Resolute Dermatology Llc

CLIA Laboratory Citation Details

2
Total Citations
8
Total Deficiencyies
4
Unique D-Tags
CMS Certification Number 17D2207992
Address 6800 Hilltop Rd Ste 102, Shawnee, KS, 66226
City Shawnee
State KS
Zip Code66226
Phone(913) 901-5001

Citation History (2 surveys)

Survey - December 18, 2024

Survey Type: Standard

Survey Event ID: YZW611

Deficiency Tags: D5217 D5401 D6046 D5217 D5401 D6046

Summary:

Summary Statement of Deficiencies D5217 EVALUATION OF PROFICIENCY TESTING PERFORMANCE CFR(s): 493.1236(c)(1) At least twice annually, the laboratory must verify the accuracy of any test or procedure it performs that is not included in subpart I of this part. This STANDARD is not met as evidenced by: Based on the review of the CMS116 test lists, lack of documentation of accuracy checks from 6/29/23 to date of survey, lack of proficiency testing enrollment, patient KOH test results, and interview with the office manager (OM), the laboratory failed to verify the accuracy of KOH testing for mycology twice a year from 6/29/23 to date of survey. Findings: 1. Review of the CMS116 test lists and accuracy check documentation revealed the laboratory did not perform accuracy checks twice a year for KOH testing from 6/29/23 (KOH testing start date) to date of survey. 2. The surveyor asked if the laboratory was enrolled in proficiency testing for KOH testing. The OM stated the laboratory was not enrolled in proficiency testing for KOH testing. 3. Review of patient testing records revealed six patients had test results for KOH testing for mycology from 6/29/23 to date of survey. 4. Interview with the OM on 12 /18/24 at 2:10 p.m. confirmed, the laboratory failed to verify the accuracy of KOH testing for mycology twice a year from 6/29/23 to date of survey. D5401 PROCEDURE MANUAL CFR(s): 493.1251(a) A written procedures manual for all tests, assays, and examinations performed by the laboratory must be available to, and followed by, laboratory personnel. Textbooks may supplement but not replace the laboratory's written procedures for testing or examining specimens. Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 2 -- This STANDARD is not met as evidenced by: Based on the review of the CMS116 test lists, lack of procedure for KOH testing, patient KOH test results and interview with the OM, the laboratory failed to have a procedure for KOH testing available to, and followed by, laboratory personnel. Findings: 1. Review of the CMS116 test lists showed KOH testing had been added since the last survey 3/23/23. 2. Review of the procedure manual revealed no procedure for KOH testing was included. 3. Review of patient testing records revealed six patients had test results for KOH testing for mycology from 6/29/23 to date of survey. 4. Interview with the OM on 12/18/24 at 2:05 p.m. confirmed, the laboratory failed to have a procedure for KOH testing available to, and followed by, laboratory personnel. D6046 TECHNICAL CONSULTANT RESPONSIBILITIES CFR(s): 493.1413(b)(8) (b) The technical consultant is responsible for-- (b)(8) Evaluating the competency of all testing personnel and assuring that the staff maintain their competency to perform test procedures and report test results promptly, accurately and proficiently. This STANDARD is not met as evidenced by: Based upon the review of the CMS209 personnel form, lack of personnel competency assessment records and interview with the OM, the technical consultant (TC) failed to assess competency for two of two moderate complexity testing personnel (TP) for 2023 and to date of survey 2024. Findings: 1. Review of the CMS 209 revealed two moderate complexity TP. 2. The surveyor requested competency assessment records for the TP. No documentation of competency assessment for 2023 and to date of survey 2024 for the two of two moderate complexity TP was made available at the time of survey. 3. Interview with the OM on 12/18/24 at 2:05 p.m. confirmed, the TC failed to assess competency for two of two moderate complexity testing personnel (TP) for 2023 and to date of survey 2024. -- 2 of 2 --

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access

Survey - July 21, 2021

Survey Type: Standard

Survey Event ID: T8DK11

Deficiency Tags: D5435 D5435

Summary:

Summary Statement of Deficiencies D5435 MAINTENANCE AND FUNCTION CHECKS CFR(s): 493.1254(b)(2) For equipment, instruments, or test systems developed in-house, commercially available and modified by the laboratory, or maintenance and function check protocols are not provided by the manufacturer, the laboratory must: (i) Define a function check protocol that ensures equipment, instrument, and test system performance that is necessary for accurate and reliable test results and test result reporting. (ii) Perform and document the function checks, including background or baseline checks, specified in paragraph (b)(2)(i) of this section. Function checks must be within the laboratory's established limits before patient testing is conducted. This STANDARD is not met as evidenced by: Based on an absence of thermometer function check records or certificates of accuracy, protocols for thermometer and hygrometer function checks and interview, the laboratory failed to define and perform a function check protocol for 2 of 2 thermometers and 1 hygrometer. Findings: 1. No documentation was available for function checks on 2 of 2 thermometers and 1 of 1 hygrometers at the time of survey. 2. No documentation was available for the certification of accuracy (NIST traceble) on 2 of 2 thermometers and 1 of 1 hygrometers at the time of survey. 3. Proctocols for the function checks of thermometers and hygrometer were not made available at the time of survey. 4. Interview with the laboratory director on 7/21/2021 at 2:45 p.m. confirmed, the laboratory failed to define and perform a function check protocol for 2 of 2 thermometers and 1 hygrometer. Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 1 --