blue or the 1 or 2 windows are red, then transit temperatures were satisfactory and the cartridges can be used" but there was no evidence the temperature was checked. 2. The TC confirmed on 3/23/2021 at 1:00 pm that the laboratory did not follow the procedure stated above. D5415 TEST SYSTEMS, EQUIPMENT, INSTRUMENTS, REAGENT CFR(s): 493.1252(c) Reagents, solutions, culture media, control materials, calibration materials, and other supplies, as appropriate, must be labeled to indicate the following: (1) Identity and when significant, titer, strength or concentration. (2) Storage requirements. (3) Preparation and expiration dates. (4) Other pertinent information required for proper use. This STANDARD is not met as evidenced by: (a) Based on surveyor observation of the Quality Control (QC) reagents and interview with the Technical Consultant (TC), the laboratory failed to put a new expiration date on the Eightchek-3wp X-TRA QC reagents in use on the Sysmex XP-300c analyzer on the date of the survey. The findings include: 1. The Manufacturers Package Insert (MPI) stated "open vial stability is 14 days after opening." 2. The laboratory did not put new expiration dates on the Eightchek-3wp X-TRA QC after opening. 3. The TC confirmed on 3/23/2021 at 11:10 am the laboratory failed to put new expiration date on the QC reagents. (b) Based on surveyor observation of the QC reagents and interview with the TC, the laboratory failed to put a new expiration date on the Bio- Rad Lyphochek Immunoassay Plus Controls QC reagents in use on the Tosoh AIA- 360 analyzer on the date of the survey. The findings include: 1. The Manufacturers Package Insert (MPI) stated "open vial stability is 7 days after opening." 2. The laboratory did not put new expiration dates on the Bio-Rad Lyphochek Immunoassay Plus Controls QC after opening. 3. The TC confirmed on 3/23/2021 at 11:10 am the laboratory failed to put new expiration date on the QC reagents. D5421 ESTABLISHMENT AND VERIFICATION OF PERFORMANCE CFR(s): 493.1253(b)(1) Each laboratory that introduces an unmodified, FDA-cleared or approved test system must do the following before reporting patient test results: (1)(i) Demonstrate that it can obtain performance specifications comparable to those established by the -- 2 of 5 -- manufacturer for the following performance characteristics: (1)(i)(A) Accuracy. (1)(i) (B) Precision. (1)(i)(C) Reportable range of test results for the test system. (1)(ii) Verify that the manufacturer's reference intervals (normal values) are appropriate for the laboratory's patient population. This STANDARD is not met as evidenced by: a) Based on surveyor review of the Performance Specification (PS) records and interview with the Technical Consultant (TS) the Laboratory Director (LD) failed to ensure that PS procedures performed on the Tosoh AIA-360 analyzer were adequate from July 2020 to the date of survey. The findings include: 1. The LD did not review and sign the PS results. 2. There was no documention evaluating accuracy, or reportable range study. 3. There were no work records to substantiate precision. 4. There was no raw data found for linearity, precision. 5. There was no documentation which showed what were acceptable results for linearity, accuracy and reportable range. 6. There was no documented evidence a reference range study was perofrmed. 7. The TC confirmed on 3/23/21 at 11:00 that PS records were not adequate. (b) Based on the lack of the PS records and interview with the TC the LD failed to ensure that any PS procedures were performed on Istat analyzers serial numbers 374151 and 374154 from Januaey 24, 2020 to the date of survey. The TC confirmed on 3/23/21 at 12:00 pm that PS were not performed on the aforementioned analyzers. D5439 CALIBRATION AND CALIBRATION VERIFICATION CFR(s): 493.1255(b) Unless otherwise specified in this subpart, for each applicable test system the laboratory must do the following: Perform and document calibration verification procedure - (b)(1) Following the manufacturer's calibration verification instructions; (b)(2) Using the criteria verified or established by the laboratory under 493.1253(b)(3) -- (b)(2)(i) Including the number, type, and concentration of the materials, as well as acceptable limits for calibration verification; and (b)(2)(ii) Including at least a minimal (or zero) value, a mid-point value, and a maximum value near the upper limit of the range to verify the laboratory's reportable range of test results for the test system; and (b)(3) At least once every 6 months and whenever any of the following occur: (b)(3)(i) A complete change of reagents for a procedure is introduced, unless the laboratory can demonstrate that changing reagent lot numbers does not affect the range used to report patient test results, and control values are not adversely affected by reagent lot number changes. (b)(3)(ii) There is major preventive maintenance or replacement of critical parts that may influence test performance. (b)(3)(iii) Control materials reflect an unusual trend or shift, or are outside of the laboratory's acceptable limits, and other means of assessing and correcting unacceptable control values fail to identify and correct the problem. (b)(3)(iv) The laboratory's established schedule for verifying the reportable range for patient test results requires more frequent calibration verification. This STANDARD is not met as evidenced by: (a) Based on lack of Calibration Verification (CV) records and interview with the Technical Consultant (TC), the laboratory failed to perform and document CV procedures at least once every six months for Endocrinology Tests performed on the Tosoh AIA-360 analyzer from July 2020 to the date of the surveyor. The TC confirmed on 3/23/21 at 12:10 pm CV was not