Summary:
Summary Statement of Deficiencies D5413 TEST SYSTEMS, EQUIPMENT, INSTRUMENTS, REAGENT CFR(s): 493.1252(b) (b) The laboratory must define criteria for those conditions that are essential for proper storage of reagents and specimens, accurate and reliable test system operation, and test result reporting. The criteria must be consistent with the manufacturer's instructions, if provided. These conditions must be monitored and documented and, if applicable, include the following: (b)(1) Water quality. (b)(2) Temperature. (b)(3) Humidity. (b)(4) Protection of equipment and instruments from fluctuations and interruptions in electrical current that adversely affect patient test results and test reports. This STANDARD is not met as evidenced by: Based on record review and interview with the laboratory's technical consultant, the laboratory did not record daily temperature readings for two of three refrigerators in use. Findings: 1. One of the three refrigerators used by the laboratory is located within the laboratory and two other refrigerators are located in a room outside of the laboratory. The laboratory did not document the daily temperatures (taken each day of patient testing) for the two refrigerators that were located outside of the laboratory. 2. Review of the monthly temperature and humidity logs from December 2024 to June 2025 showed that the laboratory documented the temperature reading of the refrigerator located within the laboratory, but the temperature readings of the other two refrigerators were not recorded on the log. 1. During interview with the technical consultant on July 16, 2025 at 12:00 pm, it was confirmed that the temperature readings of the two refrigerators located outside of the laboratory, used for refrigerated storage by the laboratory, were not recorded each day of patient testing. D5431 MAINTENANCE AND FUNCTION CHECKS CFR(s): 493.1254(a)(2) Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 4 -- (a)(2) Function checks as defined by the manufacturer and with at least the frequency specified by the manufacturer. Function checks must be within the manufacturers established limits before patient testing is conducted. (b) Equipment, instruments, or test systems developed in-house, commercially available and modified by the laboratory, or maintenance and function check protocols are not provided by the manufacturer. The laboratory must do the following: This STANDARD is not met as evidenced by: Based on record review and interview, the laboratory preventive maintenance records were incomplete. Findings: 1. The laboratory documented daily, weekly and monthly preventive maintenance for the Medica chemistry analyzer on the daily cleaning and inspection log. Review of the log showed that the air filter cleaning (required monthly) was not dated or initialed as being performed for June thru November of 2024. 2. The laboratory documented daily, weekly and monthly preventive maintenance for the Tosoh chemistry analyzer on the TOSOH AIA Maintenance Schedule log. Review of twelve monthly logs from June 2024 to June 2025 showed that weekly maintenance was performed for only two of the four weeks in each month of August 2024, February 2025, and April 2025. 3. The laboratory documented monthly cleaning of the red cell and white cell transducer for the XP300 hematology analyzer on the monthly preventive maintenance log. Review of twelve logs showed that this was not performed or initialed for the months of June, April and January of 2025. D5437 CALIBRATION AND CALIBRATION VERIFICATION CFR(s): 493.1255(a) (a )Unless otherwise specified in this subpart, for each applicable test system the laboratory must perform and document calibration procedures-- (a)(1) Following the manufacturer's test system instructions, using calibration materials provided or specified, and with at least the frequency recommended by the manufacturer; (a)(2) Using the criteria verified or established by the laboratory as specified in 493.1253(b) (3)-- (a)(2)(i) Using calibration materials appropriate for the test system and, if possible, traceable to a reference method or reference material of known value; and (a) (2)(ii) Including the number, type, and concentration of calibration materials, as well as acceptable limits for and the frequency of calibration; and (a)(3) Whenever calibration verification fails to meet the laboratory's acceptable limits for calibration verification. This STANDARD is not met as evidenced by: Based on record review and interview, the lab did not have calibration documentation for the routine chemistry testing performed on the Medica. The calibration dates and records including data, calibration lot number and identification of testing staff could not be retrieved from the analyzer to show that calibrations were performed in a reliable and accurate manner. Findings: 1. The manufacturer requires different calibration intervals, depending on the analyte. For example: albumin requires calibration every sixty days, total bilirubin requires calibration every seven days and magnesium requires calibration every three days. 2. Records for previous calibrations performed in 2024 and 2025 were requested from the technical consultant on July 16, 2025, the records were not available and this was confirmed during interview with the technical consultant on July 16, 2023 at 11:30 am, the technical consultant stated that analyzer was calibrated as needed and every Monday, but was unable to provide -- 2 of 4 -- documentation showing that the calibrations were performed according to the performance standards for each test reagent. D5779