Forefront Dermatology, Sc

CLIA Laboratory Citation Details

2
Total Citations
4
Total Deficiencyies
4
Unique D-Tags
CMS Certification Number 01D0918760
Address 8573 County Road 64, Daphne, AL, 36526
City Daphne
State AL
Zip Code36526
Phone(251) 621-2244

Citation History (2 surveys)

Survey - June 25, 2025

Survey Type: Standard

Survey Event ID: CK9U11

Deficiency Tags: D5217 D5413 D5417

Summary:

Summary Statement of Deficiencies D5217 EVALUATION OF PROFICIENCY TESTING PERFORMANCE CFR(s): 493.1236(c)(1) At least twice annually, the laboratory must verify the accuracy of any test or procedure it performs that is not included in subpart I of this part. This STANDARD is not met as evidenced by: Based on reviews of the Quality Control (QC) MOHS Micrographic Surgery Peer Review (MSPR) records, the Proficiency Testing (PT) Policy and Procedure (P&P) manual, and an interview with Office Manager (OM), the laboratory failed to document PT accuracy verification, at least bi-annually. The surveyor noted one of the five bi-annual PT events reviewed was missing documentation of the peer review from the pathologist and the final accuracy verification review of the LD in 2024. The findings include: 1. A review of the 2023-2025 QC Mohs MSPR records revealed an incomplete documentation of accuracy verification assessments for 2024 PT 2 event, 3 cases. 2. A review of the "Quality Control for Mohs Micrographic" log sheets revealed the Mohs Tech completed the "Date", "Patient Name", "Mohs Number", and "Site" for three patients biannually. The surveyor noted the Mohs Tech also included the Mohs Surgeon's diagnosis with the "Site" information. The case slides and the sheet are sent to a second MOHS surgeon for "Peer Review". 3. A review of the "Quality Control for Mohs Micrographic" Bi-annual Event 2024 #2 sheet revealed the Peer Reviewer failed to document his assessment of the "Stain Technique"," Adequate Specimen Preparation" and the "Diagnostic Interpretation". The second Mohs surgeon only signed and dated the bottom of the form. There was no evidence of the Laboratory Director's review of the returned results to determine completeness or whether there were discrepancies in the any of the cases. 2. A review of the PT P&P for Mohs Micrographic Surgery cases revealed the following instructions were not followed by the laboratory based on the incomplete documentation of assessments and final review. ..."Three cases will be sent internally or externally to another Mohs surgeon to be evaluated for any inconsistencies in the quality and interpretation." ..." Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 3 -- Upon receipt of the pathology report from the Pathologist/Mohs surgeon diagnosis of the slide specimen will be match to the in-house diagnosis by the physician." 3. During the exit conference on 6-25-2025 at 5:30 PM, the OM confirmed the above findings. D5413 TEST SYSTEMS, EQUIPMENT, INSTRUMENTS, REAGENT CFR(s): 493.1252(b) (b) The laboratory must define criteria for those conditions that are essential for proper storage of reagents and specimens, accurate and reliable test system operation, and test result reporting. The criteria must be consistent with the manufacturer's instructions, if provided. These conditions must be monitored and documented and, if applicable, include the following: (b)(1) Water quality. (b)(2) Temperature. (b)(3) Humidity. (b)(4) Protection of equipment and instruments from fluctuations and interruptions in electrical current that adversely affect patient test results and test reports. This STANDARD is not met as evidenced by: Based on a review of the Room Temperature (RT) and Humidity logs, the laboratory analyzers' user manuals and an interview with the Office Manager (OM), the staff failed to document the RT and Humidity for four of the four days of patient testing in October 2023. The findings include: 1) A review of the RT and Humidity logs revealed the staff failed to monitor and document the RT and Humidity of the laboratory when patient testing was performed in October 2023 for the following days 12, 13, 19, and 20. 2) The Thermo Scientific Stainer, Avantik and Leica Cryostat user manuals revealed the following manufacturers' requirement. A) Themo Scientific Linistat Linear Stainer, "Temperature: The operating limits for the Linistat are +5C to +40C (+41F to +104F)". "Relative Humidity: The maximum relative humidity allowed is 80% RH up to 31C, decreasing linearly to 50% RH at 40C." B) Avantik Cryostat QS 11 "recommended an operating condition include room RT range of +5C to +35C (41F to 95F) and a maximum relative humidity of 60%." C) Leica Cryostat, "Operating range: Leica cryostats generally have an operating ambient temperature range of 18C to 35C, though some models like the CM1510 may have a range of 18C to 40C." Maximum relative humidity, the maximum recommended relative humidity for operation is generally 60%, non-condensing. 3) During the exit conference on 06- 25-2025 at 5:30 PM, the OM confirmed the above findings. D5417 TEST SYSTEMS, EQUIPMENT, INSTRUMENTS, REAGENT CFR(s): 493.1252(d) (d) Reagents, solutions, culture media, control materials, calibration materials, and other supplies must not be used when they have exceeded their expiration date, have deteriorated, or are of substandard quality. This STANDARD is not met as evidenced by: Based on direct observations, reviews of the Quality Control (QC) MOHS Tissue Staining Procedure logs, the patient Provider Performed Microscopy (PPM) logs and an interview with Office Manager (OM) and the MOHS Technician, the laboratory utilized three reagents after expiry for patient testing from the date of the last survey on 12-22-2022 through the current survey on 06-25-2025. The findings include: 1. During the laboratory tour with the OM, at approximately 1:44 PM, the surveyor -- 2 of 3 -- observed Potassium Hydroxide (KOH) reagent, Lot 1033 with an Expiration date of 02-02-2023. No open date was written on the bottle. The expired reagent was used for 16 patient PPM examinations from 02-23-2023 through 12-17-2024. The OM confirmed during the tour, the laboratory did not have any unexpired KOH solution on site. 2. Reviews of the QC Mohs Tissue Staining Procedure logs and patient logs revealed the following reagents were used for patient testing after expiration: A) Hematoxylin Lot 2230833 expired 11-10-2024 and was utilized for nine days after expiration. 135 patient specimens were stained using the expired reagent. B) Bluing Agent Lot 163928 expired 03-31-2023 and was utilized for 22 days after expiration. 242 patient specimens were stained using the expired reagent. C) Xylene Substitute Lot 1278321 expired 05-31-2023 and was utilized for 45 days after expiration. 495 patient specimens were stained using the expired reagent. 3. During the exit conference on 06-25-2025 at 5:30 PM, OM and the Mohs Technician confirmed the above findings. -- 3 of 3 --