performed every six months. (b) Based on lack of CV records and interview with the TC, the laboratory failed to perform and -- 3 of 5 -- document CV procedures at least once every six months for Chemistry Tests performed on the i-Stat analyzers from February 2020 to the date of the surveyor. The TC confirmed on 3/23/201 at 12:30 pm CV was not performed every six months. D5469 CONTROL PROCEDURES CFR(s): 493.1256(d)(10)(g) Unless CMS Approves a procedure, specified in Appendix C of the State Operations Manual (CMS Pub. 7), that provides equivalent quality testing, the laboratory must-- Establish or verify the criteria for acceptability of all control materials. (i) When control materials providing quantitative results are used, statistical parameters (for example, mean and standard deviation) for each batch and lot number of control materials must be defined and available. (ii) The laboratory may use the stated value of a commercially assayed control material provided the stated value is for the methodology and instrumentation employed by the laboratory and is verified by the laboratory. (iii) Statistical parameters for unassayed control materials must be established over time by the laboratory through concurrent testing of control materials having previously determined statistical parameters. (g) The laboratory must document all control procedures performed. This STANDARD is not met as evidenced by: (a) Based on the lack of Quality Control (QC) records and interview with the Technical Consultant (TC), the laboratory failed to verify commercially assayed QC material with each new lot and/or shipment of Bio-Rad Lyphochek immunoassay plus control material performed on the Tosoh AIA-360 from July 2020 until the date of survey. The TC confirmed on 3/23/21 at 10:15 am that the assayed values of QC material were not verified before putting in use. (b) Based on the lack of QC records and interview with the TC, the laboratory failed to verify commercially assayed QC material with each new lot and/or shipment of I-Stat Tricontrols CHEM8+ Level one and two performed on the i-stat from February 2020 until the date of survey. The TC confirmed on 3/23/21 at 11:15 am that the assayed values of QC material were not verified before putting in use. D5775 COMPARISON OF TEST RESULTS CFR(s): 493.1281(a)(c) (a) If a laboratory performs the same test using different methodologies or instruments, or performs the same test at multiple testing sites, the laboratory must have a system that twice a year evaluates and defines the relationship between test results using the different methodologies, instruments, or testing sites. (c) The laboratory must document all test result comparison activities. This STANDARD is not met as evidenced by: Based on surveyor review of the Laboratory Records and interview with the Technical Consultant (TC), the laboratory failed to perform and document comparison studies for the two i-Stat analyzers twice per year from February 2020 to the date of the survey. The TC confirmed on 3/23/21 at 12:20 pm that the laboratory did not do comparison studies. D5791 ANALYTIC SYSTEMS QUALITY ASSESSMENT CFR(s): 493.1289(a)(c) -- 4 of 5 -- (a) The laboratory must establish and follow written policies and procedures for an ongoing mechanism to monitor, assess, and when indicated, correct problems identified in the analytic systems specified in 493.1251 through 493.1283. (c) The laboratory must document all analytic systems assessment activities. This STANDARD is not met as evidenced by: Based on surveyor review of the Procedure Manual and interview with the Technical Consultant (TC) the laboratory failed to have a procedure to verify new lots of controls before they were put in use from 4/30/19 to the date of the survey. The TC confirmed on 3/23/21 at 11:30 am that laboratory did not have the above procedure. D6013 LABORATORY DIRECTOR RESPONSIBILITIES CFR(s): 493.1407(e)(3)(ii) The laboratory director is responsible for the overall operation and administration of the laboratory, including the employment of personnel who are competent to perform test procedures, and record and report test results promptly, accurate, and proficiently and for assuring compliance with the applicable regulations. (e) The laboratory director must-- (e)(3) Ensure that-- (e)(3)(ii) Verification procedures used are adequate to determine the accuracy, precision, and other pertinent performance characteristics of the method; This STANDARD is not met as evidenced by: Based on surveyor review of the Performance Specification (PS) records and interview with the Technical Consultant (TC) the Laboratory Director (LD) failed to ensure that verification procedures were performed for Human chorionic gonadotropin performed on the Tosoh AIA-3600 and Sodium, Potassium, Cholride and Glucose perfromed on the i-stat anlyzers from 1/24/20 to the date of survey. The TC confirmed on 3/23/2021 at 12:10 pm that verification studies were not performed. D6074 TESTING PERSONNEL RESPONSIBILITIES CFR(s): 493.1425(b)(5) Each individual performing moderate complexity testing must be capable of identifying problems that may adversely affect test performance or reporting of test results and either must correct the problems or immediately notify the technical consultant, clinical consultant or director. This STANDARD is not met as evidenced by: Based on surveyor review of the Quality Control (QC) records and interview with the Technical Contultant (TC), the TC failed to identify problems that may affect test performance by not reviewing and evaluating trends and/or shifts for tests performed on the Tosoh AIA-360 and i-stat analyzers from 1/24/20 the date of survey. The TC confirmed on 3/23/21 at 12:45 pm that trends and shifts were not reviewed. -- 5 of 5 --

