Laboratory Of Parasitic Disease

CLIA Laboratory Citation Details

2
Total Citations
10
Total Deficiencyies
6
Unique D-Tags
CMS Certification Number 21D0957207
Address Bldg 4/ 211, Bethesda, MD
City Bethesda
State MD
Phone(301) 496-4000

Citation History (2 surveys)

Survey - April 9, 2025

Survey Type: Standard

Survey Event ID: JE1O11

Deficiency Tags: D5413 D5435 D5435 D5209 D5413

Summary:

Summary Statement of Deficiencies D5209 PERSONNEL COMPETENCY ASSESSMENT POLICIES CFR(s): 493.1235 As specified in the personnel requirements in subpart M, the laboratory must establish and follow written policies and procedures to assess employee and, if applicable, consultant competency. This STANDARD is not met as evidenced by: Based on a review of laboratory policies and procedures, CMS Laboratory Personnel Report (Form CMS-209), competency assessment records, and an interview with the Laboratory Director, the laboratory failed to establish and implement written policies and procedures for assessing competency based on position responsibilities for one of one individual fulfilling the roles of Clinical Consultant, Technical Supervisor, and General Supervisor. Findings: 1. On April 9, 2025, at approximately 12:30 p.m., a review of the laboratory 's policy titled QC-QA Plan revealed the absence of any documented process for assessing competency specific to the responsibilities of personnel serving as Clinical Consultant, Technical Supervisor, and General Supervisor. 2. On April 9, 2025, at approximately 10:30 a.m., a review of the CMS Laboratory Personnel Report (Form CMS-209) indicated that one individual was listed as fulfilling the roles of Clinical Consultant, Technical Supervisor, and General Supervisor. 3. On April 9, 2025, at approximately 12:00 p.m., a review of the laboratory 's competency assessment records revealed that no competency documentation was available for the individual identified as serving in the aforementioned roles. 4. On April 9, 2025, at approximately 12:40 p.m., an interview with the Laboratory Director confirmed that the laboratory does not have policies or procedures in place for assessing competency of individuals in roles other than testing personnel. D5413 TEST SYSTEMS, EQUIPMENT, INSTRUMENTS, REAGENT CFR(s): 493.1252(b) Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 3 -- (b) The laboratory must define criteria for those conditions that are essential for proper storage of reagents and specimens, accurate and reliable test system operation, and test result reporting. The criteria must be consistent with the manufacturer's instructions, if provided. These conditions must be monitored and documented and, if applicable, include the following: (b)(1) Water quality. (b)(2) Temperature. (b)(3) Humidity. (b)(4) Protection of equipment and instruments from fluctuations and interruptions in electrical current that adversely affect patient test results and test reports. This STANDARD is not met as evidenced by: Based on a review of the laboratory 's policies and procedures, direct observation and staff interview conducted during a laboratory tour, the laboratory failed to monitor and document ambient room temperature and humidity, as well as internal temperatures for two freezers and two refrigerators used to store laboratory reagents and patient specimens. Findings: 1. On April 9, 2025, at approximately 1:00 p.m., a review of the laboratory 's policy titled QC-QA Plan revealed a section labeled Equipment Monitoring, which stated: "All instruments such as freezers, refrigerators, and incubators are monitored manually or electronically by alarm system and recorded on each day of use. Refrigerator temperature should be between 2 and 8 C. Freezer temperature should be between -25 and -15 C. Ambient room temperature between 50 and 90 F is appropriate for testing procedures." 2. On April 9, 2025, at approximately 11:00 a.m., a physical inspection of Room 11S216, where laboratory testing was performed, revealed that ambient room temperature and humidity were not being monitored or documented. No temperature or humidity logs were available for review at the time of inspection. 3. On April 9, 2025, at approximately 11:00 a.m., observation of laboratory cold storage units revealed the presence of: -Two freezers labeled #8 (NIH ID# 02229735) and #13 (NIH ID# 02309916) -Two refrigerators, one labeled with NIH ID# 02269071 and another labeled probe #14 (no visible NIH ID#) All four units contained laboratory reagents and patient specimens. No internal temperature probes were observed inside any of the units. Furthermore, no temperature monitoring logs (manual or electronic) were available for review at the time of inspection. 4. On April 9, 2025, at approximately 11:15 a.m., an interview with the Laboratory Director confirmed that the laboratory had not been monitoring or documenting room temperature, humidity, or the temperatures of the two freezers and two refrigerators since relocating to its current location on February 14, 2025. D5435 MAINTENANCE AND FUNCTION CHECKS CFR(s): 493.1254(b)(2) (b)(2)(i) Define a function check protocol that ensures equipment, instrument, and test system performance that is necessary for accurate and reliable test results and test result reporting. (b)(2)(ii) Perform and document the function checks, including background or baseline checks, specified in paragraph (b)(2)(i) of this section. Function checks must be within the laboratory's established limits before patient testing is conducted. This STANDARD is not met as evidenced by: Based on a review of the laboratory 's policies and procedures, equipment maintenance records, and an interview with the Laboratory Director, the laboratory failed to perform and document annual pipette calibrations for 96 out of 96 pippetes as -- 2 of 3 -- required by its own policy since March 20, 2023. Findings: 1. On April 9, 2025, at approximately 1:00 p.m., a review of the laboratory ' s policy titled QC-QA Plan revealed a section labeled Equipment Monitoring, which states: "Pipettes are calibrated every 12 months." 2. On April 9, 2025, at approximately 1:00 p.m., a review of the laboratory ' s equipment maintenance records showed that 78 single- channel pipettes, 15 multichannel pipettes, and 3 multichannel electronic pipettes were last calibrated on March 20, 2023. No documentation of subsequent pipette calibrations was available for review at the time of inspection. 3. On April 9, 2025, at approximately 1:15 p.m., an interview with the Laboratory Director confirmed that pipette calibrations had not been performed annually, as required by the laboratory ' s own policy. -- 3 of 3 --

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access

Survey - April 6, 2023

Survey Type: Standard

Survey Event ID: YQ8O11

Deficiency Tags: D0000 D5217 D5217 D5423 D5423

Summary:

Summary Statement of Deficiencies D0000 The laboratory of Parasitic Disease was surveyed under 42 CFR part 493 CLIA regulations. The following standard deficiencies were found during the announced routine CLIA recertification survey performed on April 6, 2023: D5217 EVALUATION OF PROFICIENCY TESTING PERFORMANCE CFR(s): 493.1236(c)(1) At least twice annually, the laboratory must verify the accuracy of any test or procedure it performs that is not included in subpart I of this part. This STANDARD is not met as evidenced by: Based on proficiency documentation review and an interview with testing personnel #1 (TP1) and technical supervisor #2 (TS2), the laboratory failed to verify the accuracy of tests performed that is not included in subpart I of this part at least twice annually. Findings include: 1. On 04/6/2023 at 11:45 AM, a review of the Quality Assurance Plan Version 6 (7/28/2021), section Proficiency/Accuracy Testing required the following procedures: Once a year, 8 blinded - but previously tested - specimens are repeated to ensure that similar results are obtained. 2. On 04/6/2023 at 12:25 PM, a review of the proficiency testing sheets, showed that proficiency testing / Accuracy testing had been performed once a year for these analytes. a. Antifilarial IgG and Antifilarial IgG4: 3/15/21, and 2/28/22 b. Tsol1R13 qPCR: 8/7/20, 7/30/21, 1/12/22, and 3/31/23 3. On 04/6/2023 at 1:10 PM, the laboratory TP1 and TS2 confirmed the above findings. D5423 ESTABLISHMENT AND VERIFICATION OF PERFORMANCE CFR(s): 493.1253(b)(2) Each laboratory that modifies an FDA-cleared or approved test system, or introduces a test system not subject to FDA clearance or approval (including methods developed in-house and standardized methods such as text book procedures), or uses a test Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 2 -- system in which performance specifications are not provided by the manufacturer must, before reporting patient test results, establish for each test system the performance specifications for the following performance characteristics, as applicable: (2)(i) Accuracy. (2)(ii) Precision. (2)(iii) Analytical sensitivity. (2)(iv) Analytical specificity to include interfering substances. (2)(v) Reportable range of test results for the test system. (2)(vi) Reference intervals (normal values). (2)(vii) Any other performance characteristic required for test performance. This STANDARD is not met as evidenced by: Based on observation, a lack of documentation, and an interview with testing personnel #1 (TP1), the laboratory failed to demonstrate that they had established test performance specifications for their Antifilarial IgG and Antifilarial IgG4 laboratory- developed tests (LDT) before patient testing in new Bio Plex 200 System Reader to detect the identity and determine the quantity of the analytes. Findings include: 1. A tour of the laboratory on April 6, 2023, at 10:00 AM, revealed it was one Bio Plex 200 System Reader (Serial# LX10021091421) in the laboratory for performing Antifilarial IgG and Antifilarial IgG4 tests. 2. The laboratory did not establish test performance specifications for Antifilarial IgG and Antifilarial IgG4 tests, and there was no documentation available at the time of the survey. 3. Interview with the TP1 on 04/6 /2023 at 11:30 AM confirmed the above finding and the fact the laboratory started the patient testing for Antifilarial IgG and Antifilarial IgG4 tests on 7/9/2021. -- 2 of 2 --

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access