James Richards, Md

CLIA Laboratory Citation Details

4
Total Citations
16
Total Deficiencyies
11
Unique D-Tags
CMS Certification Number 05D0544111
Address 6222 W Manchester Ave, Ste A, Los Angeles, CA, 90045
City Los Angeles
State CA
Zip Code90045
Phone310 670-1840
Lab DirectorJAMES RICHARDS

Citation History (4 surveys)

Survey - August 30, 2023

Survey Type: Standard

Survey Event ID: 32G511

Deficiency Tags: D2094 D6004 D6019 D2087 D5211 D6018 D6023

Summary:

Summary Statement of Deficiencies D2087 ROUTINE CHEMISTRY CFR(s): 493.841(a) Failure to attain a score of at least 80 percent of acceptable responses for each analyte in each testing event is unsatisfactory analyte performance for the testing event. This STANDARD is not met as evidenced by: Based on Surveyor review of laboratory's proficiency testing (PT) records, and interview with the laboratory technical consultant on August 30, 2023, at 3:00 pm, the laboratory failed to attain a score of at least 80 percent of acceptable responses for chloride at the 2nd event in 2023 and is unsatisfactory analyte performance for the testing event. The findings include: 1. The laboratory participated in the API PT program for the year 2023. The laboratory scored 40% for chloride at the 2nd event which resulted in an unsatisfactory analyte performance. Therefore, the accuracy of the patient test results reported by the laboratory during the proficiency testing event cannot be assured and might have harmed patients. 2. The laboratory technical consultant on August 30, 2023, at 3:00 pm, affirmed that the laboratory scored 40% for chloride at the 2nd event in 2023. 3. The laboratory's testing declaration form, signed by the laboratory director on 8/30/2023 stated that the laboratory performs approximately 59,248 tests in routine chemistry, annually. D2094 ROUTINE CHEMISTRY CFR(s): 493.841(e) (1) For any unsatisfactory analyte or test performance or testing event for reasons other than a failure to participate, the laboratory must undertake appropriate training and employ the technical assistance necessary to correct problems associated with a proficiency testing failure. (2) For any unacceptable analyte or testing event score, remedial action must be taken and documented, and the documentation must be maintained by the laboratory for two years from the date of participation in the Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 4 -- proficiency testing event. This STANDARD is not met as evidenced by: Based on Surveyor review of laboratory's proficiency testing (PT) records, and interview with the laboratory technical consultant on August 30, 2023, at 3:00 pm, the laboratory failed to take remedial action after receiving the unsatisfactory analyte performance at the 2nd event in 2023. The findings include: 1. The laboratory participated in the API PT program for the year 2023. The laboratory scored 40% at the 2nd event which resulted in an unsatisfactory analyte performance. However, the laboratory did not take any remedial action for the failure except re-running the PT sample. The remedial action must address the underlying cause, future prevention and impact on the patients results. Therefore, the accuracy of the patient test results reported by the laboratory during the proficiency testing event cannot be assured and might have harmed patients. 2. The laboratory technical consultant on August 30, 2023, at 3:00 pm, affirmed that the laboratory did not take any remedial or

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access

Survey - April 5, 2022

Survey Type: Special

Survey Event ID: QUJZ11

Deficiency Tags: D2016 D6000 D2096 D6016

Summary:

Summary Statement of Deficiencies D2016 SUCCESSFUL PARTICIPATION CFR(s): 493.803(a)(b)(c) (a) Each laboratory performing nonwaived testing must successfully participate in a proficiency testing program approved by CMS, if applicable, as described in subpart I of this part for each specialty, subspecialty, and analyte or test in which the laboratory is certified under CLIA. (b) Except as specified in paragraph (c) of this section, if a laboratory fails to participate successfully in proficiency testing for a given specialty, subspecialty, analyte or test, as defined in this section, or fails to take remedial action when an individual fails gynecologic cytology, CMS imposes sanctions, as specified in subpart R of this part. (c) If a laboratory fails to perform successfully in a CMS- approved proficiency testing program, for the initial unsuccessful performance, CMS may direct the laboratory to undertake training of its personnel or to obtain technical assistance, or both, rather than imposing alternative or principle sanctions except when one or more of the following conditions exists: (1) There is immediate jeopardy to patient health and safety. (2) The laboratory fails to provide CMS or a CMS agent with satisfactory evidence that it has taken steps to correct the problem identified by the unsuccessful proficiency testing performance. (3) The laboratory has a poor compliance history. This CONDITION is not met as evidenced by: Based on desk review of CMS proficiency testing (PT) records (i.e. CMS CASPER Reports 0155D entitled, "Individual Laboratory Profile" and CMS CASPER Report 0153D entitled, "Unsuccessful (2 of 3) Report"), it was determined that the laboratory failed to successfully participate in a PT program approved by CMS for each analyte or test in which the laboratory is certified under CLIA. The findings included: The laboratory failed to achieve satisfactory performance for the same analyte or test in two out of three consecutive testing events in the specialty of Routine Chemistry constituting unsuccessful PT performance. (See D2096) Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 2 -- D2096 ROUTINE CHEMISTRY CFR(s): 493.841(f) Failure to achieve satisfactory performance for the same analyte or test in two consecutive testing events or two out of three consecutive testing events is unsuccessful performance. This STANDARD is not met as evidenced by: Based on desk review of CMS PT records (CMS CASPER Report 0155D and 0153D, it was determined that the laboratory failed to achieve satisfactory performance for the same analyte or test in two out of three consecutive PT events for the analyte, Chloride (Cl), as follows: 2021 Q3 2022 Q1 Chloride 0% 40% Q1 = First testing event Q3 = Third testing event b. Failure to achieve satisfactory performance for the same analyte or test in two of three consecutive PT resulted in an initial unsuccessful performance for the analyte, Cl.. D6000 MODERATE COMPLEXITY LABORATORY DIRECTOR CFR(s): 493.1403 The laboratory must have a director who meets the qualification requirements of 493. 1405 of this subpart and provides overall management and direction in accordance with 493.1407 of this subpart. This CONDITION is not met as evidenced by: Based on the severity of the deficiencies cited herein, the Condition: Laboratories Performing Moderate Complexity Testing: Laboratory director was not met. The laboratory director, moderate complexity testing, failed to ensure that PT samples were tested as required under Subpart H of this part. (See D6016) D6016 LABORATORY DIRECTOR RESPONSIBILITIES CFR(s): 493.1407(e)(4)(i) The laboratory director is responsible for the overall operation and administration of the laboratory, including the employment of personnel who are competent to perform test procedures, and record and report test results promptly, accurate, and proficiently and for assuring compliance with the applicable regulations. (e) The laboratory director must-- (e)(4)(i) Ensure that the proficiency testing samples are tested as required under Subpart H of this part; This STANDARD is not met as evidenced by: Based on a desk review of CMS PT records, it was determined the laboratory director, moderate complexity testing, failed to ensure that PT samples were tested as required under subpart H. of this part. The findings included: For the analyte, Cl, Total, the laboratory repeatedly failed to achieve satisfactory performance for the same analyte or test in two out of three consecutive testing events, resulting in unsuccessful PT performance. (see D2016 and D2096) -- 2 of 2 --

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access

Survey - January 7, 2021

Survey Type: Standard

Survey Event ID: KDHX11

Deficiency Tags: D2087

Summary:

Summary Statement of Deficiencies D2087 ROUTINE CHEMISTRY CFR(s): 493.841(a) Failure to attain a score of at least 80 percent of acceptable responses for each analyte in each testing event is unsatisfactory analyte performance for the testing event. This STANDARD is not met as evidenced by: Based on the surveyor's review of the laboratory's records for evaluation of proficiency testing performance from CMS, laboratory proficiency reporting from API, and an interview with laboratory technical consultant (TC) on 1/7/21 between 2: 30 p.m. . and 4:00 p.m., it was determined that there were several unacceptable (below 80 %) proficiency testing (PT) results for Chloride (Cl) for Cycle Q3 (2019). Findings include: 1. On 1/7/2021, an inspection was conducted between 2:30 p.m. and 4:00 p. m. 2. During a review of the laboratory documentation from API (agency providing the proficiency specimens), it was noted at approximately 3:30 p.m. that there were several unacceptable PT results for cycle Q3 Chloride on the 2019 report. The laboratory utilizes the Envoy 500 instrument for general chemistry testing The TC recognized these atypical results. 3. The findings and acceptable ranges were as follows: Analyte: Chloride (5/5 Unacceptable) Sample Actual Result Expected Result (range) CH11 101 88-99 CH12 103 91-102 CH13 115 100-112 CH14 116 103-115 CH15 122 109-121 4. The TC affirmed the unacceptable results listed above. Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 1 --

