Northwest Ga Hematology & Oncology,

CLIA Laboratory Citation Details

1
Total Citation
5
Total Deficiencyies
5
Unique D-Tags
CMS Certification Number 11D0676409
Address 1504 North Thornton Avenue Suite 102, Dalton, GA, 30720
City Dalton
State GA
Zip Code30720
Phone706 226-0508
Lab DirectorMONICA VERMA

Citation History (1 survey)

Survey - May 28, 2024

Survey Type: Special

Survey Event ID: NPB611

Deficiency Tags: D0000 D2130 D6016 D2016 D6000

Summary:

Summary Statement of Deficiencies D0000 A proficiency testing desk review was completed on May 28, 2024. At the time of the review, the laboratory was not in compliance with the Clinical Laboratory Improvement Amendments of 1988, 42 CFR 493.1 through 42 CFR 493.1780. The following Condition level deficiencies were cited: D2016, D6000 D2016 SUCCESSFUL PARTICIPATION CFR(s): 493.803(a)(b)(c) (a) Each laboratory performing nonwaived testing must successfully participate in a proficiency testing program approved by CMS, if applicable, as described in subpart I of this part for each specialty, subspecialty, and analyte or test in which the laboratory is certified under CLIA. (b) Except as specified in paragraph (c) of this section, if a laboratory fails to participate successfully in proficiency testing for a given specialty, subspecialty, analyte or test, as defined in this section, or fails to take remedial action when an individual fails gynecologic cytology, CMS imposes sanctions, as specified in subpart R of this part. (c) If a laboratory fails to perform successfully in a CMS- approved proficiency testing program, for the initial unsuccessful performance, CMS may direct the laboratory to undertake training of its personnel or to obtain technical assistance, or both, rather than imposing alternative or principle sanctions except when one or more of the following conditions exists: (1) There is immediate jeopardy to patient health and safety. (2) The laboratory fails to provide CMS or a CMS agent with satisfactory evidence that it has taken steps to correct the problem identified by the unsuccessful proficiency testing performance. (3) The laboratory has a poor compliance history. This CONDITION is not met as evidenced by: Based on review of the Centers for Medicare and Medicaid (CMS) CASPER 155 report and review of the American Proficiency Institute (API) reports, the laboratory Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 2 -- failed to successfully participate in proficiency testing (PT) in 2 of 3 testing events for White Blood Cell (WBC) Differential (Diff) resulting in the initial unsuccessful participation for WBC Diff. Refer to D 2130 D2130 HEMATOLOGY CFR(s): 493.851(f) Failure to achieve satisfactory performance for the same analyte in two consecutive events or two out of three consecutive testing events is unsuccessful performance. This STANDARD is not met as evidenced by: Based on review of the CMS CASPER 155 report and review of API PT reports, the laboratory failed to demonstrate satisfactory performance in 2 of 3 testing events for WBC Diff, resulting in the initial unsuccessful participation for WBC Diff. Findings: 1. A review of Casper Report 155 disclosed the laboratory failed for WBC Diff on the following: 2023 Event 2 Score 40% 2024 Event 1 Score 28% 2. A review of the laboratory's API Reports confirmed the laboratory failed WBC Diff with the aforementioned scores. D6000 MODERATE COMPLEXITY LABORATORY DIRECTOR CFR(s): 493.1403 The laboratory must have a director who meets the qualification requirements of 493. 1405 of this subpart and provides overall management and direction in accordance with 493.1407 of this subpart. This CONDITION is not met as evidenced by: Based on review of the CMS CASPER 155 report and review of the API reports, the laboratory director failed to provide overall management and direction for successful participation in PT. The laboratory director failed to ensure PT samples were tested as required. Refer to D6016 D6016 LABORATORY DIRECTOR RESPONSIBILITIES CFR(s): 493.1407(e)(4)(i) The laboratory director is responsible for the overall operation and administration of the laboratory, including the employment of personnel who are competent to perform test procedures, and record and report test results promptly, accurate, and proficiently and for assuring compliance with the applicable regulations. (e) The laboratory director must-- (e)(4)(i) Ensure that the proficiency testing samples are tested as required under Subpart H of this part; This STANDARD is not met as evidenced by: Based on review of the CMS CASPER Report 155 and the API PT evaluation reports, the laboratory director failed to ensure successful PT participation in 2 of 3 testing events. Refer to D 2130 -- 2 of 2 --

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access