Campus Health Service

CLIA Laboratory Citation Details

3
Total Citations
6
Total Deficiencyies
6
Unique D-Tags
CMS Certification Number 03D0859649
Address Arizona State Universty Health Services, Tempe, AZ, 85281
City Tempe
State AZ
Zip Code85281
Phone(480) 965-3346

Citation History (3 surveys)

Survey - December 14, 2023

Survey Type: Standard

Survey Event ID: 9YZM11

Deficiency Tags: D5433 D5205

Summary:

Summary Statement of Deficiencies D5205 COMPLAINT INVESTIGATIONS CFR(s): 493.1233 The laboratory must have a system in place to ensure that it documents all complaints and problems reported to the laboratory. The laboratory must conduct investigations of complaints, when appropriate. This STANDARD is not met as evidenced by: Based on review of the laboratory's policy and procedure manual and interview with the testing personnel (TP-1), the laboratory failed to have a system in place to document complaints and problems reported to the laboratory. Findings include: 1. The laboratory failed to provide evidence to indicate that it has a system in place to ensure that it documents all complaints and problems reported to the laboratory 2. The TP-1 interviewed on 12/14/2023 at 3:15 PM confirmed the laboratory did not have an established system in place to address complaints and problems reported to the laboratory. 3. The laboratory's reported annual test volume is 10,179. D5433 MAINTENANCE AND FUNCTION CHECKS CFR(s): 493.1254(b)(1) For equipment, instruments, or test systems developed in-house, commercially available and modified by the laboratory, or maintenance and function check protocols are not provided by the manufacturer, the laboratory must establish a maintenance protocol that ensures equipment, instrument, and test system performance that is necessary for accurate and reliable test results and test result reporting. The laboratory must perform and document the maintenance activities specified in paragraph (b)(1)(i) of this section. This STANDARD is not met as evidenced by: Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 2 -- Based on lack of maintenance and function check documentation for the pipettes utilized by the laboratory and interview with the Testing Personnel (TP-1), the laboratory failed to establish a maintenance and calibration protocol for the pipettes that ensures equipment performance which is necessary for accurate and reliable test results and test result reporting. Findings include: 1. The laboratory performs patient testing in the specialties of Microbiology, Diagnostic Immunology, Chemistry and Hematology, with a reported annual test volume of 10,179. 2. The laboratory utilizes two pipettes in conjunction with patient testing, a 100 Microliter (L) pipette, serial# EH92326 and a 200 Microliter (L) pipette, serial# EH83600. 3. No documentation was presented for review during the survey conducted on 12/14/2023 to indicate the laboratory established a maintenance and function protocol for the pipettes listed above. 4. No evidence was provided for review to indicate the laboratory calibrated the pipettes from January 2021 through the date of the survey, 12/14/2023. 5. The TP- 1 interviewed on 12/14/2023 at 3:02 PM confirmed the laboratory failed to establish a maintenance and calibration protocol for the pipettes, and confirmed the pipettes were not routinely calibrated during the timeframe indicated above. -- 2 of 2 --

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access

Survey - August 3, 2021

Survey Type: Standard

Survey Event ID: O4KU11

Deficiency Tags: D5407

Summary:

Summary Statement of Deficiencies D5407 PROCEDURE MANUAL CFR(s): 493.1251(d) Procedures and changes in procedures must be approved, signed, and dated by the current laboratory director before use. This STANDARD is not met as evidenced by: Based on review of the laboratory's policy and procedure manual and interview with the facility personnel, the laboratory failed to have the current laboratory director approve, sign and date test procedures before use. Findings include: 1. The current laboratory director indicated in the CLIA Federal Database and assigned on the CMS- 209, Laboratory Personnel Form presented for review during the survey has been listed as laboratory director since 12/28/2020. 2. The policy and procedure manual presented for review during the survey conducted on August 3, 2021 was not approved, signed and dated by the current laboratory director. 3. The facility personnel confirmed that the policy and procedure manual indicated above was not approved, signed and dated by the current laboratory director. Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 1 --

