Schweiger Dermatology Group

CLIA Laboratory Citation Details

3
Total Citations
18
Total Deficiencyies
8
Unique D-Tags
CMS Certification Number 39D2171014
Address 50 Monument Rd, Ste #301, Bala Cynwyd, PA, 19004
City Bala Cynwyd
State PA
Zip Code19004
Phone(610) 668-2570

Citation History (3 surveys)

Survey - September 4, 2025

Survey Type: Standard

Survey Event ID: XUIS11

Deficiency Tags: D0000 D5400 D5400 D5449 D0000 D3009 D5449 D3009 D5407 D5407

Summary:

Summary Statement of Deficiencies D0000 A recertification survey conducted by the Pennsylvania State Agency on 09/04/2025 found the Schweiger Dermatology Group laboratory to be out of compliance with the following condition: 493.1250 Condition: Analytic systems D3009 FACILITIES CFR(s): 493.1101(c) The laboratory must be in compliance with applicable Federal, State, and local laboratory requirements. This STANDARD is not met as evidenced by: A. Based on review of laboratory policy, temperature records, and interview with the General Manager (GM), the laboratory failed to monitor and document room temperature for 43 of 673 days to ensure substantial proof of efficacy was provided for reagents and operating conditions were met for reliable test system operation when chemistry testing was performed from 11/01/2023 to the day of survey as required per PA regulations. Findings include: 1. The PA State regulation 5.46 states: "Reagents, procedures or equipment which have been demonstrated to be inadequate for clinical laboratory use as evidenced by reliable data from generally acceptable scientific testing and evaluating sources shall be prohibited for use by clinical laboratories upon specific notification by the Department. Also, reagents, equipment and procedures which do not have substantial proof of efficacy either by trial or extended use experience shall be prohibited for routine use". 2. The laboratory's Quality Control of Laboratory Equipment policy states, "Temperature and humidity devices are maintained in laboratory and ranges are recorded daily when testing is performed." 3. On the day of survey, 09/04/2025 at 10:12 am, review of the laboratory's temperature logs revealed the laboratory failed to monitor and document room temperatures (laboratory's acceptable range 20 C to 25 C) for 43 of 673 days to ensure proper storage of reagents and operating conditions were met for the following reagents used in the laboratory from 11/01/2023 to 09/04/2025: - Clarity Diagnostics Pregnancy Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 3 -- One Step Rapid test cassettes (manufacturer's acceptable range 2C to 30 C) 4. Review of the laboratory's pregnancy test logs revealed that the laboratory performed 44 urine pregnancy tests from 11/01/2023 to 09/04/2025. 5. The GM confirmed the above findings on 09/04/25 at 11:29 am. B. Based on review of laboratory procedures, lack of documentation and interview with the General Manager (GM), the laboratory failed to follow and document quality control (QC) procedures to establish the accuracy, specificity and precision of 1 of 1 chemistry test performed from 11/01/2023 to the date of survey as required per PA regulations. Findings include: 1. The PA State regulation 5.61(b) states: "A degree of accuracy, specificity and precision satisfactory to the Department shall be shown in quality control records at all times". 2. The laboratory's Urine Pregnancy procedure states, "When opening a new box of pregnancy tests, run a control test with a known pregnant specimen and a known non- pregnant specimen, and record." 