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access

Survey - April 12, 2018

Survey Type: Special

Survey Event ID: 32LM11

Deficiency Tags: D2096 D6016 D2016 D6000

Summary:

Summary Statement of Deficiencies D2016 SUCCESSFUL PARTICIPATION CFR(s): 493.803(a)(b)(c) (a) Each laboratory performing nonwaived testing must successfully participate in a proficiency testing program approved by CMS, if applicable, as described in subpart I of this part for each specialty, subspecialty, and analyte or test in which the laboratory is certified under CLIA. (b) Except as specified in paragraph (c) of this section, if a laboratory fails to participate successfully in proficiency testing for a given specialty, subspecialty, analyte or test, as defined in this section, or fails to take remedial action when an individual fails gynecologic cytology, CMS imposes sanctions, as specified in subpart R of this part. (c) If a laboratory fails to perform successfully in a CMS- approved proficiency testing program, for the initial unsuccessful performance, CMS may direct the laboratory to undertake training of its personnel or to obtain technical assistance, or both, rather than imposing alternative or principle sanctions except when one or more of the following conditions exists: (1) There is immediate jeopardy to patient health and safety. (2) The laboratory fails to provide CMS or a CMS agent with satisfactory evidence that it has taken steps to correct the problem identified by the unsuccessful proficiency testing performance. (3) The laboratory has a poor compliance history. This CONDITION is not met as evidenced by: Based on desk review of CMS proficiency testing (PT) records (i.e. CMS CASPER Reports 0155D entitled, "Individual Laboratory Profile" and CMS CASPER Report 0153D entitled, "Unsuccessful (2 of 3) Report"), it was determined that the laboratory failed to successfully participate in a PT program approved by CMS for each analyte or test in which the laboratory is certified under CLIA. The findings included: The laboratory failed to achieve satisfactory performance for the same analyte or test in two out of three consecutive testing events in the subspecialty of Routine Chemistry constituting unsuccessful PT performance. (See D2096) Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 2 -- D2096 ROUTINE CHEMISTRY CFR(s): 493.841(f) Failure to achieve satisfactory performance for the same analyte or test in two consecutive testing events or two out of three consecutive testing events is unsuccessful performance. This STANDARD is not met as evidenced by: Based on desk review of CMS PT records (CMS CASPER Report 0155D and 0153D, it was determined that the laboratory failed to achieve satisfactory performance for the same analyte or test in two out of three consecutive PT events for the analyte, Chloride (CL), resulting in an "initial" (first) unsuccessful performance. The findings include: a. The laboratory failed to maintain successful performance with the PT program by failing to obtain a score of 80% of acceptable responses in two out of three consecutive PT events for the analyte, CL, as follows: 2017 Q3 (20%) 2018 Q1 (40%) Q1 = First Testing Event Q3 = Third Testing Event b. Failure to achieve satisfactory performance for the same analyte or test in two out of three consecutive PT resulted in an initial unsuccessful performance for the analyte, CL. D6000 MODERATE COMPLEXITY LABORATORY DIRECTOR CFR(s): 493.1403 The laboratory must have a director who meets the qualification requirements of 493. 1405 of this subpart and provides overall management and direction in accordance with 493.1407 of this subpart. This CONDITION is not met as evidenced by: Based on the severity of the deficiencies cited herein, the Condition: Laboratories Performing Moderate Complexity Testing: Laboratory director was not met. The laboratory director, moderate complexity testing, failed to ensure that PT samples were tested as required under Subpart H of this part. (See D6016) D6016 LABORATORY DIRECTOR RESPONSIBILITIES CFR(s): 493.1407(e)(4)(i) The laboratory director is responsible for the overall operation and administration of the laboratory, including the employment of personnel who are competent to perform test procedures, and record and report test results promptly, accurate, and proficiently and for assuring compliance with the applicable regulations. (e) The laboratory director must-- (e)(4)(i) Ensure that the proficiency testing samples are tested as required under Subpart H of this part; This STANDARD is not met as evidenced by: Based on a desk review of CMS PT records, it was determined the laboratory director, moderate complexity testing, failed to ensure that PT samples were tested as required under subpart H. of this part. The findings included: For the analyte, CL, the laboratory repeatedly failed to achieve satisfactory performance for the same analyte or test in two out of three consecutive testing events, resulting in unsuccessful PT performance. (See D2016 and D2096) -- 2 of 2 --

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access