Powell Valley Healthcare

CLIA Laboratory Citation Details

4
Total Citations
14
Total Deficiencyies
8
Unique D-Tags
CMS Certification Number 53D0681667
Address 777 Avenue H, Powell, WY, 82435
City Powell
State WY
Zip Code82435
Phone(307) 754-2267

Citation History (4 surveys)

Survey - February 18, 2026

Survey Type: Standard

Survey Event ID: 9KLQ11

Deficiency Tags: D5411 D5555 D5209 D5421

Summary:

Summary Statement of Deficiencies D5209 PERSONNEL COMPETENCY ASSESSMENT POLICIES CFR(s): 493.1235 As specified in the personnel requirements in subpart M, the laboratory must establish and follow written policies and procedures to assess employee and, if applicable, consultant competency. This STANDARD is not met as evidenced by: Based on review of personnel files, review of the Centers for Medicare and Medicaid Services (CMS) 209 Laboratory Personnel Report, lack of documentation, review of policy and procedure, and staff interview, the technical/general supervisor failed to complete an initial competency assessment for serology testing for 1 of 9 testing personnel (TP #8) reviewed; failed to complete an annual competency assessment in the specialty of microbiology for 1 of 9 testing personnel (TP #7) reviewed; failed to ensure an annual competency assessment was completed for the position of general supervisor for 2 of 2 years reviewed (2024, 2025); and failed to ensure an annual competency assessment was completed for the position of technical supervisor for 1 of 2 years reviewed (2024). The findings were: 1. Review of the personnel file for TP #8 showed she was hired on 6/2/25. There was no evidence an initial competency assessment was completed for serology testing. 2. Review of the personnel file for TP #7 failed to show a competency assessment for the specialty of microbiology was completed in 2025. 3. Review of the CMS-209 Laboratory Personnel Report showed the duties of the general supervisor and technical supervisor were performed by the laboratory manager. Review of the personnel record for the laboratory showed a competency assessment was not completed for the duties of the technical supervisor in 2024. In addition, there was no evidence a competency assessment had been completed for the duties of the general supervisor in 2024 or 2025. 4. Interview with the technical supervisor on 2/18/26 at 10:48 AM confirmed no further documentation could be located. 5. Review of the "Personnel Competency Review In the Lab" policy, effective November 1996, showed "Technical Supervisor, General Supervisor, Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 3 -- Medical Technologists and Phlebotomists will be given an annual competency assessment that will be reviewed by the Director of Laboratory Services or qualified designee. This list will include the knowledge and skills required to perform their job. New Employees competence will be reviewed upon start, at six months, and at one year during their first year of employment. After the first year of employment, each staff member's competency is assessment on an annual basis..." D5411 TEST SYSTEMS, EQUIPMENT, INSTRUMENTS, REAGENT CFR(s): 493.1252(a) (a) Test systems must be selected by the laboratory. The testing must be performed following the manufacturer's instructions and in a manner that provides test results within the laboratory's stated performance specifications for each test system as determined under 493.1253. This STANDARD is not met as evidenced by: Based on review of patient test reports, review of the VITROS XT 7600 analyzer manufacturer's instructions for use, and staff interview, the laboratory failed to follow the manufacturer's instructions to include the prostate specific antigen (PSA) test assay method on 1 of 1 PSA patient tests reports (patient #1) reviewed. The laboratory performed 874 PSA patient tests annually. The findings were: 1. Review of the VITROS XT 7600 analyzer manufacturer's instructions stated "Different test methods cannot be used interchangeably. PSA results in a given patient sample determined with different tests and from different manufacturers can vary due to differences in test methods and reagent specificity. A change to a different method during serial monitoring of a patient should be accompanied by additional sequential testing to confirm baseline values. The results reported by the laboratory to the physician must include the identity of the PSA test used." The following concerns were identified: a. Review of the test report for patient #1, dated 2/17/26, failed to include the PSA test methodology. 