Paul Maller Md

CLIA Laboratory Citation Details

3
Total Citations
22
Total Deficiencyies
9
Unique D-Tags
CMS Certification Number 33D0142626
Address 22 South Tyson Ave, Floral Park, NY, 11001
City Floral Park
State NY
Zip Code11001
Phone516 437-5151
Lab DirectorPAUL MALLER

Citation History (3 surveys)

Survey - September 12, 2024

Survey Type: Special

Survey Event ID: DYOV11

Deficiency Tags: D0000 D2016 D2130 D6000 D6016 D0000 D2016 D2130 D6000 D6016

Summary:

Summary Statement of Deficiencies D0000 Based on a proficiency testing (PT) desk review survey performed on September 12, 2024, the laboratory was found to be out of compliance based on the following CONDITION LEVEL DEFICIENCIES: D2016 - 42 C.F.R. 493.803 Condition: Successful participation. D6000 - 42 C.F.R. 493.1403 Condition: Laboratory Director, moderate complexity. D2016 SUCCESSFUL PARTICIPATION CFR(s): 493.803(a)(b)(c) (a) Each laboratory performing nonwaived testing must successfully participate in a proficiency testing program approved by CMS, if applicable, as described in subpart I of this part for each specialty, subspecialty, and analyte or test in which the laboratory is certified under CLIA. (b) Except as specified in paragraph (c) of this section, if a laboratory fails to participate successfully in proficiency testing for a given specialty, subspecialty, analyte or test, as defined in this section, or fails to take remedial action when an individual fails gynecologic cytology, CMS imposes sanctions, as specified in subpart R of this part. (c) If a laboratory fails to perform successfully in a CMS- approved proficiency testing program, for the initial unsuccessful performance, CMS may direct the laboratory to undertake training of its personnel or to obtain technical assistance, or both, rather than imposing alternative or principle sanctions except when one or more of the following conditions exists: (1) There is immediate jeopardy to patient health and safety. (2) The laboratory fails to provide CMS or a CMS agent with satisfactory evidence that it has taken steps to correct the problem identified by the unsuccessful proficiency testing performance. (3) The laboratory has a poor compliance history. This CONDITION is not met as evidenced by: Based on review of Centers for Medicare & Medicaid Services (CMS) Proficiency Testing (PT) Certification and Survey Provider Enhanced Reporting system (CASPER 0155D) and American Proficiency Institute (API) PT summary reports, the Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 2 -- laboratory failed to successfully participate in the CMS approved PT program for two of three consecutive testing events in the Hematology subspecialty for the Platelets test analyte in 2024, resulting in non-initial unsuccessful performance. Refer to D2130. D2130 HEMATOLOGY CFR(s): 493.851(f) Failure to achieve satisfactory performance for the same analyte in two consecutive events or two out of three consecutive testing events is unsuccessful performance. This STANDARD is not met as evidenced by: Based on CMS PT CASPER 0155D and API PT summary reports from 2024, the laboratory failed to achieve satisfactory performance (80% or greater) for two of three consecutive testing events in the Hematology subspecialty for the Platelets test analyte. FINDINGS: 1. A review of the CASPER 155 report revealed the following unsatisfactory scores: Platelets Test Analyte: 2024 First Event = 60% 2024 Second Event = 40% 2. A review of the proficiency testing scores from API (2024) confirmed the above findings D6000 MODERATE COMPLEXITY LABORATORY DIRECTOR CFR(s): 493.1403 The laboratory must have a director who meets the qualification requirements of 493. 1405 of this subpart and provides overall management and direction in accordance with 493.1407 of this subpart. This CONDITION is not met as evidenced by: Based on review of CMS PT CASPER 0155D and API PT summary reports from 2024, the laboratory director (LD) failed to failed to provide overall management and direction of the laboratory services. Refer to D2016. D6016 LABORATORY DIRECTOR RESPONSIBILITIES CFR(s): 493.1407(e)(4)(i) The laboratory director is responsible for the overall operation and administration of the laboratory, including the employment of personnel who are competent to perform test procedures, and record and report test results promptly, accurate, and proficiently and for assuring compliance with the applicable regulations. (e) The laboratory director must-- (e)(4)(i) Ensure that the proficiency testing samples are tested as required under Subpart H of this part; This STANDARD is not met as evidenced by: Based on review of CMS PT CASPER 0155D and API PT 2024-1 and 2024-2 summary reports, the LD failed to ensure successful participation in an HHS approved proficiency testing program. Refer to D2130. -- 2 of 2 --

