Summary:
Summary Statement of Deficiencies D5217 EVALUATION OF PROFICIENCY TESTING PERFORMANCE CFR(s): 493.1236(c)(1) At least twice annually, the laboratory must verify the accuracy of any test or procedure it performs that is not included in subpart I of this part. This STANDARD is not met as evidenced by: Based on review of proficiency testing (PT) records and interview with the general supervisor, the laboratory failed to verify the accuracy of glycogen enzymatic pathways (GEP) test at least twice annually during 2018. Findings: 1. Review of PT records revealed the laboratory did not have documentation to show it verified the accuracy of the GEP test at least twice annually during 2018. 2. Interview with the general supervisor on September 25, 2019 at 12:15 PM confirmed the laboratory failed to verify the accuracy of GEP test at least twice annually during 2018. D5403 PROCEDURE MANUAL CFR(s): 493.1251(b) The procedure manual must include the following when applicable to the test procedure: (1) Requirements for patient preparation; specimen collection, labeling, storage, preservation, transportation, processing, and referral; and criteria for specimen acceptability and rejection as described in 493.1242. (2) Microscopic examination, including the detection of inadequately prepared slides. (3) Step-by-step performance of the procedure, including test calculations and interpretation of results. (4) Preparation of slides, solutions, calibrators, controls, reagents, stains, and other materials used in testing. (5) Calibration and calibration verification procedures. (6) The reportable range for test results for the test system as established or verified in 493.1253. (7) Control procedures. (8)