East Texas Family Medicine Pa

CLIA Laboratory Citation Details

7
Total Citations
81
Total Deficiencyies
33
Unique D-Tags
CMS Certification Number 45D0673541
Address 4201 South Loop 256, Palestine, TX, 75801
City Palestine
State TX
Zip Code75801
Phone(903) 723-8533

Citation History (7 surveys)

Survey - June 27, 2024

Survey Type: Standard

Survey Event ID: YQGF11

Deficiency Tags: D0000 D5217 D5401 D5787 D5805 D6000 D6007 D6029 D6053 D6054 D6063 D6066 D0000 D5217 D5401 D5787 D5805 D6000 D6007 D6029 D6053 D6054 D6063 D6066

Summary:

Summary Statement of Deficiencies D0000 An onsite survey conducted 06/27/2024 found the laboratory out of compliance with 42 CFR Part 493, Requirements for Laboratories for the following conditions: D6000 - 42 C.F.R. 493.1403 Condition: Laboratories performing moderate complexity testing; laboratory director; D6063 - 42 C.F.R. 493.1421 Condition: Laboratories performing moderate complexity testing; testing personnel; D5217 EVALUATION OF PROFICIENCY TESTING PERFORMANCE CFR(s): 493.1236(c)(1) At least twice annually, the laboratory must verify the accuracy of any test or procedure it performs that is not included in subpart I of this part. This STANDARD is not met as evidenced by: Based on surveyor observation, review of laboratory documents, and confirmed in interview, the laboratory failed to perform twice annual accuracy evaluation for direct wet mount procedures being performed for records reviewed in 2023. The findings included: 1. During a tour of the facility space on 6/27/2024 at 11:30 hours the surveyor observed a microscope next to a box of slides, slide covers, saline, and pipettes. Surveyor asked what the microscope was utilized for and testing personnel (TP) 3 stated, on 6/27/2024 at 11:34 in the nursing area, that the laboratory director and TP11 performed vaginal wet preps on patients. 2. Review of the laboratory policy titled "Quality Assurance Policies and Procedures" did not include instructions for a twice annual accuracy evaluation for non-regulated analytes. 3. Review of laboratory Current Procedural Terminology (CPT) code billing summary with for "Wet mount for infection agents" for 2023 and January 2024 - June 27, 2024 included 27 patients that had a direct wet mount procedure performed and billed for. 4. In an interview on 6 /27/2024 at 12:00 hours, in the office, the practice manager confirmed that the laboratory did not perform twice annual accuracy evaluations for direct wet mount procedures. Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 6 -- D5401 PROCEDURE MANUAL CFR(s): 493.1251(a) A written procedures manual for all tests, assays, and examinations performed by the laboratory must be available to, and followed by, laboratory personnel. Textbooks may supplement but not replace the laboratory's written procedures for testing or examining specimens. This STANDARD is not met as evidenced by: Based on surveyor observation, review of laboratory documents, billing records, and confirmed in interview, the laboratory failed to have a written policy for the direct wet mount procedures being performed on patient specimens for records reviewed in 2023 and January to June 27, 2024. The findings included: 1. During a tour of the facility space on 6/27/2024 at 11:30 hours the surveyor observed a microscope next to a box of slides, slide covers, saline, and pipettes. Surveyor asked what the microscope was utilized for and testing personnel (TP) 3 stated, on 6/27/2024 at 11:34 in the nursing area, that the laboratory director and TP11 performed vaginal wet preps on patients. 2. Surveyor asked for the laboratory policy for the direct wet mount procedures being performed and none was provided. 3. Review of laboratory Current Procedural Terminology (CPT) code billing summary with for "Wet mount for infection agents" for 2023 and January 2024 - June 27, 2024 included 27 patients that had a direct wet mount procedure performed and billed for. 4. In an interview on 6/27/2024 at 11:45, in the office, the practice manager confirmed that the laboratory did not have a written policy in place for direct wet mount procedures being performed. D5787 TEST RECORDS CFR(s): 493.1283(a) The laboratory must maintain an information or record system that includes the following: (a)(1) The positive identification of the specimen. (a)(2) The date and time of specimen receipt into the laboratory. (a)(3) The condition and disposition of specimens that do not meet the laboratory's criteria for specimen acceptability. (a)(4) The records and dates of all specimen testing, including the identity of the personnel who performed the test(s). This STANDARD is not met as evidenced by: Based on surveyor observation, review of laboratory documents, and confirmed in interview, the laboratory failed to maintain information or record for the positive identification of specimens, date and time the specimen was received and tested, and the identity of the personnel performing the test for direct wet mount procedures being performed in 2023 and 2024. The findings included: 1. During a tour of the facility space on 6/27/2024 at 11:30 hours the surveyor observed a microscope next to a box of slides, slide covers, saline, and pipettes. Surveyor asked what the microscope was utilized for and testing personnel (TP) 3 stated, on 6/27/2024 at 11:34 in the nursing area, that the laboratory director and TP11 performed vaginal wet preps on patients. 2. Surveyor asked how the laboratory tracked specimens being tested to ensure proper patient identification, test performance, results, and personnel performing the test. The practice manager stated, on 6/27/2024 at 11:36 hours in the nursing pod, they didn't even know the test was being performed. 3. Review of laboratory Current Procedural Terminology (CPT) code billing summary with for "Wet mount for infection agents" for 2023 and January 2024 - June 27, 2024 included 27 patients that had a direct wet -- 2 of 6 -- mount procedure performed and billed for. 2023: 13 patients Patient 1, Patient 2, Patient 3, Patient 4, Patient 5, Patient 6, Patient 7, Patient 8, Patient 9, Patient 10, Patient 11, Patient 12, Patient 13 (See Crosswalk) January 1, 2024 - June 27, 2024: 14 patients Patient 14, Patient 15, Patient 16, Patient 17, Patient 18, Patient 19, Patient 20, Patient 21, Patient 22, Patient 23, Patient 24, Patient 25, Patient 26, Patient 27 (See Crosswalk) 4. In an interview on 6/27/2024 at 11:55 hours, in the office, the practice manager confirmed that the laboratory did not have record for the positive identification of specimens, date and time the specimen was received and tested, and the identity of the personnel performing the test for direct wet mount procedures being performed in 2023 and 2024. D5805 TEST REPORT CFR(s): 493.1291(c) The test report must indicate the following: (c)(1) For positive patient identification, either the patient's name and identification number, or a unique patient identifier and identification number. (c)(2) The name and address of the laboratory location where the test was performed. (c)(3) The test report date. (c)(4) The test performed. (c)(5) Specimen source, when appropriate. (c)(6) The test result and, if applicable, the units of measurement or interpretation, or both. (c)(7) Any information regarding the condition and disposition of specimens that do not meet the laboratory's criteria for acceptability. This STANDARD is not met as evidenced by: Based on surveyor observation, review of laboratory documents, and confirmed in interview, the laboratory failed to ensure that the patient final report included the test results for direct wet mount procedures being performed for records reviewed in 2023 and 2024. The findings included: 1. During a tour of the facility space on 6/27/2024 at 11:30 hours the surveyor observed a microscope next to a box of slides, slide covers, saline, and pipettes. Surveyor asked what the microscope was utilized for and testing personnel (TP) 3 stated, on 6/27/2024 at 11:34 in the nursing area, that the laboratory director and TP11 performed vaginal wet preps on patients. 2. Review of laboratory Current Procedural Terminology (CPT) code billing summary with for "Wet mount for infection agents" for 2023 and January 2024 - June 27, 2024 included 27 patients that had a direct wet mount procedure performed and billed for. Review of patient final reports did not include the results of the direct wet mount procedure. 3. In an interview on 6/27/2024 at 12:10 hours, in the office, the practice manager confirmed that the patient final reports did not include the test results for the direct wet mount procedures being performed. D6000 MODERATE COMPLEXITY LABORATORY DIRECTOR CFR(s): 493.1403 The laboratory must have a director who meets the qualification requirements of 493. 1405 of this subpart and provides overall management and direction in accordance with 493.1407 of this subpart. This CONDITION is not met as evidenced by: Based on surveyor observation, review of laboratory personnel records, laboratory documents, patient charts, and confirmed in interview the laboratory director failed to ensure proper management and direction for personnel performing direct wet mount -- 3 of 6 -- procedures, from preanalytical phases of testing through the analytical and postanalytical phases of testing (D6007), and ensuring testing personnel had the appropriate training and continued competencies to ensure accurate and reliable testing (D6029) for records reviewed from January 2023 to June 27, 2024. D6007 LABORATORY DIRECTOR RESPONSIBILITIES CFR(s): 493.1407(e)(1) The laboratory director is responsible for the overall operation and administration of the laboratory, including the employment of personnel who are competent to perform test procedures, and record and report test results promptly, accurate, and proficiently and for assuring compliance with the applicable regulations. (E) The laboratory director must-- (E)(1) Ensure that testing systems developed and used for each of the tests performed in the laboratory provide quality laboratory services for all aspects of test performance, which includes the preanalytic, analytic, and postanalytic phases of testing; This STANDARD is not met as evidenced by: Based on surveyor observation, review of laboratory documents, and confirmed in interview, the laboratory director failed to ensure that all aspects of the direct wet mount procedures, performed by providers and mid-level practitioners, were met for records reviewed in from January 2023 to June 27, 2024. The findings included: 1. During a tour of the facility space on 6/27/2024 at 11:30 hours the surveyor observed a microscope next to a box of slides, slide covers, saline, and pipettes. Surveyor asked what the microscope was utilized for and testing personnel (TP) 3 stated that the laboratory director and TP11 performed vaginal wet preps on patients. 2. Review of laboratory documents did not include a written policy for the direct wet mount procedures being performed on patient specimens for records reviewed in 2023 and January to June 2024. Refer to D5401. 3. Review of laboratory documents did not include record for the positive identification of specimens, date and time the specimen was received and tested, and the identity of the personnel performing the test for direct wet mount procedures being performed in 2023 and 2024. Refer to D5787. 4. Review of laboratory documents failed to ensure that the patient final report included the test results for direct wet mount procedures being performed for records reviewed in 2023 and 2024. Refer to D5805. D6029 LABORATORY DIRECTOR RESPONSIBILITIES CFR(s): 493.1407(e)(11) The laboratory director is responsible for the overall operation and administration of the laboratory, including the employment of personnel who are competent to perform test procedures, and record and report test results promptly, accurate, and proficiently and for assuring compliance with the applicable regulations. (e) The laboratory director must-- (e)(11) Ensure that prior to testing patients' specimens, all personnel have the appropriate education and experience, receive the appropriate training for the type and complexity of the services offered, and have demonstrated that they can perform all testing operations reliably to provide and report accurate results. This STANDARD is not met as evidenced by: Based on review of laboratory personnel files and confirmed in interview, the laboratory director failed to ensure that all personnel had the appropriate training -- 4 of 6 -- (refer to D6066) and continued competency to ensure accurate and reliable test results (refer to D6053 and D6054) for records reviewed in 2023. D6053 TECHNICAL CONSULTANT RESPONSIBILITIES CFR(s): 493.1413(b)(9) The technical consultant is responsible for evaluating and documenting the performance of individuals responsible for moderate complexity testing at least semiannually during the first year the individual tests patient specimens. This STANDARD is not met as evidenced by: Based on review of personnel records, laboratory policy, and confirmed in interview, the technical consultant failed to ensure two of two new testing personnel received two competency assessments in the first year of testing, for records reviewed in 2023 and 2024. The findings included: 1. Review of laboratory personnel records included the following two testing personnel who began testing in 2023: TP Hire Date 1st Competency TP8 3/12/2023 3/28/2023 TP9 4/12/2023 4/29/2023 Surveyor asked for the above testing personnel had their second annual competency assessment and none was provided. 2. Review of the laboratory policy titled "Quality Assurance Policies and Procedures", section 6. "Personnel Assessment" did not include instructions for semiannual competency assessments in the first year of testing for new testing personnel. 3. In an interview on 6/27/2024 at 09:20 hours, in the break room, the practice manager confirmed that the technical consultant did not perform a second competency assessment for the above two testing personnel during the first year of testing. D6054 TECHNICAL CONSULTANT RESPONSIBILITIES CFR(s): 493.1413(b)(9) The technical consultant is responsible for evaluating and documenting the performance of individuals responsible for moderate complexity testing at least annually, after the first year. This STANDARD is not met as evidenced by: Based on review of laboratory policy, personnel records, and confirmed in interview, the technical consultant failed to ensure annual competency evaluations were performed in 2023 for seven of seven testing personnel performing moderate complexity testing. The findings included: 1. Review of the laboratory policy titled "Quality Assurance Policies and Procedures", section 6. "Personnel Assessment" stated the following: "The laboratory director will use the results of Proficiency testing and direct observation to perform an annual evaluation of all testing personnel to ensure competency in job performance." 2. Review of personnel records included the following six testing personnel, performing testing on the Sysmex XN-330 hematology analyzer, who did not have a 2023 competency assessment performed: TP1, TP2, TP3, TP4, TP5, TP6, TP7 Review of personnel records included the following one personnel, performing direct wet mount preparations, who did not have a 2023 competency assessment performed: TP11 3. In an interview on 6/27/2024 at 09:20 hours, in the breakroom, the practice manager confirmed that the technical consultant failed to perform annual competency assessments in 2023. D6063 LABORATORY TESTING PERSONNEL -- 5 of 6 -- CFR(s): 493.1421 The laboratory must have a sufficient number of individuals who meet the qualification requirements of 493.1423, to perform the functions specified in 493. 1425 for the volume and complexity of tests performed. This CONDITION is not met as evidenced by: Based on review of laboratory personnel records, manufacturers' instructions for use, and confirmed in interview, the laboratory failed to ensure three of three new testing personnel met the qualification requirements at 493.1423 (b)(4)(ii) for records reviewed in 2023 and 2024. Refer to D6066. D6066 TESTING PERSONNEL QUALIFICATIONS CFR(s): 493.1423(b)(4)(ii) Have documentation of training appropriate for the testing performed prior to analyzing patient specimens. This STANDARD is not met as evidenced by: Based on review of laboratory personnel records, manufacturers' instructions for use, and confirmed in interview, the laboratory failed to ensure three of three new testing personnel had documentation of training on the Sysmex XN-330 hematology analyzer before being patient testing. The findings included: 1. Review of laboratory personnel records included the following three testing personnel (TP) who were qualified under 493.1423(b)(4)(i) without documentation of training for the testing performed on the Sysmex XN-330 hematology analyzer: TP8, hire date 3/12/2023 TP9, hire date 4/12 /2023 TP10, hire date 4/23/2023 2. Review of the "Sysmex XN-330 General Information" instructions for use, chapter 2 "Safety Information", subsection 2.9 "Operators" had the following statement: "Caution! The instrument must only be used by properly trained personnel." 3. In an interview on 6/27/2024 at 09:15 hours, in the breakroom, the practice manager confirmed that training for the new employees had not been documented. -- 6 of 6 --

