Summary:
Summary Statement of Deficiencies D5417 TEST SYSTEMS, EQUIPMENT, INSTRUMENTS, REAGENT CFR(s): 493.1252(d) (d) Reagents, solutions, culture media, control materials, calibration materials, and other supplies must not be used when they have exceeded their expiration date, have deteriorated, or are of substandard quality. This STANDARD is not met as evidenced by: Based on observation and interview with laboratory staff, the laboratory had supplies available for use after their expiration date. Findings follow: A) During a tour of the laboratory on 3/9/26 at 1:17 pm, one (of three) box containing ten kits of Para-Pak ZN- PVA/Formalin, lot# 538000T, expiration date 3/7/26, REF 301012, Meridian Bioscience Inc, was observed in the laboratory, available for use beyond the expiration date. B) In an interview on 3/7/26 at 1:17 pm testing person #1 (as listed on the CMS 209 form) confirmed that the items, identified above, had exceeded their expiration date and were available for use in the laboratory. Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 1 --