Summary:
Summary Statement of Deficiencies D0000 The recertification survey was performed on 02/20,21,22/2024. The laboratory was found out of compliance with the following CLIA Conditions: 493.1250; D5400: Analytic Systems 493.1405; D6000: Laboratory Director 493.1409; D6033: Technical Consultant The findings were reviewed with the technical consultant, testing person #1, and testing person #2 during an exit conference performed at the conclusion of the survey. D5400 ANALYTIC SYSTEMS CFR(s): 493.1250 Each laboratory that performs nonwaived testing must meet the applicable analytic systems requirements in 493.1251 through 493.1283, unless HHS approves a procedure, specified in Appendix C of the State Operations Manual (CMS Pub.7), that provides equivalent quality testing. The laboratory must monitor and evaluate the overall quality of the analytic systems and correct identified problems as specified in 493.1289 for each specialty and subspecialty of testing performed. This CONDITION is not met as evidenced by: Based on a review of records, written policies and procedures, manufacturer's instructions, and interview with the technical consultant, the laboratory failed to monitor and evaluate the overall quality of analytic systems and correct identified problems for each specialty and subspecialty of testing performed during the review period of June 2022 through the current date. Findings include: (1) The laboratory failed to follow their written policy for performing quality control for CBC testing for one of one day of patient testing reviewed from 07/28/2023 through 08/15/2023. Refer to D5401; (2) The laboratory failed to have complete written procedures for one of one procedure reviewed. Refer to D5403; (3) The laboratory failed to ensure one of one policy had been approved, signed, and dated by the laboratory director. Refer to D5407; (4) The laboratory failed to ensure the laboratory humidity was maintained as required by the manufacturer of the TOSOH ALA-360 analyzer for three of three Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 14 -- months reviewed. Refer to D5413; (5) The laboratory failed to ensure the manufacturer's instructions were followed for performing weekly maintenance procedures on the Sysmex XN-330 analyzer and the Siemens Viva Pro E analyzer during the review period of January 2023 through December 2023. Refer to D5429; (6) The laboratory failed to perform calibration verification procedures at least once every six months for the Piccolo test system during the review period of January 2023 through the current date. Refer to D5439; (7) The laboratory failed to perform quality control procedures each day of patient testing for eight of 40 days reviewed. Refer to D5447; (8) The laboratory failed to have a system that twice a year evaluated and defined the relationship between test results using different Piccolo reagent discs for three of 12 analytes reviewed from June 2022 through the current date. Refer to D5775; (9) The laboratory failed to have an ongoing mechanism for performing quality assessment. Refer to D5791. NOTE: D5400 was cited on the initial survey performed on 01/28/2022. D5401 PROCEDURE MANUAL CFR(s): 493.1251(a) A written procedures manual for all tests, assays, and examinations performed by the laboratory must be available to, and followed by, laboratory personnel. Textbooks may supplement but not replace the laboratory's written procedures for testing or examining specimens. This STANDARD is not met as evidenced by: Based on a review of records, written policies and procedures, and interview with the technical consultant, the laboratory failed to follow their written policy for performing QC (quality control) for CBC testing for one of one day of patient testing reviewed between 07/28/2023 through 08/16/2023. Findings include: (1) On 02/20/2024 at 03: 00 pm, the technical consultant stated the following: (a) The laboratory performed CBC (Complete Blood Count) testing using the Sysmex XN-330 analyzer; (b) Three levels of XN-L CHECK QC (Quality Control) materials were tested each day of patient testing. (2) A review of the policy titled, "Complete Blood Count (CBC)" under section titled "Quality Control Procedure - QC Frequency" stated, "Test three levels of QC samples once per day using Commercial Control"; (3) A review of QC and patient test reports from 07/28/2023 through 08/16/2023 identified the laboratory had tested two levels of QC materials instead of three levels (no evidence the laboratory performed level 1 QC) for the following patients: (a) Patient #512714332831745, tested on 08/11/2023 at 11:52 am; (b) Patient #268257629503489, tested on 08/11/2023 at 01:41 pm. (4) The findings were reviewed with the technical consultant who stated on 02/21/2024 at 02:15 pm, the laboratory had not followed their policy for performing QC procedures as stated above. D5403 PROCEDURE MANUAL CFR(s): 493.1251(b) The procedure manual must include the following when applicable to the test procedure: (1) Requirements for patient preparation; specimen collection, labeling, storage, preservation, transportation, processing, and referral; and criteria for specimen acceptability and rejection as described in 493.1242. (2) Microscopic examination, including the detection of inadequately prepared slides. (3) Step-by-step performance of the procedure, including test calculations and interpretation of results. -- 2 of 14 -- (4) Preparation of slides, solutions, calibrators, controls, reagents, stains, and other materials used in testing. (5) Calibration and calibration verification procedures. (6) The reportable range for test results for the test system as established or verified in 493.1253. (7) Control procedures. (8)