Northfield Hospital

CLIA Laboratory Citation Details

4
Total Citations
11
Total Deficiencyies
7
Unique D-Tags
CMS Certification Number 24D0667985
Address 2000 North Avenue, Northfield, MN, 55057
City Northfield
State MN
Zip Code55057
Phone(507) 646-1000

Citation History (4 surveys)

Survey - July 28, 2026

Survey Type: Standard

Survey Event ID: V7IG11

Deficiency Tags: D0000 D5775

Summary:

Summary Statement of Deficiencies D0000 The Northfield Hospital laboratory was found to be out of compliance with the regulations of the Clinical Laboratory Improvement Amendments of 1988 (42 C.F.R. part 493) upon completion of the recertification survey completed on July 23, 2026. The following standard-level deficiencies were cited: 493.1281 Comparison of test results . D5775 COMPARISON OF TEST RESULTS CFR(s): 493.1281(a)(c) (a) If a laboratory performs the same test using different methodologies or instruments, or performs the same test at multiple testing sites, the laboratory must have a system that twice a year evaluates and defines the relationship between test results using the different methodologies, instruments, or testing sites. This STANDARD is not met as evidenced by: . Based on observation, document review, and interview with laboratory personnel, the laboratory failed to evaluate and document the relationship between four of five duplicate testing methods at least twice annually in 2024 and 2025. Findings are as follows: 1. The laboratory performed Chemistry, Hematology, and Immunohematology testing as confirmed by the General Supervisor (GS) during a tour of the laboratory at 10:45 a..m. on 7/22/26. 2. The following test systems were observed as present and available for use during the tour: Chemistry: QuidelOrtho Vitros 7600 XT (primary implemented in December 2025) Abbott iStat (high sensitivity troponin and creatinine) Radiometer ABL90-Flex Plus (lactate) Hematology: Sysmex XN-550 (autodifferential) Nikon Exlipse E400 (manual differential) Immunohematology: QuidelOrtho VisionSwift (primary) Manual Ortho Workstation Manual Tube Method A QuidelOrtho Vitros 5600 was used for the primary Chemsitry testing prior to the implementation of the QuidelOrtho Vitros 7600 XT. 3. Twice annual comparison of test results obtained from multiple non-waived test methods was required as etablished in the Comparison of Test Results from Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 2 -- Different Platforms policy found in the laboratory's Quality Control and Quality Assurance manual reviewed on the date of survey. 4. Laboratory records included twice annual comparison for hemoglobin between the Radiometer ABL90-Flex Plus and the Sysmex XN-550. Comparison documentation of the following non-waived test systems were not found in laboratory records. See below. Chemistry: High sensitivity troponin: Abott iStat vs. Vitros 7600 XT Creatinine: Abott iStat vs. Vitros 7600 XT Lactate: Vitros 7600 XT vs. Radiometer ABL90-Flex Plus Hematology: Differential: manual (Nikon Eclipse E400) vs. automated Sysmex XN-550 Immunohematology: Type and Screen: QuidelOrtho VisionSwift vs Manual Ortho Workstation vs. tube method Crossmatch: QuidelOrtho VisionSwift vs Manual Ortho Workstation vs. tube method The laboratory was unable to provide missing documentation upon request. 5. In an interview at 3:25 p.m. on 7/23/26, the GS confirmed the above findings. . -- 2 of 2 --

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access

Survey - April 4, 2024

Survey Type: Standard

Survey Event ID: JKUS11

Deficiency Tags: D5403 D5445 D5429

Summary:

Summary Statement of Deficiencies D5403 PROCEDURE MANUAL CFR(s): 493.1251(b) The procedure manual must include the following when applicable to the test procedure: (1) Requirements for patient preparation; specimen collection, labeling, storage, preservation, transportation, processing, and referral; and criteria for specimen acceptability and rejection as described in 493.1242. (2) Microscopic examination, including the detection of inadequately prepared slides. (3) Step-by-step performance of the procedure, including test calculations and interpretation of results. (4) Preparation of slides, solutions, calibrators, controls, reagents, stains, and other materials used in testing. (5) Calibration and calibration verification procedures. (6) The reportable range for test results for the test system as established or verified in 493.1253. (7) Control procedures. (8)

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access

Survey - May 19, 2022

Survey Type: Standard

Survey Event ID: YD0B11

Deficiency Tags: D5445 D5215 D5775

Summary:

Summary Statement of Deficiencies D5215 EVALUATION OF PROFICIENCY TESTING PERFORMANCE CFR(s): 493.1236(b)(2) The laboratory must verify the accuracy of any analyte, specialty or subspecialty assigned a proficiency testing score that does not reflect laboratory test performance (that is, when the proficiency testing program does not obtain the agreement required for scoring as specified in subpart I of this part, or the laboratory receives a zero score for nonparticipation, or late return or results). This STANDARD is not met as evidenced by: . Based on document review and interview with laboratory personnel, the laboratory failed to verify the accuracy of one 2021 non-graded proficiency testing (PT) results for a non-regulated analyte when the PT program did not obtain the agreement required for scoring. Findings are as follows: 1. The laboratory performed Bacteriology, Mycology, and Bacteriology testing as confirmed by the General Supervisor (GS) during a tour of the laboratory at 8:05 a.m. on 05/19/22. 2. The laboratory performed PT using the American Proficiency Institute (API) PT provider. 3. One Vaginal Wet Preparation (VWP) result from the third 2021 Hematology /Coagulation PT event was not graded by API due to lack of consensus. See below. Sample ID Analyte VKP-03 VWP 4. The API report referred the laboratory to the expected result data summary for evaluation of the non-graded test result. The data summary for the above sample was not present in laboratory records. An evaluation of the non-graded result was not found in laboratory records. The laboratory was unable to provide an evaluation of the non-graded result upon request. 5. The laboratory's Proficiency Testing Policy found in the Laboratory Standard Operating Procedures manual did not include direction to evaluate non-graded PT results. 6. In an interview at 12:20 p.m. on 05/19/22, the GS confirmed the above finding. . D5445 CONTROL PROCEDURES CFR(s): 493.1256(d)(1)(2)(g) Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 2 -- Unless CMS Approves a procedure, specified in Appendix C of the State Operations Manual (CMS Pub. 7), that provides equivalent quality testing, the laboratory must-- (d)(1) Perform control procedures as defined in this section unless otherwise specified in the additional specialty and subspecialty requirements at 493.1261 through 493.1278. (d)(2) For each test system, perform control procedures using the number and frequency specified by the manufacturer or established by the laboratory when they meet or exceed the requirements in paragraph (d)(3) of this section. (g) The laboratory must document all control procedures performed. This STANDARD is not met as evidenced by: . Based on observation, document review, and interview with laboratory personnel, the laboratory failed to perform minimum quality control activities required for a Microbiology test system. Findings are as follows: 1. The laboratory performed Microbiology testing as confirmed by the General Supervisor (GS) during a tour of the laboratory at 8:05 a.m. on 05/19/22. 2. A Cepheid GeneXpert System was observed as present and available for use during the tour. The GS stated the SARS- CoV-2/Flu/RSV test was performed on this system. 3. Quality control (QC) testing with positive and negative control materials was required with each new lot and shipment as indicated in the SARS-CoV-2/Flu/RSV on the Cepheid GeneXpert System procedure located in the Procedure Manual/GeneXpert manual. 4. The laboratory's QC log indicated QC was performed upon receipt of new lots and shipments in the timeframe reviewed, July - September 2021. 5. The laboratory did not establish an Individual Quality Control Plan to reduce the frequency of QC performance from 2 levels of control material each day of patient testing. 6. In an interview at 4:00 p.m. on 05/19/22, the GS confirmed the above finding. . D5775 COMPARISON OF TEST RESULTS CFR(s): 493.1281(a)(c) (a) If a laboratory performs the same test using different methodologies or instruments, or performs the same test at multiple testing sites, the laboratory must have a system that twice a year evaluates and defines the relationship between test results using the different methodologies, instruments, or testing sites. (c) The laboratory must document all test result comparison activities. This STANDARD is not met as evidenced by: . Based on observation, document review, and interview with laboratory personnel, the laboratory failed to evaluate the relationship between test results obtained from two Immunohematology testing methods at least twice annually 2021. Findings are as follows: 1. The laboratory performed Immunohematology testing as confirmed by the General Supervisor (GS) during a tour of the laboratory at 8:05 a.m. on 05/19/22. 2. The following Immunohematology methods were observed as present and available for use during the tour: Ortho Vision automated analyzer - primary method Bio-Rad IH-Centrifuge L and IH-Incubator L manual system - back up method 3. A twice annual process for comparison of test results obtained from multiple methods was not found in the laboratory's procedure manuals. 4. Comparison of test results obtained from the two Immunohematology methods was not found in laboratory records from 2021. The laboratory was unable to provide documentation of test comparisons upon request. 5. In an interview at 4:25 p.m. on 05/19/22, the GS confirmed the above finding. -- 2 of 2 --

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access

Survey - May 31, 2018

Survey Type: Standard

Survey Event ID: G3MK11

Deficiency Tags: D5431 D5403 D5445

Summary:

Summary Statement of Deficiencies D5403 PROCEDURE MANUAL CFR(s): 493.1251(b) The procedure manual must include the following when applicable to the test procedure: (1) Requirements for patient preparation; specimen collection, labeling, storage, preservation, transportation, processing, and referral; and criteria for specimen acceptability and rejection as described in 493.1242. (2) Microscopic examination, including the detection of inadequately prepared slides. (3) Step-by-step performance of the procedure, including test calculations and interpretation of results. (4) Preparation of slides, solutions, calibrators, controls, reagents, stains, and other materials used in testing. (5) Calibration and calibration verification procedures. (6) The reportable range for test results for the test system as established or verified in 493.1253. (7) Control procedures. (8)

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access