Summary:
Summary Statement of Deficiencies D0000 An announced CLIA recertification survey was conducted at Millennium Medical Management LLC on April 20, 2026 to May 4, 2026. The laboratory is not in compliance with 42 CFR Part 493, Requirement for Laboratories. The following Conditions were cited: D5400 493.1250 Condition: Analytic Systems D6033 493.1409 Condition: Technical Consultant - Moderate Complexity D5400 ANALYTIC SYSTEMS CFR(s): 493.1250 Each laboratory that performs nonwaived testing must meet the applicable analytic systems requirements in 493.1251 through 493.1283, unless HHS approves a procedure, specified in Appendix C of the State Operations Manual (CMS Pub.7), that provides equivalent quality testing. The laboratory must monitor and evaluate the overall quality of the analytic systems and correct identified problems as specified in 493.1289 for each specialty and subspecialty of testing performed. This CONDITION is not met as evidenced by: Based on review of calibration records and interview, the laboratory failed to perform calibration verification for the Horiba Micro 60 hematology analyzer at least every six months in 2025 for the calibration records reviewed from 01/02/2024 to 04/15/2026. (See D5439) D5439 CALIBRATION AND CALIBRATION VERIFICATION CFR(s): 493.1255(b) (b)(1) Following the manufacturer's calibration verification instructions; (b)(2) Using the criteria verified or established by the laboratory under 493.1253(b)(3)-- (b)(2)(i) Including the number, type, and concentration of the materials, as well as acceptable limits for calibration verification; and (b)(2)(ii) Including at least a minimal (or zero) value, a mid-point value, and a maximum value near the upper limit of the range to Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 3 -- verify the laboratory's reportable range of test results for the test system; and (b)(3) At least once every 6 months and whenever any of the following occur: (b)(3)(i) A complete change of reagents for a procedure is introduced, unless the laboratory can demonstrate that changing reagent lot numbers does not affect the range used to report patient test results, and control values are not adversely affected by reagent lot number changes. (b)(3)(ii) There is major preventive maintenance or replacement of critical parts that may influence test performance. (b)(3)(iii) Control materials reflect an unusual trend or shift, or are outside of the laboratory's acceptable limits, and other means of assessing and correcting unacceptable control values fail to identify and correct the problem. (b)(3)(iv) The laboratory's established schedule for verifying the reportable range for patient test results requires more frequent calibration verification. This STANDARD is not met as evidenced by: Based on review of allegation of compliance, calibration records, and interview, the laboratory failed to perform calibration verification for the Horiba Micro 60 hematology analyzer at least every six months in 2025 for the calibration records review from 01/02/2024 to 04/15/2026. This is a repeat deficiency from recertification survey performed on 12/27/2023. Findings: 1. Review of the allegation of compliance (signed by the Laboratory Director on ) stated "The lab consultant and Lab Director will be responsible for making sure that calibration is completed every 6 months, using the manual calendar and the Outlook email reminder. They will also use the monthly document sheet to check for calibration results if present, and if due, can investigate." 1. Review of calibration records for the hematology analyzer revealed calibrations were performed on 01/02/2024, 07/26/2024, 12/19/2024, 05/09/2025, 05 /19/2025, and 04/15/2026. 2. Review of the document titled