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access

Survey - January 3, 2024

Survey Type: Special

Survey Event ID: IVJS11

Deficiency Tags: D0000 D2016 D2131 D6000 D6016 D0000 D2016 D2131 D6000 D6016

Summary:

Summary Statement of Deficiencies D0000 Based on a proficiency testing desk review survey performed on January 3, 2024, the laboratory was found to be out of compliance based on the following CONDITION LEVEL DEFICIENCIES: D2016 - 42 C.F.R. 493.803 Condition: Successful participation D6000 - 42 C.F.R. 493.1403 Condition: Laboratory Director, moderate complexity D2016 SUCCESSFUL PARTICIPATION CFR(s): 493.803(a)(b)(c) (a) Each laboratory performing nonwaived testing must successfully participate in a proficiency testing program approved by CMS, if applicable, as described in subpart I of this part for each specialty, subspecialty, and analyte or test in which the laboratory is certified under CLIA. (b) Except as specified in paragraph (c) of this section, if a laboratory fails to participate successfully in proficiency testing for a given specialty, subspecialty, analyte or test, as defined in this section, or fails to take remedial action when an individual fails gynecologic cytology, CMS imposes sanctions, as specified in subpart R of this part. (c) If a laboratory fails to perform successfully in a CMS- approved proficiency testing program, for the initial unsuccessful performance, CMS may direct the laboratory to undertake training of its personnel or to obtain technical assistance, or both, rather than imposing alternative or principle sanctions except when one or more of the following conditions exists: (1) There is immediate jeopardy to patient health and safety. (2) The laboratory fails to provide CMS or a CMS agent with satisfactory evidence that it has taken steps to correct the problem identified by the unsuccessful proficiency testing performance. (3) The laboratory has a poor compliance history. This CONDITION is not met as evidenced by: Based on desk review of the Certification and Survey Provider Enhanced Reporting (CASPER) Report 155 Individual Laboratory Profile, American Proficiency Institute (API) testing records, the laboratory failed to achieve successful performance in two Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 3 -- of three testing events in 2023 for the specialty of Hematology, resulting in unsuccessful performance. Refer to D2131. D2131 HEMATOLOGY CFR(s): 493.851(g) Failure to achieve an overall testing event score of satisfactory performance for two consecutive testing events or two out of three consecutive testing events is unsuccessful performance. This STANDARD is not met as evidenced by: Based on a desk review of the Certification and Survey Provider Enhanced Reporting (CASPER) Report 155 Individual Laboratory Profile, American Proficiency Institute (API) testing records from 2023, the laboratory failed to achieve an overall testing event score of satisfactory performance (80% or greater) for two of three testing events for the specialty of Hematology. Two out of three overall testing event scores of unsatisfactory performance result in unsuccessful PT performance. Findings were: 1. A review of the CASPER Report 155 listed the following scores for the specialty of Hematology in 2023: Hematology 2023 Event 1: 0% Hematology 2023 Event 3: 0% 2. A proficiency desk review of API proficiency testing records from 2023 confirmed that the laboratory received an overall score of 0% for Hematology 1st and 3rd events. D6000 MODERATE COMPLEXITY LABORATORY DIRECTOR CFR(s): 493.1403 The laboratory must have a director who meets the qualification requirements of 493. 1405 of this subpart and provides overall management and direction in accordance with 493.1407 of this subpart. This CONDITION is not met as evidenced by: Based on a desk review of the Certification and Survey Provider Enhanced Reporting (CASPER) Report 155 Individual Laboratory Profile, American Proficiency Institute (API) testing records, the laboratory director failed to ensure successful participation in an HHS approved proficiency testing program for the specialty of Hematology for two of three events in 2023. Refer to D6016. D6016 LABORATORY DIRECTOR RESPONSIBILITIES CFR(s): 493.1407(e)(4)(i) The laboratory director is responsible for the overall operation and administration of the laboratory, including the employment of personnel who are competent to perform test procedures, and record and report test results promptly, accurate, and proficiently and for assuring compliance with the applicable regulations. (e) The laboratory director must-- (e)(4)(i) Ensure that the proficiency testing samples are tested as required under Subpart H of this part; This STANDARD is not met as evidenced by: Based on a desk review of the Certification and Survey Provider Enhanced Reporting (CASPER) Report 155 Individual Laboratory Profile American Proficiency Institute (API) testing records, the laboratory director failed to ensure successful participation -- 2 of 3 -- in a HHS approved proficiency testing program for the specialty of Hematology for two of three events in 2023. Refer to D2131. -- 3 of 3 --

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access

Survey - January 5, 2023

Survey Type: Standard

Survey Event ID: GVJK11

Deficiency Tags: D5413 D6045 D0000 D5413 D6045

Summary:

Summary Statement of Deficiencies D0000 An onsite survey conducted 01/05/2023 found the laboratory in compliance with 42 CFR Part 493, Requirements for Laboratories. D5413 TEST SYSTEMS, EQUIPMENT, INSTRUMENTS, REAGENT CFR(s): 493.1252(b) The laboratory must define criteria for those conditions that are essential for proper storage of reagents and specimens, accurate and reliable test system operation, and test result reporting. The criteria must be consistent with the manufacturer's instructions, if provided. These conditions must be monitored and documented and, if applicable, include the following: (1) Water quality. (2) Temperature. (3) Humidity. (4) Protection of equipment and instruments from fluctuations and interruptions in electrical current that adversely affect patient test results and test reports. This STANDARD is not met as evidenced by: Based on surveyor observations, review of manufacturers instructions for use, laboratory documents, and confirmed in interview the laboratory failed to monitor the storage temperature for SysmexXN-330 hematology quality controls (QC) for six of six months from June to December 2022. The findings include: 1. Review of the XN- L Check QC instructions for use (IFU) section "Storage and shelf life of unopened product" stated the following: "When stored at 2-8(degrees) Celsius(C), the unopened vial is stable for 100 days." 2. In a tour of the laboratory on 1/5/2023 at 1030a, testing person (TP) 3 lead the surveyor to a silver-faced miniature refrigerator labeled "PRIVATE PAY" as the storage location for QC intended to be used for the SysmexXN-330 hematology analyzer. Surveyor noted the following 2 boxes of controls in the door of the refrigerator: 1 unopened box of XN-L Check, Lot 2323, expiration 2/28/2023 Level 1 - 4 vials Level 2 - 4 vials Level 3 - 4 Vials 1 opened box of XN-L Check controls, Lot2323, expiration 2/28/2023 Level 1 - 1 vial Level 2 - 1 vial Level 3 - 1 vial 3. Review of temperature documentation for June to December 2022 did not include a record of the refrigerator labeled "PRIVATE PAY". 4. In an Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 2 -- interview on 1/5/2023 at 10:55 hours, in the conference room, TP3 and the practice consultant confirmed that there was no record of the monitoring of acceptable temperatures for the refrigerator where QC for the SysmexXN-330 hematology analyzer was kept. D6045 TECHNICAL CONSULTANT RESPONSIBILITIES CFR(s): 493.1413(b)(7) (b) The technical consultant is responsible for-- (b)(7) Identifying training needs and assuring that each individual performing tests receives regular in-service training and education appropriate for the type and complexity of the laboratory services performed; This STANDARD is not met as evidenced by: Based on a review of personnel records, laboratory policy, and confirmed in an interview the technical consultant (TC) failed to have documentation of training for one of one new testing personnel, hired in 2022, performing testing on the SysmexXN- 330 hematology analyzer. The findings include: 1. Review of new hire personnel records for testing person (TP) 6, as listed on the Centers for Medicare and Medicaid (CMS) personnel form 209, with a hire date of 4/5/2022. Further review showed documentation of TP6's first annual competency assessment for the SysmexXN-330 hematology analyzer by the technical consultant (TC) on 6/31/2022. Surveyor queried for documentation of training on the SysmexXN-330 hematology analyzer prior to the competency assessment performed on 6/31/2022, and none was provided. 2. Review of the laboratory policy titled "Quality Assurance Policies and Procedures", subsection "General Quality Policies" stated the following: "1. Quality in the entire test system is of foremost importance to this laboratory" "2. All laboratory personnel are trained properly commensurate with their positions, duties, and responsibilities." 3. In an interview on 1/5/2023 at 09:35 hours, in the breakroom, the practice consultant confirmed that the TC did not have documented training for TP6 on the SysmexXN-330. -- 2 of 2 --

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access

Survey - May 10, 2022

Survey Type: Standard

Survey Event ID: 7ZZS12

Deficiency Tags: D1001 D5413

Summary:

Summary Statement of Deficiencies D1001 CERTIFICATE OF WAIVER TESTS CFR(s): 493.15(e) Laboratories eligible for a certificate of waiver must-- (1) Follow manufacturers' instructions for performing the test; and (2) Meet the requirements in subpart B, Certificate of Waiver, of this part. This STANDARD is not met as evidenced by: Based on a review of instrument instruction manuals and confirmed in interview, the laboratory failed to follow manufacturers' instructions for quality control (QC) requirements for two of two new McKesson 120 Urine Analyzers since they were put in use for patient testing in March 2022. The findings include: 1. In a tour of the laboratory on 5/10/2022 at 09:20 hours, the practice manager stated the laboratory received two new McKesson 120 Urine Analyzers for patient testing. The urine analyzer in the patient "Pod A" was received 3/23/2022, and the urine analyzer placed in the laboratory was received 5/3/2022. 2. Review of the McKesson 120 Urine Analyzer Instruction manual Section 7 "Quality Control" stated: "Test known positive and negative specimen/controls at the following events in accordance with local, state, and/or federal regulations or accreditation requirements. - A new canister of strips is opened - A new operator uses the analyzer - Test results seem inaccurate - After performing maintenance or service on the analyzer" 3. Surveyor queried for the quality control records for the two McKesson 120 Urine Analyzers and none was provided. 4. In an interview on 5/10/2022 at 09:50 hours, in the hallway, the practice manager stated that the laboratory did not maintain a record of the QC performed on the two McKesson 120 Urine analyzers. D5413 TEST SYSTEMS, EQUIPMENT, INSTRUMENTS, REAGENT CFR(s): 493.1252(b) The laboratory must define criteria for those conditions that are essential for proper Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 2 -- storage of reagents and specimens, accurate and reliable test system operation, and test result reporting. The criteria must be consistent with the manufacturer's instructions, if provided. These conditions must be monitored and documented and, if applicable, include the following: (1) Water quality. (2) Temperature. (3) Humidity. (4) Protection of equipment and instruments from fluctuations and interruptions in electrical current that adversely affect patient test results and test reports. This STANDARD is not met as evidenced by: . Based on surveyor observation, manufacturer requirements, and confirmed in interview, the laboratory failed to monitor and document the temperature for the storage of 40 of 40 BD vacutainer tubes in use for patient testing since May 1st, 2022. The findings include: 1. In a tour of the facility on 5/10/2022 at 09:10 hours the surveyor observed a drawer full of the following 40 vacutainer blood collection tubes in administration pod B since May 1st, 2022: - 3 Red Top BD Vacutainer Tubes - Lot 106827, EXP 3/31/2023 - 37 Greiner bio-one K2E K2EDT Purple Top tubes - Lot B21123LJ, EXP 4/13/2023 2. Surveyor queried 5/10/2022 at 09:12 hours for the temperature records for pod B, and none was provided. 3. Review of manufacture instructions had the following temperature storage requirements. - BD Red Top Vacutainer Tubes: 4(degrees) Celsius(C) - 25(degrees) C - Greiner bio-one storage: 4 (degrees)C - 25(degrees) C 4. In an interview on 5/10/2022 at 09:15 hours, the practice manager stated that the laboratory was not monitoring temperatures in all areas where laboratory supplies were being stored. . -- 2 of 2 --

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access

Survey - January 19, 2022

Survey Type: Standard

Survey Event ID: 7ZZS11

Deficiency Tags: D0000 D3007 D5018 D5311 D5400 D5411 D5421 D5435 D5439 D5787 D6033 D6035 D6040 D6046 D6046 D0000 D3007 D5018 D5311 D5400 D5411 D5421 D5435 D5439 D5461 D5461 D5787 D6033 D6035 D6040

Summary:

Summary Statement of Deficiencies D0000 The laboratory was found to be out of compliance based on the following CONDITION LEVEL DEFICIENCIES: D3000 - 42 C.F.R. 493.1101 Condition: Facility Administration; D3000 FACILITY ADMINISTRATION CFR(s): 493.1100 Each laboratory that performs nonwaived testing must meet the applicable requirements under 493.1101 through 493.1105, unless HHS approves a procedure that provides equivalent quality testing as specified in Appendix C of the State Operations Manual (CMS Pub. 7). (a) Reporting of SARS-CoV-2 test results During the Public Health Emergency, as defined in 400.200 of this chapter, each laboratory that performs a test that is intended to detect SARS-CoV-2 or to diagnose a possible case of COVID-19 (hereinafter referred to as a "SARS-CoV-2 test") must report SARS-CoV-2 test results to the Secretary in such form and manner, and at such timing and frequency, as the Secretary may prescribe. This CONDITION is not met as evidenced by: Based on review of the QuickVue SARS Antigen test instructions for use, Abbott ID NOW test instructions for use, patient records, and interview with facility personnel, the laboratory failed to report negative SARS-COV-2 antigen test results as required for twenty-one of twenty-one days of testing performed between December 20, 2021 and January 19, 2022. The findings included: 1. Based on review of the QuickVue SARS Antigen test instructions for use (Ref 20387), under CONDITIONS OF AUTHORIZATION FOR THE LABORATORY AND PATIENT CARE SETTINGS, the instructions stated "Authorized laboratories using your product must have a process in place for reporting test results to healthcare providers and relevant public health authorities, as appropriate." 2. Based on review of the Abbott ID NOW test instructions for use ( Rev. 8, 2021/08), under Intended Use, the instructions stated, "Testing facilities within the United States and its territories are required to report all Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 10 -- results to the appropriate public health authorities." 3. Based on review of patient records, the laboratory performed 492 patient tests under SARS-COV-2 COVID19W /OPTIC and 34 patient tests with IADNA SARS-COV-2 COVID-19 AMPLIFIED PROBE TQ for twenty-one testing days between December 20, 2021 and January 19, 2022. 4. In an interview at 13:54 hours on January 19, 2022 hours in the office, the Practice Manager stated the laboratory had not reported negative SARS-COV-2 test results to HHS. The Practice Manager stated, "we send all of our positive tests to the state, but we don't even keep track of the negative tests." Key: SARS-COV-2 - severe acute respiratory syndrome coronavirus 2 HHS - Health and Human Services D3007 FACILITIES CFR(s): 493.1101(b) The laboratory must have appropriate and sufficient equipment, instruments, reagents, materials, and supplies for the type and volume of testing it performs. This STANDARD is not met as evidenced by: - Based on surveyor observation, review of laboratory policy, instrument manual, review of the Center for Medicare and Medicaid Services (CMS) form 116, and confirmed in interview, the laboratory failed to have an appropriate centrifuge for the centrifugation of urine sediment for one of one provider performed microscopy (PPM) test; "Urine Microscopy". The findings include: 1. During a tour of the laboratory on 1 /19/2022 at 9:20 hours, surveyor observed the fixed speed "VanGuard Centrifuge" placed near a sink in a patient processing area, surrounded by open patient urine containers. Surveyor queried if this was the only centrifuge in use for the centrifugation of urine specimens and the practice manager stated, "I think so,". Further query with the medical assistants in the processing area confirmed that the "VanGuard Centrifuge" was the one used for urine centrifugation. 2. Review of the policy titled "Urine Microscopy", section "Procedure", bullet point 3. stated: "Centrifuge urine specimen according to manufacturer's directions at a RCF (relative centrifugal force) of 400 x g for 5-10 minutes. To calculate the RPM (rotations per minute) for a specific centrifuge ..." 3. Review of the "VanGuard Compact Centrifuge Operator Manual", section "VanGuard Centrifuge Models", stated the following: "NOTE: ALL VANGUARD CENTRIFUGE MODELS PERFORM AT 3400 RPM FULLY LOADED AND RELATIVE CENTRIFUGAL FORCE 1318." 4. Review of the CMS116, section VII "PPM Testing" lists the estimated total annual volume for the PPM test "Urine Sediment Examinations" as 500. 5. In an interview on 1/19/2022 at 09:50 hours, the practice manager confirmed the laboratory did not have a centrifuge that could be used for the processing of urine specimens at 400 RCF. . D5018 URINALYSIS CFR(s): 493.1211 If the laboratory provides services in the subspecialty of Urinalysis, the laboratory must meet the requirements specified in 493.1230 through 493.1256, and 493.1281 through 493.1299. This CONDITION is not met as evidenced by: . Based on review of the laboratory's policy and procedures, procedure manuals, the Centers for Medicare and Medicaid Services (CMS) form 116, surveyor observation, and staff interview the laboratory failed to meet the requirements specified in in 493. -- 2 of 10 -- 1230 through 493.1256, and 493.1281 through 493.1299 for the subspecialty of Urinalysis. The findings included: 1. The laboratory failed to follow its own policy for specimen processing for the centrifugation of urine sediment for one of one provider performed microscopy (PPM) test; "Urine Microscopy". Refer to D5311. 2. The laboratory failed to define a function check protocol to ensure the laboratory equipment was accurate for reliable testing for one of one centrifuged used for the Provider Performed Microscopy (PPM) test; Urine sediment examinations. Refer to D5435. 3. The laboratory failed to record patient results for one of one provider performed microscopy (PPM) test; Urine sediment examination. Refer to D5787. . D5311 SPECIMEN SUBMISSION, HANDLING, AND REFERRAL CFR(s): 493.1242(a) The laboratory must establish and follow written policies and procedures for each of the following, if applicable: (1) Patient preparation. (2) Specimen collection. (3) Specimen labeling, including patient name or unique patient identifier and, when appropriate, specimen source. (4) Specimen storage and preservation. (5) Conditions for specimen transportation. (6) Specimen processing. (7) Specimen acceptability and rejection. (8) Specimen referral. This STANDARD is not met as evidenced by: . Based on surveyor observation, review of laboratory policy, instrument manual, review of the Center for Medicare and Medicaid Services (CMS) form 116, and confirmed in interview, the laboratory failed to follow its own policy for specimen processing for the centrifugation of urine sediment for one of one provider performed microscopy (PPM) test; "Urine Microscopy". The findings include: 1. During a tour of the laboratory on 1/19/2022 at 9:20 hours, surveyor observed the fixed speed "VanGuard Centrifuge" placed near a sink in a patient processing area, surrounded by open patient urine containers. Surveyor queried if this was the only centrifuge in use for the centrifugation of urine specimens and the practice manager stated, "I think so,". Further query with the medical assistants in the processing area confirmed that the "VanGuard Centrifuge" was the one used for urine centrifugation. 2. Review of the policy titled "Urine Microscopy", section "Procedure", bullet point 3. stated: "Centrifuge urine specimen according to manufacturer's directions at a RCF (relative centrifugal force) of 400 x g for 5-10 minutes. To calculate the RPM (rotations per minute) for a specific centrifuge ..." 3. Review of the "VanGuard Compact Centrifuge Operator Manual", section "VanGuard Centrifuge Models", stated the following: "NOTE: ALL VANGUARD CENTRIFUGE MODELS PERFORM AT 3400 RPM FULLY LOADED AND RELATIVE CENTRIFUGAL FORCE 1318." 4. Review of the CMS116, section VII "PPM Testing" lists the estimated total annual volume for the PPM test "Urine Sediment Examinations" as 500. 5. In an interview on 1/19/2022 at 09:50 hours, the practice manager confirmed the laboratory was not following its own policy for the processing of urine specimens for the urine sediment examinations. . D5400 ANALYTIC SYSTEMS CFR(s): 493.1250 Each laboratory that performs nonwaived testing must meet the applicable analytic systems requirements in 493.1251 through 493.1283, unless HHS approves a procedure, specified in Appendix C of the State Operations Manual (CMS Pub.7), that provides equivalent quality testing. The laboratory must monitor and evaluate the -- 3 of 10 -- overall quality of the analytic systems and correct identified problems as specified in 493.1289 for each specialty and subspecialty of testing performed. This CONDITION is not met as evidenced by: Based on quality control records, reagent records, laboratory policies and procedures, manufacturer instructions,operator's manuals, and staff interview, the laboratory failed to meet the analytic system requirements in 493.1251 through 493.1283. The findings included: 1. The laboratory failed to follow the manufacturer's instructions for the processing of Interpretive Program (IP) messages on the Sysmex XM-330 hematology analyzer, for eight of ten patient complete blood cell (CBC) tests reviewed in January 2022. Refer to D5411. 2. The laboratory failed to verify the patient normal ranges for complete blood cell (CBC) tests prior to performing patient testing for the Sysmex XM-330 hematology analyzer. Refer to D5421. 3. The laboratory failed to follow manufacturer calibration instructions for forty-one of forty-one months for the Sysmex XN-330 hematology analyzer, since its installation in August 2018. Refer to D5439. 4. The laboratory failed to perform QC after a change of reagent for the Sysmex XN-330 for sixty-seven of sixty-seven reagent changes reviewed in 2020 and 2021. Refer to D5461. D5411 TEST SYSTEMS, EQUIPMENT, INSTRUMENTS, REAGENT CFR(s): 493.1252(a) Test systems must be selected by the laboratory. The testing must be performed following the manufacturer's instructions and in a manner that provides test results within the laboratory's stated performance specifications for each test system as determined under 493.1253. This STANDARD is not met as evidenced by: . Based on review of manufacturer's instruction, patient printouts, and confirmed in interview, the laboratory failed to follow the manufacturer's instructions for the processing of Interpretive Program (IP) messages on the Sysmex XM-330 hematology analyzer, for eight of ten patient complete blood cell (CBC) tests reviewed in January 2022. The findings include: 1. Review of the "XN-530/XN-430/XN-330 Troubleshooting" manual (BW212660), chapter 5 "IP Messages", stated the following: a: Section 5.1 "Overview of IP Messages" had the following statement: "A Positive or Error Judgement indicated the possibility of an abnormality. It is not a diagnosis of the patient. If a Positive or Error judgement occurs, check the data and repeat the analysis, or examine carefully in accordance with the protocol of your laboratory. b: Section 5.2 "Lists of IP messages" had the following statement: "Message types" There are 2 types of IP messages that may be displayed for WBC, RBC/RET, and PLT: abnormal messages and suspect messages. Abnormal messages: Indicates that the sample is clearly abnormal. Suspect messages: Indicates a possibility that the sample is abnormal. "Positive/Negative judgment" [Positive] Indicates that a blood cell analysis value or cell morphology exceeds the preset criteria for the IP message (abnormal sample). A Positive judgement is classified into the 3 types shown below. - [Diff. Abnormal]: Indicates an abnormal blood cell differentiation value. - [Morph. Abnormal]: Indicates an abnormal cell morphology. - [Count Abnormal]: Indicates an abnormal blood cell count. [Negative] Indicates that there were no analysis errors or abnormalities, and that there is no IP message (normal sample). 2. Review of patient printouts from the Sysmex XN-330 has the following eight of ten patients with positive IP messages that were not repeated or assessed -- 4 of 10 -- before reporting in January 2022. 1/3/2022 Patient ID: 0000016884 - Positive: Diff. 1 /4/2022 Patient ID: 45087 - Positive: Morph Patient ID: 0000015521 - Positive: Morph. Patient ID: 00000000479 - Positive: Diff. Morph. 1/6/2022 Patient ID: 0000015907 - Positive: Diff. 1/11/2022 Patient ID: 0000004334 - Positive: Morph. 1 /12/2022 Patient ID: 0000015750 - Positive: Diff. Morph. Count Patient ID: 47666 - Positive: Diff 3. In an interview on 1/19/2022 at 10:40 hours in the laboratory, the practice manager confirmed that it was not policy to rerun CBC results with positive IP messages as defined by the manufacturer. Key: WBC - White blood cell RBC - Red blood cell RET - Reticulocyte PLT - Platelet Diff. - Differential Morph. - Morphology . D5421 ESTABLISHMENT AND VERIFICATION OF PERFORMANCE CFR(s): 493.1253(b)(1) Each laboratory that introduces an unmodified, FDA-cleared or approved test system must do the following before reporting patient test results: (1)(i) Demonstrate that it can obtain performance specifications comparable to those established by the manufacturer for the following performance characteristics: (1)(i)(A) Accuracy. (1)(i) (B) Precision. (1)(i)(C) Reportable range of test results for the test system. (1)(ii) Verify that the manufacturer's reference intervals (normal values) are appropriate for the laboratory's patient population. This STANDARD is not met as evidenced by: Based on review of the manufacturer instructions, verification study records, laboratory procedures, and interview with facility personnel, the laboratory failed to verify the patient normal ranges for complete blood cell (CBC) tests prior to performing patient testing for the Sysmex XM-330 hematology analyzer for one of one hematology analzyers. The findings included: 1. Based on review of the Sysmex "Method Verification Manual" (Rev 2, Dec 2016), in Section 3 Method Verification Protocols, the document stated "It is the customer's responsibility to perform additional studies, following the requirements of their accrediting agency. The follow protocols are provided: Correlation Studies, Sensitivity Study, Reference Range Verification, Stability Study, Mixing Study". In Section 5 of the Application Manual, under Reference Range Study, the document stated, "Reference ranges describe analyte levels associated with person who are considered "healthy". Multiple factors can influence a reference range including, population selection (geography and demographics) and, specimen collection and handling. Assuming the study that resulted in the current reference range was performed correctly, that the population has not changed, and the new analyzer is determined to be comparable (correlation), a study to verify the current reference ranges with the new analyzer is appropriate. The outcome of this study either verifies the current reference range or determines whether a new reference ranges needs to be established." 2. Based on review of the hematology analyzer verification studies, the Sysmex XN was approved for use on 8 /29/2018. 