2. Interview with the technical supervisor on 2/18/26 at 12:20 PM confirmed the PSA test method was not included on the patient test report. D5421 ESTABLISHMENT AND VERIFICATION OF PERFORMANCE CFR(s): 493.1253(b)(1) (b) Each laboratory that introduces an unmodified, FDA-cleared or approved test system must do the following before reporting patient test results: (b)(1)(i) Demonstrate that it can obtain performance specifications comparable to those established by the manufacturer for the following performance characteristics: (b)(1)(i) (A) Accuracy. (b)(1)(i)(B) Precision. (b)(1)(i)(C) Reportable range of test results for the test system. (b)(1)(ii) Verify that the manufacturer's reference intervals (normal values) are appropriate for the laboratory's patient population. This STANDARD is not met as evidenced by: Based on new instrument and new test method verification study review, lack of documentation, review of policy and procedure, and staff interview, the laboratory failed to verify the reportable range and confirm the manufacturer's normal values were appropriate for the laboratory's patient population for 1 of 6 (VITROS XT 7600) verification studies reviewed. The findings were: 1. Review of the 5/8/24 new instrument verification study for the VITROS XT 7600 analyzer failed to show documentation the laboratory had verified the reportable range and had confirmed the -- 2 of 3 -- reference intervals used for the test system were appropriate for the laboratory's population. 2. Interview with the technical supervisor on 2/18/26 at 12:20 PM confirmed no further documentation was available. In addition, the technical supervisor revealed the laboratory began using the VITROS XT 7600 analyzer on 5/13 /24. 3. Review of the "Verification of Tests, methods, and Instruments" policy and procedure, effective February 2010, showed "...1. When adding or replacing a test, method, or instrument, PVHC laboratory shall verify the manufacturer's performance specifications including accuracy, precision and reportable ranges. 2. When replacing an old test, method, or instrument PVHC laboratory shall verify correlation between the old and new test, method, or instrument. 3. For each new test, method, or instrument, PVHC laboratory shall verify and establish reference ranges and QC ranges for the test, method or instrument and population served, prior to reporting patient results..." D5555 IMMUNOHEMATOLOGY CFR(s): 493.1271(c)(f) (c) Blood shall be stored in a clean and orderly environment in a manner to prevent mix-ups. Expired blood must not be in the routine inventory. Unacceptable units must be segregated from routine inventory. (c)(1) An audible alarm system must monitor proper blood and blood product storage temperature over a 24-hour period. (c)(2) Inspections of the alarm system must be documented. This STANDARD is not met as evidenced by: Based on staff interview, review of the laboratory's immunohematology documentation, and policy and procedure review, the laboratory failed to ensure the audible alarm was inspected and functioning appropriately for 2 of 2 blood product storage units for 1 of 2 years reviewed (2025). The findings were: 1. Review of the 2025 Blood Bank refrigerator temperature and alarm check (to be performed quarterly) log sheet showed the audible alarm was checked on 3/31/25 and 10/14/25. Review of the 2025 plasma freezer temperature and alarm check (to be done quarterly) log sheet showed the audible alarm was checked on 3/31/25, 7/20/25, and 10 /14/25. 2. Interview with the technical supervisor on 2/18/26 at 3 PM confirmed the audible alarm systems on the blood and blood product storage units had not been performed as required. 3. Review of the "Plasma Freezer Alarm Checks" policy and procedure, effective June 1997, showed "The alarm system on the plasma freezer should be checked quarterly to be sure that it is functioning properly. The results of the check will be recorded on the log sheet provided for this purpose." 4. Review of the "Blood Bank Refrigerator Alarm Checks and Quarterly Maintenance" policy and procedure, effective February 2007, showed "To provide a means of checking that the alarm system on the blood bank refrigerator functions properly by testing both low and high temperature alarm activations. The alarm checks on the blood bank refrigerator should be done quarterly." -- 3 of 3 --

