Summary:
Summary Statement of Deficiencies D5413 TEST SYSTEMS, EQUIPMENT, INSTRUMENTS, REAGENT CFR(s): 493.1252(b) The laboratory must define criteria for those conditions that are essential for proper storage of reagents and specimens, accurate and reliable test system operation, and test result reporting. The criteria must be consistent with the manufacturer's instructions, if provided. These conditions must be monitored and documented and, if applicable, include the following: (1) Water quality. (2) Temperature. (3) Humidity. (4) Protection of equipment and instruments from fluctuations and interruptions in electrical current that adversely affect patient test results and test reports. This STANDARD is not met as evidenced by: Based on surveyor review of laboratory records and procedures and interview with Testing Personnel (Staff A), the laboratory did not define the acceptable humidity requirements and did not document the room temperature or humidity to ensure appropriate storage of reagents and reliable operation of the analyzers in the laboratory for the last two of two years. Findings include: 1. Review of laboratory records showed no documentation of the room temperature or humidity in the laboratory. 2. Review of the procedure, "Daily Laboratory Quality Control", showed the procedure required daily documentation of the room temperature between 20 - 26 degrees C (Celsius). Further review showed no evidence the laboratory evaluated the humidity requirements for the test systems in the laboratory. 3. Interview with Staff A on September 5, 2024, at 11:50 AM confirmed staff did not monitor or document the room temperature and humidity in the laboratory. D6000 MODERATE COMPLEXITY LABORATORY DIRECTOR CFR(s): 493.1403 The laboratory must have a director who meets the qualification requirements of 493. 1405 of this subpart and provides overall management and direction in accordance Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number (X3) Date Survey Completed Name of Provider or Supplier Street Address, City, State -- 1 of 4 -- with 493.1407 of this subpart. This CONDITION is not met as evidenced by: Based on surveyor review of laboratory records and interview with testing personnel, the laboratory director, who is also the technical consultant, did not provide overall management and direction in accordance with 493.1407 of this subpart. Findings include: 1. The laboratory director did not ensure the quality control program was established and maintained for the Pentra C400 analyzer to identify issues with the test system when they occurred in December 2023. See D 6020. 2. The laboratory director did not ensure testing personnel did not report patient test results when the Pentra C400 test system was not functioning properly. See D 6025. D6020 LABORATORY DIRECTOR RESPONSIBILITIES CFR(s): 493.1407(e)(5) The laboratory director is responsible for the overall operation and administration of the laboratory, including the employment of personnel who are competent to perform test procedures, and record and report test results promptly, accurate, and proficiently and for assuring compliance with the applicable regulations. (e) The laboratory director must-- (e)(5) Ensure that the quality control program is established and maintained to assure the quality of laboratory services provided. This STANDARD is not met as evidenced by: Based on surveyor review of quality control (QC) records from the Pentra C400 analyzer and procedures and interview with Testing Personnel (Staff A), the director did not ensure the quality control program for the Pentra C400 was established and maintained to allow testing personnel to identify issues with the test system as they occurred on six of six days in December 2023. Findings include: 1. Review of the 'Monthly QC results December 2023' and the 'Current Worklist QC Results' records for December 18 through 22 and 26, 2023, showed testing personnel performed multiple repeated testing runs over several days for each the following six analytes. The number of runs on each day with each control is shown below. The report identified the two external controls as 'N Multicontrol' and 'P Multicontrol'. Glucose: Level N Multicontrol December 18, 2 runs December 19, 3 runs December 20, 5 runs Level P Multicontrol December 20, 2 runs Potassium: Level N Multicontrol December 18, 5 runs, no acceptable result December 19, 3 runs Level P Multicontrol December 18, 5 runs, no acceptable result December 19, 2 runs December 21, 2 runs Chloride: Level N Multicontrol December 18, 5 runs, no acceptable result Level P Multicontrol December 18, 5 runs, no acceptable result December 21, 2 runs Sodium: Level P Multicontrol December 18, 5 runs, no acceptable result December 21, 2 runs High Density Lipoprotein (HDL) Level N Multicontrol December 21, 3 runs December 22, 2 runs December 26, 5 runs Low Density Lipoprotein (LDL) Level N Multicontrol December 22, 4 runs, no acceptable result December 26, 9 runs Level P Multicontrol December 26, 6 runs The 'Current Worklist QC Results' Daily control report showed the following