3. Based on review of the laboratory's policy, "Complete Blood Count: Whole Blood on the Sysmex XN-350/Xn-330 Automated Hematology Analyzer", under section VIII Report Results, the laboratory Adult Reference Ranges were listed. Examples: Hemoglobin (HGB) 11.7 - 17.1 Hematocrit (HCT) 35.0 - 50.0 4. Based on review of the Sysmex XN530/XN-430/XN-330 General Information manual, on page 5-19, the manufacturer provided the following "Reference intervals (normal population reference ranges) were developed using normal individuals. The range for each parameter is calculated for 95% confidence intervals. The table below shows the Normal Population Reference Ranges". Examples: Female Hemoglobin (HGB) 11.2 - -- 5 of 10 -- 15.7 Female Hematocrit (HCT) 34.1 to 44.9 Male Hemoglobin (HGB) 13.7 to 17.5 Male Hematocrit (HCT) 40.1 to 51.0 5. In an interview at 14:56 hours on 1/19/2022, the surveyor requested documentation of the verification of patient normal ranges for the Sysmex XN hematology analyzer. The Practice Manager stated the only studies that were performed were the ones that the Sysmex representatives performed at the time of installation. D5435 MAINTENANCE AND FUNCTION CHECKS CFR(s): 493.1254(b)(2) For equipment, instruments, or test systems developed in-house, commercially available and modified by the laboratory, or maintenance and function check protocols are not provided by the manufacturer, the laboratory must: (i) Define a function check protocol that ensures equipment, instrument, and test system performance that is necessary for accurate and reliable test results and test result reporting. (ii) Perform and document the function checks, including background or baseline checks, specified in paragraph (b)(2)(i) of this section. Function checks must be within the laboratory's established limits before patient testing is conducted. This STANDARD is not met as evidenced by: . Based on review of the operator's manual and confirmed in interview, the laboratory failed to define a function check protocol to ensure the laboratory equipment was accurate for reliable testing for one of one centrifuged used for the Provider Performed Microscopy (PPM) test; Urine sediment examinations. The findings include: 1. Review of the "VanGuard Compact Centrifuge Operator Manual", section "VanGuard Centrifuge Models", stated the following: "NOTE: ALL VANGUARD CENTRIFUGE MODELS PERFORM AT 3400 RPM FULLY LOADED AND RELATIVE CENTRIFUGAL FORCE 1318." 2. On 1/19/2022 at 10:40 hours, surveyor queried the practice manager for documentation of maintenance, timer checks, and tachometer checks for the VanGuard centrifuge in use for the urine sediment examination. The practice manager stated that no maintenance or function checks had ever been performed on the centrifuge. . D5439 CALIBRATION AND CALIBRATION VERIFICATION CFR(s): 493.1255(b) Unless otherwise specified in this subpart, for each applicable test system the laboratory must do the following: Perform and document calibration verification procedure - (b)(1) Following the manufacturer's calibration verification instructions; (b)(2) Using the criteria verified or established by the laboratory under 493.1253(b)(3) -- (b)(2)(i) Including the number, type, and concentration of the materials, as well as acceptable limits for calibration verification; and (b)(2)(ii) Including at least a minimal (or zero) value, a mid-point value, and a maximum value near the upper limit of the range to verify the laboratory's reportable range of test results for the test system; and (b)(3) At least once every 6 months and whenever any of the following occur: (b)(3)(i) A complete change of reagents for a procedure is introduced, unless the laboratory can demonstrate that changing reagent lot numbers does not affect the range used to report patient test results, and control values are not adversely affected by reagent lot number changes. (b)(3)(ii) There is major preventive maintenance or replacement of critical parts that may influence test performance. (b)(3)(iii) Control materials reflect an unusual trend or shift, or are outside of the laboratory's acceptable limits, and other means of assessing and correcting unacceptable control values fail to -- 6 of 10 -- identify and correct the problem. (b)(3)(iv) The laboratory's established schedule for verifying the reportable range for patient test results requires more frequent calibration verification. This STANDARD is not met as evidenced by: . Based on review of procedure, laboratory documents, and confirmed in interview, the laboratory failed to follow manufacturer calibration instructions for forty-one of forty-one months for the Sysmex XN-330 hematology analyzer, since its installation in August 2018. The findings include: 1. Review of the "Sysmex XN-350/330 CLSI Procedure" (1236-LLS), section IV. "Calibration and Precision" stated the following: "The laboratory must verify calibration every six months or on an "as-needed" basis to ensure the accuracy of system." 2. Review of the "Sysmex Service Maintenance Certificate" for a contract date range of 8/16/2019 to 8/15/2023 has the following date for calibration of the Sysmex XN-330: Calibration - Completed - 8/17/2018 3. In an interview on 1/19/2022 at 1340, the practice manager confirmed that calibration was not performed every six months on the Sysmex XN-330 hematology analyzer, and that they "thought the Sysmex service technicians were performing it for them". . D5461 CONTROL PROCEDURES CFR(s): 493.1256(d)(6)(g) Unless CMS Approves a procedure, specified in Appendix C of the State Operations Manual (CMS Pub. 7), that provides equivalent quality testing, the laboratory must-- Perform control material testing as specified in this paragraph before resuming patient testing when a complete change of reagents is introduced; major preventive maintenance is performed; or any critical part that may influence test performance is replaced. (g) The laboratory must document all control procedures performed. This STANDARD is not met as evidenced by: . Based on review of manufacturer's instructions, review of reagent change logs, review of the Centers for Medicare and Medicaid Services (CMS) form 116, and confirmed in interview, the laboratory failed to perform QC after a change of reagent for the Sysmex XN-330 for sixty-seven of sixty-seven reagent changes reviewed in 2020 and 2021. The findings include: 1. Review of the "XN-L Series XN-530/XN-430 /XN-330 Basic Operation" manual, chapter 3 "Performing Quality Control (QC)", section 3.2.2 "When QC analysis is performed" stated: "QC is performed at the following times. - Before sample analysis - After replacement/replenishment of reagents - After instrument maintenance - When there is a concern about the accuracy of analysis values" 2. Review of the reagent change log for January through December 2020, and January through December 2021 list the following sixty-seven times a reagent was changed on the Sysmex XN-330 hematology analyzer. 2020 - January through December Sulfolyser - 8 times CellPack DCL - 13 times Fluorocell WDF - 6 times Lysercell WDF - 3 times 2021 - January through December Sulfolyser - 8 times CellPack DCL - 16 times Fluorocell WDF - 9 times Lysercell WDF -4 times 3. Review of the CMS-116 section VII. "Non-Waived Testing" list the total annual volume for the specialty Hematology at 2,500. 4. In an interview at 1/19/2022 at 13: 30 hours the practice manager confirmed that QC was not being performed after reagent change on the Sysmex XM-330 hematology analyzer. . D5787 TEST RECORDS CFR(s): 493.1283(a) -- 7 of 10 -- The laboratory must maintain an information or record system that includes the following: (a)(1) The positive identification of the specimen. (a)(2) The date and time of specimen receipt into the laboratory. (a)(3) The condition and disposition of specimens that do not meet the laboratory's criteria for specimen acceptability. (a)(4) The records and dates of all specimen testing, including the identity of the personnel who performed the test(s). This STANDARD is not met as evidenced by: . Based on review of the Centers for Medicare and Medicaid Services (CMS) form 116, review of laboratory policy, and confirmed in interview, the laboratory failed to record patient results for one of one provider performed microscopy (PPM) test; Urine sediment examination. The findings include: 1: Review of the CMS-116 section VII. "PPM Testing" lists the following PPM procedure being performed: "Urine sediment examinations" With an estimated total annual test volume of 500. 2. Review of laboratory policy titled "Urine Microscopy" signed by the lab director 10/30/2018, section "Reporting Results" states: "Results should be recorded in the patient record done with the date/time the sample was collected and the name of the provider performing the examination." 3. On 1/19/2022 at 10:20 hours, in the breakroom, surveyor queried the practice manager for urine microscopic patient final reports. The practice manager informed the surveyor that they could not provide any patient final reports, because they were not reported, and that the providers "just looked at the urine sediment to asses the next step of treatment" for the patient. 4. In an interview on 1/19/2022 at 15:00, in the practice managers office, the practice manager confirmed that the urinalysis sediment exams we're being performed to make clinical decisions, but not reported in the patient's chart. . D6033 TECHNICAL CONSULTANT-MODERATE COMPEXITY CFR(s): 493.1409 The laboratory must have a technical consultant who meets the qualification requirements of 493.1411 of this subpart and provides technical oversight in accordance with 493.1413 of this subpart. This CONDITION is not met as evidenced by: Based on review of testing personnel competency assessments and hematology analzyer verification studies, one of two the technical consultants failed to provide technical and scientific oversight of the laboratory for two of two years. Refer to D6035, D6040, and D6046. D6035 TECHNICAL CONSULTANT QUALIFICATIONS CFR(s): 493.1411 (a) The technical consultant must be qualified and must possess a current license issued by the State in which the laboratory is located, if such licensing is required. (b) The technical consultant must-- (b)(1)(i) Be a doctor of medicine or doctor of osteopathy licensed to practice medicine or osteopathy in the State in which the laboratory is located; and (b)(1)(ii) Be certified in anatomic or clinical pathology, or both, by the American Board of Pathology or the American Osteopathic Board of Pathology or possess qualifications that are equivalent to those required for such certification; or (b)(2)(i) Be a doctor of medicine, doctor of osteopathy, or doctor of -- 8 of 10 -- podiatric medicine licensed to practice medicine, osteopathy, or podiatry in the State in which the laboratory is located; and (b)(2)(ii) Have at least one year of laboratory training or experience, or both in non-waived testing, in the designated specialty or subspecialty areas of service for which the technical consultant is responsible (for example, physicians certified either in hematology or hematology and medical oncology by the American Board of Internal Medicine are qualified to serve as the technical consultant in hematology); or (b)(3)(i) Hold an earned doctoral or master's degree in a chemical, physical, biological or clinical laboratory science or medical technology from an accredited institution; and (b)(3)(ii) Have at least one year of laboratory training or experience, or both in non-waived testing, in the designated specialty or subspecialty areas of service for which the technical consultant is responsible; or (b)(4)(i) Have earned a bachelor's degree in a chemical, physical or biological science or medical technology from an accredited institution; and (b)(4)(ii) Have at least 2 years of laboratory training or experience, or both in non-waived testing, in the designated specialty or subspecialty areas of service for which the technical consultant is responsible. Note: The technical consultant requirements for "laboratory training or experience, or both" in each specialty or subspecialty may be acquired concurrently in more than one of the specialties or subspecialties of service, excluding waived tests. For example, an individual who has a bachelor's degree in biology and additionally has documentation of 2 years of work experience performing tests of moderate complexity in all specialties and subspecialties of service, would be qualified as a technical consultant in a laboratory performing moderate complexity testing in all specialties and subspecialties of service. This STANDARD is not met as evidenced by: Based on review of the laboratory personnel credentials and interview with facility personnel, one of two Technical Consultants did not meet the education and experience requiments to perform duties of a Technical Consultant for two of two years. The findings included: 1. Based on review of the laboratory personnel credentials, the individual performing testing personnel competeny assessments did not have documentation of education to qualify at 493.1411. 2. In an interview at 11: 56 hours on 1/19/2022, the Practice Manager confirmed the signatures attesting to the testing personnel competency were that of the previous Practice Manager, listed as Techincal Consultant 2 and that this individual did not have the qualification necessary to qualify for Technical Consultant. D6040 TECHNICAL CONSULTANT RESPONSIBILITIES CFR(s): 493.1413(b)(2) The technical consultant is responsible for-- (b)(2) Verification of the test procedures performed and the establishment of the laboratory's test performance characteristics, including the precision and accuracy of each test and test system. This STANDARD is not met as evidenced by: Based on review of the manufacturer instructions, verification study records, patient final reports, and interview with facility personnel, the Technical Consultant failed to verify the patient normal ranges for complete blood cell (CBC) tests prior to performing patient testing for the Sysmex XM-330 hematology analyzer for one of one hematology analzyers. Refer to D5421. D6046 TECHNICAL CONSULTANT RESPONSIBILITIES -- 9 of 10 -- CFR(s): 493.1413(b)(8) (b) The technical consultant is responsible for-- (b)(8) Evaluating the competency of all testing personnel and assuring that the staff maintain their competency to perform test procedures and report test results promptly, accurately and proficiently. This STANDARD is not met as evidenced by: Based on review of policies and procedures, testing personnel competency assessments, and interview with facility personnel, the Technical Consultant failed to assess the competency of seven of seven testing personnel for 2020 and 2021. The findings included: 1. Based on review of the laboratory policy "East Texas Family Medicine, PA Quality Assurance Policies and Procedures" under 6. PERSONNEL ASSESSMENT" stated the following: "The Laboratory Director will use the results of Proficiency testing and direct observation to perform an annual evaluation of all testing personnel to ensure competency in job performance" 2. Based on a review of seven of seven testing personnel, all competency assessments for 2020 and 2021 were signed by the individual listed as Technical Consultant 2 on the CMS-209 Laboratory Personnel Report. The individual listed as Technical Consultant 2 does not meet the qualifications for a Technical Consultant at 493. 1411. Refer to D6035. 3. In an interview at 11:56 hours on 1/19/2022, the Practice Manager confirmed the signatures attesting to the testing personnel competency were that of the previous Practice Manager, listed as Techincal Consultant 2 and that this individual did not have the qualification necessary to qualify for Technical Consultant at 493.1411. -- 10 of 10 --

