Karmanos Cancer Institute Bloomfield

CLIA Laboratory Citation Details

2
Total Citations
3
Total Deficiencyies
2
Unique D-Tags
CMS Certification Number 23D0371168
Address 43097 Woodward Avenue Ste 100, Bloomfield Hills, MI, 48302
City Bloomfield Hills
State MI
Zip Code48302
Phone(248) 334-0050

Citation History (2 surveys)

Survey - July 12, 2021

Survey Type: Standard

Survey Event ID: O3GV11

Deficiency Tags: D5439 D5439

Summary:

Summary Statement of Deficiencies D5439 CALIBRATION AND CALIBRATION VERIFICATION CFR(s): 493.1255(b) Unless otherwise specified in this subpart, for each applicable test system the laboratory must do the following: Perform and document calibration verification procedure - (b)(1) Following the manufacturer's calibration verification instructions; (b)(2) Using the criteria verified or established by the laboratory under 493.1253(b)(3) -- (b)(2)(i) Including the number, type, and concentration of the materials, as well as acceptable limits for calibration verification; and (b)(2)(ii) Including at least a minimal (or zero) value, a mid-point value, and a maximum value near the upper limit of the range to verify the laboratory's reportable range of test results for the test system; and (b)(3) At least once every 6 months and whenever any of the following occur: (b)(3)(i) A complete change of reagents for a procedure is introduced, unless the laboratory can demonstrate that changing reagent lot numbers does not affect the range used to report patient test results, and control values are not adversely affected by reagent lot number changes. (b)(3)(ii) There is major preventive maintenance or replacement of critical parts that may influence test performance. (b)(3)(iii) Control materials reflect an unusual trend or shift, or are outside of the laboratory's acceptable limits, and other means of assessing and correcting unacceptable control values fail to identify and correct the problem. (b)(3)(iv) The laboratory's established schedule for verifying the reportable range for patient test results requires more frequent calibration verification. This STANDARD is not met as evidenced by: . Based on record review and interview with Technical Consultant #1 (TC1), the laboratory failed to perform calibration for hematology testing on the Horiba ABX Micros 60 analyzer at least every 6 months for 2 (July 2019 to July 2021) of two years. Findings include: 1. A review of the laboratory's calibration documentation for the Horiba ABX Micros 60 hematology analyzer revealed calibration was performed Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 2 -- on the following dates between July 2019 and July 2021: a. 3/4/21 (9 months after 6/3 /20) b. 6/3/20 (8 months after 9/19/20) c. 9/19/19 2. A review of the laboratory's "ABX Micros 60" policy and procedure revealed a section titled "Calibration" stating, "Routine calibration must be performed every 6 months and may also need to be performed after: Initial instrument installation. Major maintenance is performed on the instrument. Any major instrument (parts) are repaired or changed. And if controls are consistently out of assay range after all other attempts have been made to clean the instrument and eliminate the controls as the problem." 3. An interview on 7/12/21 at 9: 54 am with TC1 confirmed the laboratory did not perform calibrations at least every 6 months and patient testing had not seized during these timeframes. -- 2 of 2 --

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access

Survey - June 5, 2019

Survey Type: Standard

Survey Event ID: L75U11

Deficiency Tags: D5445

Summary:

Summary Statement of Deficiencies D5445 CONTROL PROCEDURES CFR(s): 493.1256(d)(1)(2)(g) Unless CMS Approves a procedure, specified in Appendix C of the State Operations Manual (CMS Pub. 7), that provides equivalent quality testing, the laboratory must-- (d)(1) Perform control procedures as defined in this section unless otherwise specified in the additional specialty and subspecialty requirements at 493.1261 through 493.1278. (d)(2) For each test system, perform control procedures using the number and frequency specified by the manufacturer or established by the laboratory when they meet or exceed the requirements in paragraph (d)(3) of this section. (g) The laboratory must document all control procedures performed. This STANDARD is not met as evidenced by: . Based on procedure review, record review, and interview with the office manager and Testing Personnel #1 (TP1), the laboratory failed to perform quality control as required for the hematology complete blood cell count test for 1 (medical record number 781309) of 10 patient charts audited. Findings include: 1. Review of the "ABX Micros 60 Procedure" states that three levels of controls must be run and acceptable each day of patient testing. 2. Record review of the monthly quality control records revealed for patient medical record 781309 testing performed on January 2, 2019, no quality control was performed prior to testing patient specimens. 3. During the interview on June 5, 2019 at approximately 1:00 pm, the office manager and TP1 acknowledged that no quality control was performed prior to testing patient specimens on January 2, 2019. Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 1 --

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access