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access

Survey - September 11, 2018

Survey Type: Standard

Survey Event ID: F2OH11

Deficiency Tags: D5473

Summary:

Summary Statement of Deficiencies D5473 CONTROL PROCEDURES CFR(s): 493.1256(e)(2)(g) (e) For reagent, media, and supply checks, the laboratory must do the following: (e) (2) Each day of use (unless otherwise specified in this subpart), test staining materials for intended reactivity to ensure predictable staining characteristics. Control materials for both positive and negative reactivity must be included, as appropriate. (g) The laboratory must document all control procedures performed. This STANDARD is not met as evidenced by: Based on a lack of documentation for the daily slide control quality, and an interview with Mohs Technician #1, the surveyor determined one of two Mohs Surgeons (Clinical Consultant #2) failed to assess and document the slide stain quality on 33 days of patient testing from September 2017 thru August 2018. The findings include: 1. During the entrance interview and review of the CMS Form-209 at approximately 11:30 AM on 9/11/2018, Mohs Technician #1 stated the facility has had a second physician (Clinical Consultant #2) performing Mohs surgical procedures since 9/15 /2017. [Mohs surgical procedures include on-site preparation, reading and interpretation of Histopathology slides from frozen section specimens.] Mohs Tech #1 explained the Laboratory Director scheduled his Mohs surgery cases every Monday, and Clinical Consultant (CC) #2 scheduled his cases on Thursdays or Fridays two to four days a month. 2. A review of the MOHS Procedure Manual in Section 5 under "Daily Quality Control" revealed, "A daily control is done each surgery day. The first slide of the day is evaluated by the physician for acceptable staining and microtomy ... '. 3. A review of the Histology records revealed documentation of the daily slide stain quality only on days when the Laboratory Director performed Mohs surgical procedures There were no records of daily slide quality control on days when the CC #2 read the Mohs Histopathology slides. 4. A review of the Mohs patient log revealed the CC #2 performed Mohs surgery on 84 patients on 8 days (9/15 thru 12/19/2017) in 2017, and on 393 patients on 25 days (1/16 thru 8/31/2018) in 2018, for a total of 33 Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 2 -- days with no documentation of daily slide quality control. 5. During an interview on 9 /11/2018 at 2:15 PM, Mohs Tech #1 confirmed the laboratory had no documentation of daily slide quality control for the CC #2. Thus the above noted findings were confirmed. SURVEYOR: Laura T. Williams, BS, MT (ASCP) Licensure and Certification Surveyor -- 2 of 2 --

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access