3. On the day of survey 09/04/2025, the laboratory failed to provide documentation for the positive and negative external QC performed when a new box was opened for 1 of 1 chemistry test (Urine human chorionic gonadotropin) performed from 11/01/2023 to 09/04/2025. 4. Review of the laboratory's pregnancy test logs revealed that the laboratory performed 44 urine pregnancy tests from 11/01/2023 to 09/04/2025. 5. The GM confirmed the above findings on 09/04/25 at 11:29 am. D5400 ANALYTIC SYSTEMS CFR(s): 493.1250 Each laboratory that performs nonwaived testing must meet the applicable analytic systems requirements in 493.1251 through 493.1283, unless HHS approves a procedure, specified in Appendix C of the State Operations Manual (CMS Pub.7), that provides equivalent quality testing. The laboratory must monitor and evaluate the overall quality of the analytic systems and correct identified problems as specified in 493.1289 for each specialty and subspecialty of testing performed. This CONDITION is not met as evidenced by: Based on lack of documentation and interview with the General Manager (GM), the laboratory failed to ensure a positive and negative control was performed and documented each day of patient testing as required under 493.1256 when Potassium Hydroxide (KOH) microscopic examinations were performed for 16 of 16 months from 11/01/2023 to 03/05/2025. Refer to D5449 D5407 PROCEDURE MANUAL CFR(s): 493.1251(d) (d) Procedures and changes in procedures must be approved, signed, and dated by the current laboratory director before use. This STANDARD is not met as evidenced by: Based on review of the laboratory's procedures and interview with the General Manager (GM), the laboratory failed to ensure that 1 of 1 laboratory procedure manual in use for Histopathology was approved, signed and dated by the current Laboratory Director (LD) for 5 of 5 months from 04/01/2025 to 09/04/2025. Findings include: 1. On the day of survey 09/04/2025 at 09:49 am, review of the laboratory procedure manual in use for Histopathology revealed the laboratory failed to ensure all procedures were approved, signed and dated by the current LD when -- 2 of 3 -- histopathology testing was performed for 5 of 5 months from 04/01/2025 to 09/04 /2025. 2. The laboratory performed 138 histopathology examinations in 2024 (CMS- 116, estimated annual volume, dated 08/20/2025). 3. The GM confirmed the above findings on 09/04/2025 at 11:29 am. D5449 CONTROL PROCEDURES CFR(s): 493.1256(d)(3)(ii)(g) (d)(3)(ii) Each qualitative procedure, include a negative and positive control material; This STANDARD is not met as evidenced by: Based on lack of documentation and interview with the General Manager (GM), the laboratory failed to document a positive and negative control each day of patient testing for Potassium Hydroxide (KOH) microscopic examinations performed for 16 of 16 months from 11/01/2023 to 03/05/2025. Findings include: 1. The laboratory's KOH examination procedure states, "On the KOH log sheet, each slide must be determined to be Acceptable or Unacceptable by circling "Good" or " Poor" (Quality Control). The provider will then sign each test entry". 2. On the day of survey, 09/04 /2025 at 10:04 am, the laboratory failed to provide documentation of the positive and negative quality control performed every day of patient testing for KOH microscopic examinations performed from 11/01/2023 to 03/05/2025. 3. The laboratory performed 130 KOH microscopic examinations from 11/01/2023 to 03/05/2025. 4. The GM confirmed the above findings on 09/04/2025 at 11:29 am. *REPEAT DEFICIENCY -- 3 of 3 --

