Summary:
Summary Statement of Deficiencies D0000 Based on an on-site recertification survey conducted on 6/11/18, a deficiency was cited for Bozeman Health Belgrade Clinic in Belgrade, MT. D3031 RETENTION REQUIREMENTS CFR(s): 493.1105(a)(3) Analytic systems records. Retain quality control and patient test records (including instrument printouts, if applicable) and records documenting all analytic systems activities specified in 493.1252 through 493.1289 for at least 2 years. This STANDARD is not met as evidenced by: Based on observation, record review, and interview, the laboratory failed to retain quality control values for the hematology analyzer from 3/25/17 through 1/7/18. The findings include: 1. On 6/11/18 at 9:30 a.m. a Beckman Coulter DxH 600 hematology analyzer was observed in the laboratory. 2. A review on 6/11/18 at 11:45 a.m. of the quality control results between 3/25/17 and 1/7/18 were not retained on paper, saved electronically, retained on the analyzer, or on the quality control website. 3. On 6/11 /18 at 11:45 a.m., staff member A stated the results were not kept. 4. A review on 6/11 /18 at 12:50 p.m. of the November 2017 UniCel DxH Series Cleaning Checklist included a box daily to check for process controls. The boxes were all checked. Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 1 --