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access

Survey - April 30, 2019

Survey Type: Standard

Survey Event ID: 1MW311

Deficiency Tags: D5411 D5803 D6074 D6030 D6074 D5803 D5891 D6030 D5891

Summary:

Summary Statement of Deficiencies D5411 TEST SYSTEMS, EQUIPMENT, INSTRUMENTS, REAGENT CFR(s): 493.1252(a) Test systems must be selected by the laboratory. The testing must be performed following the manufacturer's instructions and in a manner that provides test results within the laboratory's stated performance specifications for each test system as determined under 493.1253. This STANDARD is not met as evidenced by: Based on surveyor review of the Immucor Blood Grouping Reagent Package Insert (RPI) and interview with the Technical Supervisor (TS), the laboratory failed to follow RPI instructions for Anti-D (Series 4) tests from 2/24/17 to the date of survey. The findings include: 1. The RPI stated "Using a transfer pipette, add 1 drop of 2-5% suspension of Red Blood Cells (RBC)" but the laboratory did not ensure the suspension added was 2-5% RBC. 2. Approximately 15 patients are run each day. 4. The TS confirmed on 4/30/19 at 2:10 pm that the RPI was not followed. D5803 TEST REPORT CFR(s): 493.1291(b) Test report information maintained as part of the patient's chart or medical record must be readily available to the laboratory and to CMS or a CMS agent upon request. This STANDARD is not met as evidenced by: Based on the surveyor review of the Patients Charts (PC), Test Results (TR) and interview with the Technical Supervisor (TS), the laboratory failed to have all tests perfomed on the PC from 2/24/17 to the date of the survey. The findings include: 1. Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 3 -- The laboratory performed Rubella and Human Immunodeficiency Virus (HIV) tests for an insurance client but did not report the TR in the PC. 2. The TS confirmed on 4 /30/19 at 2:10 pm that the TR was not in the PC. D5891 POSTANALYTIC SYSTEMS QUALITY ASSESSMENT CFR(s): 493.1299(a) The laboratory must establish and follow written policies and procedures for an ongoing mechanism to monitor, assess and, when indicated, correct problems identified in the postanalytic systems specified in 493.1291. This STANDARD is not met as evidenced by: Based on surveyor review of the Procedure Manual (PM) and interview with the Technical Supervisor (TS), the laboratory failed to establish a procedure for verifying manually entered results from 2/24/17 to the date of survey. The TS confirmed on 4/30 /19 at 2:00 pm that the laboratory did not have the procedure mentioned above. D6030 LABORATORY DIRECTOR RESPONSIBILITIES CFR(s): 493.1407(e)(12) The laboratory director is responsible for the overall operation and administration of the laboratory, including the employment of personnel who are competent to perform test procedures, and record and report test results promptly, accurate, and proficiently and for assuring compliance with the applicable regulations. (e) The laboratory director must-- (e)(12) Ensure that policies and procedures are established for monitoring individuals who conduct preanalytical, analytical, and postanalytical phases of testing to assure that they are competent and maintain their competency to process specimens, perform test procedures and report test results promptly and proficiently, and whenever necessary, identify needs for remedial training or continuing education to improve skills; This STANDARD is not met as evidenced by: Based on surveyor review of the Procedure Manual, Competency Assessment (CA) records and interview with the Technical Supervisor (TS), the Laboratory Director (LD) failed to establish a CA procedure with all the required elements to ensure Testing Personnel are competent to perform tests from 2/24/17 to the date of the survey. The findings include: 1. The CA did not include: a. Monitoring the recording and reporting of results. b. Review of worksheets, quality control records, and preventative maintenance records. c. Direct observation of performance of instrument maintenance and function checks. d. Assessment of test performance through Proficiency Testing or internal blind samples. e. Assessment of problem solving skills. 2. The TS confirmed on 4/30/19 at 2:30 pm that a CA procedure was not adequately established. D6074 TESTING PERSONNEL RESPONSIBILITIES CFR(s): 493.1425(b)(5) Each individual performing moderate complexity testing must be capable of identifying problems that may adversely affect test performance or reporting of test results and either must correct the problems or immediately notify the technical consultant, clinical consultant or director. -- 2 of 3 -- This STANDARD is not met as evidenced by: Based on surveyor review of the Quality Control (QC) records and interview with the Technical Supervisor (TS), the Testing Personnel failed to identify problems that may affect test performance by not reviewing and evaluating trends and/or shifts for tests performed on the Sysmex XP-300 and the BioMrieux Mini Vidas analyzer from 2/24 /17 to the date of the survey. The TS confirmed on 4/30/19 at 1:45 pm that trends and shifts were not reviewed. -- 3 of 3 --

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access