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access

Survey - January 9, 2024

Survey Type: Standard

Survey Event ID: ZCWC11

Deficiency Tags: D5209 D5417

Summary:

Summary Statement of Deficiencies D5209 PERSONNEL COMPETENCY ASSESSMENT POLICIES CFR(s): 493.1235 As specified in the personnel requirements in subpart M, the laboratory must establish and follow written policies and procedures to assess employee and, if applicable, consultant competency. This STANDARD is not met as evidenced by: Based on review of the CMS (Centers for Medicare and Medicaid Services) 209 Laboratory Personnel Report, review of personnel records, policy and procedure review, and staff interview, the laboratory failed to ensure the competency assessment for the technical supervisor (TS) and the general supervisor (GS) was completed for 1 of 2 years (2023) reviewed. The findings were: 1. Review of the CMS 209 Laboratory Personnel Report showed the laboratory employed one staff member which performed the duties of the TS and the GS. The following concerns were identified: a. Review of the personnel record for the TS/GS showed no evidence a competency assessment was completed in 2023. b. Interview with the TS on 1/9/24 at 10:55 AM confirmed the competency assessment for the duties of the technical supervisor and general supervisor was not completed in 2023. c. Review of the policy and procedure titled "Personnel Competency Review In the Lab" last reviewed by the laboratory director on 1/19/23 failed to include a procedure for assessing the competency of the technical supervisor and the general supervisor. . D5417 TEST SYSTEMS, EQUIPMENT, INSTRUMENTS, REAGENT CFR(s): 493.1252(d) Reagents, solutions, culture media, control materials, calibration materials, and other supplies must not be used when they have exceeded their expiration date, have deteriorated, or are of substandard quality. Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 2 -- This STANDARD is not met as evidenced by: . Based on observation and staff interview, the laboratory failed to ensure the Blood Agar/MacConkey Biplate (BAP/MAC) media used for the cultivation, isolation, and enumeration of all bacterial and yeast urinary pathogens was not used beyond its expiration date for 9 of 9 urine cultures performed between 1/3/24 and 1/9/24. This failure affected 9 of 9 patients (#1, #2, #3, #4, #5, #6, #7, #8, and #9). The findings were: 1. Observation on 1/9/24 at 12:50 PM of the bacteriology storage refrigerator showed the BAP/MAC media, lot #752139 with an expiration date of 1/2/24, was available for use. 2. Observation of the urine cultures currently in progress and the finalized cultures from 1/3/24 to 1/9/24 showed a urine specimen for patient #1 was inoculated on 1/3/24; patients #2, #4, and #3 were inoculated on 1/4/24; patients #5, #6 and #7 were inoculated on 1/5/24, patient #8 was inoculated on 1/8/24; and patient #9 was inoculated on 1/9/24. 3. Interview with the technical supervisor on 1/9/24 at 1: 03 PM confirmed the media in use was expired and should not have been used. -- 2 of 2 --

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access

Survey - January 20, 2022

Survey Type: Standard

Survey Event ID: P9B211

Deficiency Tags: D5421 D6086 D5421 D6086

Summary:

Summary Statement of Deficiencies D5421 ESTABLISHMENT AND VERIFICATION OF PERFORMANCE CFR(s): 493.1253(b)(1) Each laboratory that introduces an unmodified, FDA-cleared or approved test system must do the following before reporting patient test results: (1)(i) Demonstrate that it can obtain performance specifications comparable to those established by the manufacturer for the following performance characteristics: (1)(i)(A) Accuracy. (1)(i) (B) Precision. (1)(i)(C) Reportable range of test results for the test system. (1)(ii) Verify that the manufacturer's reference intervals (normal values) are appropriate for the laboratory's patient population. This STANDARD is not met as evidenced by: Based on review of the new instrumentation and new test methodology verification studies, policy and procedure review, and staff interview, the laboratory failed to verify precision prior to testing patient samples for 2 of 4 new instrument and/or test methodology verification studies reviewed (Sysmex XN-550, Alcor miniiSED). The laboratory performed approximately 72,000 hematology tests annually and 130 erythocyte sedimentation rates from 8/30/21 through 1/20/22. The findings were: 1. Review of the Sysmex XN-550 new instrument verification study completed on 1/25 /21 showed an analysis of the performance specification of precision had not been performed prior to testing patient samples. 2. Review of the Alcor miniiSED new test methodology verification study completed on 8/30/21 showed the performance specification of precision had not been verified. In addition there was no evidence the laboratory director had reviewed the verification study documentation and approved the test procedure prior to testing patient samples. 3. Interview with the technical consultant on 1/20/22 at 12:15 PM confirmed the verification studies were incomplete. 4. Review of the "New Test, Method or Instrument Procedure" reviewed by the laboratory director on 1/18/22 showed "...When adding or replacing a test, method or instrument, PVHC laboratory shall verify the manufacturer's performance specifications regarding accuracy, precision and reportable ranges." Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 2 -- D6086 LABORATORY DIRECTOR RESPONSIBILITIES CFR(s): 493.1445(e)(3)(ii) The laboratory director must ensure that verification procedures used are adequate to determine the accuracy, precision, and other pertinent performance characteristics of the method. This STANDARD is not met as evidenced by: Based on review of the Sysmex XN-550 new instrument and the Alcor miniiSED new test method verification documentation, policy and procedure review, and staff interview, the laboratory director failed to ensure the test verifications of accuracy, precision, analytic specificity, and analytic sensitivity was complete and met the laboratory's performance characteristics specified by the manufacturer. Refer to D5421. -- 2 of 2 --

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access

Survey - August 29, 2020

Survey Type: Standard

Survey Event ID: 7DE911

Deficiency Tags: D5407 D6128 D5407 D6128

Summary:

Summary Statement of Deficiencies D5407 PROCEDURE MANUAL CFR(s): 493.1251(d) Procedures and changes in procedures must be approved, signed, and dated by the current laboratory director before use. This STANDARD is not met as evidenced by: Based on direct observation, procedure manual review, and interview with staff, the laboratory failed to ensure the change to the blood bank specimen and unit labeling process was approved, signed and dated by the director. The laboratory performed approximately 3-6 compatibility tests per week. Findings include: 1. The laboratory changed the procedure for positive identification of blood recipient's specimens and tagging of crossmatched donor specimens in 2020. 2. The laboratory failed to change the established procedure and failed to have the director's signature and date of approval on the revised procedure. 3. In an interview conducted on 08/29/2020 at approximately 1:10 P.M., the laboratory manager confirmed the laboratory had not revised the procedure with the approval of the director by signature and date of approval. D6128 TECHNICAL SUPERVISOR RESPONSIBILITIES CFR(s): 493.1451(b)(9) The technical supervisor is responsible for evaluating and documenting the performance of individuals responsible for high complexity testing at least annually after the first year, unless test methodology or instrumentation changes, in which case, prior to reporting patient test results, the individual's performance must be reevaluated to include the use of the new test methodology or instrumentation. This STANDARD is not met as evidenced by: Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 2 -- Based on testing personnel competency evaluations reviewed, lack of documentation and confirmation by the laboratory manager, the technical supervisor failed to ensure seven of seven high complexity testing personnel were evaluated for competency in 4 of 5 specialties at least annually for 1 of 2 years of testing reviewed, 2019. Findings include: 1. The laboratory failed to document competency evaluations for Hematology, Chemistry, Microbiology, and General Immunology for the seven high complexity testing personnel in 2019. 2. In an interview conducted on 08/29/2020 at approximately 1:00 P.M., the laboratory manager confirmed competency was not documented for 2019 for specialties other than immunohematology. -- 2 of 2 --

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access