Care Plus Family Medical

CLIA Laboratory Citation Details

2
Total Citations
4
Total Deficiencyies
4
Unique D-Tags
CMS Certification Number 01D2061968
Address 8914 U S Hwy 431, Albertville, AL, 35950
City Albertville
State AL
Zip Code35950
Phone256 279-7200
Lab DirectorBENJAMIN FAVIS

Citation History (2 surveys)

Survey - October 15, 2024

Survey Type: Standard

Survey Event ID: 4XTF11

Deficiency Tags: D6033 D6034

Summary:

Summary Statement of Deficiencies D6033 TECHNICAL CONSULTANT-MODERATE COMPEXITY CFR(s): 493.1409 The laboratory must have a technical consultant who meets the qualification requirements of 493.1411 of this subpart and provides technical oversight in accordance with 493.1413 of this subpart. This CONDITION is not met as evidenced by: Based on a review of the personnel records and an interview with Testing Personnel (TP) #1, the laboratory failed to fill the position of Technical Consultant (TC). The TC position had been unfilled since 09-22-2023 to the date of the current survey 10- 15-2024. The findings include: 1. Refer to D6034. D6034 TECHNICAL CONSULTANT QUALIFICATIONS CFR(s): 493.1411 The laboratory must employ one or more individuals who are qualified by education and either training or experience to provide technical consultation for each of the specialties and subspecialties of service in which the laboratory performs moderate complexity tests or procedures. The director of a laboratory performing moderate complexity testing may function as the technical consultant provided he or she meets the qualifications specified in this section. This STANDARD is not met as evidenced by: Based on a review of the personnel records and an interview with Testing Personnel (TP) #1, the laboratory failed to employ a qualified Technical Consultant (TC) who can provide technical consultation for the moderate complexity testing performed in the laboratory. The surveyor noted the TC position had been vacant for 13 out of the 17 months surveyed. The findings include: 1. A review of the personnel records and Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 2 -- interview with TP#1 revealed the laboratory does not have a qualified TC from 09-22- 2023 to the date of the current survey 10-15-2024. 2. During the exit interview on 10- 15-2024 at 12:38 PM, TP#1 confirmed the above findings. -- 2 of 2 --

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access

Survey - February 21, 2019

Survey Type: Standard

Survey Event ID: YZLX11

Deficiency Tags: D5445 D6029

Summary:

Summary Statement of Deficiencies D5445 CONTROL PROCEDURES CFR(s): 493.1256(d)(1)(2)(g) Unless CMS Approves a procedure, specified in Appendix C of the State Operations Manual (CMS Pub. 7), that provides equivalent quality testing, the laboratory must-- (d)(1) Perform control procedures as defined in this section unless otherwise specified in the additional specialty and subspecialty requirements at 493.1261 through 493.1278. (d)(2) For each test system, perform control procedures using the number and frequency specified by the manufacturer or established by the laboratory when they meet or exceed the requirements in paragraph (d)(3) of this section. (g) The laboratory must document all control procedures performed. This STANDARD is not met as evidenced by: Based on a review of the laboratory's test menu, a review of policies and procedures, including the IQCP (Individualized Quality Control Plan) for microalbumin and creatinine, a review of 2016-2018 quality control records for Chemistry, and an interview with Testing Personnel (TP) #1, the surveyor determined the laboratory failed to perform the quality control testing for the microalbumin and creatinine, as established in the quality control plan of the IQCP. This affected January of 2017, one month in the year. The findings include: 1. During the entrance tour of the laboratory on 2/21/19 at 8:40 AM, the laboratory supervisor (also TP #1) stated microalbumin and creatinine were performed on the Siemens DCA Vantage and were considered moderate complexity tests. 2. The quality control plan of the IQCP for microalbumin (ualb) and creatinine indicated the following: QC: two levels (low and high) are run with each new lot of reagent and to train and confirm performance of new analysts, or when results do not match patient's clinical condition; one level of control, either a low or high, is performed on each box of reagents within the same lot number; two levels of control are performed once monthly whether the same or different lot number, or as an added quality assurance. 3. A review of the quality control records revealed only one level was tested in January of 2017. 4. At 12:38 PM on 2/21/19, the Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 2 -- surveyor and TP #1 reviewed the quality control records for ualb and creatinine. TP #1 confirmed that only one level of quality control was tested in January, 2017; although the IQCP indicated two levels would be tested once per month. 5. The surveyor discussed the above noted findings with TP #1 again during the exit summation on 2/21/19 at 2:00 PM. D6029 LABORATORY DIRECTOR RESPONSIBILITIES CFR(s): 493.1407(e)(11) The laboratory director is responsible for the overall operation and administration of the laboratory, including the employment of personnel who are competent to perform test procedures, and record and report test results promptly, accurate, and proficiently and for assuring compliance with the applicable regulations. (e) The laboratory director must-- (e)(11) Ensure that prior to testing patients' specimens, all personnel have the appropriate education and experience, receive the appropriate training for the type and complexity of the services offered, and have demonstrated that they can perform all testing operations reliably to provide and report accurate results. This STANDARD is not met as evidenced by: Based on a review of the personnel and training records and an interview with Testing Personnel (TP) #1, the surveyor determined the Laboratory Director failed to ensure two testing personnel of moderate complexity testing presented with the appropriate educational credentials, prior to testing patient specimens and reporting the results. This affected two of seven listed testing personnel on the CMS form #209 (Laboratory Personnel Report). The findings include: 1. A review of the personnel records revealed certifications for TP #6 (Medical Assistant) and #7 (Licensed Practical Nurse). Note: The minimum requirement for personnel of moderate complexity testing is a high school diploma. A science-based degree may be used as educational credential. 2. Further review of the training records revealed both testing personnel had been trained on moderate complexity testing: TP #6 was trained and oriented in the laboratory on 2/4/19 and #7, on 1/21/19. 3. In an interview on 2/21/19 at 12:38 PM, TP #1 (also the Laboratory Supervisor) confirmed the two testing personnel performed moderate complexity testing. TP #1 also reviewed the personnel records and noted only certifications of the employees had been obtained. 4. During the exit summation on 2/21/19 at 2:00 PM, the surveyor stated to TP #1 the laboratory could fax the educational documents to the State Agency by 5:00 Friday, February 22. The educational document for TP #6 was received by the State Agency via fax on 2/22/19. No additional documentation of education was received for TP #7. -- 2 of 2 --