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access

Survey - December 20, 2018

Survey Type: Standard

Survey Event ID: NU4I11

Deficiency Tags: D5221 D2009 D5445

Summary:

Summary Statement of Deficiencies D2009 TESTING OF PROFICIENCY TESTING SAMPLES CFR(s): 493.801(b)(1) The individual testing or examining the samples and the laboratory director must attest to the routine integration of the samples into the patient workload using the laboratory's routine methods. This STANDARD is not met as evidenced by: Based on review of proficiency testing (PT) records from 2018 and interview with the facility personnel, the laboratory director failed to sign the PT attestation statement. Findings include: 1. The laboratory participates in PT for testing performed in the specialties of Microbiology, Diagnostic Immunology, Chemistry and Hematology. 2. The PT attestation statements presented for review for the second and third events of 2018 lacked the director's signature. 3. The facility personnel confirmed that the PT attestation statements indicated above were not signed by the laboratory director. D5221 EVALUATION OF PROFICIENCY TESTING PERFORMANCE CFR(s): 493.1236(d) All proficiency testing evaluation and verification activities must be documented. This STANDARD is not met as evidenced by: Based on review of verification activities for Measles testing and interview with the facility personnel, the laboratory failed to provide documentation showing the verification activities were performed by the laboratory and acceptable. Findings include: 1. The laboratory performs Measles testing on the Mini-Vidas analyzer. It is the practice of the laboratory to send 5 samples at least twice annually to another laboratory to verify the accuracy of the test. 2. No documentation was provided for review during the survey conducted on December 20, 2018 to indicate the laboratory Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 2 -- verified the accuracy of the Measles test performed on the Mini-Vidas analyzer during 2017 and 2018. During the survey, the laboratory presented a spread sheet for review for each year which contained the test results of the 5 samples tested on the Mini- Vidas and the test results from the reference laboratory, but failed to retain the instrument printouts from either laboratory indicating the samples were tested by the laboratory and the reference laboratory. 3. No documentation was provided for review to indicate the laboratory evaluated the verification results referenced above for acceptability. 4. The facility personnel confirmed that the laboratory failed to retain the analyzer printouts/reference lab reports and evaluate the verification results for acceptability for Measles testing during 2017 and 2018. D5445 CONTROL PROCEDURES CFR(s): 493.1256(d)(1)(2)(g) Unless CMS Approves a procedure, specified in Appendix C of the State Operations Manual (CMS Pub. 7), that provides equivalent quality testing, the laboratory must-- (d)(1) Perform control procedures as defined in this section unless otherwise specified in the additional specialty and subspecialty requirements at 493.1261 through 493.1278. (d)(2) For each test system, perform control procedures using the number and frequency specified by the manufacturer or established by the laboratory when they meet or exceed the requirements in paragraph (d)(3) of this section. (g) The laboratory must document all control procedures performed. This STANDARD is not met as evidenced by: Based on lack of quality control (QC) documentation and interview with the facility personnel, the laboratory failed to perform and document control procedures using the number and frequency as required. Findings include: 1. The laboratory performs serum hCG testing using the Henry Schein combo pregnancy test kit under the sub- specialty of Endocrinology.. On the date of the survey, December 20, 2018, the laboratory's quality control procedure consisted of performing two levels of external control material, once each month and/or each new lot of test kit. 2. No QC documentation was provided for review during the survey to indicate the laboratory performed two levels of control material of different concentrations, each day of patient testing as required since January 1, 2016. 3. During the survey, review of QC records from January 1, 2016 through December 20, 2018 indicated the laboratory performed and documented QC with the number and frequency as described in their policy above, and as of January 1, 2016, the laboratory had not implemented an Individualized Quality Control Plan (IQCP) for this test kit. 4. The facility personnel confirmed that the laboratory did not perform and document controls as required since January 1, 2016 and confirmed that the laboratory had not implemented an Individualized Quality Control Plan (IQCP) for testing performed on the serum pregnancy test. 5. The number of patients tested during the time period indicated above could not be determined at the time of the survey. -- 2 of 2 --

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access