Mrh Med Grp-Verona Family Medicine & Urgent Care

CLIA Laboratory Citation Details

4
Total Citations
13
Total Deficiencyies
9
Unique D-Tags
CMS Certification Number 25D2160281
Address 5024 Raymond Ave, Tupelo, MS, 38801
City Tupelo
State MS
Zip Code38801
Phone662 432-0782
Lab DirectorMONICA MAY

Citation History (4 surveys)

Survey - April 10, 2025

Survey Type: Standard

Survey Event ID: U8N611

Deficiency Tags: D5429 D5437 D6019 D6041 D6049

Summary:

Summary Statement of Deficiencies D5429 MAINTENANCE AND FUNCTION CHECKS CFR(s): 493.1254(a)(1) (a)(1) Maintenance as defined by the manufacturer and with at least the frequency specified by the manufacturer. This STANDARD is not met as evidenced by: Based on review of preventive maintenance logs for the CDS Medonic M-Series Hematology analyzer from 5/1/2023 through 3/31/2025, interviews with Laboratory Director/Technical Consultant (LD/TC) and Testing Personnel #1 (TP #1) and lack of documentation of preventive maintenance, the laboratory failed to perform and document monthly maintenance for five of twenty-three months. Findings include: 1. Review of preventive maintenance logs for the CDS Medonic M-Series Hematology analyzer from 5/1/2023 through 3/31/2025, revealed no documentation of the following monthly maintenance procedures for five months, from November 2024 through March 2025. a. Monthly maintenance procedures- "Monthly Cleaning" and "Clot Prevention." 2. An interview with the LD/TC and TP #1 on 4/10/2025 at 1:30 p. m., confirmed monthly maintenance was not performed as required by the manufacturer. THIS IS A REPEAT DEFICIENCY D5437 CALIBRATION AND CALIBRATION VERIFICATION CFR(s): 493.1255(a) (a )Unless otherwise specified in this subpart, for each applicable test system the laboratory must perform and document calibration procedures-- (a)(1) Following the manufacturer's test system instructions, using calibration materials provided or specified, and with at least the frequency recommended by the manufacturer; (a)(2) Using the criteria verified or established by the laboratory as specified in 493.1253(b) (3)-- (a)(2)(i) Using calibration materials appropriate for the test system and, if Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 3 -- possible, traceable to a reference method or reference material of known value; and (a) (2)(ii) Including the number, type, and concentration of calibration materials, as well as acceptable limits for and the frequency of calibration; and (a)(3) Whenever calibration verification fails to meet the laboratory's acceptable limits for calibration verification. This STANDARD is not met as evidenced by: Based on review of the CDS Medonic M-Series Hematology analyzer records including quality control, maintenance, and calibration and confirmation with the Laboratory Director/Technical Consultant (LD/TC) and Testing Personnel #1 (TP #1) as listed on the Centers for Medicare and Medicaid Services 209 personnel form, the laboratory failed to perform calibration on complete blood count (CBC) performed on the CDS Medonic M-Series every six months as required by the instrument's manufacturer instructions for one of four six-month calibrations. Findings include: 1. Review of the CDS Medonic M-Series calibration records revealed calibration was performed on 6/7/2023, 2/21/2024, and 7/12/2024. There has been no calibration performed on the Medonic Hematology analyzer since 7/12/2024. This timeframe of calibrations performed has exceeded the requirements of the manufacturer. 2. Review of the CDS Medonic M-Series calibration records revealed one of four calibrations from 6/7/2023 to 4/10/2025 were not performed every six months as required by the manufacturer. 3. An interview with the LD/TC and TP #1 on 4/10/2025 at 2:00 p.m., confirmed CBC calibrations were not performed every six months on the CDS Medonic M-Series Hematology analyzer. THIS IS A REPEAT DEFICIENCY D6019 LABORATORY DIRECTOR RESPONSIBILITIES CFR(s): 493.1407(e)(4)(iv) (e)(4)(iv) An approved

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access

Survey - March 23, 2023

Survey Type: Standard

Survey Event ID: MNZ311

Deficiency Tags: D5411 D5429 D5437 D6041

Summary:

Summary Statement of Deficiencies D5411 TEST SYSTEMS, EQUIPMENT, INSTRUMENTS, REAGENT CFR(s): 493.1252(a) Test systems must be selected by the laboratory. The testing must be performed following the manufacturer's instructions and in a manner that provides test results within the laboratory's stated performance specifications for each test system as determined under 493.1253. This STANDARD is not met as evidenced by: Based on review of manufacturer's specifications for the Clinical Diagnostic Solutions (CDS) Medonic hematology analyzer, laboratory temperature records since the last survey on 9/3/2021, and interview with the technical consultant on 3/23/2023 at 1:00 p.m., the laboratory failed to follow the manufacturer's instructions for operation of the CDS Medonic hematology analyzer at the proper temperature and humidity for sixteen of nineteen months reviewed. Findings include: 1. Review of manufacturer's specifications for the CDS Medonic hematology analyzer revealed the operating temperature is 18 - 32 degrees Celsius (64 - 90 degrees Fahrenheit) and the humidity should be 80 percent or below. 2. Review of laboratory temperature records since the last survey on 9/3/2021 revealed no documentation of room temperature and humidity for sixteen of nineteen months for the room in which patient complete blood count (CBC) testing was performed on the CDS Medonic hematology analyzer from 9/3 /2021 until 1/19/2023, to ensure the proper operating temperature and humidity was maintained. 3. In an interview on 3/23/2023 at 1:00 p.m., the technical consultant stated there was no documentation of room temperature and humidity from the last survey on 9/3/2021 until 1/19/2023. D5429 MAINTENANCE AND FUNCTION CHECKS CFR(s): 493.1254(a)(1) Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 3 -- For unmodified manufacturer's equipment, instruments, or test systems, the laboratory must perform and document maintenance as defined by the manufacturer and with at least the frequency specified by the manufacturer. This STANDARD is not met as evidenced by: Based on review of preventive maintenance logs for the CDS Medonic hematology analyzer from 9/3/2021 through 3/23/2023 and lack of documentation of preventive maintenance for six months from November 2021 through April 2022, the laboratory failed to perform and document daily and monthly maintenance for six of nineteen months and six-month maintenance for eleven of nineteen months. Findings include: 1. Review of preventive maintenance logs for the CDS Medonic hematology analyzer since the last survey on 9/3/2021 through 3/23/2023 revealed no documentation of the following daily and monthly maintenance procedures for six months, from November 2021 through April 2022, of the nineteen months reviewed. a. Daily maintenance procedure "Clean Probe with alcohol." Daily maintenance was not performed for six of nineteen months (182 days) reviewed. b. Monthly maintenance procedures "Monthly Cleaning" and "Clot Prevention." Monthly maintenance was not performed for six of nineteen months reviewed. 2. Review of preventive maintenance logs for the CDS Medonic hematology analyzer since the last survey on 9/3/2021 through 3/23 /2023 revealed no documentation of the following six-month maintenance procedure for eleven months, from 9/3/2021 until 8/8/2022. a. Six-month maintenance "Boule Cleaning Kit Procedure." Six-month maintenance was not performed for one of one six-month periods during the timeframe reviewed. D5437 CALIBRATION AND CALIBRATION VERIFICATION CFR(s): 493.1255(a) Unless otherwise specified in this subpart, for each applicable test system the laboratory must perform and document calibration procedures-- (1) Following the manufacturer's test system instructions, using calibration materials provided or specified, and with at least the frequency recommended by the manufacturer; (2) Using the criteria verified or established by the laboratory as specified in 493.1253(b) (3)-- (2)(i) Using calibration materials appropriate for the test system and, if possible, traceable to a reference method or reference material of known value; and (2)(ii) Including the number, type, and concentration of calibration materials, as well as acceptable limits for and the frequency of calibration; and (3) Whenever calibration verification fails to meet the laboratory's acceptable limits for calibration verification. This STANDARD is not met as evidenced by: Based on review of the CDS M-Series Hematology Analyzer Procedure Manual, calibration records for the CDS Medonic hematology analyzer since the last survey on 9/3/2021, and interview with the technical consultant on 3/23/2023 at 12:10 p.m., the laboratory failed to perform calibration at a minimum of every six months, as recommended by the manufacturer, from 9/3/2021 until 1/17/2023 for three of three six-month periods. Findings include: 1. Review of the CDS M-Series Hematology Analyzer Procedure Manual revealed the calibration procedure states, "Calibration must be performed upon setup of the instrument and then at a minimum of every six months." 2. Review of calibration records for the CDS Medonic hematology analyzer from 9/3/2021 through 3/23/2023 revealed calibration was not performed for sixteen months, from 9/3/2021 until 1/17/2023. Calibration was not performed for three of three six-month periods reviewed. 3. In an interview on 3/23/2023 at 12:10 p.m., the -- 2 of 3 -- technical consultant confirmed that calibration was not performed since the last survey on 9/3/2021 until 1/17/2023. D6041 TECHNICAL CONSULTANT RESPONSIBILITIES CFR(s): 493.1413(b)(3) (b) The technical consultant is responsible for-- (b)(3) Enrollment and participation in an HHS approved proficiency testing program commensurate with the services offered; This STANDARD is not met as evidenced by: Based on review of the laboratory's proficiency testing (PT) records since the last survey on 9/3/2021, the Centers for Medicare and Medicaid Services (CMS) data system CASPER Report 0155D, and interview with the current technical consultant on 3/23/2023 at 10:30 a.m., the former technical consultant failed to ensure that the laboratory participated in a Health and Human Services (HHS) approved proficiency testing program for complete blood count (CBC) testing for one of three PT events for 2021 (2021-3rd event) and failed to enroll the laboratory in an HHS approved PT program for the year of 2022. Findings include: 1. Review of the laboratory's PT records since the last survey on 9/3/2021 revealed no documentation of the performance of PT for the third hematology event of 2021 and no PT records for the year of 2022. 2. There was no documentation of the performance of PT after the second event of 2021 on 8/2/2021 until the first event of 2023 on 3/9/23. a. Review of the CMS data system CASPER Report 0155D revealed a score of zero for the third hematology PT event of 2021 for failure to participate (one of three events in 2021). b. Review of the CMS data system CASPER Report 0155D revealed no scores for three of three hematology PT events in 2022. 3. In an interview on 3/23/2023 at 10:30 a.m., the current technical consultant confirmed that hematology PT was not performed for the third event of 2021 and stated the former technical consultant failed to enroll the laboratory in an HHS approved proficiency testing program for CBC testing for the year of 2022. -- 3 of 3 --

