Franey Medical Laboratory

CLIA Laboratory Citation Details

3
Total Citations
19
Total Deficiencyies
14
Unique D-Tags
CMS Certification Number 22D0081111
Address 52 Mercantile Way, Mashpee, MA, 02649
City Mashpee
State MA
Zip Code02649
Phone(508) 888-7546

Citation History (3 surveys)

Survey - July 19, 2023

Survey Type: Standard

Survey Event ID: UBPH11

Deficiency Tags: D0000 D6127 D5439 D6128

Summary:

Summary Statement of Deficiencies D0000 A CLIA recertification survey was conducted for the Franey Medical Laboratory pursuant to the Clinical Laboratory Improvement Amendments (CLIA) of 1988 and CLIA regulations at 42 CFR 493. D5439 CALIBRATION AND CALIBRATION VERIFICATION CFR(s): 493.1255(b) Unless otherwise specified in this subpart, for each applicable test system the laboratory must do the following: Perform and document calibration verification procedure - (b)(1) Following the manufacturer's calibration verification instructions; (b)(2) Using the criteria verified or established by the laboratory under 493.1253(b)(3) -- (b)(2)(i) Including the number, type, and concentration of the materials, as well as acceptable limits for calibration verification; and (b)(2)(ii) Including at least a minimal (or zero) value, a mid-point value, and a maximum value near the upper limit of the range to verify the laboratory's reportable range of test results for the test system; and (b)(3) At least once every 6 months and whenever any of the following occur: (b)(3)(i) A complete change of reagents for a procedure is introduced, unless the laboratory can demonstrate that changing reagent lot numbers does not affect the range used to report patient test results, and control values are not adversely affected by reagent lot number changes. (b)(3)(ii) There is major preventive maintenance or replacement of critical parts that may influence test performance. (b)(3)(iii) Control materials reflect an unusual trend or shift, or are outside of the laboratory's acceptable limits, and other means of assessing and correcting unacceptable control values fail to identify and correct the problem. (b)(3)(iv) The laboratory's established schedule for verifying the reportable range for patient test results requires more frequent calibration verification. This STANDARD is not met as evidenced by: . Based on record review and interview with the Laboratory Director (LD) and General Supervisor (GS) on 7/19/2023, the laboratory failed to perform calibration Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 3 -- verifications every six months or, as appropriate, as evidenced by the following: The surveyor reviewed the quality control records for calendar years 2021, 2022, and 2023. The review revealed that calibration verifications of at least three points were not performed once every six months for three (3) out of three (3) analytes requiring calibration verification on the Beckman Coulter AU680 analyzer. Six-month calibration verifications for Ethyl Alcohol, Specific Gravity, and Urine Creatinine were performed on 8/24/2021 and 1/28/2022. The calibration verifications were not performed every six months. The LD and GS confirmed in an interview on 7/19/2023 at 12:00 P.M. that calibration verifications had not been performed every six months for three (3) out of three (3) analytes requiring calibration verification. The laboratory performs 137,800 toxicology tests, 13,000 urinalysis tests, and 26,000 routine chemistry tests annually. This is a repeat deficiency. . D6127 TECHNICAL SUPERVISOR RESPONSIBILITIES CFR(s): 493.1451(b)(9) The technical supervisor is responsible for evaluating and documenting the performance of individuals responsible for high complexity testing at least semiannually during the first year the individual tests patient specimens. This STANDARD is not met as evidenced by: . Based on record review and interview with the Laboratory Director (LD) and General Supervisor (GS) on 7/19/2023, the Technical Supervisor (TS) failed to evaluate and document the performance of individuals responsible for high complexity testing at least semiannually during the first year the individual tested patient specimens as evidenced by the following: The surveyor asked to review the personnel competency records. The review revealed that a semiannual competency evaluation was not performed and documented for the (1) out of one (1) Testing Person (TP) performing high complexity toxicology testing. The LD and GS confirmed in an interview on 7/19/2023 at 10:48 AM that the TS failed to perform and document a semiannual competency evaluation for one (1) out of one (1) TP in the first year of performing high complexity toxicology testing. The laboratory performs 137,800 toxicology tests annually. . D6128 TECHNICAL SUPERVISOR RESPONSIBILITIES CFR(s): 493.1451(b)(9) The technical supervisor is responsible for evaluating and documenting the performance of individuals responsible for high complexity testing at least annually after the first year, unless test methodology or instrumentation changes, in which case, prior to reporting patient test results, the individual's performance must be reevaluated to include the use of the new test methodology or instrumentation. This STANDARD is not met as evidenced by: . Based on record review and interview with the Laboratory Director (LD) and General Supervisor (GS) on 7/19/2023, the Technical Supervisor (TS) failed to evaluate and document the performance of individuals responsible for high complexity testing at least annually, after the first year the individual tested patient specimens as evidenced by the following: The surveyor asked to review the personnel competency records. The review revealed that a annual competency evaluation was not performed and documented for one (1) out of one (1) Testing Person (TP) -- 2 of 3 -- performing high complexity toxicology testing. The LD and GS confirmed in an interview on 7/19/2023 at 10:48 AM that the TS failed to perform and document an annual competency evaluation for one (1) out of one (1) TP in 2022 after the first year of performing high complexity toxicology testing. The laboratory performs 137,800 toxicology tests annually. . -- 3 of 3 --

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access

Survey - July 16, 2021

Survey Type: Standard

Survey Event ID: P6W411

Deficiency Tags: D0000 D5401 D6084 D2003 D5439

Summary:

Summary Statement of Deficiencies D0000 A CLIA recertification survey was conducted for the Franey Medical Lab, Inc laboratory pursuant to the Clinical Laboratory Improvement Amendments (CLIA) of 1988 and CLIA regulations at 42 CFR 493. D2003 ENROLLMENT CFR(s): 493.801(a)(2)(ii) For those tests performed by the laboratory that are not included in subpart I of this part, a laboratory must establish and maintain the accuracy of its testing procedures, in accordance with 493.1236(c)(1) This STANDARD is not met as evidenced by: Based on record review and interview on 7/16/21 with the technical supervisor (TS) and laboratory director (LD), the laboratory failed to establish and maintain the accuracy of its testing procedures in accordance with 493.1236(c)(1). Findings include: The surveyor reviewed proficiency testing (PT) records from calendar years 2019, 2020, and 2021on 7/16/21. The review revealed that the laboratory failed to establish and maintain accuracy of Ethylglucuronide (EtG) on the Olympus AU680 analyzer. The TS and LD confirmed in an interview on 7/16/21 at 11:15 AM that the laboratory failed to establish and maintain accuracy of EtG on the Olympus AU680 urine drug screen analyzer. The laboratory performs approximately 3,353 EtG screens annually. . D5401 PROCEDURE MANUAL CFR(s): 493.1251(a) A written procedures manual for all tests, assays, and examinations performed by the laboratory must be available to, and followed by, laboratory personnel. Textbooks may supplement but not replace the laboratory's written procedures for testing or examining specimens. Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 3 -- This STANDARD is not met as evidenced by: . Based on record review and interview on 7/16/21 with the technical supervisor (TS) and laboratory director (LD), the laboratory failed to follow policies and procedures in place for all tests, assays, and examinations performed by the laboratory as evidenced by the following: The surveyor reviewed the "Basic Operation of LC-MS Instruments" standard operating procedure (SOP) on 7/16/21. The SOP states "Maintenance: Semiannually (or on need basis): Preventative maintenance (P.M.) will be performed by Waters field service engineers under the terms of our service contracts." The surveyor asked to review the terms of their service contract with Waters for preventative maintenance (PM). The TS and LD were not able to produce a service contract. The surveyor toured the laboratory space on 7/16/21. The surveyor observed 4 Liquid Chromatography-Mass Spectrometers (LC/MS) utilized for urine toxicology confirmation testing. Each LC/MS was equipped with a sticker indicating when the next PM is due. 3 out of 4 LC/MS were overdue for their PM indicating "Next PM due" 8/2020, 9/2020, and 10/2020. The TS and LD confirmed in an interview on 7/16/21 at 2 PM that PMs on 3 out of 4 LC/MS were overdue and the laboratory did not have a current service contract per their SOP. . D5439 CALIBRATION AND CALIBRATION VERIFICATION CFR(s): 493.1255(b) Unless otherwise specified in this subpart, for each applicable test system the laboratory must do the following: Perform and document calibration verification procedure - (b)(1) Following the manufacturer's calibration verification instructions; (b)(2) Using the criteria verified or established by the laboratory under 493.1253(b)(3) -- (b)(2)(i) Including the number, type, and concentration of the materials, as well as acceptable limits for calibration verification; and (b)(2)(ii) Including at least a minimal (or zero) value, a mid-point value, and a maximum value near the upper limit of the range to verify the laboratory's reportable range of test results for the test system; and (b)(3) At least once every 6 months and whenever any of the following occur: (b)(3)(i) A complete change of reagents for a procedure is introduced, unless the laboratory can demonstrate that changing reagent lot numbers does not affect the range used to report patient test results, and control values are not adversely affected by reagent lot number changes. (b)(3)(ii) There is major preventive maintenance or replacement of critical parts that may influence test performance. (b)(3)(iii) Control materials reflect an unusual trend or shift, or are outside of the laboratory's acceptable limits, and other means of assessing and correcting unacceptable control values fail to identify and correct the problem. (b)(3)(iv) The laboratory's established schedule for verifying the reportable range for patient test results requires more frequent calibration verification. This STANDARD is not met as evidenced by: . Based on record review and interview on 7/16/21 with the technical consultant (TC) and laboratory director (LD), , the laboratory failed to perform calibration verifications as appropriate as evidenced by the following: The surveyor reviewed quality control (QC) records for calendar years 2019, 2020, and 2021 on 7/16/21. The review revealed that calibration verifications of at least 3 points were not performed every six months for specific gravity, urine creatinine, and alcohol on the Olympus -- 2 of 3 -- AU 680 urine drug screen analyzer. The TC and LD confirmed through interview on 7 /16/21 at 11:45 AM that calibration verifications of at least 3 points had not been performed at least once every six months. This is a repeat deficiency. . D6084 LABORATORY DIRECTOR RESPONSIBILITIES CFR(s): 493.1445(e)(2) The laboratory director must ensure that the physical plant and environmental conditions provide a safe environment in which employees are protected from physical, chemical, and biological hazards. This STANDARD is not met as evidenced by: . Based on observation, record review, and interview on 7/16/21 with the technical supervisor (TS) and laboratory director (LD), the LD failed to ensure that the physical plant and environmental conditions provided a safe environment in which employees were protected from physical, chemical, and biological hazards as evidenced by the following: The surveyor toured the laboratory space on 7/16/21. The surveyor observed maintenance documentation for the eye wash station. The documentation stated "Need to check each week." The latest documented eye wash check was 3/9/21. The TS stated that the person who checked the eye wash each week is no longer employed at the laboratory and no one took over the responsibility. The TS and LD confirmed in an interview on 7/16/21 at 1:30 PM that LD failed to ensure a safe environment. . -- 3 of 3 --

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access

Survey - November 19, 2018

Survey Type: Standard

Survey Event ID: 1XBI11

Deficiency Tags: D0000 D2003 D5400 D6018 D6120 D5411 D5421 D5439 D5805 D6091

Summary:

Summary Statement of Deficiencies D0000 A CLIA recertification survey was conducted for Franey Medical Laboratories, Inc. pursuant to the Clinical Laboratory Improvement Amendments (CLIA) of 1988 and CLIA regulations at 42 CFR 493. D2003 ENROLLMENT CFR(s): 493.801(a)(2)(ii) For those tests performed by the laboratory that are not included in subpart I of this part, a laboratory must establish and maintain the accuracy of its testing procedures, in accordance with 493.1236(c)(1) This STANDARD is not met as evidenced by: Based on record review and interview, the laboratory failed to establish and maintain the accuracy of its testing procedures in accordance with 493.1236(c)(1). Findings include: Review of proficiency testing records from calendar years 2017 and 2018 revealed that the laboratory failed to establish and maintain accuracy of Liquid Chromatography Mass Spectrometery (LCMS) urine confirmation toxicology testing for 2018. Interview with the technical supervisor and laboratory director on 11/19/18 at 10:30 AM confirmed that the laboratory failed to establish and maintain accuracy of LCMS urine toxicology confirmation testing for 2018. The laboratory performs 223,317 LCMS urine toxicology confirmation tests per year. D5400 ANALYTIC SYSTEMS CFR(s): 493.1250 Each laboratory that performs nonwaived testing must meet the applicable analytic systems requirements in 493.1251 through 493.1283, unless HHS approves a procedure, specified in Appendix C of the State Operations Manual (CMS Pub.7), that provides equivalent quality testing. The laboratory must monitor and evaluate the overall quality of the analytic systems and correct identified problems as specified in Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 5 -- 493.1289 for each specialty and subspecialty of testing performed. This CONDITION is not met as evidenced by: Based on the deficiencies cited herein, the laboratory failed to meet the the applicable analytic systems requirements in 493.1251 through 493.1283 as evidenced by the following: 1. The laboratory failed to perform testing following the manufacturer's instructions for prothrombin time (PT) testing by not updating the International Sensitivity Index (ISI) with each new lot of PT reagent. (Refer to D5411) 2. The laboratory failed to provide validation studies for the Sysmex CA 620 coagulation analyzer after it was moved. (Refer to D5421) D5411 TEST SYSTEMS, EQUIPMENT, INSTRUMENTS, REAGENT CFR(s): 493.1252(a) Test systems must be selected by the laboratory. The testing must be performed following the manufacturer's instructions and in a manner that provides test results within the laboratory's stated performance specifications for each test system as determined under 493.1253. This STANDARD is not met as evidenced by: Based on observation and record review, the laboratory failed to perform testing following the manufacturer's instructions as evidenced by the following: The manufacturer of the Sysmex CA 620 coagulation analyzer indicates that the International Sensitivity Index (ISI) value be changed whenever a new lot of prothrombin time (PT) reagent is used. The technical consultant indicated that the analyzer was discontinued and moved in June 2018 until September 10, 2018. The laboratory began using reagent lot number 549708 on 9/10/18. The ISI value listed on the package insert for the CA 600 system was 1.00. The surveyor observed that the analyzer indicated a lot number of 539305 with an ISI of 0.99. The technical consultant and laboratory director confirmed through interview on 11/19/18 at 4:00 PM that the incorrect ISI value was entered into the analyzer. The laboratory director handed the surveyor a letter stated that the laboratory would cease PT testing effective 11/19/18. Twelve (12) PT tests were performed between 9/10/18 and 11/19/18. D5421 ESTABLISHMENT AND VERIFICATION OF PERFORMANCE CFR(s): 493.1253(b)(1) Each laboratory that introduces an unmodified, FDA-cleared or approved test system must do the following before reporting patient test results: (1)(i) Demonstrate that it can obtain performance specifications comparable to those established by the manufacturer for the following performance characteristics: (1)(i)(A) Accuracy. (1)(i) (B) Precision. (1)(i)(C) Reportable range of test results for the test system. (1)(ii) Verify that the manufacturer's reference intervals (normal values) are appropriate for the laboratory's patient population. This STANDARD is not met as evidenced by: Based on record review and interview the laboratory failed to demonstrate that it can obtain performance specifications comparable to those established by the manufacturer of a new test system before reporting patient test results as evidenced by the following: The technical consultant indicated that the Sysmex CA 620 coagulation -- 2 of 5 -- analyzer was discontinued and moved in June 2018. The analyzer was moved back into the laboratory and used for patient testing starting 9/10/18. Record review revealed that the laboratory did not perform validation studies prior to patient testing. The technical consultant confirmed through interview on 11/19/18 at 4:00 PM that the laboratory did not demonstrate the Sysmex CA 620 can obtain performance specifications comparable to those established by the manufacturer for prothrombin time (PT) and partial prothrombin time (PTT) prior to reporting patient test results. The laboratory director handed the surveyor a letter stated that the laboratory would cease PT and PTT testing effective 11/19/18. Twelve (12) PT and two (2) PTT tests were performed between 9/10/18 and 11/19/18. D5439 CALIBRATION AND CALIBRATION VERIFICATION CFR(s): 493.1255(b) Unless otherwise specified in this subpart, for each applicable test system the laboratory must do the following: Perform and document calibration verification procedure - (b)(1) Following the manufacturer's calibration verification instructions; (b)(2) Using the criteria verified or established by the laboratory under 493.1253(b)(3) -- (b)(2)(i) Including the number, type, and concentration of the materials, as well as acceptable limits for calibration verification; and (b)(2)(ii) Including at least a minimal (or zero) value, a mid-point value, and a maximum value near the upper limit of the range to verify the laboratory's reportable range of test results for the test system; and (b)(3) At least once every 6 months and whenever any of the following occur: (b)(3)(i) A complete change of reagents for a procedure is introduced, unless the laboratory can demonstrate that changing reagent lot numbers does not affect the range used to report patient test results, and control values are not adversely affected by reagent lot number changes. (b)(3)(ii) There is major preventive maintenance or replacement of critical parts that may influence test performance. (b)(3)(iii) Control materials reflect an unusual trend or shift, or are outside of the laboratory's acceptable limits, and other means of assessing and correcting unacceptable control values fail to identify and correct the problem. (b)(3)(iv) The laboratory's established schedule for verifying the reportable range for patient test results requires more frequent calibration verification. This STANDARD is not met as evidenced by: Based on record review and interview, the laboratory failed to perform calibration verifications as appropriate as evidenced by the following: Quality control (QC) record review for calendar years 2017 and 2018 revealed that calibration verifications of at least 3 points were not performed every six months for twenty (20) out of twenty (20) routine chemistry analytes on the Olympus AU 680e. Calibration verifications were performed on 8/15/17 and 8/16/18. The technical consultant and laboratory director interviewed on 11/19/18 at 2:00 PM confirmed that calibration verifications of at least 3 points had not been performed at least once every six months. The laboratory performs a total of 106,869 routine chemistry tests annually. D5805 TEST REPORT CFR(s): 493.1291(c) The test report must indicate the following: (c)(1) For positive patient identification, either the patient's name and identification number, or a unique patient identifier and identification number. (c)(2) The name and address of the laboratory location where the test was performed. (c)(3) The test report date. (c)(4) The test performed. (c)(5) -- 3 of 5 -- Specimen source, when appropriate. (c)(6) The test result and, if applicable, the units of measurement or interpretation, or both. (c)(7) Any information regarding the condition and disposition of specimens that do not meet the laboratory's criteria for acceptability. This STANDARD is not met as evidenced by: Based on record review and interview the laboratory failed to have a final report that included the address of the laboratory location where the testing was performed as evidenced by the following: The laboratory unitilizes LabCorp as their reference laboratory. The technical supervisor stated the laboratory results from LabCorp are manually enetered into the laboratory's Orchard laboratory information system (LIS). Record review revealed that the final laboratory reports generated in the LIS do not contain the address of the laboratory location where the testing was performed. The final report indicated "Test performed at LabCorp" as the test performing site. The technical superisor and laboratory director confrimed through interview on 11/19/18 at 5:30 PM that the final test report did not indicate the address of the laboratory where the testing was performed. D6018 LABORATORY DIRECTOR RESPONSIBILITIES CFR(s): 493.1407(e)(4)(iii) The laboratory director is responsible for the overall operation and administration of the laboratory, including the employment of personnel who are competent to perform test procedures, and record and report test results promptly, accurate, and proficiently and for assuring compliance with the applicable regulations. (e) The laboratory director must-- (e)(4)(iii) Ensure that all proficiency testing reports received are reviewed by the appropriate staff to evaluate the laboratory's performance and to identify any problems that require

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access