Marco Gutierrez Md And Associates

CLIA Laboratory Citation Details

2
Total Citations
6
Total Deficiencyies
3
Unique D-Tags
CMS Certification Number 45D1003540
Address 401 S Alamo, Alamo, TX, 78516
City Alamo
State TX
Zip Code78516
Phone956 782-9111
Lab DirectorMARCO MD

Citation History (2 surveys)

Survey - July 26, 2022

Survey Type: Standard

Survey Event ID: ATSM11

Deficiency Tags: D3011 D5421 D3011 D5421

Summary:

Summary Statement of Deficiencies D3011 FACILITIES CFR(s): 493.1101(d) Safety procedures must be established, accessible, and observed to ensure protection from physical, chemical, biochemical, and electrical hazards, and biohazardous materials. This STANDARD is not met as evidenced by: Based on surveyor observation, review of laboratory policies, and interview with laboratory personnel, the laboratory failed to follow its infection control policy for one of one observed events occurring on July 26,2022. The findings included: 1. Surveyor observation on July 26, 2022 at 09:10 hours in the laboratory found 2 personal drinking items, one located next to the computer and one located on top of a laboratory refrigerator. 2. Review of the laboratory's policy title 'Infection Control Policy' approved by the laboratory director on September 1, 2016 stated, "Infection Control Policy is to ensure the avoidance of spreading infectious disease. Universal Precautions apply to blood and other body fluids containing blood..." 3. The findings were confirmed in interview with the technical consultant on July 26, 2022 at 11:30 hours in the radiology room. D5421 ESTABLISHMENT AND VERIFICATION OF PERFORMANCE CFR(s): 493.1253(b)(1) Each laboratory that introduces an unmodified, FDA-cleared or approved test system must do the following before reporting patient test results: (1)(i) Demonstrate that it can obtain performance specifications comparable to those established by the manufacturer for the following performance characteristics: (1)(i)(A) Accuracy. (1)(i) (B) Precision. (1)(i)(C) Reportable range of test results for the test system. (1)(ii) Verify that the manufacturer's reference intervals (normal values) are appropriate for the laboratory's patient population. Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 2 -- This STANDARD is not met as evidenced by: Based on review of the laboratory's verification studies performed on the Medonic M hematology analyzer, and confirmed in staff interviews, laboratory failed to have documentation of completing verification studies on 1 of 1 new analyzers implemented in November 2021. The findings included: 1. A review of the laboratory's verification studies performed on the Medonic M Series hematology analyzer in November 2021 found the laboratory failed to have documentation of verification of patient normal ranges. As of July 22, 2022, the day of the survey, patient normal ranges had not been verified. 2. The laboratory was asked to provide documentation of the missing studies. No documentation was provided. 3. An interviews with the technical consultant on July 26, 2022 at 11:30 hours in the radiology room confirmed the findings. -- 2 of 2 --

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access

Survey - August 29, 2018

Survey Type: Standard

Survey Event ID: J9N911

Deficiency Tags: D5213 D5213

Summary:

Summary Statement of Deficiencies D5213 EVALUATION OF PROFICIENCY TESTING PERFORMANCE CFR(s): 493.1236(b)(1) The laboratory must verify the accuracy of any analyte or subspecialty without analytes listed in subpart I of this part that is not evaluated or scored by a CMS- approved proficiency testing program. This STANDARD is not met as evidenced by: Based on review of the laboratory's American Proficiency Institute (API) proficiency testing results from 2016, 2017, and 2018, and confirmed in interview of facility personnel, the laboratory failed to provide documentation of evaluating proficiency testing results returned as 'not graded' by the proficiency testing agency. The findings included: 1. A review of the laboratory's API proficiency testing results from 2016 (event 3), 2017 (events 1, 2, and 3), and 2018 (events 1 and 2) revealed the proficiency agency returned the following results as "Not Graded": 2017 (event 2) Analyte: Monocytes Sample ID: HEM-07, HEM-10 Grade: Not Graded-2 2. The proficiency testing agency defined code "2" as: No Consensus. 3. An interview with the primary testing person on 08/29/2018 at 09:15 hours in the office confirmed the findings. Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 1 --

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access