Paragon Skin Dermatology

CLIA Laboratory Citation Details

1
Total Citation
5
Total Deficiencyies
3
Unique D-Tags
CMS Certification Number 31D2312374
Address 1616 Route 72 West, Manahawkin, NJ, 08050
City Manahawkin
State NJ
Zip Code08050
Phone(732) 234-7546

Citation History (1 survey)

Survey - December 17, 2025

Survey Type: Standard

Survey Event ID: G1LM11

Deficiency Tags: D5217 D5417 D5417 D5601 D5601

Summary:

Summary Statement of Deficiencies D5217 EVALUATION OF PROFICIENCY TESTING PERFORMANCE CFR(s): 493.1236(c)(1) At least twice annually, the laboratory must verify the accuracy of any test or procedure it performs that is not included in subpart I of this part. This STANDARD is not met as evidenced by: Based on the lack of the Biannual Assessments (BA) records and interview with the Office Manager (OM) the laboratory failed to verify the accuracy of Histopathology testing biannually from 10/10/24 to 12/17/25. The finding includes: 1. There was no documented evidence the laboratory verified the accuracy of histopathology testing. 2. The OM confirmed on 12/17/25 at 12:30 pm, the laboratory did not verify the accuracy of Histopathology testing. D5417 TEST SYSTEMS, EQUIPMENT, INSTRUMENTS, REAGENT CFR(s): 493.1252(d) (d) Reagents, solutions, culture media, control materials, calibration materials, and other supplies must not be used when they have exceeded their expiration date, have deteriorated, or are of substandard quality. This STANDARD is not met as evidenced by: Based on surveyor observation of Histology reagents and interview with the Office Manager (OM), the laboratory usedf and failed to discard expired Histopathology reagents from 10/10/24 to 12/17/25. The findings include: 1. Platinumline Eosin Y Stain Solution 1% W/V in Alcohol Lot 1924210 Exp 2021-09-04. 2. Platinumline Hematoxylin Stain Soluaiton,GILL III Lot: 2002408 EXP: 2022-01-28 3. Platinumline Hematoxyulin Stain solution, Gill III Lot: 2105416 EXP: 2023-03-04 4. Platinumline Eosin Y Stain Solution 1% W/V in Alcohol Lot 2110308 Exp 2023-04- Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 2 -- 16. 5. Approximately 200 patients were tested with expired reagent. 6. The OM confirmed on 12/17/25 at 1:00 pm that the laboratory used and failed to discard expired reagents. D5601 HISTOPATHOLOGY CFR(s): 493.1273(a)(f) (a) As specified in 493.1256(e)(3), fluorescent and immunohistochemical stains must be checked for positive and negative reactivity each time of use. For all other differential or special stains, a control slide of known reactivity must be stained with each patient slide or group of patient slides. Reactions of the control slide with each special stain must be documented. This STANDARD is not met as evidenced by: Based on the lack of Quality Control (QC) records and interview with the Office Manager (OM), the laboratory failed to document Hematoxylin and Eosin (H&E) control slide reactions for Histopathology testing from 10/10/24 to 12/19/25. The findings include: 1. The laboratory did not document H&E stain QC reactions during the time period stated above. 2. The laboratory read and reported approximately 200 patients in the above time period. 3. The OM confirmed on 12/19/25 at 12:30 pm that the laboratory did not document H&E QC stain reaction. -- 2 of 2 --

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access