Drs Ruggiero Landinsky Stambler Northor

CLIA Laboratory Citation Details

1
Total Citation
10
Total Deficiencyies
5
Unique D-Tags
CMS Certification Number 33D0156859
Address 126 East Main St, East Islip, NY, 11730
City East Islip
State NY
Zip Code11730
Phone631 581-0090
Lab DirectorJAY STAMBLER

Citation History (1 survey)

Survey - July 7, 2022

Survey Type: Special

Survey Event ID: SZ7G11

Deficiency Tags: D2016 D2087 D2087 D2107 D6000 D2016 D2107 D6000 D6016 D6016

Summary:

Summary Statement of Deficiencies D2016 SUCCESSFUL PARTICIPATION CFR(s): 493.803(a)(b)(c) (a) Each laboratory performing nonwaived testing must successfully participate in a proficiency testing program approved by CMS, if applicable, as described in subpart I of this part for each specialty, subspecialty, and analyte or test in which the laboratory is certified under CLIA. (b) Except as specified in paragraph (c) of this section, if a laboratory fails to participate successfully in proficiency testing for a given specialty, subspecialty, analyte or test, as defined in this section, or fails to take remedial action when an individual fails gynecologic cytology, CMS imposes sanctions, as specified in subpart R of this part. (c) If a laboratory fails to perform successfully in a CMS- approved proficiency testing program, for the initial unsuccessful performance, CMS may direct the laboratory to undertake training of its personnel or to obtain technical assistance, or both, rather than imposing alternative or principle sanctions except when one or more of the following conditions exists: (1) There is immediate jeopardy to patient health and safety. (2) The laboratory fails to provide CMS or a CMS agent with satisfactory evidence that it has taken steps to correct the problem identified by the unsuccessful proficiency testing performance. (3) The laboratory has a poor compliance history. This CONDITION is not met as evidenced by: Based on Proficiency Test (PT) desk review of the Center for Medicaid & Medicare Services (CMS) PT data reports and PT summary reports from American Proficiency Institute (API) PT program, the laboratory failed to participate successfully in CMS approved proficiency testing program, for the test analyte Thyroid Stimulating Hormone (TSH) The following scores were assigned: 2022 first event = 40% 2022 second event = 40% This is considered unsuccessful PT performance. Refer to D2107 D2087 ROUTINE CHEMISTRY CFR(s): 493.841(a) Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 3 -- Failure to attain a score of at least 80 percent of acceptable responses for each analyte in each testing event is unsatisfactory analyte performance for the testing event. This STANDARD is not met as evidenced by: Based on PT desk review of the CMS PT data reports and PT records from the API PT program, the laboratory failed to participate successfully in proficiency testing for the test analyte's Asparatate Aminotransferase (AST) Total Chlesterol and Sodium (Na) The following scores were assigned: AST 2022 first event = 60% Total Cholesterol 2022 first event = 0% Na 2022 first event = 20% This is considered unsatisfactory PT performance. D2107 ENDOCRINOLOGY CFR(s): 493.843(f) Failure to achieve satisfactory performance for the same analyte or test in two consecutive testing events or two out of three consecutive testing events is unsuccessful performance. This STANDARD is not met as evidenced by: Based on PT desk review of the CMS PT data reports and PT summary reports from API PT program, the laboratory failed to participate successfully in CMS approved proficiency testing program, the test analyte TSH. The following scores were assigned: 2022 first event = 40% 2022 second event = 40% This is considered unsuccessful PT performance D6000 MODERATE COMPLEXITY LABORATORY DIRECTOR CFR(s): 493.1403 The laboratory must have a director who meets the qualification requirements of 493. 1405 of this subpart and provides overall management and direction in accordance with 493.1407 of this subpart. This CONDITION is not met as evidenced by: Based on PT desk review of the CMS PT data reports and API PT program records, the laboratory director failed to fulfill the laboratory director's responsibilities and ensure that the laboratory achieved a satisfactory performance and successfully participate in a PT program, approved by CMS, for the test analyte's TSH, AST, Total Chlesterol and Na. Refer to D6016 D6016 LABORATORY DIRECTOR RESPONSIBILITIES CFR(s): 493.1407(e)(4)(i) The laboratory director is responsible for the overall operation and administration of the laboratory, including the employment of personnel who are competent to perform test procedures, and record and report test results promptly, accurate, and proficiently and for assuring compliance with the applicable regulations. (e) The laboratory director must-- (e)(4)(i) Ensure that the proficiency testing samples are tested as required under Subpart H of this part; -- 2 of 3 -- This STANDARD is not met as evidenced by: Based on PT desk review of the CMS PT data reports and API PT program records, the laboratory director failed to fulfill the laboratory director's responsibilities and ensure that the laboratory achieved a satisfactory performance and successfully participate in a PT program, approved by CMS, for the test analyte's TSH, AST, Total Chlesterol and Na. The following scores were assigned: 2022 first event = 40% 2022 second event = 40% This is considered unsuccessful PT performance. AST 2022 first event = 60% Total Cholesterol 2022 first event = 0% Na 2022 first event = 20% This is considered unsatisfactory PT performance. -- 3 of 3 --

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access