Montana Skin Cancer And Dermatology Center (Mscdc)

CLIA Laboratory Citation Details

2
Total Citations
10
Total Deficiencyies
5
Unique D-Tags
CMS Certification Number 27D0410607
Address 1905 W College Street, Bozeman, MT, 59718
City Bozeman
State MT
Zip Code59718
Phone(406) 587-4432

Citation History (2 surveys)

Survey - April 23, 2024

Survey Type: Standard

Survey Event ID: FM6L11

Deficiency Tags: D5217 D5421 D5475 D5217 D5421 D5475

Summary:

Summary Statement of Deficiencies D5217 EVALUATION OF PROFICIENCY TESTING PERFORMANCE CFR(s): 493.1236(c)(1) At least twice annually, the laboratory must verify the accuracy of any test or procedure it performs that is not included in subpart I of this part. This STANDARD is not met as evidenced by: Based on a review of testing records, personnel records, and an interview with testing personnel (TP) #8, the laboratory failed to perform twice yearly accuracy verification of its method to detect the presence or absence of fungi, MOHS surgery procedure, and immunohistochemical (IHC) test method for years 2022 and 2023. Findings: 1. A review of personnel records revealed the laboratory failed to perform one of the two accuracy verifications for its method to detect the presence or absence of fungi and MOHS surgery procedures for years 2022 and 2023. 2. A review of personnel records lacked documentation of twice annually accuracy verification for IHC testing for QXS-Enh Mart-1 and QXS-Enh CK5 for years 2022 and 2023. 3. A review of the laboratory's procedures failed to define its twice yearly accuracy verification and evaluation parameters. 4. Interview on April 23, 2024, at 10:30 AM with testing personnel (TP) #8 confirmed the laboratory failed to perform twice annually accuracy verification of its method to detect the presence or absence of fungi, MOHS surgery procedure, and immunohistochemical (IHC) staining method for years 2022 and 2023. D5421 ESTABLISHMENT AND VERIFICATION OF PERFORMANCE CFR(s): 493.1253(b)(1) Each laboratory that introduces an unmodified, FDA-cleared or approved test system must do the following before reporting patient test results: (1)(i) Demonstrate that it can obtain performance specifications comparable to those established by the manufacturer for the following performance characteristics: (1)(i)(A) Accuracy. (1)(i) Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 2 -- (B) Precision. (1)(i)(C) Reportable range of test results for the test system. (1)(ii) Verify that the manufacturer's reference intervals (normal values) are appropriate for the laboratory's patient population. This STANDARD is not met as evidenced by: Based on a review of testing records, procedures and an interview with testing personnel (TP) #8, the laboratory failed to verify the performance specifications of two out of two immunohistochemical (IHC) tests before reporting patient test results from April 23, 2022, to April 23, 2024. Findings: 1. A review of testing records revealed the laboratory testing menu included IHC testing for QXS-Enh Mart-1 and QXS-Enh CK5 from April 23, 2022, to April 23, 2024. 2. The laboratory lacked documentation of the IHC tests' performance specifications (accuracy and precision) that were comparable to those established by the manufacturer of Novodiax Q-Stain QXS-Enh Mart-1 Antibody Reagent and Novodiax Q-Stain QXS-Enh CK5 Antibody Reagent prior to reporting patient test results. 3. A review of the laboratory's procedures failed to define a process to verify new instruments or assays prior to reporting patient test results. 4. Interview on April 23, 2024, at 11:30 AM with testing personnel (TP) #8 confirmed the laboratory failed to verify the manufacturer's performance specifications of Mart-1 and CK5 immunohistochemical (IHC) tests prior to reporting patient test results from April 23, 2022, to April 23, 2024. D5475 CONTROL PROCEDURES CFR(s): 493.1256(e)(3)(g) (e) For reagent, media, and supply checks, the laboratory must do the following: (e) (3) Check fluorescent and immunohistochemical stains for positive and negative reactivity each time of use. (g) The laboratory must document all control procedures performed. This STANDARD is not met as evidenced by: Based on a review of testing records, procedures, and an interview with testing personnel (TP) #8, the laboratory failed to perform a positive and negative quality control check for immunohistochemical (IHC) stains each time of use before reporting three out of three patient test results and failed to document the results of negative controls performed from April 23, 2022, to April 23, 2024. Findings: 1. A review of patient testing records for the IHC test (QXS-Enh Mart-1) revealed the laboratory failed to perform a positive and negative quality control for each time the test was performed on: a. October 3, 2022 (AB22-512) lacked a negative control. b. March 8, 2023 (AB23-125 and AB23-129) lacked one positive and one negative control. c. March 24, 2023 (MB23-109 and MB23-110) lacked one negative control. 2. A review of the "Stain Quality Assurance Log" lacked documentation of reactions for each IHC negative and positive control performed from April 23, 2022, to April 23, 2024. 3. An interview on April 23, 2024, at 1:30 PM with testing personnel (TP) #8 confirmed the laboratory failed to perform a positive and negative quality control check for each time the IHC test (QXS-Enh Mart-1) was performed for three out of three patient tests and document the results of negative controls performed from April 23, 2022, to April 23, 2024. -- 2 of 2 --