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access

Survey - January 19, 2022

Survey Type: Special

Survey Event ID: US0P11

Deficiency Tags: D0000 D3000 D0000 D3000

Summary:

Summary Statement of Deficiencies D0000 A routine recertification survey conducted on January 19, 2022 found the laboratory out of compliance with the following CONDITION LEVEL DEFICIENCIES: D5018 - 42 C.F.R. 493.1211 Condition: Urinalysis; D5400 - 42 C.F.R. 493.1250 Condition: Analytic systems; D6033 - 42 C.F.R. 493.1409 Condition: Laboratories performing moderate complexity testing; technical consultant; D3000 FACILITY ADMINISTRATION CFR(s): 493.1100 Each laboratory that performs nonwaived testing must meet the applicable requirements under 493.1101 through 493.1105, unless HHS approves a procedure that provides equivalent quality testing as specified in Appendix C of the State Operations Manual (CMS Pub. 7). (a) Reporting of SARS-CoV-2 test results During the Public Health Emergency, as defined in 400.200 of this chapter, each laboratory that performs a test that is intended to detect SARS-CoV-2 or to diagnose a possible case of COVID-19 (hereinafter referred to as a "SARS-CoV-2 test") must report SARS-CoV-2 test results to the Secretary in such form and manner, and at such timing and frequency, as the Secretary may prescribe. This CONDITION is not met as evidenced by: Based on review of the QuickVue SARS Antigen test instructions for use, Abbott ID NOW test instructions for use, patient records, and interview with facility personnel, the laboratory failed to report negative SARS-COV-2 antigen test results as required for twenty-one of twenty-one days of testing performed between December 20, 2021 and January 19, 2022. The findings included: 1. Based on review of the QuickVue SARS Antigen test instructions for use (Ref 20387), under CONDITIONS OF AUTHORIZATION FOR THE LABORATORY AND PATIENT CARE SETTINGS, the instructions stated "Authorized laboratories using your product must have a process in place for reporting test results to healthcare providers and relevant public health authorities, as appropriate." 2. Based on review of the Abbott ID NOW test Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 10 -- instructions for use ( Rev. 8, 2021/08), under Intended Use, the instructions stated, "Testing facilities within the United States and its territories are required to report all results to the appropriate public health authorities." 3. Based on review of patient records, the laboratory performed 492 patient tests under SARS-COV-2 COVID19W /OPTIC and 34 patient tests with IADNA SARS-COV-2 COVID-19 AMPLIFIED PROBE TQ for twenty-one testing days between December 20, 2021 and January 19, 2022. 4. In an interview at 13:54 hours on January 19, 2022 hours in the office, the Practice Manager stated the laboratory had not reported negative SARS-COV-2 test results to HHS. The Practice Manager stated, "we send all of our positive tests to the state, but we don't even keep track of the negative tests." Key: SARS-COV-2 - severe acute respiratory syndrome coronavirus 2 HHS - Health and Human Services D3007 FACILITIES CFR(s): 493.1101(b) The laboratory must have appropriate and sufficient equipment, instruments, reagents, materials, and supplies for the type and volume of testing it performs. This STANDARD is not met as evidenced by: - Based on surveyor observation, review of laboratory policy, instrument manual, review of the Center for Medicare and Medicaid Services (CMS) form 116, and confirmed in interview, the laboratory failed to have an appropriate centrifuge for the centrifugation of urine sediment for one of one provider performed microscopy (PPM) test; "Urine Microscopy". The findings include: 1. During a tour of the laboratory on 1 /19/2022 at 9:20 hours, surveyor observed the fixed speed "VanGuard Centrifuge" placed near a sink in a patient processing area, surrounded by open patient urine containers. Surveyor queried if this was the only centrifuge in use for the centrifugation of urine specimens and the practice manager stated, "I think so,". Further query with the medical assistants in the processing area confirmed that the "VanGuard Centrifuge" was the one used for urine centrifugation. 2. Review of the policy titled "Urine Microscopy", section "Procedure", bullet point 3. stated: "Centrifuge urine specimen according to manufacturer's directions at a RCF (relative centrifugal force) of 400 x g for 5-10 minutes. To calculate the RPM (rotations per minute) for a specific centrifuge ..." 3. Review of the "VanGuard Compact Centrifuge Operator Manual", section "VanGuard Centrifuge Models", stated the following: "NOTE: ALL VANGUARD CENTRIFUGE MODELS PERFORM AT 3400 RPM FULLY LOADED AND RELATIVE CENTRIFUGAL FORCE 1318." 4. Review of the CMS116, section VII "PPM Testing" lists the estimated total annual volume for the PPM test "Urine Sediment Examinations" as 500. 5. In an interview on 1/19/2022 at 09:50 hours, the practice manager confirmed the laboratory did not have a centrifuge that could be used for the processing of urine specimens at 400 RCF. . D5018 URINALYSIS CFR(s): 493.1211 If the laboratory provides services in the subspecialty of Urinalysis, the laboratory must meet the requirements specified in 493.1230 through 493.1256, and 493.1281 through 493.1299. This CONDITION is not met as evidenced by: . Based on review of the laboratory's policy and procedures, procedure manuals, the -- 2 of 10 -- Centers for Medicare and Medicaid Services (CMS) form 116, surveyor observation, and staff interview the laboratory failed to meet the requirements specified in in 493. 1230 through 493.1256, and 493.1281 through 493.1299 for the subspecialty of Urinalysis. The findings included: 1. The laboratory failed to follow its own policy for specimen processing for the centrifugation of urine sediment for one of one provider performed microscopy (PPM) test; "Urine Microscopy". Refer to D5311. 2. The laboratory failed to define a function check protocol to ensure the laboratory equipment was accurate for reliable testing for one of one centrifuged used for the Provider Performed Microscopy (PPM) test; Urine sediment examinations. Refer to D5435. 3. The laboratory failed to record patient results for one of one provider performed microscopy (PPM) test; Urine sediment examination. Refer to D5787. . D5311 SPECIMEN SUBMISSION, HANDLING, AND REFERRAL CFR(s): 493.1242(a) The laboratory must establish and follow written policies and procedures for each of the following, if applicable: (1) Patient preparation. (2) Specimen collection. (3) Specimen labeling, including patient name or unique patient identifier and, when appropriate, specimen source. (4) Specimen storage and preservation. (5) Conditions for specimen transportation. (6) Specimen processing. (7) Specimen acceptability and rejection. (8) Specimen referral. This STANDARD is not met as evidenced by: . Based on surveyor observation, review of laboratory policy, instrument manual, review of the Center for Medicare and Medicaid Services (CMS) form 116, and confirmed in interview, the laboratory failed to follow its own policy for specimen processing for the centrifugation of urine sediment for one of one provider performed microscopy (PPM) test; "Urine Microscopy". The findings include: 1. During a tour of the laboratory on 1/19/2022 at 9:20 hours, surveyor observed the fixed speed "VanGuard Centrifuge" placed near a sink in a patient processing area, surrounded by open patient urine containers. Surveyor queried if this was the only centrifuge in use for the centrifugation of urine specimens and the practice manager stated, "I think so,". Further query with the medical assistants in the processing area confirmed that the "VanGuard Centrifuge" was the one used for urine centrifugation. 2. Review of the policy titled "Urine Microscopy", section "Procedure", bullet point 3. stated: "Centrifuge urine specimen according to manufacturer's directions at a RCF (relative centrifugal force) of 400 x g for 5-10 minutes. To calculate the RPM (rotations per minute) for a specific centrifuge ..." 3. Review of the "VanGuard Compact Centrifuge Operator Manual", section "VanGuard Centrifuge Models", stated the following: "NOTE: ALL VANGUARD CENTRIFUGE MODELS PERFORM AT 3400 RPM FULLY LOADED AND RELATIVE CENTRIFUGAL FORCE 1318." 4. Review of the CMS116, section VII "PPM Testing" lists the estimated total annual volume for the PPM test "Urine Sediment Examinations" as 500. 5. In an interview on 1/19/2022 at 09:50 hours, the practice manager confirmed the laboratory was not following its own policy for the processing of urine specimens for the urine sediment examinations. . D5400 ANALYTIC SYSTEMS CFR(s): 493.1250 Each laboratory that performs nonwaived testing must meet the applicable analytic systems requirements in 493.1251 through 493.1283, unless HHS approves a -- 3 of 10 -- procedure, specified in Appendix C of the State Operations Manual (CMS Pub.7), that provides equivalent quality testing. The laboratory must monitor and evaluate the overall quality of the analytic systems and correct identified problems as specified in 493.1289 for each specialty and subspecialty of testing performed. This CONDITION is not met as evidenced by: Based on quality control records, reagent records, laboratory policies and procedures, manufacturer instructions,operator's manuals, and staff interview, the laboratory failed to meet the analytic system requirements in 493.1251 through 493.1283. The findings included: 1. The laboratory failed to follow the manufacturer's instructions for the processing of Interpretive Program (IP) messages on the Sysmex XM-330 hematology analyzer, for eight of ten patient complete blood cell (CBC) tests reviewed in January 2022. Refer to D5411. 2. The laboratory failed to verify the patient normal ranges for complete blood cell (CBC) tests prior to performing patient testing for the Sysmex XM-330 hematology analyzer. Refer to D5421. 3. The laboratory failed to follow manufacturer calibration instructions for forty-one of forty-one months for the Sysmex XN-330 hematology analyzer, since its installation in August 2018. Refer to D5439. 4. The laboratory failed to perform QC after a change of reagent for the Sysmex XN-330 for sixty-seven of sixty-seven reagent changes reviewed in 2020 and 2021. Refer to D5461. D5411 TEST SYSTEMS, EQUIPMENT, INSTRUMENTS, REAGENT CFR(s): 493.1252(a) Test systems must be selected by the laboratory. The testing must be performed following the manufacturer's instructions and in a manner that provides test results within the laboratory's stated performance specifications for each test system as determined under 493.1253. This STANDARD is not met as evidenced by: . Based on review of manufacturer's instruction, patient printouts, and confirmed in interview, the laboratory failed to follow the manufacturer's instructions for the processing of Interpretive Program (IP) messages on the Sysmex XM-330 hematology analyzer, for eight of ten patient complete blood cell (CBC) tests reviewed in January 2022. The findings include: 1. Review of the "XN-530/XN-430/XN-330 Troubleshooting" manual (BW212660), chapter 5 "IP Messages", stated the following: a: Section 5.1 "Overview of IP Messages" had the following statement: "A Positive or Error Judgement indicated the possibility of an abnormality. It is not a diagnosis of the patient. If a Positive or Error judgement occurs, check the data and repeat the analysis, or examine carefully in accordance with the protocol of your laboratory. b: Section 5.2 "Lists of IP messages" had the following statement: "Message types" There are 2 types of IP messages that may be displayed for WBC, RBC/RET, and PLT: abnormal messages and suspect messages. Abnormal messages: Indicates that the sample is clearly abnormal. Suspect messages: Indicates a possibility that the sample is abnormal. "Positive/Negative judgment" [Positive] Indicates that a blood cell analysis value or cell morphology exceeds the preset criteria for the IP message (abnormal sample). A Positive judgement is classified into the 3 types shown below. - [Diff. Abnormal]: Indicates an abnormal blood cell differentiation value. - [Morph. Abnormal]: Indicates an abnormal cell morphology. - [Count Abnormal]: Indicates an abnormal blood cell count. [Negative] Indicates that there were no analysis errors or abnormalities, and that there is no IP message (normal -- 4 of 10 -- sample). 2. Review of patient printouts from the Sysmex XN-330 has the following eight of ten patients with positive IP messages that were not repeated or assessed before reporting in January 2022. 1/3/2022 Patient ID: 0000016884 - Positive: Diff. 1 /4/2022 Patient ID: 45087 - Positive: Morph Patient ID: 0000015521 - Positive: Morph. Patient ID: 00000000479 - Positive: Diff. Morph. 1/6/2022 Patient ID: 0000015907 - Positive: Diff. 1/11/2022 Patient ID: 0000004334 - Positive: Morph. 1 /12/2022 Patient ID: 0000015750 - Positive: Diff. Morph. Count Patient ID: 47666 - Positive: Diff 3. In an interview on 1/19/2022 at 10:40 hours in the laboratory, the practice manager confirmed that it was not policy to rerun CBC results with positive IP messages as defined by the manufacturer. Key: WBC - White blood cell RBC - Red blood cell RET - Reticulocyte PLT - Platelet Diff. - Differential Morph. - Morphology . D5421 ESTABLISHMENT AND VERIFICATION OF PERFORMANCE CFR(s): 493.1253(b)(1) Each laboratory that introduces an unmodified, FDA-cleared or approved test system must do the following before reporting patient test results: (1)(i) Demonstrate that it can obtain performance specifications comparable to those established by the manufacturer for the following performance characteristics: (1)(i)(A) Accuracy. (1)(i) (B) Precision. (1)(i)(C) Reportable range of test results for the test system. (1)(ii) Verify that the manufacturer's reference intervals (normal values) are appropriate for the laboratory's patient population. This STANDARD is not met as evidenced by: Based on review of the manufacturer instructions, verification study records, laboratory procedures, and interview with facility personnel, the laboratory failed to verify the patient normal ranges for complete blood cell (CBC) tests prior to performing patient testing for the Sysmex XM-330 hematology analyzer for one of one hematology analzyers. The findings included: 1. Based on review of the Sysmex "Method Verification Manual" (Rev 2, Dec 2016), in Section 3 Method Verification Protocols, the document stated "It is the customer's responsibility to perform additional studies, following the requirements of their accrediting agency. The follow protocols are provided: Correlation Studies, Sensitivity Study, Reference Range Verification, Stability Study, Mixing Study". In Section 5 of the Application Manual, under Reference Range Study, the document stated, "Reference ranges describe analyte levels associated with person who are considered "healthy". Multiple factors can influence a reference range including, population selection (geography and demographics) and, specimen collection and handling. Assuming the study that resulted in the current reference range was performed correctly, that the population has not changed, and the new analyzer is determined to be comparable (correlation), a study to verify the current reference ranges with the new analyzer is appropriate. The outcome of this study either verifies the current reference range or determines whether a new reference ranges needs to be established." 2. Based on review of the hematology analyzer verification studies, the Sysmex XN was approved for use on 8 /29/2018. 