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access

Survey - July 2, 2018

Survey Type: Special

Survey Event ID: CZFN11

Deficiency Tags: D6000 D6016 D6000 D6016 D2016 D2096

Summary:

Summary Statement of Deficiencies D2016 SUCCESSFUL PARTICIPATION CFR(s): 493.803(a)(b)(c) (a) Each laboratory performing nonwaived testing must successfully participate in a proficiency testing program approved by CMS, if applicable, as described in subpart I of this part for each specialty, subspecialty, and analyte or test in which the laboratory is certified under CLIA. (b) Except as specified in paragraph (c) of this section, if a laboratory fails to participate successfully in proficiency testing for a given specialty, subspecialty, analyte or test, as defined in this section, or fails to take remedial action when an individual fails gynecologic cytology, CMS imposes sanctions, as specified in subpart R of this part. (c) If a laboratory fails to perform successfully in a CMS- approved proficiency testing program, for the initial unsuccessful performance, CMS may direct the laboratory to undertake training of its personnel or to obtain technical assistance, or both, rather than imposing alternative or principle sanctions except when one or more of the following conditions exists: (1) There is immediate jeopardy to patient health and safety. (2) The laboratory fails to provide CMS or a CMS agent with satisfactory evidence that it has taken steps to correct the problem identified by the unsuccessful proficiency testing performance. (3) The laboratory has a poor compliance history. This CONDITION is not met as evidenced by: Based on a proficiency testing (PT) desk review of the Center for Medicare and Medicaid Services (CMS) PT reports and PT records from the American Proficiency Institute (API) PT program, the laboratory failed to participate and perform successfully in a PT program, approved by CMS, for the test analyte Total Iron. The following scores were assigned: 2018 first event = 40% 2018 second event = 40% This is considered unsuccessful PT performance. Refer to D2096 D2096 ROUTINE CHEMISTRY CFR(s): 493.841(f) Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 2 -- Failure to achieve satisfactory performance for the same analyte or test in two consecutive testing events or two out of three consecutive testing events is unsuccessful performance. This STANDARD is not met as evidenced by: Based on a PT desk review of the CMS PT reports and PT records from the API PT program, the laboratory failed to participate and perform successfully in a PT program, approved by CMS, for the test analyte Total Iron. The following scores were assigned: 2018 first event = 40% 2018 second event = 40% This is considered unsuccessful PT performance. D6000 MODERATE COMPLEXITY LABORATORY DIRECTOR CFR(s): 493.1403 The laboratory must have a director who meets the qualification requirements of 493. 1405 of this subpart and provides overall management and direction in accordance with 493.1407 of this subpart. This CONDITION is not met as evidenced by: Based on PT desk review of the CMS PT and API PT program records, the laboratory director failed to fulfill the laboratory director's responsibilities and ensure that the laboratory achieved a satisfactory performance and successfully participate in a PT program, approved by CMS, for the test analyte Total Iron. Refer to D6016 D6016 LABORATORY DIRECTOR RESPONSIBILITIES CFR(s): 493.1407(e)(4)(i) The laboratory director is responsible for the overall operation and administration of the laboratory, including the employment of personnel who are competent to perform test procedures, and record and report test results promptly, accurate, and proficiently and for assuring compliance with the applicable regulations. (e) The laboratory director must-- (e)(4)(i) Ensure that the proficiency testing samples are tested as required under Subpart H of this part; This STANDARD is not met as evidenced by: Based on PT desk review of the CMS PT and API PT program records, the laboratory director failed to fulfill the laboratory director's responsibilities and ensure that the laboratory achieved a satisfactory performance and successfully participate in a PT program, approved by CMS, for the test analyte Total Iron. The following scores were assigned: 2018 first event = 40% 2018 second event = 40% This is considered unsuccessful PT performance. -- 2 of 2 --

