Summary:
Summary Statement of Deficiencies D0000 A CLIA recertification survey was conducted for the Reliant Medical Group Neponset St. Dermatology laboratory pursuant to the Clinical Laboratory Improvement Amendments (CLIA) of 1988 and CLIA regulations at 42 CFR 493. . D5407 PROCEDURE MANUAL CFR(s): 493.1251(d) Procedures and changes in procedures must be approved, signed, and dated by the current laboratory director before use. This STANDARD is not met as evidenced by: Based on procedure manual review and interview, the laboratory director failed to approve, sign, and date laboratory procedures as evidenced by the following: * A new laboratory director was hired for the laboratory in February of 2023 according to the Supervisor of Ancillary Support interviewed on 8/22/23 at 9:45 AM. A review of the clinical laboratory procedure manual revealed that the current laboratory director had not reviewed and approved the laboratory policy and procedure manual. The laboratory performs approximately 80 Physician Performed Microscopy Procedures (PPMP) and 200 histopathology (MOHS slide exams) annually. Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 1 --