Summary:
Summary Statement of Deficiencies D5209 PERSONNEL COMPETENCY ASSESSMENT POLICIES CFR(s): 493.1235 As specified in the personnel requirements in subpart M, the laboratory must establish and follow written policies and procedures to assess employee and, if applicable, consultant competency. This STANDARD is not met as evidenced by: Based on surveyor review of the Competency Assessment (CA) records, Procedure Manual (PM) and interview with the Technical Consultant (TC), the laboratory failed to follow its policies for assessing the competency of Testing Personnel (TP) from 1/1 /24 to 6/17/25. The findings include: 1. TP#4 and TP #5 as listed on the CMS 209 form, had incomplete CA for calendar year 2024. 2. TP #7 and TP#8 as listed on the CMS 209 form, did not have a CA performed in 2024. 3. The TC confirmed on 6/17 /25 at 11:25 am, the laboratory failed to follow its policies for assessing the competency of TP. D5437 CALIBRATION AND CALIBRATION VERIFICATION CFR(s): 493.1255(a) (a )Unless otherwise specified in this subpart, for each applicable test system the laboratory must perform and document calibration procedures-- (a)(1) Following the manufacturer's test system instructions, using calibration materials provided or specified, and with at least the frequency recommended by the manufacturer; (a)(2) Using the criteria verified or established by the laboratory as specified in 493.1253(b) (3)-- (a)(2)(i) Using calibration materials appropriate for the test system and, if possible, traceable to a reference method or reference material of known value; and (a) (2)(ii) Including the number, type, and concentration of calibration materials, as well as acceptable limits for and the frequency of calibration; and (a)(3) Whenever calibration verification fails to meet the laboratory's acceptable limits for calibration Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 2 -- verification. This STANDARD is not met as evidenced by: Based on surveyor review of the Calibration records, Procedure Manual (PM), Medonic User Manual (MUM) and interview with the Technical Consultant (TC) the laboratory failed to perform and document Calibration procedures at least once every six months for Hematology Tests performed on the Medonic analyzer from 7/1/24 to 1 /1/25. The findings include: 1. A review of calibration records revealed that the laboratory did not perform calibration every 6 months as per the manufacturer. 2. The TC confirmed on 6/17/25 at 11:15 am, the laboratory failed to perform and document calibration of the analyzer at least every six months. D5775 COMPARISON OF TEST RESULTS CFR(s): 493.1281(a)(c) (a) If a laboratory performs the same test using different methodologies or instruments, or performs the same test at multiple testing sites, the laboratory must have a system that twice a year evaluates and defines the relationship between test results using the different methodologies, instruments, or testing sites. This STANDARD is not met as evidenced by: Based on surveyor review of the Procedure Manual (PM), Instrument Correlation Records (ICR) and interview with the Technical Consultant (TC), the laboratory did not have acceptable criteria to evaluate and define the relationship for Hematology test results between the two Medonic analyzers from 1/7/24 to 6/17/25. The finding includes: 1. The procedure did not have written acceptability or rejection criteria for the delta values between the Medonic analyzers. 2. The TC confirmed on 6/17/25 at 11:50 am, the laboratory failed to have written criteria for acceptable differences in test values for correlation studies performed on the two Medonic analyzers. D5779