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access

Survey - November 1, 2023

Survey Type: Standard

Survey Event ID: BEOV11

Deficiency Tags: D5217 D6054 D5449 D6054 D5217 D5449

Summary:

Summary Statement of Deficiencies D5217 EVALUATION OF PROFICIENCY TESTING PERFORMANCE CFR(s): 493.1236(c)(1) At least twice annually, the laboratory must verify the accuracy of any test or procedure it performs that is not included in subpart I of this part. This STANDARD is not met as evidenced by: Based on lack of documentation, review of the laboratory's procedure manual, and interview with the Assistant General Manager (AGM), the laboratory failed to perform twice annually the verification of accuracy of Mohs microscopic examinations from 12/02/2021 to the date of survey. Findings include: 1.On the day of survey, 11/01/2023 at 10:45 AM, the laboratory could not provide the biannual documentation for the verification of accuracy of Mohs microscopic examinations from 12/02/2021 to the date of survey. 2.According to the laboratory's proficiency testing policy, "peer review will be conducted twice a year. For each test performed, a set of representative samples (at least 2 for Mohs) will be reviewed in a blinded fashion". 3.The laboratory reported an annual volume of 119 microscopic examinations/tests performed in histopathology (CMS 116). 4.The AGM confirmed the above findings on 11/01/2023 at 12:00 PM. D5449 CONTROL PROCEDURES CFR(s): 493.1256(d)(3)(ii)(g) Unless CMS Approves a procedure, specified in Appendix C of the State Operations Manual (CMS Pub. 7), that provides equivalent quality testing, the laboratory must-- At least once a day patient specimens are assayed or examined perform the following for-- Each qualitative procedure, include a negative and positive control material; (g) The laboratory must document all control procedures performed. Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 2 -- This STANDARD is not met as evidenced by: Based on review of quality control (QC) records and interview with the Assistant General Manager (AGM), the laboratory failed to document a positive and negative control each day of patient testing for Potassium Hydroxide (KOH) microscopic examinations from 12/02/2021 to 11/01/2023. Findings include: 1. On the day of survey 11/01/2023 at 11:00 AM, the laboratory failed to provide positive and negative visual control for KOH examination for each day of patient testing from 12/02/2021 to 11/01/2023. 2. According to laboratory's KOH examination procedure, each KOH slide must be determined to be acceptable or unacceptable by circling "Good" or "Poor" quality control. The provider will then sign each test entry. 3. The laboratory performed 137 mycology tests in 2022 (CMS116). 4. The AGM confirmed the above findings on 11/01/2023 at 12:00 PM. D6054 TECHNICAL CONSULTANT RESPONSIBILITIES CFR(s): 493.1413(b)(9) The technical consultant is responsible for evaluating and documenting the performance of individuals responsible for moderate complexity testing at least annually, after the first year. This STANDARD is not met as evidenced by: Based on lack of documentation and interview with the Assistant General Manager (AGM), the Technical Consultant (TC) failed to evaluate the competency assessment for 3 of 3 Testing Personnel (TP) who performed moderate complexity testing in mycology from 12/02/2021 to 11/01/2023. Findings include: 1. On the day of survey 11/01/2023 at 11:00 AM, the laboratory failed to provide competency assessment records for 3 of 3 TP (CMS 209 personnel #1, #3, #4) who performed moderate complexity testing in mycology from 12/02/2021 to the date of survey. 2. AGM confirmed the findings above on 11/01/2023 around 12:00 PM. -- 2 of 2 --

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access

Survey - December 2, 2021

Survey Type: Standard

Survey Event ID: 7SW511

Deficiency Tags: D5449 D5209

Summary:

Summary Statement of Deficiencies D5209 PERSONNEL COMPETENCY ASSESSMENT POLICIES CFR(s): 493.1235 As specified in the personnel requirements in subpart M, the laboratory must establish and follow written policies and procedures to assess employee and, if applicable, consultant competency. This STANDARD is not met as evidenced by: Based on review of the laboratory's procedure and interview with the General Manager (GM), the laboratory failed to establish a complete procedure that includes all six components required for competency assessment for 3 of 3 testing personnel (TP) who performed Potassium Hydroxide (KOH) microscopic examinations from 06 /11/2021 to the date of survey. Findings Include: 1. On the day of survey, 12/02/2021 at 9:55 a.m., The laboratory failed to provide a complete written policy that reviews how to assess the competency for 3 of 3 TP performing KOH microscopic examinations from 06/11/2021 to 12/02/2021. 2. The GM confirmed the finding above on 12/02/2021 at 12:00 p.m. D5449 CONTROL PROCEDURES CFR(s): 493.1256(d)(3)(ii)(g) Unless CMS Approves a procedure, specified in Appendix C of the State Operations Manual (CMS Pub. 7), that provides equivalent quality testing, the laboratory must-- At least once a day patient specimens are assayed or examined perform the following for-- Each qualitative procedure, include a negative and positive control material; (g) The laboratory must document all control procedures performed. This STANDARD is not met as evidenced by: Based on review of quality control records and interview with General Manager Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 2 -- (GM), the laboratory failed to include a negative and positive control material, each day of patient testing for Potassium Hydroxide (KOH) microscopic examinations from 6/11/2021 to the day of survey. Findings include: 1. On the day of survey, 12/02 /2021 at 10:08 a.m. The laboratory could not provide documentation of daily quality control performed for KOH microscopic examinations from 06/11/2021 to the day of survey. 2. The laboratory performed 68 KOH specimens from 06/11/2021 to 12/02 /2021. 3. The GM confirmed the findings above on 12/02/2021 at 12:00 p.m. -- 2 of 2 --

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility