Grundy County Memorial Hospital

CLIA Laboratory Citation Details

3
Total Citations
11
Total Deficiencyies
10
Unique D-Tags
CMS Certification Number 16D0647988
Address 201 East J Avenue, Grundy Center, IA, 50638
City Grundy Center
State IA
Zip Code50638
Phone(319) 824-5421

Citation History (3 surveys)

Survey - August 28, 2025

Survey Type: Standard

Survey Event ID: YS0T11

Deficiency Tags: D5435 D5445 D5555 D6128

Summary:

Summary Statement of Deficiencies D5435 MAINTENANCE AND FUNCTION CHECKS CFR(s): 493.1254(b)(2) (b)(2)(i) Define a function check protocol that ensures equipment, instrument, and test system performance that is necessary for accurate and reliable test results and test result reporting. (b)(2)(ii) Perform and document the function checks, including background or baseline checks, specified in paragraph (b)(2)(i) of this section. Function checks must be within the laboratory's established limits before patient testing is conducted. This STANDARD is not met as evidenced by: Based on review of the laboratory's blood bank logs and records and confirmed by Technical Supervisor #1 (TS #1) at 2:30 pm on 08/28/2025, the laboratory failed to perform and document volume checks on the Ortho MTS saline dispenser from 01/01 /2024- 08/28/2025. In addition, the laboratory did not have a policy/procedure for performing and documenting Ortho MTS saline dispenser volume checks that included the frequency with which they must be performed. The findings include: 1. The laboratory uses an Ortho MTS saline dispenser to perform immunohematology procedures. 3. At the time of the survey, TS #1 confirmed the laboratory did not document Ortho MTS saline dispenser checks from 01/01/2024- 08/28/2025. In addition, TS #1 confirmed the laboratory did not have a policy/procedure for performing and documenting Ortho MTS saline dispenser volume checks that included the frequency with which they must be performed D5445 CONTROL PROCEDURES CFR(s): 493.1256(d)(1)(2)(g) (d) Unless CMS Approves a procedure, specified in Appendix C of the State Operations Manual (CMS Pub. 7), that provides equivalent quality testing, the laboratory must-- (d)(1) Perform control procedures as defined in this section unless Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 3 -- otherwise specified in the additional specialty and subspecialty requirements at 493. 1261 through 493.1278. (d)(2) For each test system, perform control procedures using the number and frequency specified by the manufacturer or established by the laboratory when they meet or exceed the requirements in paragraph (d)(3) of this section. (d)(3) At least once each day patient specimens are assayed or examined perform the following for: This STANDARD is not met as evidenced by: Based on the laboratory's Individualized Quality Control Plan (IQCP), observations made during the survey, lack of quality control (QC) records, and confirmed by interview with Technical Supervisor #1 (TS #1) at 1:30 pm on 08/28/2025, the laboratory failed to perform a positive and negative control in accordance with its IQCP for the C. Diff Quik Chek Complete test kit for one out of one lot number (lot 824035, expiration 07/01/2026). The findings include: 1. Review of the laboratory's IQCP indicated that the laboratory intended to perform QC with each new lot and shipment of C. Diff Quik Chek Complete test kits and monthly. 2. Observations made during the survey indicated the laboratory had in use one C. Diff Quik Chek Complete test kit (lot 824035, expiration 07/01/2026). The laboratory wrote on the outside of the box "Received 12/11/2024". The laboratory did not document the date on which it began using the kit. 3. At the time of the survey, TS #1 confirmed the laboratory did not know when it began using lot 824035 (expiration 07/01/2026) of the C. Diff Quik Chek Complete test kit it had in use nor did it document performance of QC according to its IQCP. D5555 IMMUNOHEMATOLOGY CFR(s): 493.1271(c)(f) (c) Blood shall be stored in a clean and orderly environment in a manner to prevent mix-ups. Expired blood must not be in the routine inventory. Unacceptable units must be segregated from routine inventory. (c)(1) An audible alarm system must monitor proper blood and blood product storage temperature over a 24-hour period. (c)(2) Inspections of the alarm system must be documented. This STANDARD is not met as evidenced by: Based on review of the laboratory's policies and procedures, lack of plasma freezer alarm check records, and confirmed by interview with Technical Supervisor #1 (TS #1) at 2:30 pm on 08/28/2025, the laboratory failed to inspect, perform, and document alarm system checks for the plasma storage freezer from 01/01/2024- 08/28/2025. In addition, the laboratory did not have a policy/procedure for performing and documenting alarm system checks that included the frequency with which they must be performed. The findings include: 1. The laboratory stored fresh frozen plasma units in the laboratory's Helmer freezer. 2. At the time of the survey, TS #1 confirmed that the laboratory did not perform alarm checks on the plasma storage freezer from 01/01 /2024- 08/28/2025. In addition, TS #1 confirmed that the laboratory did not have a policy/procedure for performing and documenting alarm system checks that included the frequency with which they must be performed. D6128 TECHNICAL SUPERVISOR RESPONSIBILITIES CFR(s): 493.1451(b)(9) (b)(9) Thereafter, evaluations must be performed at least annually unless test -- 2 of 3 -- methodology or instrumentation changes, in which case, prior to reporting patient test results, the individuals performance must be reevaluated to include the use of the new test methodology or instrumentation. This STANDARD is not met as evidenced by: Based on review of the CMS-209 Laboratory Personnel Report, personnel records, and confirmed by interview with Technical Supervisor #1 (TS #1) at 9:50 am on 08/28 /2025, the technical supervisor failed to assess and document the competency of individuals performing high complexity testing at least annually for nine out of nine testing personnel in 2023 and 2024. The findings include: 1. The CMS-209 Laboratory Personnel Report listed testing personnel (TP) #1- #9 as performing high complexity testing. 2. At the time of the survey, TS #1 confirmed TP #1- #9 did not have annual competency assessments performed and documented for high complexity testing in 2023 or 2024. -- 3 of 3 --

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access

Survey - July 18, 2023

Survey Type: Standard

Survey Event ID: NY4I11

Deficiency Tags: D5217 D5429 D5435 D5477 D5775

Summary:

Summary Statement of Deficiencies D5217 EVALUATION OF PROFICIENCY TESTING PERFORMANCE CFR(s): 493.1236(c)(1) At least twice annually, the laboratory must verify the accuracy of any test or procedure it performs that is not included in subpart I of this part. This STANDARD is not met as evidenced by: A. Based on review of the Laboratory Test List and Annual volume form and confirmed by personnel identifier #3 (refer to the Laboratory Personnel Report) at approximately 9:15 am on 07/18/2023, the laboratory failed to verify the accuracy for performing erythrocyte sedimentation rate (ESR) twice annually for three out of three time periods from 01/01/2022 - 07/18/2023. At the time of the survey, the laboratory did not enroll in proficiency testing for ESR testing and did not verify the accuracy of the testing by another method from 01/01/2022- 07/18/2023. B. Based on review of the Laboratory Test List and Annual volume form and confirmed by personnel identifier #3 (refer to the Laboratory Personnel Report) at approximately 2:00 pm on 07/18/2023, the laboratory failed to verify the accuracy for performing Sars- CoV-2 testing on the Biosystems T2 test system twice annually for three out of three time periods from 03/01/2022 - 07/18/2023. The findings include: 1. The laboratory began using the Biosystems T2 test system to perform Sars-CoV-2 patient testing in March 2022. 2. At the time of the survey, the laboratory did not enroll in proficiency testing for Sars- CoV-2 testing on the Biosystems T2 test system and did not verify the accuracy of the testing by another method from 01/01/2022- 07/18/2023. D5429 MAINTENANCE AND FUNCTION CHECKS CFR(s): 493.1254(a)(1) For unmodified manufacturer's equipment, instruments, or test systems, the laboratory must perform and document maintenance as defined by the manufacturer and with at least the frequency specified by the manufacturer. Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 4 -- This STANDARD is not met as evidenced by: Based on review of the Cepheid GeneXpert instrument maintenance records and confirmed by laboratory personnel identifier #3 (refer to the Laboratory Personnel Report) at approximately 10:30 am on 07/18/2023, the laboratory failed to perform and document monthly and quarterly maintenance as defined by the manufacturer on the Cepheid GeneXpert instrument for six out of six months and two out of two quarters of patient testing from 01/01/2023- 06/30/2023. The findings include: 1. The Cepheid GeneXpert System Maintenance Log indicated that the manufacturer requires the laboratory to perform the following monthly maintenance: *Archive tests *Purge tests *Replace fan filters 2. Records showed that the laboratory did not perform and document all monthly maintenance activities from January 2023- June 2023. 3. The GeneXpert System Maintenance Log indicated that the manufacturer requires the laboratory to perform the following quarterly maintenance: *Clean plunger rod and cartridge bays *Clean instrument surfaces 4. Records showed that the laboratory did not perform and document any quarterly maintenance at any time between 01/01 /2023- 06/30/2023. 5. At the time of the survey, personnel identifier #3 confirmed that the laboratory failed to perform and document maintenance on the Cepheid GeneXpert as required by the manufacturer. D5435 MAINTENANCE AND FUNCTION CHECKS CFR(s): 493.1254(b)(2) For equipment, instruments, or test systems developed in-house, commercially available and modified by the laboratory, or maintenance and function check protocols are not provided by the manufacturer, the laboratory must: (i) Define a function check protocol that ensures equipment, instrument, and test system performance that is necessary for accurate and reliable test results and test result reporting. (ii) Perform and document the function checks, including background or baseline checks, specified in paragraph (b)(2)(i) of this section. Function checks must be within the laboratory's established limits before patient testing is conducted. This STANDARD is not met as evidenced by: Based on review of the laboratory's policy and procedure manual, pipette function check records, and confirmed by laboratory personnel identifier #3 (refer to the Laboratory Personnel Report) at approximately 10:45 am on 07/18/2023, the laboratory failed to perform and document semiannual pipette function checks for one out of one immunohematology pipette and three out of three semi-annual periods from 01/01/2022- 07/18/2023. The findings include: 1. The laboratory uses an ID Tipmaster (12.5, 25, 50 uL) pipette to perform immunohematology testing. 2. The laboratory's "Quality Control for Equipment" policy stated that, "Semi-automated pipettes and diluters are checked semi-annually for accuracy." 3. At the time of the survey, the laboratory did not have documentation of function checks performed on the immunohematology ID Tipmaster pipette from 01/01/2022- 07/18/2023. D5477 CONTROL PROCEDURES CFR(s): 493.1256(e)(4)(g) (e) For reagent, media, and supply checks, the laboratory must do the following: (e) (4) Before, or concurrent with the initial use-- (e)(4)(i) Check each batch of media for sterility if sterility is required for testing; (e)(4)(ii) Check each batch of media for its -- 2 of 4 -- ability to support growth and, as appropriate, select or inhibit specific organisms or produce a biochemical response; and (e)(4)(iii) Document the physical characteristics of the media when compromised and report any deterioration in the media to the manufacturer. (g) The laboratory must document all control procedures performed. This STANDARD is not met as evidenced by: Based on review of the Test List and Annual Volume form, lack of media quality control (QC) records, lack of an Individualized Quality Control Plan (IQCP), and confirmed by laboratory personnel identifier #3 (refer to the Laboratory Personnel Report) at approximately 2:45 pm on 07/18/2023, the laboratory failed to check each lot of Cary Blair and viral transport media for sterility, ability to support growth, and ability to select or inhibit specific organisms or produce a biochemical response, as appropriate, from the last survey on 10/21/2021 to the time of the survey on 07/18 /2023. The findings include: 1. Review of the Test List and Annual Volume form indicated that the laboratory uses the BioFire FilmArray test system to perform gastrointestinal (GI) panel testing. 2. The laboratory's BioFire GI Panel testing procedure indicated that stool specimens must be placed in Cary Blair transport media prior to testing. 3. Review of the Test List and Annual Volume form indicated that the laboratory uses the Cepheid GeneXpert test system to perform Sars-CoV-2, respiratory syncytial virus (RSV), and Influenza A and B testing. 4. The laboratory's Cepheid GeneXpert Sars-CoV-2, RSV, and Influenza A and B testing procedures indicated that respiratory specimens must be placed in viral transport media prior to testing. 5. Review of the Test List and Annual Volume form indicated that the laboratory uses the Biosystems T2 test system to perform Sars- CoV-2 testing. 6. The laboratory's Biosystems T2 Sars-CoV-2 testing procedure indicated that respiratory specimens must be placed in viral transport media prior to testing. 7. Personnel identifier #3 indicated that the laboratory intended to retain the manufacturer's QC certificates and not perform additional QC. 8. At the time of the survey, the laboratory did not retain the manufacturer's QC certificates and did not have an IQCP for Cary Blair and viral transport media. D5775 COMPARISON OF TEST RESULTS CFR(s): 493.1281(a)(c) (a) If a laboratory performs the same test using different methodologies or instruments, or performs the same test at multiple testing sites, the laboratory must have a system that twice a year evaluates and defines the relationship between test results using the different methodologies, instruments, or testing sites. (c) The laboratory must document all test result comparison activities. This STANDARD is not met as evidenced by: A. Based on lack of comparison testing records and confirmed by laboratory personnel #3 (refer to the Laboratory Personnel Report) at approximately 2:00 pm on 07/18/2023, the laboratory failed to perform comparison studies between the Cepheid GeneXpert and Biosystems T2 test systems twice a year for three out of three time periods from 03/01/2022- 07/18/2023. The findings include: 1. The laboratory began using the Biosystems T2 test system to perform Sars-CoV-2 patient testing in March 2022. 2. The laboratory also performs Sars-CoV-2 testing on the Cepheid GeneXpert test system. 3. At the time of the survey, the laboratory did not have documentation of comparison testing for the analyte, Sars-CoV-2, on the Cepheid GeneXpert and Biosystems T2 test systems. B. Based on lack of comparison testing records and -- 3 of 4 -- confirmed by laboratory personnel #3 (refer to the Laboratory Personnel Report) at approximately 2:00 pm on 07/18/2023, the laboratory failed to perform comparison studies between the BioFire Gastrointestinal (GI) Panel and C Diff Quik Chek test systems twice a year for three out of three time periods from 01/01/2022- 07/18/2023. The findings include: 1. The laboratory performs Clostridium difficile toxin testing on both the BioFire GI Panel and C Diff Quik Chek test system. 2. At the time of the survey, the laboratory did not have documentation of comparison testing for the analyte Clostridium difficile toxin testing on the the BioFire GI Panel and C Diff Quik Chek test system. -- 4 of 4 --

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access

Survey - January 4, 2018

Survey Type: Standard

Survey Event ID: SXGI12

Deficiency Tags: D3031 D6055

Summary:

Summary Statement of Deficiencies No Tags No deficiency details available. Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 1 --

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access