Summary:
Summary Statement of Deficiencies D0000 An announced CLIA recertification survey was conducted at Midstate Skin Institute LLC on 04/29/25 - 05/02/2025. The laboratory is not in compliance with 42 CFR Part 493, Requirements for Laboratories. The following is a description of the standard level deficiency: D5803 TEST REPORT CFR(s): 493.1291(b) (b) Test report information maintained as part of the patient's chart or medical record must be readily available to the laboratory and to CMS or a CMS agent upon request. This STANDARD is not met as evidenced by: Based on record review and interview, the laboratory was unable to provide two (Patient #2, #3) of three patient test reports (Patients #1- #3). Findings included: Patient test reports were reviewed. Patient #2 and #3's reports were unavailable. The lab Histology Technician confirmed the two reports were not available on 04/29/2025 at 11:30 a.m. Electronic communication, received 05/02/2025 at 3:15 pm, confirmed Patient #2 and #3's reports were unavailable. Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 1 --