Summary:
Summary Statement of Deficiencies D2007 TESTING OF PROFICIENCY TESTING SAMPLES CFR(s): 493.801(b)(1) (b)(1) The samples must be examined or tested with the laboratory's regular patient workload by personnel who routinely perform the testing in the laboratory, using the laboratory's routine methods. This STANDARD is not met as evidenced by: Based on laboratory observation, review of the Department of Health and Human Services Centers for Medicare and Medicaid Services Laboratory Personnel Report (CLIA) (Form CMS-209), American Association of Bioanalysts Medical Laboratory Evaluation (AAB-MLE) proficiency testing (PT) records and staff interviews, the laboratory failed to ensure that three of four testing personnel (TP) who performed hematology patient testing also participated in PT testing in 2025 and 2026. The findings include: 1. Laboratory observation on 04/29/2026 at 09:30 a.m. revealed a Sysmex pocH-100i (Serial Number: A4176) used for complete blood count (CBC) testing. 2. A review of the Form CMS-209 revealed a total of four testing personnel (TP1, TP2, TP3, and TP4) who routinely perform moderately complex CBC patient testing. 3. A review of the laboratory's AAB-MLE proficiency testing records revealed that TP2, TP3 and TP4 did not participate in any hematology PT events (0 of 4 reviewed) in 2025 and 2026. 4. The laboratory director and nursing supervisor confirmed the survey findings in an interview on 04/29/2026 at 12:00 p.m. . D5417 TEST SYSTEMS, EQUIPMENT, INSTRUMENTS, REAGENT CFR(s): 493.1252(d) (d) Reagents, solutions, culture media, control materials, calibration materials, and other supplies must not be used when they have exceeded their expiration date, have deteriorated, or are of substandard quality. Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 3 -- This STANDARD is not met as evidenced by: Based on laboratory observation and staff interviews, the laboratory failed to ensure it did not use blood collection tubes for reference laboratory patient testing past their expiration date on the survey date, 04/29/2026. The findings include: 1. Laboratory observation on 04/29/2026 at 09:30 a.m. revealed the following: - (3) 2.7 milliliter (mL) Becton Dickinson (BD) Vacutainer light blue top Sodium (Na) Citrate (0.109M, 3.2%) blood collection tubes, lot number: 4222854, expiration date: 2025-05-31. - (1) 6 mL BD Vacutainer royal blue top K2EDTA (dipotassium ethylenediaminetetraacetic acid) Trace Element blood collection tube, lot number: 4229079, expiration date: 2025-07-31. 2. The laboratory director and nursing supervisor confirmed the survey findings in an interview on 04/29/2026 at 12:00 p.m. . D5781