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access

Survey - July 7, 2022

Survey Type: Standard

Survey Event ID: 6IGV11

Deficiency Tags: D5209 D5209 D5601 D5601

Summary:

Summary Statement of Deficiencies D5209 PERSONNEL COMPETENCY ASSESSMENT POLICIES CFR(s): 493.1235 As specified in the personnel requirements in subpart M, the laboratory must establish and follow written policies and procedures to assess employee and, if applicable, consultant competency. This STANDARD is not met as evidenced by: Based on review of personnel files for 2020, 2021 and 2022 Competency Review and Personnel Training, and interview with laboratory Testing Personnel (TP) #9 , the laboratory failed to perform competency for year 2021. Findings: 1. Review of Personnel Records lacked documentation to prove either semiannually competency during the 1st year of patient testing or annual competency of testing personnel for year 2021 for 6 out of 10 testing personnel (TP#3, TP#5, TP#6, TP#7, TP#8 and TP#9). 2. Interview on July 7, 2022 at 10:00 AM with TP#9, confirmed the lack of documentation to prove competency assessments for 6 of 10 testing personnel for year 2021. D5601 HISTOPATHOLOGY CFR(s): 493.1273(a)(f) (a) As specified in 493.1256(e)(3), fluorescent and immunohistochemical stains must be checked for positive and negative reactivity each time of use. For all other differential or special stains, a control slide of known reactivity must be stained with each patient slide or group of patient slides. Reactions of the control slide with each special stain must be documented. (f) The laboratory must document all control procedures performed, as specified in this section. Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 2 -- This STANDARD is not met as evidenced by: Based on record review and interview with Testing Personnel (TP) #9, the laboratory failed to perform quality control (QC) tissue slides to check for positive and negative reactivity of immunohistochemical stains used from April 1, 2022 to July 7, 2022. Findings: 1. A review of Mohs logs lacked record of QC positive and negative reactivity for Mart-1 and CK5 IHC staining each day of use on patient's frozen tissues. 2. Review of Novodiax instruction for use (IFU) for Mart-1 and CK5 Antibody Reagent revealed, "Quality Control Procedure: Positive and negative controls should be run simultaneously with patient specimens." 3. Review of laboratory procedure "Novodaix Protocol for frozen Tissue" lacked quality controls processing and documentation instructions. 4. Interview with TP #9 on July 7, 2022 at 10:30 AM, confirmed the laboratory failed to perform positive and negative quality control (QC) each day of IHC staining with Mart-1 or CK5 Antibody Reagents from April 1, 2022 to July 7, 2022. -- 2 of 2 --

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access