Wiener And Daniels Dpm Pa

CLIA Laboratory Citation Details

1
Total Citation
15
Total Deficiencyies
9
Unique D-Tags
CMS Certification Number 21D2198912
Address 20 Crossroads Drive #15, Owings Mills, MD, 21117
City Owings Mills
State MD
Zip Code21117
Phone(410) 363-4343

Citation History (1 survey)

Survey - May 1, 2023

Survey Type: Standard

Survey Event ID: 3B3411

Deficiency Tags: D5403 D5417 D6094 D5805 D6086 D6093 D6094 D6102 D6120 D5805 D6086 D6093 D5445 D6102 D6120

Summary:

Summary Statement of Deficiencies D5403 PROCEDURE MANUAL CFR(s): 493.1251(b) The procedure manual must include the following when applicable to the test procedure: (1) Requirements for patient preparation; specimen collection, labeling, storage, preservation, transportation, processing, and referral; and criteria for specimen acceptability and rejection as described in 493.1242. (2) Microscopic examination, including the detection of inadequately prepared slides. (3) Step-by-step performance of the procedure, including test calculations and interpretation of results. (4) Preparation of slides, solutions, calibrators, controls, reagents, stains, and other materials used in testing. (5) Calibration and calibration verification procedures. (6) The reportable range for test results for the test system as established or verified in 493.1253. (7) Control procedures. (8)