Sante Chandler, Llc

CLIA Laboratory Citation Details

1
Total Citation
4
Total Deficiencyies
4
Unique D-Tags
CMS Certification Number 03D2025037
Address 825 S 94th St, Chandler, AZ, 85224
City Chandler
State AZ
Zip Code85224
Phone(480) 361-6636

Citation History (1 survey)

Survey - November 29, 2023

Survey Type: Standard

Survey Event ID: UZCJ11

Deficiency Tags: D5301 D6053 D5407 D6054

Summary:

Summary Statement of Deficiencies D5301 TEST REQUEST CFR(s): 493.1241(a) The laboratory must have a written or electronic request for patient testing from an authorized person. This STANDARD is not met as evidenced by: Based on lack of test requisition documentation for review and interview with the technical consultant (TC-1), the laboratory failed to have a written or electronic request for patient testing for one out of five patient records reviewed during the survey. Findings include: 1. The laboratory began using the i-Stat analyzer for patient testing in December 2022. 2. No written or electronic request for testing was presented for review for one out of five patient records reviewed during the survey (patient ID: 010364, tested on 12/19/22 at 2:06 PM). 3. The TC-1 interviewed on 11 /29/23 at 11:10 AM confirmed the laboratory failed to have an electronic or written test requisition for testing for the patient indicated above. 4. The laboratory's annual reported test volume in the specialties of Chemistry and Hematology is 216. D5407 PROCEDURE MANUAL CFR(s): 493.1251(d) Procedures and changes in procedures must be approved, signed, and dated by the current laboratory director before use. This STANDARD is not met as evidenced by: Based on review of the laboratory's policy and procedure manual and interview with the Technical Consultant (TC-1), the laboratory failed to have the current laboratory director approve, sign and date test procedures before use. Findings include: 1. The current laboratory director indicated in the CLIA Federal Database and assigned on Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 2 -- the CMS-209, Laboratory Personnel Form presented for review during the survey has been listed as the laboratory director since 06/07/22. 2. The laboratory began patient testing on 12/19/22 in the specialities of Hematology and Chemistry with an annual test volume of 216. 3. The policy and procedure manual presented for review during the survey conducted on 11/29/23 was not approved, signed and dated by the current laboratory director before use. 4. The TC-1 interviewed on 11/29/23 at 10:55 AM confirmed that the policy and procedure manual indicated above was not approved, signed and dated by the current laboratory director prior to performing patient testing. D6053 TECHNICAL CONSULTANT RESPONSIBILITIES CFR(s): 493.1413(b)(9) The technical consultant is responsible for evaluating and documenting the performance of individuals responsible for moderate complexity testing at least semiannually during the first year the individual tests patient specimens. This STANDARD is not met as evidenced by: Based on lack of performance evaluation documentation and interview with the technical consultant (TC-1), the technical consultant failed to evaluate and document the performance of one testing personnel, at least semiannually during the first year the individual tested patient specimens. Findings include: 1. No documentation of semiannual competency evaluation was presented for review during the survey conducted on 11/29/2023 for one out of one testing personnel who underwent initial training on 01/20/2022. 2. The TC-1 interviewed on 11/29/23 at 10:10 AM confirmed that the technical consultant failed to perform and document a semiannual competency evaluation for the testing personnel indicated above. 3. The laboratory uses the i-Stat analyzer to perform patient testing in the specialties of Chemistry and Hematology with a reported annual test volume of 216. D6054 TECHNICAL CONSULTANT RESPONSIBILITIES CFR(s): 493.1413(b)(9) The technical consultant is responsible for evaluating and documenting the performance of individuals responsible for moderate complexity testing at least annually, after the first year. This STANDARD is not met as evidenced by: Based on lack of performance evaluation documentation and interview with the technical consultant (TC-1), the technical consultant failed to evaluate and document the performance of one individual responsible for moderate complexity testing at least annually. Findings include: 1. No documentation of annual competency evaluation was presented for review for one out of one testing personnel who underwent initial training for i-Stat testing on 1/20/22. 2. The TC-1 interviewed on 11/29/23 at 10:10 AM confirmed the technical consultant failed to evaluate and document the performance of one out of one testing personnel at least annually. 3. The laboratory uses the i-Stat analyzer to perform patient testing with a reported annual test volume of 216. -- 2 of 2 --

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access