Summary:
Summary Statement of Deficiencies D0000 A recertification survey was conducted on February 24, 2022. Dan H Meirson MD PA clinical laboratory was not in compliance with 42 CFR 493, requirements for clinical laboratories. D5217 EVALUATION OF PROFICIENCY TESTING PERFORMANCE CFR(s): 493.1236(c)(1) At least twice annually, the laboratory must verify the accuracy of any test or procedure it performs that is not included in subpart I of this part. This STANDARD is not met as evidenced by: Based on record review and interview, the laboratory failed to verify accuracy of the reading and interpretation of the Hematoxylin and Eosin (H&E) stain at least twice annually in 2021. Findings: Review of "Quality Assurance Proficiency Testing" showed there were four case reviewed for proficiency testing in 2021. The "Date received by the Mohs technologist And Material received, reviewed by" the Physician on the proficiency testing form was dated 10/20/2021 for all four cases. On 02/24 /2022 at 1:40 PM, the Histology Technician acknowledged the four cases were all dated 10/20/2021. Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 1 --