3. Based on review of the laboratory's policy, "Complete Blood Count: Whole Blood on the Sysmex XN-350/Xn-330 Automated Hematology Analyzer", under section VIII Report Results, the laboratory Adult Reference Ranges were listed. Examples: Hemoglobin (HGB) 11.7 - 17.1 Hematocrit (HCT) 35.0 - 50.0 4. Based on review of the Sysmex XN530/XN-430/XN-330 General Information manual, on page 5-19, the manufacturer provided the following "Reference intervals (normal population reference ranges) were developed using normal individuals. The range for -- 5 of 10 -- each parameter is calculated for 95% confidence intervals. The table below shows the Normal Population Reference Ranges". Examples: Female Hemoglobin (HGB) 11.2 - 15.7 Female Hematocrit (HCT) 34.1 to 44.9 Male Hemoglobin (HGB) 13.7 to 17.5 Male Hematocrit (HCT) 40.1 to 51.0 5. In an interview at 14:56 hours on 1/19/2022, the surveyor requested documentation of the verification of patient normal ranges for the Sysmex XN hematology analyzer. The Practice Manager stated the only studies that were performed were the ones that the Sysmex representatives performed at the time of installation. D5435 MAINTENANCE AND FUNCTION CHECKS CFR(s): 493.1254(b)(2) For equipment, instruments, or test systems developed in-house, commercially available and modified by the laboratory, or maintenance and function check protocols are not provided by the manufacturer, the laboratory must: (i) Define a function check protocol that ensures equipment, instrument, and test system performance that is necessary for accurate and reliable test results and test result reporting. (ii) Perform and document the function checks, including background or baseline checks, specified in paragraph (b)(2)(i) of this section. Function checks must be within the laboratory's established limits before patient testing is conducted. This STANDARD is not met as evidenced by: . Based on review of the operator's manual and confirmed in interview, the laboratory failed to define a function check protocol to ensure the laboratory equipment was accurate for reliable testing for one of one centrifuged used for the Provider Performed Microscopy (PPM) test; Urine sediment examinations. The findings include: 1. Review of the "VanGuard Compact Centrifuge Operator Manual", section "VanGuard Centrifuge Models", stated the following: "NOTE: ALL VANGUARD CENTRIFUGE MODELS PERFORM AT 3400 RPM FULLY LOADED AND RELATIVE CENTRIFUGAL FORCE 1318." 2. On 1/19/2022 at 10:40 hours, surveyor queried the practice manager for documentation of maintenance, timer checks, and tachometer checks for the VanGuard centrifuge in use for the urine sediment examination. The practice manager stated that no maintenance or function checks had ever been performed on the centrifuge. . D5439 CALIBRATION AND CALIBRATION VERIFICATION CFR(s): 493.1255(b) Unless otherwise specified in this subpart, for each applicable test system the laboratory must do the following: Perform and document calibration verification procedure - (b)(1) Following the manufacturer's calibration verification instructions; (b)(2) Using the criteria verified or established by the laboratory under 493.1253(b)(3) -- (b)(2)(i) Including the number, type, and concentration of the materials, as well as acceptable limits for calibration verification; and (b)(2)(ii) Including at least a minimal (or zero) value, a mid-point value, and a maximum value near the upper limit of the range to verify the laboratory's reportable range of test results for the test system; and (b)(3) At least once every 6 months and whenever any of the following occur: (b)(3)(i) A complete change of reagents for a procedure is introduced, unless the laboratory can demonstrate that changing reagent lot numbers does not affect the range used to report patient test results, and control values are not adversely affected by reagent lot number changes. (b)(3)(ii) There is major preventive maintenance or replacement of critical parts that may influence test performance. (b)(3)(iii) Control -- 6 of 10 -- materials reflect an unusual trend or shift, or are outside of the laboratory's acceptable limits, and other means of assessing and correcting unacceptable control values fail to identify and correct the problem. (b)(3)(iv) The laboratory's established schedule for verifying the reportable range for patient test results requires more frequent calibration verification. This STANDARD is not met as evidenced by: . Based on review of procedure, laboratory documents, and confirmed in interview, the laboratory failed to follow manufacturer calibration instructions for forty-one of forty-one months for the Sysmex XN-330 hematology analyzer, since its installation in August 2018. The findings include: 1. Review of the "Sysmex XN-350/330 CLSI Procedure" (1236-LLS), section IV. "Calibration and Precision" stated the following: "The laboratory must verify calibration every six months or on an "as-needed" basis to ensure the accuracy of system." 2. Review of the "Sysmex Service Maintenance Certificate" for a contract date range of 8/16/2019 to 8/15/2023 has the following date for calibration of the Sysmex XN-330: Calibration - Completed - 8/17/2018 3. In an interview on 1/19/2022 at 1340, the practice manager confirmed that calibration was not performed every six months on the Sysmex XN-330 hematology analyzer, and that they "thought the Sysmex service technicians were performing it for them". . D5461 CONTROL PROCEDURES CFR(s): 493.1256(d)(6)(g) Unless CMS Approves a procedure, specified in Appendix C of the State Operations Manual (CMS Pub. 7), that provides equivalent quality testing, the laboratory must-- Perform control material testing as specified in this paragraph before resuming patient testing when a complete change of reagents is introduced; major preventive maintenance is performed; or any critical part that may influence test performance is replaced. (g) The laboratory must document all control procedures performed. This STANDARD is not met as evidenced by: . Based on review of manufacturer's instructions, review of reagent change logs, review of the Centers for Medicare and Medicaid Services (CMS) form 116, and confirmed in interview, the laboratory failed to perform QC after a change of reagent for the Sysmex XN-330 for sixty-seven of sixty-seven reagent changes reviewed in 2020 and 2021. The findings include: 1. Review of the "XN-L Series XN-530/XN-430 /XN-330 Basic Operation" manual, chapter 3 "Performing Quality Control (QC)", section 3.2.2 "When QC analysis is performed" stated: "QC is performed at the following times. - Before sample analysis - After replacement/replenishment of reagents - After instrument maintenance - When there is a concern about the accuracy of analysis values" 2. Review of the reagent change log for January through December 2020, and January through December 2021 list the following sixty-seven times a reagent was changed on the Sysmex XN-330 hematology analyzer. 2020 - January through December Sulfolyser - 8 times CellPack DCL - 13 times Fluorocell WDF - 6 times Lysercell WDF - 3 times 2021 - January through December Sulfolyser - 8 times CellPack DCL - 16 times Fluorocell WDF - 9 times Lysercell WDF -4 times 3. Review of the CMS-116 section VII. "Non-Waived Testing" list the total annual volume for the specialty Hematology at 2,500. 4. In an interview at 1/19/2022 at 13: 30 hours the practice manager confirmed that QC was not being performed after reagent change on the Sysmex XM-330 hematology analyzer. . -- 7 of 10 -- D5787 TEST RECORDS CFR(s): 493.1283(a) The laboratory must maintain an information or record system that includes the following: (a)(1) The positive identification of the specimen. (a)(2) The date and time of specimen receipt into the laboratory. (a)(3) The condition and disposition of specimens that do not meet the laboratory's criteria for specimen acceptability. (a)(4) The records and dates of all specimen testing, including the identity of the personnel who performed the test(s). This STANDARD is not met as evidenced by: . Based on review of the Centers for Medicare and Medicaid Services (CMS) form 116, review of laboratory policy, and confirmed in interview, the laboratory failed to record patient results for one of one provider performed microscopy (PPM) test; Urine sediment examination. The findings include: 1: Review of the CMS-116 section VII. "PPM Testing" lists the following PPM procedure being performed: "Urine sediment examinations" With an estimated total annual test volume of 500. 2. Review of laboratory policy titled "Urine Microscopy" signed by the lab director 10/30/2018, section "Reporting Results" states: "Results should be recorded in the patient record done with the date/time the sample was collected and the name of the provider performing the examination." 3. On 1/19/2022 at 10:20 hours, in the breakroom, surveyor queried the practice manager for urine microscopic patient final reports. The practice manager informed the surveyor that they could not provide any patient final reports, because they were not reported, and that the providers "just looked at the urine sediment to asses the next step of treatment" for the patient. 4. In an interview on 1/19/2022 at 15:00, in the practice managers office, the practice manager confirmed that the urinalysis sediment exams we're being performed to make clinical decisions, but not reported in the patient's chart. . D6033 TECHNICAL CONSULTANT-MODERATE COMPEXITY CFR(s): 493.1409 The laboratory must have a technical consultant who meets the qualification requirements of 493.1411 of this subpart and provides technical oversight in accordance with 493.1413 of this subpart. This CONDITION is not met as evidenced by: Based on review of testing personnel competency assessments and hematology analzyer verification studies, one of two the technical consultants failed to provide technical and scientific oversight of the laboratory for two of two years. Refer to D6035, D6040, and D6046. D6035 TECHNICAL CONSULTANT QUALIFICATIONS CFR(s): 493.1411 (a) The technical consultant must be qualified and must possess a current license issued by the State in which the laboratory is located, if such licensing is required. (b) The technical consultant must-- (b)(1)(i) Be a doctor of medicine or doctor of osteopathy licensed to practice medicine or osteopathy in the State in which the laboratory is located; and (b)(1)(ii) Be certified in anatomic or clinical pathology, or both, by the American Board of Pathology or the American Osteopathic Board of -- 8 of 10 -- Pathology or possess qualifications that are equivalent to those required for such certification; or (b)(2)(i) Be a doctor of medicine, doctor of osteopathy, or doctor of podiatric medicine licensed to practice medicine, osteopathy, or podiatry in the State in which the laboratory is located; and (b)(2)(ii) Have at least one year of laboratory training or experience, or both in non-waived testing, in the designated specialty or subspecialty areas of service for which the technical consultant is responsible (for example, physicians certified either in hematology or hematology and medical oncology by the American Board of Internal Medicine are qualified to serve as the technical consultant in hematology); or (b)(3)(i) Hold an earned doctoral or master's degree in a chemical, physical, biological or clinical laboratory science or medical technology from an accredited institution; and (b)(3)(ii) Have at least one year of laboratory training or experience, or both in non-waived testing, in the designated specialty or subspecialty areas of service for which the technical consultant is responsible; or (b)(4)(i) Have earned a bachelor's degree in a chemical, physical or biological science or medical technology from an accredited institution; and (b)(4)(ii) Have at least 2 years of laboratory training or experience, or both in non-waived testing, in the designated specialty or subspecialty areas of service for which the technical consultant is responsible. Note: The technical consultant requirements for "laboratory training or experience, or both" in each specialty or subspecialty may be acquired concurrently in more than one of the specialties or subspecialties of service, excluding waived tests. For example, an individual who has a bachelor's degree in biology and additionally has documentation of 2 years of work experience performing tests of moderate complexity in all specialties and subspecialties of service, would be qualified as a technical consultant in a laboratory performing moderate complexity testing in all specialties and subspecialties of service. This STANDARD is not met as evidenced by: Based on review of the laboratory personnel credentials and interview with facility personnel, one of two Technical Consultants did not meet the education and experience requiments to perform duties of a Technical Consultant for two of two years. The findings included: 1. Based on review of the laboratory personnel credentials, the individual performing testing personnel competeny assessments did not have documentation of education to qualify at 493.1411. 2. In an interview at 11: 56 hours on 1/19/2022, the Practice Manager confirmed the signatures attesting to the testing personnel competency were that of the previous Practice Manager, listed as Techincal Consultant 2 and that this individual did not have the qualification necessary to qualify for Technical Consultant. D6040 TECHNICAL CONSULTANT RESPONSIBILITIES CFR(s): 493.1413(b)(2) The technical consultant is responsible for-- (b)(2) Verification of the test procedures performed and the establishment of the laboratory's test performance characteristics, including the precision and accuracy of each test and test system. This STANDARD is not met as evidenced by: Based on review of the manufacturer instructions, verification study records, patient final reports, and interview with facility personnel, the Technical Consultant failed to verify the patient normal ranges for complete blood cell (CBC) tests prior to performing patient testing for the Sysmex XM-330 hematology analyzer for one of one hematology analzyers. Refer to D5421. -- 9 of 10 -- D6046 TECHNICAL CONSULTANT RESPONSIBILITIES CFR(s): 493.1413(b)(8) (b) The technical consultant is responsible for-- (b)(8) Evaluating the competency of all testing personnel and assuring that the staff maintain their competency to perform test procedures and report test results promptly, accurately and proficiently. This STANDARD is not met as evidenced by: Based on review of policies and procedures, testing personnel competency assessments, and interview with facility personnel, the Technical Consultant failed to assess the competency of seven of seven testing personnel for 2020 and 2021. The findings included: 1. Based on review of the laboratory policy "East Texas Family Medicine, PA Quality Assurance Policies and Procedures" under 6. PERSONNEL ASSESSMENT" stated the following: "The Laboratory Director will use the results of Proficiency testing and direct observation to perform an annual evaluation of all testing personnel to ensure competency in job performance" 2. Based on a review of seven of seven testing personnel, all competency assessments for 2020 and 2021 were signed by the individual listed as Technical Consultant 2 on the CMS-209 Laboratory Personnel Report. The individual listed as Technical Consultant 2 does not meet the qualifications for a Technical Consultant at 493. 1411. Refer to D6035. 3. In an interview at 11:56 hours on 1/19/2022, the Practice Manager confirmed the signatures attesting to the testing personnel competency were that of the previous Practice Manager, listed as Techincal Consultant 2 and that this individual did not have the qualification necessary to qualify for Technical Consultant at 493.1411. -- 10 of 10 --