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access

Survey - January 2, 2018

Survey Type: Standard

Survey Event ID: GWYD11

Deficiency Tags: D5469 D5469 D5805 D5805 D6015 D6015

Summary:

Summary Statement of Deficiencies D5469 CONTROL PROCEDURES CFR(s): 493.1256(d)(10)(g) Unless CMS Approves a procedure, specified in Appendix C of the State Operations Manual (CMS Pub. 7), that provides equivalent quality testing, the laboratory must-- Establish or verify the criteria for acceptability of all control materials. (i) When control materials providing quantitative results are used, statistical parameters (for example, mean and standard deviation) for each batch and lot number of control materials must be defined and available. (ii) The laboratory may use the stated value of a commercially assayed control material provided the stated value is for the methodology and instrumentation employed by the laboratory and is verified by the laboratory. (iii) Statistical parameters for unassayed control materials must be established over time by the laboratory through concurrent testing of control materials having previously determined statistical parameters. (g) The laboratory must document all control procedures performed. This STANDARD is not met as evidenced by: Based on a surveyor's review of Quality Control (QC) records and an interview with the laboratory director and the laboratory supervisor/testing person, the laboratory failed to program the manufacturer expected ranges, as defined on the QC assay sheets, into the TOSOH AIA 360 analyzer for Endocrinology testing from 11/1/16 through the survey date. FINDINGS: A) On 1/2/2018 at approximately 11:00 AM, the laboratory director and the laboratory supervisor/testing person confirmed the surveyor's review of QC records finding that the laboratory failed to program the established QC ranges and expected means into the TOSOH AIA 360 analyzer for the Endocrinology tests performed. B) Without the established QC limits, the surveyor could not determine if the quality control results were within the acceptable ranges for the Endocrinology analytes tested on the TOSOH AIA 360 analyzer. C) Approximately 200 patients' specimens were tested and reported for Endocrinology during this time period. Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 2 -- D5805 TEST REPORT CFR(s): 493.1291(c) The test report must indicate the following: (c)(1) For positive patient identification, either the patient's name and identification number, or a unique patient identifier and identification number. (c)(2) The name and address of the laboratory location where the test was performed. (c)(3) The test report date. (c)(4) The test performed. (c)(5) Specimen source, when appropriate. (c)(6) The test result and, if applicable, the units of measurement or interpretation, or both. (c)(7) Any information regarding the condition and disposition of specimens that do not meet the laboratory's criteria for acceptability. This STANDARD is not met as evidenced by: Based on a surveyor's review of patients reports and confirmed in an interview with the laboratory director and the laboratory supervisor/testing person at 11:00 AM on 1/2 /18, five of five patient hematology test reports failed to include the name and the address of the laboratory where test was performed. D6015 LABORATORY DIRECTOR RESPONSIBILITIES CFR(s): 493.1407(e)(4) The laboratory director is responsible for the overall operation and administration of the laboratory, including the employment of personnel who are competent to perform test procedures, and record and report test results promptly, accurate, and proficiently and for assuring compliance with the applicable regulations. (e) The laboratory director must-- (e)(4) Ensure that the laboratory is enrolled in an HHS approved proficiency testing program for the testing performed. This STANDARD is not met as evidenced by: Based on a surveyor's review of laboratory's Proficiency Testing (PT) records and interview with the laboratory supervisor/testing person and the laboratory director, the laboratory failed to enroll in Proficiency Testing (PT) for chemistry for the first event in 2017. FINDINGS: The laboratory supervisor/TP and the laboratory director confirmed on 1/2/2018 at approximately 11:00 AM that the laboratory failed to enroll in PT for the specialty of chemistry for the first event in 2017. -- 2 of 2 --