Viral & Rickettsial Disease Laboratory

CLIA Laboratory Citation Details

1
Total Citation
1
Total Deficiency
1
Unique D-Tag
CMS Certification Number 05D0643850
Address 850 Marina Bay Pkwy, Richmond, CA, 94804
City Richmond
State CA
Zip Code94804
Phone(510) 620-3800

Citation History (1 survey)

Survey - May 5, 2021

Survey Type: Standard

Survey Event ID: G99S11

Deficiency Tags: D5429

Summary:

Summary Statement of Deficiencies D5429 MAINTENANCE AND FUNCTION CHECKS CFR(s): 493.1254(a)(1) For unmodified manufacturer's equipment, instruments, or test systems, the laboratory must perform and document maintenance as defined by the manufacturer and with at least the frequency specified by the manufacturer. This STANDARD is not met as evidenced by: 1. Based on general supervisor interview and Hanta EIA Plate Washer Maintenance record review on May 4, 2021, the laboratory failed to perform maintenance on an enzyme-linked immunosorbent assay (EIA) plate washer as defined by the manufacturer and with at least the frequency specified by the manufacturer. Findings included: a. In the Zoonotic & Vectorborne Diseases Section, the laboratory performed patient IgG and IgM Hanta virus testing using an EIA method that involved the use of a plate washer. b. The manufacturer of the plate washer required that for each day of use, the following maintenance be performed on the plate washer: "wash buffer prepared," "full system prime," "dispense test," 'evac. test," "wash plates," "day rinse," "rinse and soak," and "check and empty waste bottle." c. Laboratory records indicate that on April 13, 2021, the laboratory performed IgG and IgM Hanta virus testing on patient specimens V21B00003, V21S00063, and V21S00064. The laboratory maintained no documentation to indicate that plate washer maintenance had been performed on April 13, 2021. d. According to laboratory personnel, the laboratory tests approximately 20 patient specimens annually for IgG and IgM Hanta virus. 2. Based on general supervisor interviews and SARS-CoV-2 testing thermal cycler maintenance record review on May 5, 2021, the laboratory failed to perform maintenance on a thermal cycler as defined by the manufacturer and with at least the frequency specified by the manufacturer. Findings included: a. The laboratory performed patient SARS-CoV-2 testing using a polymerase chain reaction (PCR) method that involved the use of an Applied Biosystems thermal cycler. b. The manufacturer of the thermal cycler required that for Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 2 -- each month of use, the laboratory "perform a background calibration." c. Laboratory records indicate that on December 10, 2020, the laboratory performed SARS-CoV-2 PCR testing on 90 patient specimens. Within the month of the December 10, 2020 patient SARS-CoV-2 testing, the laboratory maintained no documentation to indicate that the monthly required thermal cycler maintenance had been performed. d. According to laboratory personnel, the laboratory has tested 42,976 patient specimens using their SARS-CoV-2 PCR test from March 2020 to May 5, 2021. -- 2 of 2 --

πŸ”’ Unlock Deficiency Summary

Get full access to the detailed deficiency summary for this facility

One-time payment β€’ Lifetime access