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access

Survey - October 23, 2018

Survey Type: Standard

Survey Event ID: XQ7N11

Deficiency Tags: D5401 D5407 D6066 D6066 D5401 D5407

Summary:

Summary Statement of Deficiencies D5401 PROCEDURE MANUAL CFR(s): 493.1251(a) A written procedures manual for all tests, assays, and examinations performed by the laboratory must be available to, and followed by, laboratory personnel. Textbooks may supplement but not replace the laboratory's written procedures for testing or examining specimens. This STANDARD is not met as evidenced by: . Based on review of laboratory procedures, confirmed by staff interview, the laboratory failed to utilize a written procedure for the microscopic examination of urine sediment. Findings: 1. Laboratory procedures were reviewed. In the course of an interview during the review on 10-23-2018, testing person 7 (CMS form 209) stated that providers performed microscopic urine sediment examinations on an estimated 50 patient specimens a year. A copy of the procedure was requested; none was available or could be provided during the survey. . D5407 PROCEDURE MANUAL CFR(s): 493.1251(d) Procedures and changes in procedures must be approved, signed, and dated by the current laboratory director before use. This STANDARD is not met as evidenced by: . Based on review of laboratory procedure for hematology testing using the Sysmex XN-330 analyzer, the laboratory director failed to approve, sign and date the procedure manual before use. Findings: 1. Laboratory procedure documentation for hematology testing was requested. In an interview at the site on 10-23-2018, testing person 7 stated that the manufacturer-supplied operator's manual was being used as a Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 2 -- procedure manual. 2. The operator's manual showed no evidence of laboratory director approval. No documentation of approval was available or could be provided during the survey. . D6066 TESTING PERSONNEL QUALIFICATIONS CFR(s): 493.1423(b)(4)(ii) Have documentation of training appropriate for the testing performed prior to analyzing patient specimens. This STANDARD is not met as evidenced by: . Based on review of testing personnel education and training documentation, confirmed by surveyor observation and staff interview, the laboratory failed to document training on the operation of the Sysmex XN-330 hematology analyzer for 6 of 7 testing personnel. 1. Laboratory personnel documentation was reviewed. A certificate of training for testing person 7 on the operation of the Sysmex XN-330 hematology analyzer, put in operation 08-19-2018, was found. For testing personnel 1 through 6, no evidence of training was found. 2. In an interview at the site on 10-23- 2018, testing person 7 stated that other testing personnel had attended training sessions provided by the instrument manufacturer, but documentary evidence was not available and could not be provided at the time of the survey. -- 2 of 2 --

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access