Summary:
Summary Statement of Deficiencies D0000 The laboratory was found NOT to be in compliance with the CLIA regulations found at 42 CFR 493 CLIA requirements for laboratories as a result of an announced recertification survey completed on March 10, 2026. The condition not met was: D6063 - 42 C.F.R. 493.1421 Condition: Laboratories performing moderate complexity testing; testing personnel; D5213 EVALUATION OF PROFICIENCY TESTING PERFORMANCE CFR(s): 493.1236(b)(1) (b) The laboratory must verify the accuracy of the following: (b)(1) Any analyte or subspecialty without analytes listed in subpart I of this part that is not evaluated or scored by a CMS-approved proficiency testing program. This STANDARD is not met as evidenced by: Based on review of the laboratory's American Proficiency Institute's proficiency testing records from 2025 and staff interview, the laboratory failed to have documentation of evaluating 2 of 2 results reported by the proficiency testing agency as 'ungraded'. The findings included: 1. A review of the laboratory's American Proficiency Institute's proficiency testing records from 2025 (events 1, 2 and 3) determined the proficiency agency returned two results marked as "ungraded". They were: a) Event 2 2025 Sample: HEM-09 Analyte: Hematocrit b) Event 2 2025 Sample: HEM-09 Analyte: Red Cell Count 2. The laboratory did not have documentation of evaluating the ungraded results. 3. The laboratory director confirmed the findings in an interview conducted on 03/10/2026 at 1230 hours in exam room 3. D5417 TEST SYSTEMS, EQUIPMENT, INSTRUMENTS, REAGENT CFR(s): 493.1252(d) (d) Reagents, solutions, culture media, control materials, calibration materials, and Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 5 -- other supplies must not be used when they have exceeded their expiration date, have deteriorated, or are of substandard quality. This STANDARD is not met as evidenced by: Based on review of the manufacturer's instructions for the CDS 3PD Hematology Controls, review of the laboratory's quality control records from September 2025 to March 2026, review of patient test records from September 2025 to March 2026, and staff interview, the laboratory failed to ensure controls materials were not used for 6 of 7 control changes. The findings included: 1. The manufacturer's instructions for the CDS 3PD Hematology Control (PN 202448C R06.05.18) stated: "Open Vial Stability: 14 days" 2. A review of the laboratory's quality control logs from September 2025 to March 2026 identified the laboratory changed controls on the following dates: Lot: 32508 opened: 9/11/2025 Lot: 32508 opened: 10/9/2025 elapsed time: 28 days Lot: 32508 opened: 11/4/2025 elapsed time: 25 days Lot: 32508 opened: 12/1/2025 elapsed time: 26 days Lot: 32511 opened: 12/18/2025 elapsed time: 17 days Lot: 32511 opened: 1/20/2026 elapsed time: 32 days Lot: 32511 opened: 2/16/2026 elapsed time: 26 days Lot: 32511 opened: 3/02/2026 elapsed time: 14 days 3. A review of patient test records identified the following patients tested on days when expired control materials were used: a) 9/16/2025 to 10/8/2025 45 patients tested (see patient alias list) b) 10/24/2025 to 11/3/2025 27 patients tested (see patient alias list) c) 11/19/2025 to 11/30/2026 22 patients tested (see patient alias list) d) 12/16/2026 to 12/17/2026 8 patients tested (see patient alias list) e) 01/02/2026 to 01/19/2026 47 patients tested (see patient alias list) f) 02/04/2026 to 02/15/2026 39 patients tested (see patient alias list) 4. The laboratory director confirmed the findings after his review of the records in an interview conducted on 03/10/2026 at 1230 hours in exam room 3. D5437 CALIBRATION AND CALIBRATION VERIFICATION CFR(s): 493.1255(a) (a )Unless otherwise specified in this subpart, for each applicable test system the laboratory must perform and document calibration procedures-- (a)(1) Following the manufacturer's test system instructions, using calibration materials provided or specified, and with at least the frequency recommended by the manufacturer; (a)(2) Using the criteria verified or established by the laboratory as specified in 493.1253(b) (3)-- (a)(2)(i) Using calibration materials appropriate for the test system and, if possible, traceable to a reference method or reference material of known value; and (a) (2)(ii) Including the number, type, and concentration of calibration materials, as well as acceptable limits for and the frequency of calibration; and (a)(3) Whenever calibration verification fails to meet the laboratory's acceptable limits for calibration verification. This STANDARD is not met as evidenced by: Based on a review of the hematology calibration records from 2024 and 2025, and staff interview, it was revealed that the laboratory adjusted calibration factors without documentation of performing a calibration. The findings included: 1 . The facility performed a calibration of the Coulter AcT diff 2 hematology analyzer on 10/31/2024. This calibration record showed calibration factors of: WBC (white blood cell count): 1.061 RBC (red blood cell count): 1.139 HGB (hemoglobin): 1.134 MCV (mean Corpuscular hemoglobin): 0.9247 PLT (platelets count): 1.166 MPV (mean platelet volume): 1.059 2. The facility performed the next calibration on 03/27/2025. The pre- -- 2 of 5 -- calibration records indicated the following calibration factors were in use prior to calibration: WBC: 1.061 RBC: 1.147 HGB: 1.137 MCV: 0.9247 PLT: 1.166 MPV: 1.059 There was no documentation to show where the RBC and HGB calibration factors were derived and when they were changed. There was no documentation of a calibration from 10/31/2024 to 03/27/2025. Calibration factors in use as a results of the calibration performed on 03/27/2025 were: WBC: 1.061 RBC: 1.147 HGB: 1.113 MCV: 0.9247 PLT: 1.166 MPV: 1.059 3. The facility performed the next calibration on 07/17/2025. The pre-calibration records indicated the following calibration factors were in use prior to calibration: WBC: 1.061 RBC: 1.147 HGB: 1.105 MCV: 0.9247 PLT: 1.166 MPV: 1.059 There was no documentation to show where the HGB calibration factor was derived and when it was changed. There was no documentation of a calibration being performed between 3/27/2025 to 07/17/2025. Calibration factors in use as a results of the calibration performed on 07/17/2025 were: WBC: 1.061 RBC: 1.147 HGB: 1.105 MCV: 0.9247 PLT: 1.166 MPV: 1.059 4. The facility performed the next calibration on 11/14/2025. The pre-calibration records indicated the following calibration factors were in use prior to calibration: WBC: 1.061 RBC: 1.147 HGB: 1.108 MCV: 0.9247 PLT: 1.166 MPV: 1.059 There was no documentation to show where the HGB calibration factor was derived and when it was changed. There was no documentation of a calibration from 07/17/2025 to 11/14 /2025. 5. The laboratory director confirmed the findings in an interview conducted on 03/10/2026 at 1230 hours in exam room 3. D5785