Summary:
Summary Statement of Deficiencies D5211 EVALUATION OF PROFICIENCY TESTING PERFORMANCE CFR(s): 493.1236(a) The laboratory must review and evaluate the results obtained on proficiency testing performed as specified in subpart H of this part. This STANDARD is not met as evidenced by: Based on review of the laboratory's proficiency testing (PT) policy, Wisconsin State Laboratory of Hygiene (WSLH) PT records, and interview with the Medical Assistant (MA), the laboratory failed to review and evaluate the PT results obtained for 1 of 3 WSLH Bacti_Viral events performed in 2025. Findings: 1. The laboratory's Proficiency Testing Policy stated, "The lab evaluates the accuracy of any analyte that is not scored by the lab's PT program." 2. On the day of survey, 8/13/2026 at 12:45 pm, review of the laboratory's WSLH PT results revealed the laboratory failed to review and evaluate the PT results not graded by the PT agency for the following 1 of 3 WSLH Bacti_Viral events performed in 2025: - WSLH PT 2025-Bacti_Viral Event #1: Module 5170, Group A Strep, ST-4 3. The MA confirmed the findings above on 8 /13/2026 at 12:45 pm. D5413 TEST SYSTEMS, EQUIPMENT, INSTRUMENTS, REAGENT CFR(s): 493.1252(b) (b) The laboratory must define criteria for those conditions that are essential for proper storage of reagents and specimens, accurate and reliable test system operation, and test result reporting. The criteria must be consistent with the manufacturer's instructions, if provided. These conditions must be monitored and documented and, if applicable, include the following: (b)(1) Water quality. (b)(2) Temperature. (b)(3) Humidity. (b)(4) Protection of equipment and instruments from fluctuations and interruptions in electrical current that adversely affect patient test results and test reports. Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 2 -- This STANDARD is not met as evidenced by: Based on review of the laboratory's temperature logs, lack of documentation, and interview with the Medical Assistant (MA), the laboratory failed to monitor and document room humidity to ensure operating conditions were met for 1 of 1 Boekel scientific S/N# 041201515 Incubator used to incubate throat cultures from 02/12/2025 to 04/01/2026. Findings include: 1. The Boekel scientific manufacturer's operating environment specifications stated: "80% RH maximum." 2. On the date of the survey, 8/13/2026 at 12:30 pm, the laboratory failed to provide documentation for monitoring of room humidity to ensure operating conditions were met for the following incubator used to incubate throat culture plates from 02/12/2024 to 04/01/2026: - 1 of 1 Boekel scientific S/N# 041201515 3. The laboratory performed 170 Microbiology tests in 2025. (CMS-116 estimated annual volume, dated 6/25/2026). 4. The MA confirmed the findings above on 08/13/2026 at 12:30 am. -- 2 of 2 --