Cornerstone Pediatrics Pc

CLIA Laboratory Citation Details

1
Total Citation
2
Total Deficiencyies
2
Unique D-Tags
CMS Certification Number 01D0705526
Address 701 Will Halsey Way, Madison, AL, 35758
City Madison
State AL
Zip Code35758
Phone256 461-7440
Lab DirectorK DENNY

Citation History (1 survey)

Survey - April 16, 2026

Survey Type: Standard

Survey Event ID: F39B11

Deficiency Tags: D5413 D5477

Summary:

Summary Statement of Deficiencies D5413 TEST SYSTEMS, EQUIPMENT, INSTRUMENTS, REAGENT CFR(s): 493.1252(b) (b) The laboratory must define criteria for those conditions that are essential for proper storage of reagents and specimens, accurate and reliable test system operation, and test result reporting. The criteria must be consistent with the manufacturer's instructions, if provided. These conditions must be monitored and documented and, if applicable, include the following: (b)(1) Water quality. (b)(2) Temperature. (b)(3) Humidity. (b)(4) Protection of equipment and instruments from fluctuations and interruptions in electrical current that adversely affect patient test results and test reports. This STANDARD is not met as evidenced by: Based on reviews of the temperature charts, Selective Strep Agar (SSA) plate package insert, and an interview with the Laboratory Director (LD), the laboratory failed to document the