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access

Survey - September 3, 2021

Survey Type: Standard

Survey Event ID: BHJR11

Deficiency Tags: D6031

Summary:

Summary Statement of Deficiencies D6031 LABORATORY DIRECTOR RESPONSIBILITIES CFR(s): 493.1407(e)(13) The laboratory director is responsible for the overall operation and administration of the laboratory, including the employment of personnel who are competent to perform test procedures, and record and report test results promptly, accurate, and proficiently and for assuring compliance with the applicable regulations. (e) The laboratory director must-- (e)(13) Ensure that an approved procedure manual is available to all personnel responsible for any aspect of the testing process; This STANDARD is not met as evidenced by: Based on review of the manufacturer's procedure manual for the Boule Medonic M Series hematology analyzer, the written laboratory procedure manual and the MRH policy and procedure manual put into use for patient testing, the current laboratory director failed to approve the manuals for use by testing personnel as of the day of survey 9/3/21. Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 1 --

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access

Survey - August 22, 2019

Survey Type: Standard

Survey Event ID: N6VC11

Deficiency Tags: D5403 D5413 D5429

Summary:

Summary Statement of Deficiencies D5403 PROCEDURE MANUAL CFR(s): 493.1251(b) The procedure manual must include the following when applicable to the test procedure: (1) Requirements for patient preparation; specimen collection, labeling, storage, preservation, transportation, processing, and referral; and criteria for specimen acceptability and rejection as described in 493.1242. (2) Microscopic examination, including the detection of inadequately prepared slides. (3) Step-by-step performance of the procedure, including test calculations and interpretation of results. (4) Preparation of slides, solutions, calibrators, controls, reagents, stains, and other materials used in testing. (5) Calibration and calibration verification procedures. (6) The reportable range for test results for the test system as established or verified in 493.1253. (7) Control procedures. (8)

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access