Exer Medical Corporation

CLIA Laboratory Citation Details

1
Total Citation
2
Total Deficiencyies
2
Unique D-Tags
CMS Certification Number 05D2093075
Address 19346 Nordhoff St, Northridge, CA, 91324
City Northridge
State CA
Zip Code91324
Phone(818) 727-2040

Citation History (1 survey)

Survey - July 2, 2024

Survey Type: Standard

Survey Event ID: E3QX11

Deficiency Tags: D6011 D3003

Summary:

Summary Statement of Deficiencies D3003 FACILITIES CFR(s): 493.1101(a)(2) The laboratory must be constructed, arranged, and maintained to ensure contamination of patient specimens, equipment, instruments, reagents, materials, and supplies is minimized. This STANDARD is not met as evidenced by: Based on surveyors' observation during the laboratory tour of the molecular testing lab and interview with the laboratory's technical consultant on July 2, 2024; it was determined that the laboratory failed to minimize contamination of patient specimens, equipment, and materials used during specimen processing for SARS-CoV-2 testing (COVID-19). Findings include: 1. During the laboratory tour at approximately 1:30 p. m. the surveyors observed the area assigned for COVID-19 sample processing and testing by the Sofia analyzer took place on an open bench with no protection from aerosols. In addition, computer entry and storage for other reagents and instruments were in the same testing area. 2. During an interview on July 2, 2024, at approximately 1:30 p.m. the TC confirmed that the molecular testing for the presence of COVID-19 laboratory failed to minimize contamination of patient specimens, equipment, and desk materials, when processing samples in the same area/room. 3. The laboratory's testing declaration form, signed by the laboratory director on July 2, 2024, stated that the laboratory performs approximately 1,000 waived test which include testing for COVID-19. D6011 LABORATORY DIRECTOR RESPONSIBILITIES CFR(s): 493.1407(e)(2) The laboratory director is responsible for the overall operation and administration of the laboratory, including the employment of personnel who are competent to perform test procedures, and record and report test results promptly, accurate, and proficiently Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 2 -- and for assuring compliance with the applicable regulations. (e) The laboratory director must-- (e)(2) and provide a safe environment in which employees are protected from physical, chemical, and biological hazards. This STANDARD is not met as evidenced by: Based on direct observation during the laboratory tour and interviews with technical consultant and laboratory personnel; it was determined that the laboratory director failed to provide a safe environment in which employees are protected from biological hazards and cross contamination of samples and equipment. See D3003. -- 2 of 2 --

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access